CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL AGGREGATES

Transcript of CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely...

Page 1: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL

CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL

AGGREGATES

Page 2: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL

WHY?

[1]

Page 3: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL

THE SOURCE

Spherical Liquid Water Droplet Aerosol (m=1.33)

Page 4: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL

BACKSCATTER SETUP

Laser (λ= 633nm)

CCD Detector

Aerosol Target

Beam splitter

Beam Dump

Page 5: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL

Angle (˚)

Inte

nsity

(arb

itrar

y un

its) RESULTS

Page 6: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL

Mie ripples, but polydispersity wipes them out

Guinier regime: θ~λ/(πd)

SIZING: FORWARD SCATTERING

Laser (λ= 633nm)

CCD Detector

Polarizers Lens

Iris

Aerosol Target

Mirror

Page 7: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL
Page 8: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL

RESULTS

Page 9: CALIBRATING FOR LIGHT BACKSCATTER FROM FRACTAL …DLS suggests mid-range size: ~2.5-4µm Very likely polydisperse. Backscatter apparatus likely works, but needs more testing. FINAL

Backscatter suggests smaller size: ~1.5µm

Forward scattering suggests bigger size: ~6µm

DLS suggests mid-range size: ~2.5-4µm

Very likely polydisperse.

Backscatter apparatus likely works, but needs more testing.

FINAL RESULTS