2
PULSED DYE LASER ACCESSORY Mid-Infrared Generation (MIR) The mid-infrared generation unit (MIR) is designed to generate powerful laser pulses in the wavelength range from 4.8 to 11 μm. It is operated with a Quanta-Ray ® Nd:YAG pump laser, a PrecisionScan dye laser equipped with either a single or a double grating resonator with 1800 lines / mm gratings, and an optical parametrical amplifier unit (OPA-NIR). PRINCIPLES A Quanta-Ray frequency doubled, injection seeded Nd:YAG laser is used to pump a PrecisionScan dye laser operated in the red spectral range. The dye laser output and a fraction of the residual 1064 nm pulse from the Nd:YAG laser are mixed in a LiNbO 3 crystal for difference frequency generation. The generated IR pulse passes a second LiNbO 3 crystal pumped by the remaining 1064 nm pulse, and is tuned to act as an optical parametrical amplifier. The OPA amplifies the initial IR pulse and generates signal and idler frequency. The difference frequency of these two pulses is generated in a AgGaSe 2 crystal. Due to injection seeding, the resulting linewidth is determined by the configuration of the dye laser oscillator. TUNING RANGE Oscillator (Amplifier) Nd:YAG Laser 1064 nm 532 nm PrecisionScan Dye Laser OPA-NIR Telescope Seeder Pre-Amp MIR Output Oscillator SHG Signal + Idler Dichroic Wavelength Separation Dump AgGaSe 2 Crystal Polarization Matching Optics MIR Main Amp (EBP) MIR tuning curves, when pumping with approximately 550 mJ @ 532 nm Wavelength (μm) Energy (μJ) Styryl 8 (DMSO) Styryl 8 (Ethanol) 5 6 7 8 9 10 11 12 140 120 100 80 60 40 20 0 MIR tuning curves, when pumping with approximately 550 mJ @ 532 nm

Transcript of Mid-Infrared Generation (MIR) - spectra-physics.com · PULSED DYE LASER ACCESSORY Mid-Infrared...

PULSED DYE LASER ACCESSORY

Mid-Infrared Generation (MIR)The mid-infrared generation unit (MIR) is designed to generate powerful laser pulses in the wavelength range from 4.8 to 11 μm. It is operated with a Quanta-Ray® Nd:YAG pump laser, a PrecisionScan dye laser equipped with either a single or a double grating resonator with 1800 lines / mm gratings, and an optical parametrical amplifier unit (OPA-NIR).

PRINCIPLES

A Quanta-Ray frequency doubled, injection seeded Nd:YAG laser is used to pump a PrecisionScan dye laser operated in the red spectral range. The dye laser output and a fraction of the residual 1064 nm pulse from the Nd:YAG laser are mixed in a LiNbO3 crystal for difference frequency generation. The generated IR pulse passes a second LiNbO3 crystal pumped by the remaining 1064 nm pulse, and is tuned to act as an optical parametrical amplifier. The OPA amplifies the initial IR pulse and generates signal and idler frequency. The difference frequency of these two pulses is generated in a AgGaSe2 crystal. Due to injection seeding, the resulting linewidth is determined by the configuration of the dye laser oscillator.

TUNING RANGE

Oscillator(Amplifier)

Nd:YAG Laser

1064 nm

532 nm

PrecisionScan Dye Laser OPA-NIR

Telescope

Seeder

Pre-Amp

MIR Output

Oscillator

SHG

Signal + Idler

DichroicWavelengthSeparation

Dump

AgGaSe2

Crystal

PolarizationMatching Optics

MIR

Main Amp(EBP)

MIR tuning curves, when pumping with approximately 550 mJ @ 532 nm

Wavelength (µm)

Ener

gy (µ

J)

Styryl 8 (DMSO)

Styryl 8 (Ethanol)

5 6 7 8 9 10 11 12

140

120

100

80

60

40

20

0

MIR tuning curves, when pumping with approximately 550 mJ @ 532 nm

Mid-Infrared Generation (MIR)

© 2015 Newport Corporation. All Rights Reserved. Quanta-Ray, Spectra-Physics and the Spectra-Physics logo are registered trademarks of Newport Corporation. Spectra-Physics Santa Clara, California, Stahnsdorf, Germany, Rankweil, Austria and Tel Aviv, Israel have all been certified compliant with ISO 9001.

3635 Peterson Way, Santa Clara, CA 95054, USAPHONE: 1-800-775-5273 1-408-980-4300 FAX: 1-408-980-6921 EMAIL: [email protected]

www.spectra-physics.com

China +86-10-6267-0065 [email protected] +33-(0)1-60-91-68-68 [email protected] +81-3-3794-5511 [email protected] +886 -(0)2-2508-4977 [email protected] +65-6664-0400 [email protected]

Belgium +32-(0)0800-11 257 [email protected] +31-(0)30 6592111 [email protected] Kingdom +44-1235-432-710 [email protected] / Austria / Switzerland +49-(0)6151-708-0 [email protected]

SP-DS-20150210-16

PULSED DYE LASER ACCESSORY

ENERGY OUTPUT1, 2

Pump Laser Dye Laser Output Energy550 mJ @ 532 nm PrecisionScan-G-18 100 µJ

550 mJ @ 532 nm PrecisionScan-D-18 90 µJ

430 mJ @ 532 nm PrecisionScan-G-18 70 µJ

430 mJ @ 532 nm PrecisionScan-D-18 63 µJ

1. Due to continuous product improvements, specifications are suject to change without notice. 2. MIR energy output at 7.5 µm (maximum Styryl 8 in Ethanol). See tuning curve for output energies at other wavelengths.

GENERAL CHARACTERISTICS1

Required EquipmentQuanta-Ray injection seeded Nd:YAG laserPrecisionScan dye laserOPA-NIR near-infrared unit

Maximum Pump Energy 550 mJ @ 532 nm

Dye Laser Resonator 1800 lines/mm gratings recommended1800 lines/mm double grating resonator possible

Dye Laser Amplifier Enhanced Beam Profile cell recommended

Repetition Rate 10 Hz recommended; contact factory for higher rep rates

Crystal Tuning Mode Look-up table

IR Beam Polarization Horizontal, >98%

IR Beam Diamter (typical) 5–9 mm, depending on amplifier cell type

IR Beam Divergence <0.5 mrad

IR Bandwidth <1.5 x dye laser bandwidth

Tuning Range DFM-8M 4.8–8.0 μm (standard)

1. Due to continuous product improvements, specifications are suject to change without notice.

MIR DIMENSIONS

UV Output EndDye Input EndSide View

833

52 22

207

– 24

0

160465

76932

Dimensions in mmSpecifications are subject to change without notice

Manufactured by Sirah