Milliohm Meter
FEATURES ● Microprocessor-controlled. ● Four terminal measurement. ● Measures down to 100μΩ. ● 5 Ranges from 200.0mΩ to 2000Ω. ● Maximum resolution of 100μΩ. ● Three test currents with over-temperature protection. ● Protection against inadvertent connection to over voltage. ● Large LCD. ● Potential lead resistance and current lead resistancechecks. ● "Full-featured” EnerSaveTM inside. ● EnerSaveTM Auto hold. ● EnerSaveTM Auto off. ● Lightweight, robust & compact. ● Indicators show if reading may be invalid (RP, RC, and temperature). ● "O-Ring” sealed case.
SPECIFICATIONS
Measuring ranges
0-200.0mΩ in steps of 100μΩ0-2000mΩ in steps of 1mΩ0-20.00Ω in steps of 10mΩ0-200.0Ω in steps of 100mΩ0-2000Ω in steps of 1Ω
Accuracy
±0.5% of reading ±2 digits over theoperating temperature range, -15ºC to +55ºC, with the supplied test leads
Test current1mA=>2000Ω range10mA=>200 / 20Ω ranges100mA=>2000m / 200mΩ ranges
Test current accuracy ±0.1%
Protection fusesMains = 0.5A, HBC, 5 × 20mm, DINCurrent = 0.5A, HBC, 5 × 20mm, DINVoltage = 0.5A, HBC, 5 × 20mm, DIN
Maximum outputvoltage (C1~C2) 20V
Dimensions 250(L) × 190(W) × 110(D)mm
Weight 4136 mO : 1530g4137 mO : 1500g
Safety standard EN 61010-1 CAT IV 20VEN 61326-1
Accessories Instruction manual Test leads Shoulder belt 1.5V (AA) × 8 (4137 mO)
4136 mO 4137 mOPower Source : 110V AC or 230V AC Power Source : 12V DC
4 Terminal Test
4-terminal = LED off
3-terminal = Led ON 4/3 T
C1 P1
C1 P1
DIGITAL MILLIOHM METER
LOCK
• Short Test duration if TESTis depressed for less than2Sec.
• Long Test Duration if TESTis depressed for more than3Sec.
20.002000m
200.0mOFF
ON
2000200.0
AUTO-HOLDMICROPROCESSOR CONTROLLED
EnerSave™TEST RP
RC
RP
NOTEST
Test Current Range1mA10mA10mA100mA100mA
200.020.00
2000
2000m200.0m
C1 P1 C2P2
MAX. DC20V
C1 P1 C2P2
MAX. DC20V
HOLD
kADC
AC
ms
mVHz
kM
M AX M INHOLD
kADC
AC
ms
mVHz
kM
MAX M IN
RX
RBRA
CompressionJoint
Simplified Measurement
4-terminal = LED off
3-terminal = Led ON 4/3 T
C1 P1
C1 P1
DIGITAL MILLIOHM METER
LOCK
• Short Test duration if TESTis depressed for less than2Sec.
• Long Test Duration if TESTis depressed for more than3Sec.
20.002000m
200.0mOFF
ON
2000200.0
AUTO-HOLDMICROPROCESSOR CONTROLLED
EnerSave™TEST RP
RC
RP
NOTEST
Test Current Range1mA10mA10mA100mA100mA
200.020.00
2000
2000m200.0m
C1 P1 C2P2
MAX. DC20V
C1 P1 C2P2
MAX. DC20V
HOLD
kADC
AC
ms
mVHz
kM
M AX M INHOLD
kADC
AC
ms
mVHz
kM
MAX M IN
RX
AL-34B TEL-4136C B(optional)
5F., NO.105, Jhongcheng Rd., Tucheng Dist., New Taipei City 23674, Taiwan (R.O.C.)
TEL: 886-2-22681528FAX: 886-2-22681529
e-mail: [email protected]
MEASURE WITH PRECISION