Journal of Chromatography A, 896 (2000) 31-39

16
Cryogen free automated gas chromatography for the measurement of ambient volatile organic compounds Journal of Chromatography A, 896 (2000) 31-39 Jia-Lin Wang*, Wei-Li Chen, Yen-Hung Lin, Cheng-Hsiung Tsai

description

Cryogen free automated gas chromatography for the measurement of ambient volatile organic compounds. Journal of Chromatography A, 896 (2000) 31-39. Jia-Lin Wang*, Wei-Li Chen, Yen-Hung Lin, Cheng-Hsiung Tsai. experimental. HP6890-Dual traps, FID PLOT for C3-C7 30m × 0.32mm ; 8 μ m. - PowerPoint PPT Presentation

Transcript of Journal of Chromatography A, 896 (2000) 31-39

Page 1: Journal of Chromatography A, 896 (2000) 31-39

Cryogen free automated gas chromatography for the measurement of

ambient volatile organic compounds

Journal of Chromatography A, 896 (2000) 31-39

Jia-Lin Wang*, Wei-Li Chen, Yen-Hung Lin, Cheng-Hsiung Tsai

Page 2: Journal of Chromatography A, 896 (2000) 31-39
Page 3: Journal of Chromatography A, 896 (2000) 31-39
Page 4: Journal of Chromatography A, 896 (2000) 31-39
Page 5: Journal of Chromatography A, 896 (2000) 31-39
Page 6: Journal of Chromatography A, 896 (2000) 31-39

experimentalHP6890-Dual traps, FID

PLOT for C3-C730m × 0.32mm ; 8 μm.

DB-1 for C6-C1260m × 0.32mm ; 1 μm.

Uncoated column 5 m × 0.25 mm

Page 7: Journal of Chromatography A, 896 (2000) 31-39
Page 8: Journal of Chromatography A, 896 (2000) 31-39

C3-C7 NMHCs in Taiwan EPA monitoring station

1th

250th

673th

Page 9: Journal of Chromatography A, 896 (2000) 31-39

preconcentration system

Page 10: Journal of Chromatography A, 896 (2000) 31-39

Minimizing longitudinal diffusion

0.5mm ID × 1.6mmNon- Cooling GC oven

0.5mm ID × 1.6mmCooling GC oven

Page 11: Journal of Chromatography A, 896 (2000) 31-39

0.5mm ID × 1.6mmCooling GC oven

0.25mm ID × 0.8mmNon-Cooling GC oven

Page 12: Journal of Chromatography A, 896 (2000) 31-39

Reproducibility

Page 13: Journal of Chromatography A, 896 (2000) 31-39

Propane from PLOT column

Flow rate = 50ml / min

Benzne from PLOT column

Flow rate = 50ml / min

Linearity

Page 14: Journal of Chromatography A, 896 (2000) 31-39

(a) CO by NDIR

(b) Benzene by automated system

Page 15: Journal of Chromatography A, 896 (2000) 31-39

R2> 0.92

R2> 0.25

Page 16: Journal of Chromatography A, 896 (2000) 31-39

conclusion

This work was to build an automated GC system for measuring VOCs either in the lab or the field without using cryogen in preconcentration and column separation.

Two traps were coupled to their designated columns to form suitable pairs to provide adequate resolution without cryogen use.