Journal of Chromatography A, 896 (2000) 31-39
description
Transcript of Journal of Chromatography A, 896 (2000) 31-39
Cryogen free automated gas chromatography for the measurement of
ambient volatile organic compounds
Journal of Chromatography A, 896 (2000) 31-39
Jia-Lin Wang*, Wei-Li Chen, Yen-Hung Lin, Cheng-Hsiung Tsai
experimentalHP6890-Dual traps, FID
PLOT for C3-C730m × 0.32mm ; 8 μm.
DB-1 for C6-C1260m × 0.32mm ; 1 μm.
Uncoated column 5 m × 0.25 mm
C3-C7 NMHCs in Taiwan EPA monitoring station
1th
250th
673th
preconcentration system
Minimizing longitudinal diffusion
0.5mm ID × 1.6mmNon- Cooling GC oven
0.5mm ID × 1.6mmCooling GC oven
0.5mm ID × 1.6mmCooling GC oven
0.25mm ID × 0.8mmNon-Cooling GC oven
Reproducibility
Propane from PLOT column
Flow rate = 50ml / min
Benzne from PLOT column
Flow rate = 50ml / min
Linearity
(a) CO by NDIR
(b) Benzene by automated system
R2> 0.92
R2> 0.25
conclusion
This work was to build an automated GC system for measuring VOCs either in the lab or the field without using cryogen in preconcentration and column separation.
Two traps were coupled to their designated columns to form suitable pairs to provide adequate resolution without cryogen use.