Sub-µW Signal Power Doppler Radar Heart Rate Detection
Shuhei Yamada, Mingqi Chen, and Victor Lubecke
Department of Electrical Engineering, University of Hawaii at Manoa, Honolulu, Hawaii, 96822
Tel: 1-808-956-9948, E-mail: [email protected]
Abstract — Periodic motion, such as that resultingfrom cardiopulmonary activity, can be measured bydirect-conversion microwave Doppler radar. In suchsystems, motion is measured as phase modulation and measurement sensitivity is significantly affected byphase noise and amplitude modulation sidebands in the local oscillator. Presented here are experimentalresults of heart rate measurements under varioussignal output power conditions, with consideration forthe impact of respiration. Measurements were madeusing a custom compact quadrature direct conversionDoppler radar circuit, and results indicate that heartrate can be accurately assessed with transmit powerfor levels as low as 20nW for a subject distance of one meter. This is the lowest power ever reported for ISMband CW doppler radar heart beat detection.
Index Terms — Heart rate monitoring, biomedicalsignal detection, continuous phase modulation,Doppler radar, homodyne detection.
I. INTRODUCTION
Direct-conversion microwave Doppler-radar has
been introduced for non-contact cardiopulmonary
monitoring [1-4], with potential for health
monitoring and personnel detection. Doppler radar
systems generally transmit a continuous wave signal
while receiving and demodulating the signal’s
reflection from a target. According to Doppler
theory, when the target has time-varying movement
with zero net velocity the reflected signal is phase-
modulated in proportion to the position of the target
rather than the velocity. A stationary human body
presents two independent time varying sources of
motion with zero net velocity resulting from
respiration and cardiac activity, and the largest
reflection of incident RF power occurs at the body
surface. According to range correlation theory,
performance limits in Doppler radar sensing depend
on the phase noise and amplitude modulation noise
sidebands of the local oscillator signal.
In this paper, heart rates were measured using a
Doppler radar system with incrementally reduced
transmit power levels in order to assess the lower
limits for system performance. Measurements were
made both for subjects holding their breath and for
subjects breathing normally, with the former used to
assess absolute performance limits for the radar
circuit, and the latter used to asses secondary
impacts resulting from the more complex chest
motion. The design and fabrication of the quadrature
direct-conversion circuit used is also presented.
Results measured with this circuit indicate that
subject heart rates can be accurately monitored for
transmit power levels as low as 20nW, for subjects
at one meter displacement. This is the lowest power
ever reported for ISM band CW doppler radar heart
beat detection.
II. DOPPLER RADAR SYSTEM
The power of the received signal in a Doppler
radar system can be affected by many parameters
including transmit power level, path obstructions,
target composition and cross section, and antenna
patterns. In this work, only transmit power was
varied with all other parameters held constant. A
custom compact quadrature direct-conversion
SignalGenerator
Signal Processing
RFout
DigitizationAmplification,DC block andAnti aliasing filters
2Way 0 degreePower Splitter
2Way 90 degreePower Splitter
2Way 0 degreePower Splitter
Mixer
Mixer
LO
RFin
Iout
Qout
LO
LO
RFin
RFin
PhaseShifter
Variable attenuator
Object
1[m]
Fig. 1. Block diagram of Doppler radar system with
low power measurement approach. A single external
signal source was used to provide the LO and RF
output signals, and an attenuator used to incrementally
decrease the transmitted signal. The optimal output
from the I or Q mixer was processed to assess the heart
beat-to-beat interval for a seated subject at 1m.
Proceedings of Asia-Pacific Microwave Conference 2006
Copyright 2006 IEICE
Doppler radar circuit was used with the subject
seated at one meter from the co-located transmit and
receive antennas, with unobstructed air space
between them. The Doppler radar system and
measurement approach are illustrated in Fig. 1. The
circuit local oscillator (LO) and transmit signals
originated form a stable external signal source, with
LO power maintained at an efficient, though not
optimal, level. Under these conditions it was
anticipated that the system could easily provide
accurate heart rate measurements with a nominal
2mW signal at the transmit antenna feed, and thus
allow an effective means of quantifying the
detrimental effects on radar performance of a
progressive reduction in transmit signal power.
Similar effects could be expected for systems
transmitting higher power signals with signal power
reduced by other factors mentioned above.
The manner in which the mixer output in a direct
conversion Doppler radar is related to the target
motion is known to be acutely dependent on the
nominal target distance [5]. The simplest case of
direct proportionality results for an “optimum”
position that occurs periodically for displacements
of 2/ , with correspondingly distorted conditions
occurring near the “null” positions in between. The
use of a system with both in-phase (I) and quarature
(Q) mixer output channels provides a means for
avoiding null case measurements either through
channel selection or appropriate combination. In this
work channel selection was used, with an effort to
maintain an optimum-case position for the subject at
all times. This approach also helped to restrict all
observed detrimental effects to those related to
transmit power level.
The quadrature direct conversion circuit used was
fabricated as a compact microstrip circuit with
surface mount components on a 3-cm by 4-cm PC
board, with coaxially connected commercial patch
antennas (Antenna Specialists ASPPT2988 2.4 GHz)
and an external Agilent 83640B signal source. The
block diagram of the quadrature direct conversion
circuit is shown in Fig. 1. The two LO signals were
derived from the external signal source. The signals
were divided first by a two-way 0° power splitter
(Mini-Circuits RPS-2-30) to separate the LO and RF
output signals. The latter signal was directed to the
transmit patch antenna, with the former signal
further divided by on-board two-way 90° power
splitters (Mini-Circuits QCN-27) to make quadrature
outputs. The radiated signal reflected by the subject
was received by the receiving patch antenna to
create the RF input signal, which was divided using
a two-way 0° power splitter to feed the two receiver
chains. Each chain had its own mixer (Mini-Circuits
SKY-42) and produced the separate output channel
signals as Iout and Qout at the corresponding SMA
connectors.
A photograph of the fabricated quadrature direct
conversion circuit is shown in Fig 2. The RF and LO
signals were routed via 50 microstrip lines and
the design frequency was 2.4 GHz. An FR4
substrate is used for the PC board with a dielectric
constant of 4.5, a substrate thickness of 0.787 mm, a
conductor thickness of 35µm, a metal conductivity
of 5.5e7, and a delta loss tangent of 0.018.
Fig. 2. Photograph of the custom compact
quadrature direct conversion Doppler radar
circuit. The circuit was connected to an
external signal source, antennas, and
power supply for measurements.
III. MEASUREMENT OF LOW POWER
OPERATION
The measurement setup for the low power
measurements is shown in Fig 1. The external signal
source was set to 11 dBm in order to get a 0 dBm
LO-input power for the mixer. While the ideal LO
power for the mixer is 7 dBm, a 0dBm signal was
used for this measurement to maintain circuit
simplicity. The conversion loss of the mixer is
5.4dB at a 7dBm LO power, and 7.26dB at a 0 dBm
LO power, as shown in Fig. 3. This implies that the
measured system limits might still be improved.
Two variable attenuators were used to
incrementally reduce the RF output power. Each
attenuator provided 0 to 30 dB attenuation in 1 dB
steps. In order to compensate for the phase change
introduced by each different attenuator settings, a
compensating phase shifter was introduced after the
signal generator. With this, the phase change
introduced with each increment of the attenuator
could be completely compensated. Another phase
shifter was also used to further optimize the phase to
0
0.4
0.8
1.2
1.6
2
0 5 10 15 20
Time[sec]
Inte
rval
[sec
]
Reference Radar
Fig. 6. Measured result at 0.2 µW, with
respiration.
0
0.4
0.8
1.2
1.6
2
0 5 10 15 20
Time[sec]
Inte
rval
[sec
]
Reference Radar
`
Fig. 4. Measured result at 2mW, without
respiration.
0
0.4
0.8
1.2
1.6
2
0 5 10 15 20
Time[sec]
Inte
rval
[sec
]
Reference Radar
Fig. 7. Measured result at 20 nW, without
respiration. Degradation in accuracy can be
observed.
0
0.4
0.8
1.2
1.6
2
0 5 10 15 20
Time[sec]
Inte
rval
[sec
]
Reference Radar
Fig. 5. Measured result at 0.2 µW, without
respiration.
0
10
20
30
40
50
-18 -15 -12 -9 -6 -3 0 3 6
LO Power[dBm]
Co
nve
rsio
n L
oss
[dB
]
Fig. 3. Measured LO power vs conversion
loss for the Mixer (Mini-Circuits SKY-42).
The mixer was supplied with a 0 dBm LO
avoid null points. The baseband output signals were
amplified and filtered with SR560 LNAs and then
digitized with a Tektronix 3014 digital oscilloscope.
A wired finger pressure pulse sensor was used as a
reference for comparison with the heart rate data
obtained with the Doppler radar. A peak detector
and digital band pass filter implemented in
MATLAB were used for signal processing for
signals under both breath-hold and normal
conditions.
When the subject breathes during a
measurement, there is some phase variation and
resulting frequency distortion, as described in [5].
To avoid this impacting rate detection accuracy, the
subject did not breathe during the initial
measurements. The result for the system operating at
2 mW is shown in Fig. 4. The plot indicates an
average beat-to-beat interval of about one second,
with expected agreement between finger-pulse
reference and radar results.
The result for 0.2 µW is shown in Fig. 5. At
this point the system still functions well, with no
discernable degradation. A second plot at this power
level is also shown (Fig. 6), also indicating close
agreement with reference measurements in spite of
the additional phase modulation effects introduced
by respiration related chest motion.
The result for 20 nW is shown in Fig. 7.
Here appreciable degradation can be observed in the
rate tracking for the Doppler radar signal. Still, the
rate is tracked with reasonable accuracy and may be
useful for many applications.
IV. CONCLUSION
Low signal power results for Doppler motion
sensor based heart rating monitoring has been
presented. A custom compact quadrature direct-
conversion Doppler radar has been described along
with its use in assessing low power monitoring
limits. Heart rates for subjects holding their breath
and subjects breathing normally have been
successfully tracked for signal power levels as low
as 20 nW. This is the lowest power ever reported for
ISM band CW doppler radar heart beat detection.
These results can be extrapolated to larger power
systems where obstructions and antenna gain
similarly impact the signal power available for heart
motion detection.
ACKNOWLEDGEMNTS
This research was sponsored by the NSF
under contract ECS0428975 and the U.S. army
under contract W15P7T-05-C-B202. The authors
would also like to thank Byung-Kwon Park, Dung
Nguyen, and Olga Boric-Libecke for their assistance.
REFERENCES
[1] V.M. Lubecke, O. Boric-Lubecke, G. Awater, P.-W.
Ong, P. Gammel, R.-H. Yan, J.C. Lin, “Remote Sensing
of Vital Signs with Telecommunications Signals,” (Invited)
presented at the World Congress on Medical Physics and
Biomedical Engineering (WC2000), Chicago, IL, USA,
July 2000.
[2] A. Droitcour, V. M. Lubecke, J. Lin, and O. Boric-
Lubecke, “A Microwave Radio for Doppler Radar Sensing
of Vital Signs,” IEEE MTT-S International Microwave
Symposium, Phoenix, AZ, USA, vol. 1, pp. 175 178, May
2001.
[3] A. D. Dritcour, O. Boric-Lubecke, V. M. Lubecke, J.
Lin and G. T. Kovacs, “Range Correlation and I/Q
Performance Benefits in Single Chip Silicon Doppler
Radars for Non-Contact Cardiopulmonary Monitoring”
IEEE Trans. on Microwave Theory and Tech., Vol. 52, No.
3, pp. 838-848, March 2004.
[4] Y. Xiao, J. Lin, O. Boric-Lubecke, V. Lubecke,
“Frequency tuning technique for remote detection of
heartbeat and respiration using low-power double-
sideband transmission in Ka-band” IEEE Transactions on
Microwave Theory and Techniques, Vol. 54, No. 5, pp.
2023-2032, May 2006.
[5] Byung-Kwon Park, Shuhei Yamada, Olga Boric-
Lubecke, and Victor Lubecke, “Single-Channel Receiver
Limitations in Doppler Radar Measurements of Periodic
Motion,” IEEE Radio and Wireless Symposium, San-
diego, CA, January 2006.