Film-Screen Radiography - DD v7. . Film-Screen Radiography - DD v9. · PDF file Film...
date post
11-May-2020Category
Documents
view
10download
0
Embed Size (px)
Transcript of Film-Screen Radiography - DD v7. . Film-Screen Radiography - DD v9. · PDF file Film...
1
FilmFilm--Screen RadiographyScreen Radiography
David Dubowitz MD PhDDavid Dubowitz MD PhD
Transmission, Projection Imaging Image based on:
•Transmission imaging (c.f. emission, reflection)
•X-ray attenuated as it passes through patient ∝ e-μx
•Negative image recorded on film •Optical density (darkness) ∝ log [transmission of radiation]
•Projection of 3D anatomy onto 2D image
•Divergent beam, so image is magnified ∝ SID/SOD
•Irradiation ∝ 1/(distance)2
Focal spot size Large focal spot:
•Edges of projection indistinct
•Exaggerated by magnification
•Results in geometric blurring
•Reduce geometric blurring with:
•Reduced focal spot size
•AND
•Reduced magnification
“r-squared-rule”
X-ray Exposure ∝ 1/r2
Mrs. Röntgen’s Hand
X-ray + photographic film
2
Amplification with film+screen
X-ray + photographic film + intensifying screen
Film-screen geometry Screen thickness
Thicker screens are more efficient….but….more blurred
i.e. % x-ray photons absorbed by screen
~15% conversion efficiency
~5% conversion efficiency
i.e. % x-ray photons becoming light photons
Film/Screen Efficiency • Absorption Efficiency of the Screen
– Quantum Detection Efficiency (QDE): Fraction of incident x-rays that interact with it
• Intrinsic Conversion Efficiency of the Screen – Fraction of absorbed x-ray energy converted to light photons – CaWO4 = 5%, Gd2O2S:Tb = 15%
• Total Efficiency of Film/Screen Combination – Equals Absorption Efficiency x Conversion Efficiency – Ability to convert absorbed x-ray energy into film darkening – Also called Intensification Factor – 50 times greater than film alone
We use screens because they reduce dose even though they degrade image quality
screen characteristics
As detection increases...image definition decreases & v.v.
3
Film-screen combinations
~ 1000x more photons
Hurter Driffield (HD) Plot (or “characteristic curve”)
Note: log [exposure] Note: OD is a log measurement
Log-Log plot of optical transmission vs. x-ray exposure
“Optical Density” Measuring Optical Density
Hurter Driffield (HD) Plot
Contrast ∝ slope
Exposure > or < linear region reduces contrast
For radiographic film: Gradient ~2.5 – 3.5
Increase in sensitivity, No change in contrast
A more sensitive than B (“higher speed”)…. Lt shift
Less x-rays for same OD
Slope (contrast) unchanged
4
Increase in sensitivity, Increase in contrast
A more sensitive than B (“higher speed”)…. Lt shift
Slope (contrast) increased
Film latitude reduced (i.e. reduced dynamic range of exposures)
Photosensitivity of silver halidePhotosensitivity of silver halide
film processing & characteristic plot
higher concentration / higher temp / longer developer time more metallic Ag deposited darker film (increased density + more base fog)
Affected by:
•Concentration- •Temperature- •Time-
of developer
Film Sensitometer for Quality Film Sensitometer for Quality ControlControl
•Process film strips •Check Temperature •Check Development time •Check for artifacts •Monitor results daily to detect trends
Vital for Consistent Results
Quality ControlQuality Control
~ 90 sec total time~ 90 sec total time ~ 22 sec developer time~ 22 sec developer time Need Need consistent resultsconsistent results Needs QC!Needs QC! Needs regular cleaningNeeds regular cleaning
Automated Film ProcessorAutomated Film Processor
5
Common Processing ArtifactsCommon Processing Artifacts Noise • “Noise”: local variation in OD not due to attenuation in patient. • Random variation in # xray photons interacting with screen. • Random variation in fraction of light emitted by xray photons
interacting with screen.
• “faster” (more xray sensitive) screen – Each xray photon more efficient at blackening film. Fluctuations in xray photons more readily seen on film…… INCREASE IN NOISE
• Thicker (more xray absorbent) screen – Total number of x- ray photons detected to produce a given OD is unchanged. Random fluctuations in xray photons have same effect on film…… NO CHANGE IN NOISE
Effect of mAs on Effect of mAs on Optical DensityOptical Density
70kVp
5.5mAs
70kVp
11mAs
70kVp
22mAs
70kVp
40mAs
Double mAs, Doubles Optical No change in contrast
Effect of kVp on Effect of kVp on Optical DensityOptical Density
6
55kVp
25mAs
65kVp
25mAs
75kVp
25mAs
85kVp
25mAs
Photon Energy vs. kVp
15% increase kVp ≈ 2x mAs 15% decrease kVp ≈ ½x mAs
Contrast & Scatter
Scattered photons: •Different angle •Lower energy
Grid Introduced by Gustav Bucky 1913
•Alternating Pb/Al layers (or carbon)
•Fewer scattered photons reach film
•Increases contrast
•Increases dose (Bucky factor 3-6)
•Height:interspace = grid ratio
•Typically 6:1 – 12:1 (lower ratio is more forgiving!)
•Parallel, focused, crossed
Pelvis 10” x 12” No Grid 8:1 Grid
7
Grid Errors Air Gap Fewer scattered photons reach film
Increases contrast
But……
Reduced source – object distance increases dose
Magnifies image
kVp vs. skin dose
Reducing kVp increases contrast
But……
increases skin dose
Tissue Contrast
• Improved by: Lower kVp (less scatter) Smaller x-ray field (less scatter) Contrast media (high atomic number of iodine -> PE effect) Noise reduction (“contast” is actually contrast-to noise) Grid or air gap (reduce scattered photons reaching film ) Proper positioning (reduce superposition, grid artifacts)
Film-screen combinations Choose a film-screen combination to match
the clinical need:
E.g. • Abdo / pelvis: Sensitive film-screen,
reduced dose, some loss of detail. • Chest: Single phosphor / UV screens
inproves detail • Mammo: Single phosphor / Single emulsion
high resolution film • Dental: No screen at all
Choosing a grid X-ray machine near patient (divergent beam) • Use focused grid X-ray machine far from patient (~Parallel beam) • Use parallel grid High kvP (= more scatter) needs higher ratio grid
(90kVp use 12:1) • Remember: 80kvP 8:1 grid, 120kVp 12:1 grid Grid increases radiation dose (“Bucky Factor”) • Avoid grid if not really needed (e.g. pediatrics) Static grid lines can reduce detail • Blur grid lines with moving grid (Hollis Potter, 1920)
8
Sample Q’s 2002 G39:
In an x-ray machine with a tungsten target, increasing the kVp from 100 to 150 will increase all of the following except:
A. The total number of x-rays emitted. B. The maximum energy of the x-rays. C. The average energy of the spectrum. D. The energy of the characteristic x-rays. E. The heat units generated (for the same mAs).
Sample Q’s 2002 G52:
The output of a fluoroscopic unit is 10 mR/min at 50 cm. The output at 75 cm :
A. 15.0 B. 7.5 C. 6.6 D. 5.0 E. 4.4
Sample Q’s 2002 G57:
At x-ray energies between 40 and 100 keV,_ absorbs more energy than _ per gram.
A. fat, muscle B. muscle, bone C. iodine, bone D. fat, air E. air, muscle
Sample Q’s 2002 G73:
Which graph represents the shape of an H&D curve? .
Sample Q’s 2002 G75:
A grid improves the quality of a diagnostic x-ray primarily by attenuating _ .
A. Primary photons B. Compton scattered photons C. Compton electrons D. Photoelectrons E. Coherent scattered photons
Sample Q’s 2002 G76:
Regarding geometric unsharpness, which of the following is false?
It is: A. Inversely proportional to focal spot size. B. Directly proportional to object-film distance. C. Inversely proportional to focal spot-object distance. D. Characterized by penumbra width.
9
Sample Q’s 2002 G77:
The purpose of a screen is to:
1. Convert x-rays to light photons. 2. Reduce scatter reaching the film. 3. Reduce patient's exposure. 4. Increase radiographic resolution.
A. 1,2,3,4 B. 2 only C. 2,4 D. 1,3 E. 4 only
Sample Q’s 2002 G78:
A film of optical density (OD) 0.75 is placed over another identicalfilm.The OD of the pair is_.
A. 0.75 B. 1.0 C. 1.5 D. 1.75 E. 2.25
Sample Q’s 2002 D1:
For a 70 kVp x-ray beam, the mass attenuationcoefficientwhich varies the most from muscle tissue is:
A. Air. B. Aluminum. C. Bone. D. Fat. E. Iodine.
Sample Q’s 2002 D12:
A newly installed bucky radiographic system produces abdominal images that are of acceptable density over the spine and progressively lighter toward both lateral edges of the film. The most likely reason for this finding is improper:
A. Collimator tracking. B. Focal distance for grid. C. Grid ratio. D. kVp calibration of the system. E. Programming of the AEC system.
Sample Q’s 2002 D13:
The impression of noise in an x-