M A S S S P E C T R O M E T R YP
rod
uct S
pecificatio
ns
Precise, Verifi able Isotope Ratio MeasurementsThermo Scientifi c™ Delta Ray™ analyzer with URI Connect universal reference interface is a high performance, mid-infrared laser based, isotope ratio infrared spectrometer. It offers simultaneous determination of δ13C and δ18O in CO
2 with a precision of
<0.05‰ between ambient concentrations and up to 100% with automatic dilution.
The URI Connect universal reference interface enables the analysis of discrete samples from vials in an autosampler, or, through an injection port, the analysis of samples from syringes or bags. If desired, the port can be customized by the user to fi t specifi c sample containers. The sample gas is collected in a variable volume, from which the system can automatically determine the concentration and then adjust the dilution to perform isotopic analysis under optimized conditions. Moreover, the URI interface provides fully automated referencing and calibration for verifi able measurements and long term confi dence. Smart referencing adjusts the reference gas concentration to the sample to achieve superior performance.
Isotope Ratios from Discrete Samples To-GoThe modular design, low weight and built in referencing make the Delta Ray the analyzer of choice for demanding fi eld and laboratory applications of discrete samples such as isotopic composition of dissolved inorganic carbon (DIC), carbonate analysis or microbiology and labeling applications all in combination with an optional autosampler. The autosampler is small enough to be easily deployed to the fi eld. The Delta Ray analyzer can also be used for continuous greenhouse gas monitoring, or in ecology, plant science, carbon sequestration and storage research, or for volcanic monitoring. Feature rich data is acquired autonomously 24 hours a day, 7 days a week with a time resolution as low as 1 s directly where the sample originates. A quickstart guide gets you going in minimum time.
The Thermo Scientifi c Delta Ray CO2 isotope ratio infrared spectrometer with URI Connect allows you to easily interface with samples available in amounts as little as 80 µg of CO2, such as in vials, syringes or bags and then to be analyzed in the Delta Ray spectrometer. Discover more with a mobile isotope laboratory:
• δ13C and δ18O of CO2
• CO2 concentration
• Precision < 0.05‰
• Mid-infrared laser based isotope ratio infrared spectrometer
• Up to 1 Hz data
• Portable, fi eld deployable
• URI Connect to interface discrete samples
• Easy connection to samples in autosampler, syringes or bags
• Global support
Thermo Scienti� c Delta Ray CO2 Isotope Ratio Infrared Spectrometer with
URI Connect
2 Real-Time ResultsPowered by the Thermo Scientific™ Qtegra™ Intelligent Scientific Data Solution (ISDS™), the data acquisition is driven by powerful workflows. Select sample ports, choose vial positions in a tray, drive external valves, or synchronize your measurements with external triggers all with a few clicks. Qtegra software then acquires the data into labbooks without user interaction. The sample will be diluted automatically to within the guaranteed specification range (380–1500 ppm). The fully referenced and calibrated results can be inspected immediately and are easily exported to Excel or your favorite data analysis package. Qtegra runs on the Windows 7 computer, which in connection with an internet connection allows you to control your Delta Ray system from anywhere in the world.
Simple, Robust, Powerful Mid-Infrared TechnolgyAt the core of the analyzer is a difference frequency generation (DFG) laser that operates in the mid-infrared, where the absorption of CO
2 and its isotopologues are
so strong, that 5 m path length are sufficient to achieve a precision as low as 0.05‰. It is simple, robust and powerful.
Sample gas is dried in the reference interface to prevent any interaction of water and CO
2 (isotope exchange) with the
added benefit of providing dry mole fraction concentration data. In the measurement cell, where the laser analyses the gas, pressure and temperature are precisely controlled.
Performance Specifications, δ13C δ18O CO2 Concentration in Air
Precision, 60 sec averaging, <0.15‰ 1SD <0.15‰ 1SD 70 ppb 30 replicate
External reproducibility*, 0.07‰ 1SD 0.1‰ 1SD 300 sec averaging, 10 replicates
Operating range 200–3,500 ppm
Guaranteed specifications range 300–1,500 ppm without dilution (100% CO
2 with
optional dilution)
Measurement interval 1, 10, 60 s
Discrete Sample Specifications
Ambient concentration 100 mL sample <0.15‰ <0.2‰
10 mL volume with 1%..100% CO2 <0.07‰ <0.1‰
Carbonates >200 µg 0.15‰ <0.2‰
Amount of CO2 in sample 80 µg
Throughput 100 samples/day
Enrichment 13C up to 25,000‰ 5‰ precision 60 s averages, 10 replicates
Cell exchange time (response time) 35 s @ 80 sccm sample flow
Start up time (in spec) 180 min
Measurement technique Tunable laser direct absorption in the mid infrared (4.3 µm) TDLAS
* demonstrated with Zero enrichment template: sample and reference are the same gas, 5 min sample average, 5 min reference average, repeat at least 10 times.
Analytical Specifications
Operating Conditions
Temperature range 10–35 °C
Temperature gradient 0.2 °C/min
Humidity 10–80% R.H. up to 31 °C 10–50% R.H. at 40 °C1 non condensing
Altitude <2000 m
Size 19” rack mount chassis Analyzer: 588 × 424 × 219 mm (D × W × H) URI: 588 × 424 × 131 mm (D × W × H) Autosampler: 390 × 370 × 460 mm (D × W × H)
Weight Analyzer: 25 kg URI: 12 kg Autosampler: 18 kg
Power requirements steady 100–240 V, 50–60 Hz state/warm up Analyzer: 150 W typical/500 W start up at 25 °C URI: <160 W Autosampler: 100–200 V, 47–60 Hz, 1.9 A
Sample temperature -10–45°C
Sample pressure Sensor input 700–1,200 mbar
Sample flow Sensor input 80 sccm @ 1000 mbar abs.
Pump Internal, oil free
Pro
du
ct Sp
ecification
sP
rod
uct S
pecificatio
ns
PS30309-EN 0715S
Africa +43 1 333 50 34 0Australia +61 3 9757 4300Austria +43 810 282 206Belgium +32 53 73 42 41Canada +1 800 530 8447China 800 810 5118 (free call domestic)
400 650 5118
Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0Finland +358 10 3292 200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591
Japan +81 45 453 9100Korea +82 2 3420 8600Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Norway +46 8 556 468 00
Russia/CIS +43 1 333 50 34 0Singapore +65 6289 1190Spain +34 914 845 965Sweden +46 8 556 468 00Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752
www.thermoscientific.com/DeltaRay©2015 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Microsoft and Windows are trademarks of Microsfort Corp. Swagelok is a trademark of Swagelok Company.All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.
Thermo Fisher Scientific (Bremen) GmbHManagement System Registered to ISO 9001:2008
Transport Conditions (non operating)
Temperature range -33–63 °C
Vibrations MIL-STD 810G
Calibration
Calibration gases for referencing, 2 × pure CO2 with known δ13C δ18O
linearity and delta scale calibration Pressure range 1–12 bar relative (initial supply included) Consumption: 400 µL/min, 2–4 ~mL/per calibration
Synthethic air: CO2 free air for sample CO
2 concentration <0.2 ppm (<20‰ off from sample)
dilution and reference gas mixing Consumption up to 200 sccm (user provided) Pressure range 1–12 bar rel.
Connections
Electrical Video: VGA port, DVI port 4 USB ports 1 ethernet 1 Gbit 4 digital outs for 12 V or 24 V valves max 3.3 W each 2 digital in for trigger (TTL) 3 analog outs 0.5 V (0.5 mV resolution) 1 analog in 0.5 V 1 mV resolution
Gas Sample A in (1–12 bar relative) Sample B (700–1,200 mbar (abs.)), passing through dryer Carrier out for external dilution
Fittings 1/16” Swagelok 10–32 coned port