Impulse 6000D/7000DPDefibrillator/External Pacer Analyzer
Key features
• Impulse7010DefibrillatorSelectableLoadAccessoryprovidesmultipleloadsof25Ω,50Ω,75Ω,100Ω,125Ω,150Ω,175Ω,and200Ω tocomplywithIEC60601-2-4standard(optional)
• Lown,Edmark,trapezoidal,biphasicandpulsed-biphasicdefibrillationtechnologycompatibility• AEDtechnologycompatibility• First-classmeasurementaccuracy±1%ofreading+0.1J• Intuitiveuserinterfaceandbacklight,easy-to-readdisplay• Portable,rugged,easytocarry• Long-lasting,rechargeablebattery• Pacerbrandselections• Pacerinputprotectedagainstdefibrillatoroutput(7000DPonly)• 10independentECGoutputsthatprovide12leadcombinationsforstandardizedclinicalsignals• Flexibleheart-ratesettings(1BPMstep)facilitateratemeteraccuracyandalarmtesting• DSP-basedmeasurementsenablefuturefirmwareandwaveformsupgrade• Uniqueintegratedpostsforsecureconnections• Two-yearextendedwarranty(no-costextendedwarrantyavailableafterfirst-yearcalibrationatanyFluke
Biomedicalauthorizedservicecenter)• OptionalAnsurtestautomationsoftwaretostandardizetestingprocedures,capturewaveformsandtest
results,andprintanddocumenttestresults• Designed,tested,andbuilttoincomparableFlukequalitystandards
Technical Data
TheImpulse6000DDefibrillatorAnalyzerandImpulse7000DPDefibrillator/TranscutaneousPacerAnalyzerTestSystemsarerugged,portablepreci-siontestinstrumentsthatensureproperoperationandultimateperformanceofcriticallife-supportcardiac-resuscitationequipment.
TheImpulse6000DandImpulse7000DPtestcapabilitiesencompassthespectrumofworldwide-establishedpulseshapes,showcasebreakthroughAEDtechnologycompatibility,andoutperforminaccuracyandstandards.Additionally,theImpulse7000DPincorporatesthetestsandtheextensiverangeoftestloadsandmeasurementalgorithmsneededtotestexternaltranscutaneouspacemakers.
InconjunctionwithanImpulse7000DP,theImpulse7010DefibrillatorSelectableLoadAccessoryprovidesmultipleloadsof25Ω,50Ω,75Ω,100Ω,125Ω,150Ω,175Ω,and200Ωfordefibrillatorperformancetesting.AstandardUSBinterfaceenablescomputercontrolanddatatransfer,andoptionalAnsurPC-basedautomationsoftwareincreasesproductivitybyoutfittinguserswithaneasy-to-usemethodtostandardizetestingproceduresandcapture,printanddocumentdata.
www.elso.sk
riešenia na presné meranie
2FlukeBiomedical Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzer
General specificationsOperating temperature10°Cto40°C(50°Fto104°F)
Storage temperature-20°Cto60°C(-4°Fto140°F)
Humidity10%to90%non-condensing
DisplayLCDdisplay
CommunicationsUSBdeviceportforcomputercontrol
Modes of operationManualandremote
PowerInternalrechargeableNiMHbatterypackforninehours(typical)operationafterfullchargeorthebatterychargercanoperatetheanalyzerandchargethebatterysimultaneously
Battery charger100Vto240Vinput,15V/1.5Aoutput.Forbestperformance,thebatterychargershouldbeconnectedtoaproperlygroundedacreceptacle
EnclosureABSplastichousing
Dimensions(WxDxH)32cmx24cmx13cm(13inx9.5inx5in)
Weight3.02kg(6.6lb,0.1oz)
Safety standardsCE:IEC/EN61010-12ndEdition;Pollutiondegree2CAN/CSA-C22.2No61010-1;UL61010-1C-Tick:AustralianEMC
Electromagnetic compatibility standards (EMC)EuropeanEMC:EN61326-1
Defibrillator analyzer technical specificationsEnergy output measurementCompatible defibrillator waveshapesLown,Edmark,trapezoidal,dcbiphasic,andacpulsedbiphasic
Note:ACpulsedbiphasicwaveformhasnotbeenapprovedintheUnitedStates.
Autoranged measurement0.1Jto600J
Accuracy0.1Jto360J:±(1%ofreading+0.1J)360Jto600J:±(1%ofreading+0.1J),typical
Note:Forpulsedbiphasicdefibrillator,specifiedaccuracyis±(1.5%ofreading+0.3J)onbothranges.
Load resistanceResistance:50ΩAccuracy:±1%,non-inductive(<2µH)
Pulse trigger level20V
Pulse widthRange:1msto50msAccuracy:±0.1ms
VoltageRange:20Vto5000VAccuracy:±(1%ofreading+2V)
CurrentRange:0.4Ato100AAccuracy:±(1%ofreading+0.1A)
Tilt (biphasicandpulsedbiphasic)Range:1%to99%Accuracy:±1digit
Interphase delay (biphasicandpulsedbiphasic)Range:0.1msto9.9msAccuracy:±0.1ms
Frequency (pulsedbiphasiconly)Range:2000Hzto8000HzAccuracy:±1%ofreading
Duty cycle(pulsedbiphasiconly)Range:1%to99%Accuracy:±1digit
Sample rate250kHz(4µssample)
Maximum average power12W,equivalentto10defibpulsesof360Jevery5minutes
Scope outputAutorange:2000:1,400:1,and80:1dependingonrange
Waveform playback• Output:BNC• Outputimpedance:50Ω• Amplitudeaccuracy:±5%
Charge time measurementRange:0.1sto100sAccuracy:±0.05s,typical
Synchronization test (elective cardioversion)Delaytimemeasurement• Timingwindow:ECGR-wave
peaktothedefibpulsepeak• Range:-120msto380ms;
measurestimingfrom120mspriortotheR-wavepeaktoupto380msfollowingtheR-wavepeak
• Resolution:1ms• Accuracy:±1ms
www.elso.sk
riešenia na presné meranie
Normalsinuswaves:Lead#Ref.amp.I II III V1 V2 V3 V4 V5 V6
Lead I referencePerformancewavesandRwavedetection:Lead#Ref.amp.IIIIIIV1V2V3V4V5V6
Normalsinuswaves:Lead#Ref.amp.IIIIIIV1V2V3V4V5V6
ECG normal sinusRates:10BPMto360BPMin1BPMsteps
ECG high level output (BNC jack)Amplitude:• Range:0.5VpermVof
referenceleadsetting• Accuracy±5%Outputimpedance:50Ω
ECG on defibrillator input loadSameastheLeadIIamplitudebutlimitedto±4mV
ECG performance wavesSquarewave:2Hzand0.125HzTriangularwave:2Hzand2.5HzSinewaves:0.05,0.5,5,10,40,50,60,100,150,and200HzPulse:30BPMand60BPM,60mspulsewidth
R-wave detectionWaveform:Haver-triangle
Amplitude:0.05mVto0.45mVin0.05mVstepsand0.5mVto5mVin0.5mVstepsRate:30,60,80,120,200,and250BPMWidths:8,10,12ms,and20msto200msin10msstepsAccuracy:±(1%setting+1ms)
Noise immunityWave:SineLinefrequency:50Hzor60Hz(±0.5Hz)Amplitude:•Range:0.0mVto10mVin
0.5mVsteps•Accuracy:±5%
Transvenous pacer pulse simulationWidths• Range:0.1ms,0.2ms,0.5ms,
1ms,and2ms• Accuracy:±5%ofsettingAmplitudes:• Range:0(off)and±2mV,
±4mV,±6mV,±8mV,±10mV,±12mV,±14mV,±16mV,±18mV,±20mV,±50mV,±100mV,±200,±500,and±700mV
• Accuracy:±(10%setting+0.2mV)
Amplitude of transvenous pacer pulse simulation signals relative to amplitude setting (in percent)
Lead II referenceLead#Ref.amp.IIIIIIV1V2V3V4V5V6
Lead I referenceLead#Ref.amp.IIIIIIV1V2V3V4V5V6
ECGwaves• Normalsinusrhythm(NSR):
10BPMto180BPMin1BPMsteps
• Atrialfibrillation:Coarseandfine
• Monomorphicventriculartachycardia:120BPMto240BPMin5BPMsteps
• Asystole:Flatline
Automated defibrillator test ECG wavesNormalsinus:10BPMto300BPMin1BPMstepsVentricularfibrillation:CoarseandfineMonomorphicventriculartachycardia:120BPMto300BPMin5BPMstepsPolymorphicventriculartachycardia:5typesAsystole:Flatline
ECG wavesECG generalLeadconfiguration:12-leadsimulation;RA,LL,LA,RL,V1-6withindependentoutputsLeadtoleadimpedance:1000Ω(nominal)Rateaccuracy:±1%nominal
ECG amplitudesReferencelead:Selectable,LeadII(default)orLeadISettings:0.05mVto0.45mVby0.05mVstepsand0.5mVto5mVby0.5mVstepsAccuracy(allperformancewavesandnormalsinusRwaves):• LeadII....................................±2%• Allotherleads.....................±5%• Defibpaddles.......................±5%
Amplitude of ECG signals relative to amplitude setting (in percent)
Lead II referencePerformancewavesandRwavedetection:Lead# Ref.amp.I II III V1 V2 V3 V4 V5 V6
70%100%30%
100%100%100%100%100%100%
70%100%30%24%48%
100%120%112%80%
100%150%50%
100%100%100%100%100%100%
100%150%50%24%48%
100%120%112%80%
67%100%33%67%67%67%67%67%67%
100%150%50%
100%100%100%100%100%100%
Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzerFlukeBiomedical3
www.elso.sk
riešenia na presné meranie
4FlukeBiomedical Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzer
Arrhythmia selectionsPacerinteractive(7000DPonly)• Demand:30BPMto360BPMin
1BPMsteps• Asynchronous• Non-capture• Non-function• Threshold(interactivepacing
simulationonly):10mAto250mAin10mAsteps
Supraventricular• Artrialfibrillationcourse• Atrialfibrillationfine• Atrialflutter• Sinusarrhythmia• Missedbeat• Atrialtachycardia• Paroxysmalatrialtachycardia
(PAT)• Nodalrhythm• SupraventriculartachycardiaPremature• AtrialPAC• NodalPNC• PVC1leftventricle• PVC1LVearly• PVC1LVRonT• PVC2rightventricle• PVC2RVearly• PVC2RVRonT• MultifocalPVCsVentricular• PVCs6/min• PVCs12/min• PVCs24/min• Freqmultifocal• Trigeminy• Bigeminy• PairPVCs• Run5PVCs• Run11PVCs• Monomorphicventricular
tachycardia:120BPMto300BPMin5BPMsteps
• Polymorphicventriculartachycardia:1to5
• Ventricularfibrillation:coarseandfine
• Asystole
Conduction• 1°Block• 2°BlockTypeI• 2°BlockTypeII• 3°Block• RightbundlebranchblockRBBB• LeftbundlebranchblockLBBBTransvenousPacedwithselectablepacerspikeamplitudesandwidths• Atrial80BPM• Async75BPM• Demandwithfrequentsinus
beats• Demandwithoccasionalsinus
beats• AVsequential• Non-capture• Non-function
Selections for all waves in group
Atrial pacer pulseWidth:0.1,0.2,0.5,1,2msPolarity:+or-Amplitude:0(off),2to20(by2),50,100,200,500,700mV
Ventricular pacer pulseWidth:0.1,0.2,0.5,1,2msPolarity:+or-Amplitude:0(off),2to20(by2),50,100,200,500,700mV
R-wave detectionRate:30,60,80,120,200,250BPMWidth:8,10,12,20to200(by10)msAmplitude:0.05to0.45(by0.05),0.5to5(by0.5)mV
www.elso.sk
riešenia na presné meranie
Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzerFlukeBiomedical5
Transcutaneous pacemaker analyzer technical specifications(7000DP only)
Test load Selections
Defibrillator inputFixedload:50ΩAccuracy:±1%,non-inductive(<2µH)Powerrating:10defibpulsesof360Jevery5minutes
Pacemaker inputVariableload:50Ωto1500Ωin50Ω stepsAccuracy:±2%,non-inductive(<2µH)Powerrating:5Ω(average),40Ω(peak)@1000Ω
MeasurementsManufacturer specific algorithms• GEResponder(1500and1700)• MDE300(MedicalData
Electronics)• MedtronicERS/PhysioControl
LIFEPAK• MRL(MedicalResearch
Laboratory/WelchAllyn)• Philips/Agilent/HP• SchillerMedical• ZOLLMedical(plusageneralpurposeDefaultAlgorithmselection)
CurrentRange:4mAto250mAAccuracy:±1%ofreading+0.02mA
Pulse rateRange:5PPMto800PPMAccuracy:±0.5%ofreading+0.1PPM
Pulse widthRange:1msto100msAccuracy:±0.5%ofreading+0.01ms
EnergyRange:1µJto2JAccuracy:±4%ofreading+10µJ
Demand and asynchronous mode testInput pacer pulse rates30PPMto200PPM
ECG NSR waveRate:10BPMto300BPMin1BPMstepsAmplitude:1mVUnderdriverate:10BPMminimumOverdriverate:300BPMmaximum
Sensitivity testAutomatic interactive threshold detectionCompatiblepacerrates:30PPMto120PPM
ECG R waveWaveforms:Square,triangle,sineWidth:1msto19ms(by1ms),20msto95ms(by5ms),100msto300ms(by25ms)
Accuracy:±5%ofsettingAmplitude:0.05mVto0.95mV(by0.05mV),1mVto5mV(by0.5mV)Accuracy:±5%ofsetting
Refractory period testsPaced refractory period20msto500ms
Sensed refractory period15msto500ms
Accuracy±1ms
Pacer pulse rate20PPMto200PPM
ECG Waveform:TrianglewavePulsewidth:40msAmplitude:1mV
www.elso.sk
riešenia na presné meranie
6FlukeBiomedical Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzer
Impulse 7010 Defibrillator Selectable Load AccessoryGeneral specifications
Maximum voltage 5000V
Maximum continuous power 12W,equivalentto10defibpulsesof360Jevery5minutes
Inductance<2µH,@25Ω<3µH,@50Ω<4µH,@75 Ωand100Ω<5µH,@125Ω<6µH,@150Ω<7µH,@175Ω<8µH,@200Ω
TemperatureOperating:10°Cto40°C(50°Fto104°F)Storage:-20°Cto60°C(-4°Fto140°F)
Humidity10%to90%non-condensing
Dimensions (WxDxH)154mmx272mmx138.7mm(6.07inx10.71inx5.46in)
Weight (net)1.54kg(3lb6.2oz)
Safety classComplieswithEN61010-12ndEdition,ClassIIproduct
Safety and EMC marks
WarrantyTwo-yearextendedwarranty(no-costextendedwarrantyavailableafterfirst-yearcalibrationatanyFlukeBiomedicalauthorizedservicecenter)
Calibration intervalOne-year
Electrical specifications (for Load Accessory and Analyzer together)
Load settings25Ω,50Ω,75Ω,100Ω,125Ω,150Ω,175Ω,and200Ω±1%
AccuracyEnergy(allexceptpulsedbiphasic):2%ofreading+0.1Jwith25,75Ωthough200Ωloads,1%ofreading+0.1Jwith50ΩloadEnergy(pulsedbiphasic):2.5%ofreading+0.3Jwith25,75Ω though200Ωloads,1.5%ofreading+0.3Jwith50ΩloadVoltage:1%ofreading+2Vwith25Ωand50Ωloads,2%ofreading+2Vwith75Ωthrough200ΩloadsCurrent:2%ofreading+0.1Awith25Ωload,1%ofreading+0.1Awith50Ωthrough200Ωloads
N10140
www.elso.sk
riešenia na presné meranie
Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzerFlukeBiomedical7
Ordering information
Models2811928Impulse6000DDefibrillatorAnalyzer120V(US)3077031 Impulse6000DDefibrillatorAnalyzer(Schuko)3077046 Impulse6000DDefibrillatorAnalyzer(UK)3077054 Impulse6000DDefibrillatorAnalyzer(Japan)3085270 Impulse6000D DefibrillatorAnalyzer(Australia)3085281 Impulse6000DDefibrillatorAnalyzer(India)2811919 Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer120V(US)3077005 Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer(Schuko)3077010 Impulse7000DP Defibrillator/TranscutaneousPacemakerAnalyzer(UK)3077022 Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer(Japan)3085296Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer(Australia)3085308Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer(India)3326874TA-IMP7KDPImpulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation120V(US)3326888TA-IMP7KDP-01Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(Schuko)3326895TA-IMP7KDP-02Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(UK)3326901TA-IMP7KDP-03Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(Japan)3326912TA-IMP7KDP-04Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(Australia)3326920TA-IMP7KDP-05Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(India)
Standard accessories1626219USBComputerCommunicationCable3028681UserManualCD3028662Getting-StartedGuideBattery Eliminator(countryspecific)2814980CarryingCase3156262DefibPaddleContactPlates
Optional accessories3091370 AnsurImpulse6000D/7000DPPlug-In3065489 MedtronicERS/Physio-Control(FASTPATCH)(setoftwo):4mmdefibrillatoradapters3065450 KimberlyClark/R2DaroxMRL/MDE/NK:4mmdefibrillatoradapters3065438 Internaldischargepaddlecontacts(setoftwo)3065477 MedtronicERS/Physio-Control(QUIKPACE)(setoftwo):4mmpaceradapters3065527 ZollMedicalNTP/PD1400:4mmpaceradapters3065461 MedtronicERS/Physio-Control(QUIKCOMBO):4mmdefib/paceradapters3065492 Philips/Agilent/HP(CODEMASTERSeries-Round):4mmdefib/paceradapters3065509 Philips/AgilentHEARTSTARTFR2/MRX:4mmdefib/paceradapters3065511 ZollPD-2200Multi-FunctionPD-Series,M-Series,M-SeriesCCT,AEDPROandAEDPlus™defib/paceradapters3065423 GEMarquette(RESPONDER1500/1700Series)(setoftwo):4mmdefib/paceradapters3158544 Impulse 7010DefibrillatorSelectableLoadAccessory
www.elso.sk
riešenia na presné meranie
Fluke Biomedical.Better products. More choices. One company.
Fluke Biomedical6045CochranRoadCleveland,OH44139-3303U.S.A.
Fluke Biomedical EuropeScienceParkEindhoven51105692ECSon,TheNetherlands
For more information, contact us:IntheU.S.A.(800)850-4608orFax(440)349-2307InEurope/M-East/Africa+31402675435orFax+31402675436Fromothercountries+1(440)248-9300orFax+1(440)349-2307Email:sales@flukebiomedical.comWebaccess:www.flukebiomedical.com
©2008-2010FlukeBiomedical.Specificationssubjecttochangewithoutnotice.PrintedinU.S.A.3/20103393108CD-EN-N
Modification of this document is not permitted without written permission from Fluke Corporation.
About Fluke BiomedicalFluke Biomedical is the world’s leading manufacturer of quality biomedical test and simulation products. In addition, Fluke Biomedical provides the latest medical imaging and oncology quality-assurance solutions for regulatory compliance. Highly credentialed and equipped with a NVLAP Lab Code 200566-0 accredited laboratory, Fluke Biomedical also offers the best in quality and customer service for all your equipment calibration needs.
Today, biomedical personnel must meet the increasing regulatory pressures, higher quality standards, and rapid technological growth, while performing their work faster and more efficiently than ever. Fluke Biomedical provides a diverse range of software and hardware tools to meet today’s challenges.
Fluke Biomedical Regulatory CommitmentAs a medical test device manufacturer, we recognize and follow certain quality standards and certifications when developing our products. We are ISO 9001 certified and our products are:• CE Certified, where required• NIST Traceable and Calibrated• UL, CSA, ETL Certified, where required• NRC Compliant, where required
8FlukeBiomedical Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzer
www.elso.sk
riešenia na presné meranie