2
Light source: LED: λ = 528 nm (filter) Battery: 9 V block battery (life = approx. 600 tests) Auto-OFF: Automatic switch off 15 minutes after last keypress Ambient conditions : 5-40°C 30-90% rel. humidity (non-condensing). CE: DIN EN 55 022, 61 000-4-2, 61 000-4-8, 50 082-2, 50 081-1, DIN V ENV 50 140, 50 204 Technical data Operation CW2070 Colorimeter for DEHA User messages METHOD METHOD METHOD METHOD 0.0.0 RESULT RESULT ÷Err EOI -Err LO BAT dh DEHA 0.02-0.5 mg/l 0.0.0 RESULT RESULT dh Switch the unit on using the ON/OFF switch. The display shows the following: Fill a clean vial with the water sample up to the 10 ml mark, replace the cap tightly and place the vial in the sample chamber making sure that the -mark on the vial aligned with the ∇−mark on the instrument. Press the ZERO/TEST key. The method symbol flashes for approx. 3 seconds. The display shows the following: After zero calibration is completed, remove the vial from the sample chamber. Add the appropriate reagent tablet / solution ; a colour will develop in the sample. Replace the cap tightly and place the vial in the sample chamber with the and marks aligned. Press the ZERO/TEST key. The method symbol flashes for approx. 3 seconds. The result appears in the display. Repeating the analysis: Press the ZERO/TEST key again. New zero calibration: Press the MODE key until the desired method symbol appears in the display again. Light absorption too great. Reasons: zero calibration not carried out or, possibly, dirty optics. Measuring range exceeded or excessive turbidity. Result below the lowest limit of the measuring range. Replace 9 V battery, no further analysis possible. Perform zero calibration (see “Operation”). Hold the drip bottle vertically and add evenly sized 6 drops to the 10 ml water sample by pressing slowly. Replace the cap tightly and invert the vial gently to mix the contents. Add one DEHA tablet straight from the foil to the same sample and crush using a clean stirring rod. Mix well using the stirring rod to dissolve the tablet. Replace the cap tightly and place the vial in the sample chamber with the and marks aligned. Wait for a colour reaction time (in darkness) of 10 minutes! Press the ZERO/TEST key. The method symbol flashes for approx. 3 seconds. The result is shown in the display in mg/l DEHA. Measuring tolerance: ± 0.05 mg/l DEHA Method notes Observe application options, analysis regulations and matrix effects of methods. Reagent tablets / solutions are designed for use in chemical analysis only and should be kept well out of the reach of children. If necessary, request safety data sheets. Ensure proper disposal of reagent solutions. Note Always adhere to the sequence of reagent addition. Correct filling of the vial Factors for the below listed oxygen scavengers (multiply the displayed value with the appropriate factor) Hydroquinone 5 Erythorbic Acid 7 Methylethylketoxime 7 correct wrong 10 ml 10 ml Troubleshooting: Guidelines for photometric measurements 1. Vials, caps and stirring rods should be cleaned thoroughly after each analysis to prevent errors being carried over. Even minor reagent residues can cause errors in the test results. Use the brush provided for cleaning. 2. The outside of the vial must be clean and dry before starting the analysis. Clean the outside of the vials with a towel. Fingerprints or other marks will be removed. 3. Zero calibration and test must be carried out with the same vial as there may be slight differences in optical performance between vials. 4. The vials must be positioned in the sample chamber for zero calibration and test with the -mark on the vial aligned with the ∇−mark on the instrument. 5. Zero calibration and test must be carried out with the sample chamber lid closed. 6. Bubbles on the inside of the vial may also lead to errors. In this case, fit the vial with a clean stopper and remove bubbles by swirling the contents before starting test. 7. Avoid spillage of water or reagent solution in the sample chamber. If water should leak into the photometer housing, it can damage electronic components and cause corrosion. 8. Contamination of the windows over the light source and photo sensor in the sample chamber can result in errors. If this is suspected check the condition of the windows. 9. The reagent tablets should be added to the water sample without being handled. 10.Large temperature differentials between the photometer and the operating environment can lead to incorrect measurement due to the formation of condensate in the area of the lens or on the vial (e.g). 11.Method specified at 20 °C ± 2 °C. 12. To avoid errors caused by stray-light do not use the instrument in bright sunlight. • Replacement Reagents DEHA Tablets pk 100 Ref: TT/51.32.20 DEHA Solution 15ml Ref: TT/46.11.85 Zero Test Zero Test Zero Test Camlab CW2070 DEHA 2/5/06 3:23 pm Page 1

Transcript of CW2070 Colorimeter for DEHA - Camlab · PDF fileBattery: 9 V block battery (life = approx. 600...

Page 1: CW2070 Colorimeter for DEHA - Camlab · PDF fileBattery: 9 V block battery (life = approx. 600 ... Fill a clean vial with the water sample up to ... Technical changes may occur without

Light source: LED: λ = 528 nm (filter)

Battery: 9 V block battery (life = approx. 600 tests)

Auto-OFF: Automatic switch off 15 minutes after lastkeypress

Ambient conditions: 5-40°C30-90% rel. humidity (non-condensing).

CE: DIN EN 55 022, 61 000-4-2, 61 000-4-8,

50 082-2, 50 081-1, DIN V ENV 50 140, 50 204

Technical data

Operation

CW2070 Colorimeter for DEHA

User messages

Zero

Test

M E T H O DM E T H O D

M E T H O DM E T H O D

0.0.0

R E S U L TR E S U L T

Zero

Test

÷Err

EOI

-ErrLO BAT

ON

OFF

dh

DEHA 0.02-0.5 mg/l

0.0.0

R E S U L TR E S U L T

Zero

Test

dh

Switch the unit on using the ON/OFF switch.

The display shows the following:

Fill a clean vial with the water sample up to the 10 ml

mark, replace the cap tightly and place the vial in thesample chamber making sure that the ∆-mark on the vial

aligned with the ∇−mark on the instrument.

Press the ZERO/TEST key.

The method symbol flashes for approx. 3 seconds.

The display shows the following:

After zero calibration is completed, remove the vial fromthe sample chamber.Add the appropriate reagent tablet / solution ; a colour

will develop in the sample.Replace the cap tightly and place the vial in the sample

chamber with the ∆ and ∇ marks aligned.

Press the ZERO/TEST key.

The method symbol flashes for approx. 3 seconds.

The result appears in the display.

Repeating the analysis:Press the ZERO/TEST key again.

New zero calibration:Press the MODE key until the desired method symbolappears in the display again.

Light absorption too great. Reasons: zero calibration notcarried out or, possibly, dirty optics.

Measuring range exceeded or excessive turbidity.

Result below the lowest limit of the measuring range.

Replace 9 V battery, no further analysis possible.

Perform zero calibration (see “Operation”).

Hold the drip bottle vertically and add evenly sized 6 dropsto the 10 ml water sample by pressing slowly. Replacethe cap tightly and invert the vial gently to mix the contents.

Add one DEHA tablet straight from the foil to the same

sample and crush using a clean stirring rod.

Mix well using the stirring rod to dissolve the tablet.

Replace the cap tightly and place the vial in the samplechamber with the ∆ and ∇ marks aligned.

Wait for a colour reaction time (in darkness) of 10 minutes!

Press the ZERO/TEST key.

The method symbol flashes for approx. 3 seconds.

The result is shown in the display in mg/l DEHA.

Measuring tolerance: ± 0.05 mg/l DEHA

Method notes

Observe application options, analysis regulations and matrix effects ofmethods. Reagent tablets / solutions are designed for use in chemical

analysis only and should be kept well out of the reach of children.

If necessary, request safety data sheets.

Ensure proper disposal of reagent solutions.

Note

Always adhere to the sequence of reagent addition.

Correct filling of the vial

Factors for the below listed oxygen scavengers(multiply the displayed value with the appropriate factor)

Hydroquinone 5

Erythorbic Acid 7

Methylethylketoxime 7

correct wrong

10 ml 10 ml

M E T H O D

M E T H O D

R E S U L T

Troubleshooting: Guidelines for photometric measurements

1. Vials, caps and stirring rods should be cleaned thoroughly after eachanalysis to prevent errors being carried over. Even minor reagentresidues can cause errors in the test results. Use the brush provided forcleaning.

2. The outside of the vial must be clean and dry before starting the analysis.

Clean the outside of the vials with a towel. Fingerprints or other markswill be removed.

3. Zero calibration and test must be carried out with the same vial as theremay be slight differences in optical performance between vials.

4. The vials must be positioned in the sample chamber for zero calibrationand test with the ∆-mark on the vial aligned with the ∇−mark on theinstrument.

5. Zero calibration and test must be carried out with the sample chamberlid closed.

6. Bubbles on the inside of the vial may also lead to errors. In this case, fitthe vial with a clean stopper and remove bubbles by swirling the contents

before starting test.

7. Avoid spillage of water or reagent solution in the sample chamber. If

water should leak into the photometer housing, it can damage electroniccomponents and cause corrosion.

8. Contamination of the windows over the light source and photo sensorin the sample chamber can result in errors. If this is suspected checkthe condition of the windows.

9. The reagent tablets should be added to the water sample without beinghandled.

10.Large temperature differentials between the photometer and theoperating environment can lead to incorrect measurement due to the

formation of condensate in the area of the lens or on the vial (e.g).

11.Method specified at 20 °C ± 2 °C.

12. To avoid errors caused by stray-light do not use the instrument in brightsunlight.

R E S U L T

• Replacement ReagentsDEHA Tablets pk 100Ref: TT/51.32.20DEHA Solution 15mlRef: TT/46.11.85

ZeroTest

ZeroTest

ZeroTest

Camlab CW2070 DEHA 2/5/06 3:23 pm Page 1

Page 2: CW2070 Colorimeter for DEHA - Camlab · PDF fileBattery: 9 V block battery (life = approx. 600 ... Fill a clean vial with the water sample up to ... Technical changes may occur without

User calibration : cALManufacturing calibration : CAL

Zero

TestMode

On

Off

CAL

SEL

SEL

CAL

Mode

On

Off

cAL

SEL

User notes

E 10

E 70

E 71

Calibration mode

On

Off

Mode

CAL

dh

CAL

R E S U L T + xR E S U L T + x

Mode

Zero

Test

0.0.0

Zero

Test

R E S U L TR E S U L T

CAL

Zero

Test

On

Off

: :

Note

CAL

cAL

M E T H O DM E T H O D

M E T H O DM E T H O D

Recommended calibration value

DEHA: between 0.2 and 0.4 mg/l DEHA

CAL

Technical changes without notice

Printed in Germany 03/01

Calibration factor “out of range”

Manufactoring calibration incorrect / erase

User calibration incorrect / erase

Press MODE key and keep it depressed .

Switch unit on using ON/OFF key.Release MODE key after approx. 1 second.

The display shows the following in alternating mode:

Perform zero calibration (see "Operation").Press the ZERO/TEST key.

The method symbol flashes for approx. 3 seconds.

The display shows the following in alternating mode:

Place the calibration standard to be used in the samplechamber with the ∆ and ∇ marks aligned.Press the ZERO/TEST key.

The method symbol flashes for approx. 3 seconds.

The result is shown in the display, alternating with CAL.

If the result displayed corresponds with the value of thecalibration standard (within the tolerance quoted), exitcalibration mode by pressing the ON/OFF key.

Otherwise, pressing the MODE key once increases the

displayed value by 1 digit.Pressing the ZERO/TEST key once decreases thedisplayed value by 1 digit.

Pressing the relevant key until the displayed value equals

the value of the calibration standard.

By pressing the ON/OFF key, the new correction factoris calculated and stored in the user calibration software.

Confirmation of calibration (3 seconds).

Factory calibration active.

Calibration has been set by the user.

To reset the calibration to the factory setting:

Press both the MODE and ZERO/TEST and keep themdepressed .

Switch the unit on using the ON/OFF key. Release theMODE and ZERO/TEST keys after approx. 1 second.

The following messages will appear in turn on the display:

The calibration is reset to the factory setting.(SEL stands for Select)

or:

Calibration has been set by the user. (If the user calibrationis to be retained, switch the unit off using the ON/OFFkey.)

Calibration is reset to the factory setting by pressing theMODE key. The following messages will appear in turnon the display:

Switch the unit off using the ON/OFF key.

Camlab Water CW2070DEHA Colorimeter

Reference CW/20.97.00

Camlab Limited , Camlab House, Norman Way, OverCambridge CB4 5WE

Tel: 01954 233110 • Fax: 01954 233101www.camlab.co.uk

camlabChoice, Service and Trust

Technical changes may occur without notice

Mode

ZeroTest

ZeroTest

ZeroTest

Mode

Mode

ZeroTestMode

Camlab CW2070 DEHA 2/5/06 3:23 pm Page 2