(A2LA Cert. No. 3144.01) 04/07/2017 Page 1 of 3
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SPIRENT COMMUNICATIONS EATONTOWN SERVICE CENTER 541 Industrial Way West
Eatontown, NJ 07724 Brad Ball Phone: 732 578 2335
CALIBRATION
Valid until: May 31, 2019 Certificate Number: 3144.01 In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this laboratory to perform the following calibrations1: I. Electrical – RF/Microwave
Parameter/Range
Frequency CMC2 ()
Comments
Absolute Power – Measure
(+23 to +10) dBm (+10 to -70) dBm (-10 to -50) dBm (-50 to -111) dBm (-10 to -111) dBm
10 MHz to 6 GHz 10 MHz to 6 GHz (0.35 to 3.6) GHz (0.35 to 3.6) GHz (3.5 to 6) GHz
0.26 dB 0.19 dB 0.65 dB 0.66 dB 1.8 dB
Agilent E4418B power meter & Agilent E4412A power sensor, includes M = 1.5 (type-N, 50 Ω) Agilent signal analyzer MXA N9020A-508, includes M = 1.5 (type-N, 50 Ω)
Relative Power – Measure
(0 to -80) dBm
(0.35 to 3.6) GHz (3.5 to 6) GHz
0.13 dB 0.15 dB
Agilent signal analyzer MXA N9020A-508 (type-N, 50 Ω)
(A2LA Cert. No. 3144.01) 04/07/2017 Page 2 of 3
Parameter/Range
Frequency CMC2 ()
Comments
Absolute Power – Generate
(+13 to -60) dBm
(+7 to -110) dBm
(0.35 to 3) GHz (3 to 4) GHz (4 to 6) GHz 250 kHz to 2 GHz (2 to 3) GHz
0.72 dBm 0.83 dBm 0.94 dBm 0.96 dBm 1.1 dBm
Agilent signal generator MXG N5182A (type-N, 50 Ω) Agilent signal generator, 4438C ESG with options 005, 400, 402, 503 (type-N, 50 Ω)
Insertion Loss –
(0 to -50) dBm
(0.5 to 2) GHz (2 to 8) GHz (8 to 20) GHz (0.01 to 3) GHz (3 to 6) GHz
0.066 dB 0.12 dB 0.21 dB 0.21 dB 0.24 dB
Agilent network analyzer 8720D with 85052B calibration kit (type-N, 50 Ω) Agilent network analyzer 5071C with calibration kit 85092C (type-N, 50 Ω)
Return Loss –
(0 to -98) dBm
(0.01 to 3) GHz (3 to 6) GHz
0.041 dB 0.036 dB
Agilent network analyzer 5071C with calibration kit 85092C (type-N, 50 Ω)
VSWR – Measure, (0 to 2.0) VSWR
(0 to -5) dBm (-5 to -30) dBm
10 MHz to 8.5 GHz
0.043 VSWR 0.037 VSWR
Agilent network analyzer 5071C with calibration kit 85092C (type-N, 50 Ω)
(A2LA Cert. No. 3144.01) 04/07/2017 Page 3 of 3
Parameter/Range
Frequency CMC2 ()
Comments
EVM (% RMS) (Error Vector Magnitude) – Measure
EVM Span ≤ 8 MHz
0.72 % RMS
Agilent signal analyzer MXA N9020A-508 (type-N, 50 Ω) RMS means the value of the difference between a collection of measured symbols and ideal symbols
II. Time & Frequency
Parameter/Equipment
Range CMC2 ()
Comments
Time Base Measurement
10 MHz 1.4 parts in 109
Hz/Hz Rubidium oscillator & Agilent 53131A counter
_______________________________ 1 This laboratory offers commercial calibration service. 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of
measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.
Accredited Laboratory
A2LA has accredited
SPIRENT COMMUNICATIONS EATONTOWN SERVICE CENTER Eatontown, NJ
for technical competence in the field of
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005 General requirements for the competence of testing and calibration laboratories. This laboratory also meets R205 – Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates technical competence for a
defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated 8 January 2009).
Presented this 7th day of April, 2017. _______________________ President and CEO For the Accreditation Council Certificate Number 3144.01 Valid to May 31, 2019