Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

30

Click here to load reader

description

Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM). Topographic scan of a glass surface. Types of Microscope. Smaller Scale?. Scanning Probe Microscope. AFM, atomic force microscopy BEEM, ballistic electron emission microscopy CFM, chemical force microscopy - PowerPoint PPT Presentation

Transcript of Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Page 1: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Scanning Tunneling Microscope (STM)and

Atomic Force Microscope (AFM)

Page 2: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Topographic scan of a glass surface

Page 3: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Types of Microscope

Optical Microscope

Electric Microscope

Year of invention 1644 1931

Operation environment

In air, liquid and vacuum

in vacuum

ResolutionHorizontal (x,y)Vertical (z)Magnification

0.5μmNA1-2x10³

2nmNA10-10⁶

Sample preparation Not much Sample should be frozen and dried

Sample conditions Cannot be completely transparent

Can store in vacuumSurface should be conductive

Smaller Scale?

Scanning Probe Microscope

1981

In air, liquid and vacuum

0.1-1.0nm0.01nm5x10²-10⁸

none

Surface cannot be too rough

Page 4: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Scanning Probe Microscope

AFM, atomic force microscopyBEEM, ballistic electron emission microscopyCFM, chemical force microscopyC-AFM, conductive atomic force microscopyEFM, electrostatic force microscopyESTM electrochemical scanning tunneling microscopeFMM, force modulation microscopyKPFM, kelvin probe force microscopyMFM, magnetic force microscopyMRFM, magnetic resonance force microscopyNSOM, near-field scanning optical microscopy (or SNOM, scanning near-field optical microscopy)PFM, Piezoresponse Force MicroscopyPSTM, photon scanning tunneling microscopy

PTMS, photothermal microspectroscopy/microscopySECM, scanning electrochemical microscopySCM, scanning capacitance microscopySGM, scanning gate microscopySICM, scanning ion-conductance microscopySPSM spin polarized scanning tunneling microscopySSRM, scanning spreading resistance microscopySThM, scanning thermal microscopySTM, scanning tunneling microscopySVM, scanning voltage microscopySTP, scanning tunneling potentiometrySHPM, scanning Hall probe microscopySXSTM synchrotron x-ray scanning tunneling microscopy

Of these techniques AFM and STM are the most commonly used for roughness measurements

Page 5: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

STM• a powerful instrument for imaging surfaces at the atomic level• developed in 1981• By Gerd Binnig and Heinrich Rohrer (at IBM Zürich), the Nobel Prize

winners in Physics in 1986• 0.1 nm lateral resolution and 0.01 nm depth resolution

Page 6: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Quantum Tunneling

Page 7: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Quantum Tunneling

V1 = 0

electron

x = -a

V2= Vo

A B C V3 = 0

x = ax

V

e-

Assume the electron has positive energy (E > 0) and E < V2 = Vo.

How can the electron tunnel through the energy barrier under the above assumption?

Find the wave functions in Region A, B and C respectively to get the answer!

Page 8: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Quantum Tunneling1. Write down the Schrodinger’s equations in the 3 regions:

BB

e

EVdx

d

m)(

2 02

22

CACA

e

Edx

d

m ,2

,22

2

2. Get the general solutions:

Em

k

Iex

e

ikxikxA

2

Re)(

)(2

)(

0 EVmk

BeAex

e

KxKxB

Em

k

Tex

e

ikxC

2

)(

I, R and T are related to the incident, reflection and transmission coefficients.

Page 9: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Quantum Tunneling

Consider only the transmission coefficient:

inc

trans

J

J

From )(2

**

xxmi

qJ

2

2

2

2

ikIJ

ikTJ

inc

trans

, we have

2

2

I

T

Tunneling current2

T

Page 10: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Quantum Tunneling

From boundary conditions,

KaeKk

kK

kKKaKk

kKT 2

2

2222222

22 4

)2(2sinh)(

2

, for large Ka

Kae 2current Tunneling

The thickness a of the energy barrier can be found by measuring the tunneling current!

Page 11: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Components of STM

Page 12: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Operation of STM

measure the tunneling current between a conducting tip and conducting sample when a DC bias voltage is applied.

Resolution in z (perpendicular to surface) is ~0.01 nm and in x/y is ~0.1 nm, resulting in atomic-resolution images.

Page 13: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

13

Consider two metals, one is a tip and the other is a surface, separated by a finite distance uniform potential barrier.

Electron wave function can tunnel through the barrier and leads to finite current.

The probability is

Lmetal 1 metal 2

22 2

2 2

4 KLkKT e

k K

As the two metals move apart, the probability of

electrons tunneling decreases exponentially and as they move closer the tunneling increases.

Tunneling current

Operation of STM

Page 14: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

14

Scan directiontip

Lower currentConstant height

Higher current

Atomic surface

The air gap (barrier) is several nm

I (nA)

x (nm)

The current profile duplicates the atomic surface

Operation of STM

Page 15: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

15

The resolution of bending can be as low as 0.2 nm

Operation of STM

Page 16: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

• The attention paid to the first problem and the engineering solution to it is the difference between a good microscope and a not so good microscope - it need not worry us here, sufficient to say that it is possible to accurately control the relative positions of tip and surface by ensuring good vibrational isolation of the microscope and using sensitive piezoelectric positioning devices.

• Tip preparation is a science in itself - having said that, it is largely serendipity which ensures that one atom on the tip is closer to the surface than all others.

Page 17: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Let us look at the region where the tip approaches the surface in greater detail ....

... the end of the tip will almost invariably show a certain amount of structure, with a variety of crystal facets exposed ...

Page 18: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

… and if we now go down to the atomic scale ....

... there is a reasonable probability of ending up with a truly atomic tip.

Page 19: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

• If the tip is biased with respect to the surface by the application of a voltage between them then electrons can tunnel between the two, provided the separation of the tip and surface is sufficiently small - this gives rise to a tunnelling current.

• The direction of current flow is determined by the polarity of the bias.

If the sample is biased -ve with respect to the tip, then electrons will flow from the surface to the tip as shown above, whilst if the sample is biased +ve with respect to the tip, then electrons will flow from the tip to the surface as shown below.

Page 20: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

The name of the technique arises from the quantum mechanical tunnelling-type mechanism by which the electrons can move between the tip and substrate. Quantum mechanical tunnelling permits particles to tunnel through a potential barrier which they could not surmount according to the classical laws of physics - in this case electrons are able to traverse the classically-forbidden region between the two solids as illustrated schematically on the energy diagram below.

This is an over-simplistic model of the tunnelling that occurs in STM but it is a useful starting point for understanding how the technique works.

In this model, the probability of tunnelling is exponentially-dependent upon the distance of separation between the tip and surface : the tunnelling current is therefore a very sensitive probe of this separation.

Page 21: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

• Imaging of the surface topology may then be carried out in one of two ways: – in constant height mode (in which the tunnelling current is monitored as the

tip is scanned parallel to the surface) – in constant current mode (in which the tunnelling current is maintained

constant as the tip is scanned across the surface) If the tip is scanned at what is nominally a constant height above the surface, then there is actually a periodic variation in the separation distance between the tip and surface atoms. At one point the tip will be directly above a surface atom and the tunnelling current will be large whilst at other points the tip will be above hollow sites on the surface and the tunnelling current will be much smaller.

A plot of the tunnelling current v's tip position therefore shows a periodic variation which matches that of the surface structure - hence it provides a direct "image" of the surface (and by the time the data has been processed it may even look like a real picture of the surface ! ).

Page 22: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

In practice, however, the normal way of imaging the surface is to maintain the tunnelling current constant whilst the tip is scanned across the surface. This is achieved by adjusting the tip's height above the surface so that the tunnelling current does not vary with the lateral tip position. In this mode the tip will move slightly upwards as it passes over a surface atom, and conversely, slightly in towards the surface as it passes over a hollow.

The image is then formed by plotting the tip height (strictly, the voltage applied to the z-piezo) v's the lateral tip position.

Page 23: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Piezoelectric TubePiezoelectric Material• piezoelectric effect was

discovered by Pierre Curie and Jacques Curie

• (1880)

• An applied mechanical stress will generate a voltage

• An applied voltage will change the shape of the solid by a small amount

• most well-known piezoelectric material is quartz (SiO2)

Page 24: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

BimorphTwo long, thin plates of piezoelectric material are glued together, with a metal film sandwiched in between . Two more metal films coverthe outer surfaces . Both piezoelectric plates are poled along the same direction, perpendicularto the large surface

By applying a voltage, stress of oppositesign is developed in both plates, which generates a torque

Page 25: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Tube

Page 26: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Making a sharp tip

• By electrochemical etching method

By applying different voltages, some part of tip were washed with the distilled water

Consider a gold tip prepared in 0.8 M KCN solution from a gold wire (Au),

the following electrochemical redox reaction takes place:

4Au(s)+8KCN(aq)+O2(g)+2H2O→4Au(CN)2

−(aq)+4OH−(aq)+8K+(aq)

Page 27: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Manipulation of atoms using STM

Page 28: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Lithography and micromanipulation

• The interactions between the STM tip and substrate can be used to modify the surface in a controlled way.

• This can be done in a number of ways.

• Eg Eigler and Schweizer manipulated xenon atoms on a Nickel(110) surface under UHV conditions, with everything at 4K.– They reported that under these conditions

everything was stable for days. – This allowed them to move single atoms at a

time until they achieved the following result.

Page 29: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

• Obtained by manipulating CO on a Pt(111) surface.

Page 30: Scanning Tunneling Microscope (STM) and Atomic Force Microscope (AFM)

Manipulation of atoms using STM

atom