COD LR TT
3 - 150 mg/l CODb)
Dichromate / H2SO
4
130
Lr
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.
Instrument Type Cuvette λ Measuring Range
MD 100, MD 110, MD 200, MD 600, MD 610, MD 640, MultiDirect
ø 16 mm 430 nm 3 - 150 mg/l CODb)
SpectroDirect, XD 7000, XD 7500
ø 16 mm 443 nm 3 - 150 mg/l CODb)
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
COD LR/25 25 pc. 2420720COD LR/25, mercury free 25 pc. 2420710COD LR/150 150 pc. 2420725
The following accessories are required.
Accessory Packaging Unit Part Number
Thermoreactor RD 125 1 pc. 2418940
Application List
• Raw Water Treatment • Waste Water Treatment
Notes
1. The blank is stable when stored in the dark. 2. Blanks and test vials must be from the same batch. 3. Do not place hot vials in the sample chamber. The most stable measured values
can be determined if the vials are left standing overnight.
COD LR TT / 130
NL Handboek van Methoden 11/18EN Handbook of Methods 11/18
Implementation of the provision COD LR with Vario Vial TestSelect the method on the device
Blank
Prepare two reaction vials. Mark one as a blank.
Blank
2 ml
Put 2 ml deionised water in the blank.
Sample
2 ml
Put 2 ml sample in the sample vial.
Close vial(s). Carefully invert several times to mix the contents. Note: Will get hot!
Seal the vials in the pre-heated thermoreactor for 120 minutes at 150 °C .
Remove the vial from the thermoreactor. Note: vial will be hot!)
60 °C
Allow vial(s) to cool to 60 °C. Invert several times to mix the contents.
COD LR TT / 130
NL Handboek van Methoden 11/18EN Handbook of Methods 11/18
25 ºC
15 ºC
Allow the vial to cool to room temperature and then measure.
Place blank in the sample chamber. • Pay attention to the positioning.
Zero
Press the ZERO button.
Remove vial from the sam-ple chamber.
Place sample vial in the sample chamber. • Pay attention to the positioning.
Test
Press the TEST (XD: START) button.
The result in mg/l COD appears on the display.
COD LR TT / 130
NL Handboek van Methoden 11/18EN Handbook of Methods 11/18
Chemical Method
Dichromate / H2SO4
Appendix
Interferences
Persistant Interferences
• In exceptional cases, contents, for which the oxidation capacity of the reagent is not sufficient, can lead to lower results.
Removeable Interferences
• Suspended solids in the vial can lead to incorrect measurements and so to avoid this, it is important to place the vials carefully in the sample chamber as the method necessitates a build-up of precipitate at the bottom of the vial.
• The outer walls of the vial must be clean and dry before the analysis is carried out. Fingerprints or water droplets on the vial lead to incorrect measurements.
Interference from / [mg/l]
Cl- 1000
Method Validation
Limit of Detection 3.1 mg/lLimit of Quantification 9.3 mg/lEnd of Measuring Range 150 mg/lSensitivity 0.003 mg/l / AbsConfidence Range 2 mg/lStandard Deviation 0.8 µgVariation Coefficient 1 %
Conformity
ISO 15705:2002
According to
DIN 380402 part 41
a) determination of free, combined and total | b) Reactor is necessary for COD (150 °C), TOC (120 °C) and total -chromium, - phosphate, -nitrogen, (100 °C) | c) MultiDirect: Adapter is necessary for Vacu-vials® (Order code 19 20 75) | d) Spectroquant® is a Merck KGaA Trademark | e) alternative reagent, used instead of DPD No.1/No.3 in case of turbidity in the water sample caused by high concentration of calcium and/or high conductivity | f) additionally required for determination of bromine, chlorine dioxide and ozone in the presence of chlorine | g) Reagent recovers most inso-luble iron oxides without digestion | h) additionally required for samples with hardness values above 300 mg/l CaCO3 | i) high range by dilution | # including stirring rod, 10 cm
COD LR TT / 130
NL Handboek van Methoden 11/18EN Handbook of Methods 11/18
Top Related