Post on 12-Dec-2015
description
School of Chemical and Biomedical Engineering
Page 1 of 27
BG3105 Biomedical Instrumentation
Tutorial Part I Solution
Lecture 1-3 (Introduction to Bio-instrumentation): Q1. Find the equivalent resistance of the following circuit?
Solution:
6 πΞ©
School of Chemical and Biomedical Engineering
Page 2 of 27
Q2. What is the output voltage (π0) of the following amplifier circuit?
Solution:
12 V.
As we know that no current flows into the OP-AMP or the positive and the negative terminal are at the same voltage (π1). The diagram below shows the flow of currents.
π1 =5π
5πΞ©= 1ππ
π1 = 2πΞ© Γ 1ππ = 2π
π2 =1π β π1
1πΞ©=
1π β 2π1πΞ©
= β1ππ
π0 = π1 β 10πΩà π2 = 2π + 10πΞ© Γ 1ππ = 12π
School of Chemical and Biomedical Engineering
Page 3 of 27
Q3. Given a resistor π with 5% accuracy, what will be the percent accuracy in 1π
? i) 2.5% ii) 5% iii) -5% iv)
15%.
Solution: ii
π = 1/π
ππ = β1π 2 ππ
πππ = β
ππ π
οΏ½πππ οΏ½ = οΏ½
ππ π οΏ½
School of Chemical and Biomedical Engineering
Page 4 of 27
Q4. A DC voltage across a resistor is measured with an accuracy of 3%. If the resistor is valued with a 5% accuracy, what is the percent accuracy in determining the power dissipated by the resistor from values of the resistor and measured voltage?
Solution:
π =π2
π = π Γ π Γ1π
οΏ½πππ οΏ½ = οΏ½
πππ οΏ½ + οΏ½
πππ οΏ½ + οΏ½
ππ π οΏ½ = 3% + 3% + 5% = 11%
The percent accuracy in power measurement will be 11%.
School of Chemical and Biomedical Engineering
Page 5 of 27
Q5. A Wheatstone bridge for measuring resistance is shown below, where π 1, π 2, and π 3 are resistors of the bridge circuit and π π is the resistor to be measured. The bridge circuit is balanced by adjusting π 3 such that the voltmeter indicates zero reading. Then the value of π π can be represented in terms of the bridge resistors.
(i) Show that the balanced condition for the bridge circuit is π 1
π 2= π 3
π π= π, where π is a
constant.
(ii) Find the sensitivity for measurement of π π near the balanced point. (iii) If the relative measurement accuracy of resistors π 1, π 2 and π 3 is 1%, what is the relative measurement accuracy of π π?
Solution: (i) π = π 2
π 1+π 2πΈ β π π
π 3+π ππΈ. When the bridge is balanced, π = 0.Thus, at balance:
π 2
π 1 + π 2=
π ππ 3 + π π
β π 2π 3 = π 1π π βπ 1π 2
=π 3π π
= π,π€βπππ π ππ ππππππππ
(ii)
βπ = οΏ½π 2
π 1 + π 2β
π π + βπ ππ 3 + π π + βπ π
οΏ½πΈ
= οΏ½1
π 1/π 2 + 1 β1 + βπ π/π π
π 3/π π + 1 + βπ π/π ποΏ½πΈ = οΏ½
11 + π β
1 + βπ π/π π1 + π + βπ π/π π
οΏ½πΈ
= οΏ½(1 + π + βπ π/π π)β (1 + π)(1 + βπ π/π π)
(1 + π)(1 + π + βπ π/π π) οΏ½πΈ
=βπβπ π/π π
(1 + π)(1 + π + βπ π/π π)πΈ
βπ β βπ
(1 + π)2βπ ππ π
πΈ,πππ ππππ π βπ π
(iii) π π = π 2π 3π 1
β βπ ππ π
= οΏ½βπ 1π 1οΏ½+ οΏ½βπ 2
π 2οΏ½+ οΏ½βπ 3
π 3οΏ½ = 3%
School of Chemical and Biomedical Engineering
Page 6 of 27
Q6. You are measuring temperature of a solution using a mercury thermometer. You have done measurements 10 times and obtained the following readings. 30.1, 30.2, 29.8, 32.1, 30.5, 29.9, 30.1, 29.5, 31.1, and 30.6. Calculate the mean and the standard deviation of the measured temperatures.
Solution:
Mean:
ποΏ½ = βπππ
= 30.1+30.2+29.8+32.1+30.5+29.9+30.1+29.5+31.1+30.610
= 30.39
Standard Deviation:
π = οΏ½β(ππβποΏ½)2
πβ1= οΏ½(30.1β30.39)2+(30.2β30.39)2+(29.8β30.39)2+β―
10β1= 0.7505
School of Chemical and Biomedical Engineering
Page 7 of 27
Q7. Below are some hypothetical results from mammogram for breast cancer screening. What is the specificity of the mammogram?
Breast cancer No breast cancer Total Mammogram positive 70 180 250 Mammogram negative 30 720 750
Total 100 900 1000 What are the sensitivity and specificity and prevalence of the disease? Solution: Sensitivity = 70/100 = 70% Specificity = 720/900 = 80% Prevalence = 100/1000 = 10%
School of Chemical and Biomedical Engineering
Page 8 of 27
Lecture 4 (Pressure, Flow, and Temperature): Q8. Consider the potentiometer measurement circuit on the right. (i) What is the expected reading of the voltmeter? (ii) If the internal resistance of the voltmeter is πΉπ½, what is the measured voltage?
Solution: (i) π½ = πΉπ
πΉπ½π
(ii) The equivalent circuit is
The reading is:
π½πππ =πΉπ//πΉπ
(πΉβ πΉπ) + πΉπ//πΉππ½π =
πΉππΉππΉπ + πΉπ
(πΉ β πΉπ) + πΉππΉππΉπ + πΉπ
π½π
=πΉππΉπ
πΉπΉπ + πΉπΉπ β πΉπππ½π
School of Chemical and Biomedical Engineering
Page 9 of 27
Q9. A capacitive pressure measurement scheme is shown below where the movable and fixed metal diaphragms form a capacitor πΆ. Assume that the common area π of the diaphragm, the dielectric constant π of the capacitor are known, and the distance π between the diaphragm is inversely proportional to the gas pressure.
(i) Show that the gas pressure π can be measured by measuring the capacitance value πΆ.
(ii) If the percent accuracy in measuring πΆ is 5%, the known value of the area π has a percent accuracy 3% and π and π have fixed values, what is the percent accuracy in measuring the pressure π?
Solution:
(i) Since the distance π is inversely proportional to the pressure π, there is a constant π such that π = π
π, πΆ = ππ
π= ππ
ππ, π = π
πππΆ
(ii) ππ = πππππΆ β ππ
ππ2ππ = ππ
πποΏ½ππποΏ½ β ππ
πποΏ½ππποΏ½ = π ππ
πβ π ππ
π
πππ
= πππβ ππ
π, οΏ½ππ
ποΏ½ = οΏ½ππ
ποΏ½ + οΏ½ππ
ποΏ½
Percent accuracy in measuring pressure π = 5%+3% = 8%
School of Chemical and Biomedical Engineering
Page 10 of 27
Q10. If blood flowing in a blood vessel with a radius of 2 mm cuts a magnetic field with flux density π΅ = 0.1 tesla, and the blood volume flow rate is 10 cm3/s, what is the electric potential that can be detected across the blood vessel? Solution:
π =πππ2
4 =πππ4π΅
π =4ππ΅ππ =
4 Γ 10 Γ 0.013 Γ 0.13.14 Γ 2 Γ 2 Γ 10β3 = 0.318 ππ
School of Chemical and Biomedical Engineering
Page 11 of 27
Q11. Which one is a piezoelectric material? i) PVDF (Polyvinylidene fluoride) ii) PZT (Lead Zirconium Titanate) iii) Quartz iv) All of the above
Solution: iv
School of Chemical and Biomedical Engineering
Page 12 of 27
Q12. Explain briefly the working principle of thermocouple temperature sensor? Solution:
When the junctions of two dissimilar metals are placed at two different temperatures, then there is potential difference between the two junctions. By measuring this potential difference and knowing one of the junction temperatures (reference junction temperature), the other junction temperature can be measured. Typically, junction is placed at a location whose temperature needs to be measure. The other junction is placed at a known temperature (typically at zero degree C). Thus by measuring the voltage between the two junctions one can measure the unknown temperature.
School of Chemical and Biomedical Engineering
Page 13 of 27
Q13. The resistance of a semiconductor thermistor is a function in the absolute temperature
in Kelvin (K) expressed as π (π) = π (π0) ππ½(π0βπ)/(ππ0) = π (π0) ποΏ½1πβ
1π0οΏ½π½
It is known that π½ = 3150 and π (300) = 1600 Ξ©. Find the sensitivity of the thermistor at π = 300 πΎ πππ π = 310 πΎ, respectively. Solution:
π (π) = π (π0) ποΏ½1πβ
1π0οΏ½π½
ππ (π)ππ
= π (π0) π οΏ½ποΏ½
1πβ
1π0οΏ½π½οΏ½
ππ= π (π0) ποΏ½
1πβ
1π0οΏ½π½π οΏ½οΏ½1
π β1π0οΏ½π½οΏ½
ππ
ππ (π)ππ
= π (π0) ποΏ½1πβ
1π0οΏ½π½ Γ π½ Γ
π οΏ½1π β
1π0οΏ½
ππ= π (π0) ποΏ½
1πβ
1π0οΏ½π½ Γ π½ Γ
π οΏ½1ποΏ½
ππ
ππ (π)ππ
= π (π0) ποΏ½1πβ
1π0οΏ½π½ Γ π½ Γ οΏ½β
1π2οΏ½ = π (π) π½ οΏ½β
1π2οΏ½
ππ (π)ππ
= βπ½π2
π (π)
π (300) = 1600 πΊ, π (310) = 1600 Γ ποΏ½1310β
1300οΏ½Γ3150 = 1140 πΊ
βπ βπ
(300) = β31503002
Γ π (300) = β56 πΊ/π
βπ βπ
(310) = β31503102
Γ π (310) = β37.4 πΊ/π
School of Chemical and Biomedical Engineering
Page 14 of 27
Lecture 5 (Electrical Safety): Q14. What are macroshock and microshock? Solution
macroshock: When a person touches a hot wire and the ground with his limbs.
Microschock: When a small amount of current directly passes through the heart.
School of Chemical and Biomedical Engineering
Page 15 of 27
Q15. Explain how the isolation transformer works?
Solution
An isolation transformer is used to transfer electrical power from an AC source to some equipment/device isolating the equipment/device and protect against electrical shock (macroshock). With an isolation transformer there is no βreturn pathβ so one could touch the βhotβ wire without getting shock.
School of Chemical and Biomedical Engineering
Page 16 of 27
Q16. Why cannot Ground-fault interrupter (GFI) be used on life support device circuitry?
Solution:
GFI stops the power supply when there is a fault. This may lead to death of the patient on life support device.
School of Chemical and Biomedical Engineering
Page 17 of 27
Q17. Explain how a line isolation monitor (LIM) can be used to detect a neutral line shorted to the ground.
Solution
To detect fault of the neutral line the Line isolation monitor (LIM) needs to be connected between the βhotβ and βgroundβ lines. When there is no fault the current flowing through the LIM is very small due to the very small stray capacitance. Once the neutral get shorted to the ground, the current flowing through the LIM increases. The ammeter connected in the LIM shows the fault.
Equivalent circuit
The current flowing through the LIM is πΌπΏπΏπΏ = π0
οΏ½π 2+οΏ½ 1πποΏ½
2. Since the leakage capacitance C is
very small, the current is also very small.
When the neutral gets shorted to the ground the equivalent circuit becomes,
Now the current flowing through the LIM is πΌπΏπΏπΏ = π0π
.
So the LIM is set in such a way that whenever there is a current flow higher than the set limit, it will set an alarm, which indicates that there is a fault in the neutral wire.
School of Chemical and Biomedical Engineering
Page 18 of 27
Q18. Calculate the maximal safe capacitance between a liquid-filled catheter and dc-isolated pressure sensor leads for a 230 volts, 50 Hz fault in the sensor leads if its equivalent circuit is as shown in Figure below.
Solution:
As this is a microshock hazard, so the maximal safe current is 10 Β΅A.
From the circuit, we have π = 50 πΞ© + 300 Ξ© + 1πππ
Current = π|π| = 230|π| = 10 ππ
|π| =23010β5
= 2.3 Γ 107Ξ©
As both resistances are very small, |π| β οΏ½ 1πππ
οΏ½ = 1ππ
= 2.3 Γ 107
Then, πΆ = 12ππΓ2.3Γ107
= 138 ππ
This shows that the capacitance shouldnβt be more than 138 pF for safety.
School of Chemical and Biomedical Engineering
Page 19 of 27
Q19. A receptacle tester has the circuit shown below. Indicate which LEDs will be ON or OFF for fault conditions given in Table.
Conditions 1 2 3 Hot open Neutral open H and N reversed, N open Ground open Hot and ground reversed
Solution:
Conditions 1 2 3 Hot open OFF OFF OFF Neutral open ON OFF OFF H and N reversed, N open OFF ON OFF Ground open OFF OFF ON Hot and ground reversed ON ON OFF
School of Chemical and Biomedical Engineering
Page 20 of 27
Lecture 6 (Lung, Spirometer): Q20. What is internal and external respiration? Solution: Internal respiration is the exchange of gases between the bloodstream and nearby cells. External respiration is the exchange of gases between the lungs and the bloodstream.
School of Chemical and Biomedical Engineering
Page 21 of 27
Q21. Below are three different volumes related to human lung. What are i) the definitions and ii) typical values for an adult lung? Tidal volume, Functional Residual Capacity (FRC), Dead Space. Solution: Tidal Volume: volume of gas inspired or expired during each normal respiratory cycle. 500 ml. FRC: Amount of gas remaining in the lungs at the resting expiration. 2400 ml. Dead Space: Volume of air that is not available for gas exchange. 150 ml.
School of Chemical and Biomedical Engineering
Page 22 of 27
Q22. How does spirometer work? Solution: A spirometer is used to measure lung volume. A simple spirometer can only measure the volume inspired or expired, i.e., change in volume. A spirometer has a mouthpiece. One can use the mouthpiece to breathe. The volume of the gases inside the spirometer will change as the patient breathes. This volume change is proportional to the lung volume change. There is a mechanical linkage between the volume change of the spirometer and the measurement reading output (display). A simple spirometer cannot be used to measure gas volume which remains inside the lung such as, FRC, RV etc.
School of Chemical and Biomedical Engineering
Page 23 of 27
Q23. a) Figure below shows a spirometer. In its simplest form, the bellow is mechanically articulated to a biased potentiometer such that the wiper arm voltage is proportional to volume. Derive the relationship of the voltage and volume. b) If the spirometer has a radius of 10 cm, a maximum height of 10 cm, and ππ΅π΅ = 10 π£ππ ππ, plot the air volume versus the output voltage.
Solution: (a) π is the maximal distance change that the bellows can change. So the maximum volume change is π½πππ = π πππ. Since the spirometer volume (π½π) is proportional to the output voltage (π½πππ), π½π = ππ½πππ If we adjust linkage such that at π½πππ the potentiometer output has π½πππ = π½π©π©.
π½πππ = ππ½π©π© Then the constant, π = π½πππ
π½π©π©
As the changes in distance of bellows and output of potentiometer are proportional. We have
π½π =π½ππππ½π©π©
π½πππ
(b) As π = ππ ππ,π = ππ ππ,πππ π½π©π© = ππ πππππ, we have
π =π ππππ½π©π©
=π Γ (ππ ππ)π Γ (ππ ππ)
(ππ πππππ) = π.πππ πππππ/ππππ
So, π½π = π.ππππ½πππ
School of Chemical and Biomedical Engineering
Page 24 of 27
Q24. Briefly explain nitrogen washout method for estimation of lung residual volume (RV)? Solution Nitrogen washout method for estimating lung residual volume is using a modified spirometer setup and a nitrogen analyzer. Here, the patient starts inhaling pure oxygen through a one way tube. Then he exhales air into the spirometer through another one way tube. When exhaling he expels a mixture of oxygen, nitrogen and carbon-dioxide into the spirometer. So after many breathing cycle the nitrogen content in the lung and in the spirometer will come to equilibrium. By doing a mass balance of nitrogen before and after the experiment, one can calculate the RV. Provided the person started the breathing cycle at the moment when the lung volume was at RV. Lets say at the beginning (time π1, and the lung volume is ππ π), the total number of π2 moles in the lung is given by,
ππΏπΏ2(π1)ππ πππΏ
ππ
At the beginning the total number of π2 moles in the spirometer is NIL (0). At the end (time π2, and the lung volume is ππ π), the total number of π2 moles in the lung is,
ππΏπΏ2(π2)ππ πππΏ
ππ
At the end (time π2, and the spirometer volume is ππ (π2), the total number of π2 moles in the spirometer is,
ππ πΏ2(π2)ππ (π2)ππ
ππ
By doing a mass balance (since no π2 has gone out of the lungs other than spirometer),
ππΏπΏ2(π1)ππ πππΏ
ππ β ππΏπΏ2(π2)
ππ πππΏ
ππ
= ππ πΏ2(π2)ππ (π2)ππ
ππ
ππ π =ππΏππ οΏ½
ππ πΏ2(π2)ππ (π2)ππΏπΏ2(π1) β ππΏπΏ2(π2)οΏ½
School of Chemical and Biomedical Engineering
Page 25 of 27
Lecture 7 (Heart, pacemaker, oximeter, Defibrillator): Q25. When light passes through an ink solution in water the light intensity decreases by 6.02 dB. You are asked to take this solution and modify it only using water, so that when light passes through the new solution its intensity gets attenuated only by half. What would you do? Solution
π(π) = β10π ππ10 οΏ½πΌπΌ0οΏ½ ππ ππ΅
6.02 = β10π ππ10 οΏ½πΌπΌ0οΏ½
π ππ10 οΏ½πΌπΌ0οΏ½ = β6.02/10
πΌ0 = 3.99πΌ β 4πΌ Therefore, at present the ink solution is attenuating the light by 4 times. To make a new solution which will attenuate only half, we need to dilute the ink solution by two times. So take the ink solution, mix with equal volume of water and the new solution will have half the concentration, therefore, it will attenuate light only by half.
School of Chemical and Biomedical Engineering
Page 26 of 27
Q26. When a pacemaker is needed? Solution A pacemaker is needed when 1) the SA node ceases to function, or become unreliable, 2) The pulse from the SA node cannot reach the heart muscle due to blockage by damaged tissue.
School of Chemical and Biomedical Engineering
Page 27 of 27
Q27. What is Defibrillator used for? Solution A defibrillator is used to restore a normal cardiac rhythm from a ventricle fibrillation or atrial fibrillation.