Download - Milliohm Meter - SEW · Milliohm Meter FEATURES ... 4136 mO : 1530g 4137 mO : 1500g Safety standard EN 61010-1 CAT IV 20V EN 61326-1 Accessories Instruction manual Test leads Shoulder

Transcript
Page 1: Milliohm Meter - SEW · Milliohm Meter FEATURES ... 4136 mO : 1530g 4137 mO : 1500g Safety standard EN 61010-1 CAT IV 20V EN 61326-1 Accessories Instruction manual Test leads Shoulder

Milliohm Meter

FEATURES ● Microprocessor-controlled. ● Four terminal measurement. ● Measures down to 100μΩ. ● 5 Ranges from 200.0mΩ to 2000Ω. ● Maximum resolution of 100μΩ. ● Three test currents with over-temperature protection. ● Protection against inadvertent connection to over voltage. ● Large LCD. ● Potential lead resistance and current lead resistancechecks. ● "Full-featured” EnerSaveTM inside. ● EnerSaveTM Auto hold. ● EnerSaveTM Auto off. ● Lightweight, robust & compact. ● Indicators show if reading may be invalid (RP, RC, and temperature). ● "O-Ring” sealed case.

SPECIFICATIONS

Measuring ranges

0-200.0mΩ in steps of 100μΩ0-2000mΩ in steps of 1mΩ0-20.00Ω in steps of 10mΩ0-200.0Ω in steps of 100mΩ0-2000Ω in steps of 1Ω

Accuracy

±0.5% of reading ±2 digits over theoperating temperature range, -15ºC to +55ºC, with the supplied test leads

Test current1mA=>2000Ω range10mA=>200 / 20Ω ranges100mA=>2000m / 200mΩ ranges

Test current accuracy ±0.1%

Protection fusesMains = 0.5A, HBC, 5 × 20mm, DINCurrent = 0.5A, HBC, 5 × 20mm, DINVoltage = 0.5A, HBC, 5 × 20mm, DIN

Maximum outputvoltage (C1~C2) 20V

Dimensions 250(L) × 190(W) × 110(D)mm

Weight 4136 mO : 1530g4137 mO : 1500g

Safety standard EN 61010-1 CAT IV 20VEN 61326-1

Accessories Instruction manual Test leads Shoulder belt 1.5V (AA) × 8 (4137 mO)

4136 mO 4137 mOPower Source : 110V AC or 230V AC Power Source : 12V DC

4 Terminal Test

4-terminal = LED off

3-terminal = Led ON 4/3 T

C1 P1

C1 P1

DIGITAL MILLIOHM METER

LOCK

• Short Test duration if TESTis depressed for less than2Sec.

• Long Test Duration if TESTis depressed for more than3Sec.

20.002000m

200.0mOFF

ON

2000200.0

AUTO-HOLDMICROPROCESSOR CONTROLLED

EnerSave™TEST RP

RC

RP

NOTEST

Test Current Range1mA10mA10mA100mA100mA

200.020.00

2000

2000m200.0m

C1 P1 C2P2

MAX. DC20V

C1 P1 C2P2

MAX. DC20V

HOLD

kADC

AC

ms

mVHz

kM

M AX M INHOLD

kADC

AC

ms

mVHz

kM

MAX M IN

RX

RBRA

CompressionJoint

Simplified Measurement

4-terminal = LED off

3-terminal = Led ON 4/3 T

C1 P1

C1 P1

DIGITAL MILLIOHM METER

LOCK

• Short Test duration if TESTis depressed for less than2Sec.

• Long Test Duration if TESTis depressed for more than3Sec.

20.002000m

200.0mOFF

ON

2000200.0

AUTO-HOLDMICROPROCESSOR CONTROLLED

EnerSave™TEST RP

RC

RP

NOTEST

Test Current Range1mA10mA10mA100mA100mA

200.020.00

2000

2000m200.0m

C1 P1 C2P2

MAX. DC20V

C1 P1 C2P2

MAX. DC20V

HOLD

kADC

AC

ms

mVHz

kM

M AX M INHOLD

kADC

AC

ms

mVHz

kM

MAX M IN

RX

AL-34B TEL-4136C B(optional)

5F., NO.105, Jhongcheng Rd., Tucheng Dist., New Taipei City 23674, Taiwan (R.O.C.)

TEL: 886-2-22681528FAX: 886-2-22681529

e-mail: [email protected]

MEASURE WITH PRECISION