Development of low cost Development of low cost openopen--path instrument for path instrument for

monitoring of COmonitoring of CO22 at at sequestration sites sequestration sites

Power Energy Environment Research (PEER) Center Power Energy Environment Research (PEER) Center CALTECHCALTECHMay, 2006May, 2006

DOE Award number: DEDOE Award number: DE--PS26PS26--04NT4202304NT42023--2D 2D

Page 2: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

List of commercial open-air gas monitoring instruments

SourceModel

Number/DescriptionSpecifications Approx.

PriceComments

Boreal Laser Gas Finder Path Length1 m to <1000 mgood to 1 ppm CO2

$40K Laser light/reflectorSingle channel unit

Boreal Laser Gas FinderMC Path Length1 m to <1000 mgood to 1 ppm CO2

$150K Laser light/reflectorUp to 8 channels

MSA Instruments Model 3600 Infrared Gas Monitor

Available ranges0 to 0.2% CO20 to 1.0% CO20 to 5.0% CO2

$1605Note is $4065 for explosion proof

IR detection. Capable of remote sampling from 300 feet

MSA Instruments Model 3630 Infrared Gas Monitor

Long term stability of +/5% 0 to 0.2% CO2

$520 IR detection. Designed especially for indoor use. Capable of remote sampling/300 feet

RKI Instruments Spectralert D/DR InfraRedGas Detector

CO2 version:0-2000, 5000, 10,000,50,000 ppmAccuracy +/-2% fsd on all ranges

Remote sensor option up to 50 metersOffshore stainless steel construction.

Page 3: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Goals the projectGoals the project

Extend the monitoring path up to 5 km using commercial low cost telecom amplifiersDemonstrate long range (>1 km) open path measurement of CO2 using FM tunable laser spectrometerDemonstrate detection of 1ppmV CO2 change in the air (nominal ~380ppmV of CO2 in air) Build a field test ready unit

Page 4: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Progress highlights after 6 QuartersProgress highlights after 6 Quarters

Finished proof of concept bench top demo in September, 2005Tested the bench top instrument in open air over a 100 meters distance in January, 2006Built launcher, retroreflector and collection optics for long path (>100 meters) measurements, March, 2006Moving towards 1 km test by July, 2006

Page 5: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Frequency modulation spectroscopy(homodyne detection)

DiodeLaser

RF generatorω

SamplePhoto-detector

Mixer

Low-passfilter

Injection currentmodulation

Phase shifter

Dataacquisition

Page 6: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Advantages of tunable diode laser frequency modulation spectroscopy (TDL FM)

Avoids 1/f source noise by shifting detection to frequency ωLow cost of diode lasers

Page 7: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

TDL FM spectroscopy at atmospheric pressure

At the pressure of 1 atm pressure-broadened linewidths (HWHM) are ~0.07-0.1 cm-1 = 2-3 GhzThe modulation frequency needs to be comparable with the width of the absorption features

Disadvantages:High-bandwidth detectors have small area (smaller than Ø0.3mm) which leads to low collection efficiencyRequires sophisticated high-bandwidth electronics

Page 8: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Two-tone frequency modulation

DiodeLaser

Microwave (2GHz)generator

ω

SampleLow-bandwidthPhotodetector

Mixer

Low-passfilter

Injection currentmodulation

Phaseshifter

Dataacquisition

MixerLF generator

ΩFrequency

doubler

ω±Ω 2Ω

Page 9: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Advantages of Two-tone FM

Utilizes the full width of pressure-broadened rotational linesDetection is performed at low-frequency Ω (1-10MHz) permitting to use larger photo detectors and RF amplifiers.

Page 10: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Challenges of open-path detection

Launcher

Collectionoptics +Detector

Atmospheric convection leads to power fluctuations and phase noise

powerfluctuations

δφ=δnphase fluctuations(noise)

Page 11: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Our approach to open-path detection

DiodeLaser

Microwave (2GHz)generator

ω

SampleLow-bandwidthPhotodetector

Narrowband-passfilter 2Ω

Low-passfilter

Injection currentmodulation

Dataacquisition

MixerLF generator

Ω

ω±Ω

Rectifier

Page 12: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Benefits of our approach

Phase insensitive detection --- immune to phase noise introduced by atmospheric convection in the long open pathA unique design similar to an AM radio --- phase fluctuations on the

time scale of kHz is not affecting our measurement.

Two-tone FM detection that features large detector area --- easy to align and capture signal, better stability

Our improvements will make deployment and Our improvements will make deployment and maintenances very easy and reliable!maintenances very easy and reliable!

Proprietary filing sent to USPTO.Proprietary filing sent to USPTO.

Page 13: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Proof of concept bench top demoWe designed, implemented and tested electronic circuits for open path FM absorption spectroscopy (TTFM Modulation board, TTFM detection board)

DetectorTunable laserVelocity 6300A/D board

GPIB board

CO2 Cell

Block diagram of the laboratory setup for demonstration of the viability of the proposed detection method.

10MHz Demodulation board

EOM

Two tone EOM driver2GHz+/-5MHz

Page 14: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Proof of concept bench top demo

Demonstrated 10-4 signal to noise ratio, 10x better than promised 1ppm CO2 detectable change in the air

Modulation board Detection board

Page 15: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Test of the benchtop prototype over 100 meters

Transmitter

Detector Laser Modulation & detection

Retroreflector

Page 16: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Prototype for field test over 100 meters

CO2

Page 17: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Building a field test ready unit

FM spectrometer and DC power supply are mounted in waterproof enclosures and ready for field tests

Page 18: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Focuses over the next 2 quarters

Test the instrument with EDFADevelop aligning procedure for long distance deploymentConduct field tests over a several days time and collect preliminary resultsCharacterize the long range capability of this instrument

Page 19: Development of low cost open-path instrument for ... · PDF fileopen-path instrument for monitoring of CO 2 at ... RF generator ω Sample Photo- ... Retroreflector. Prototype for field

Acknowledgement

Funding from DOE NETL and PEER CaltechJosé D. Figueroa, DOE NETLPatrick J. Shuler, PEER Caltech

http://www.peer.caltech.eduTel: 626-858-5077, Fax: 626-858-9250