Fluke Biomedical Impulse 7000DP - ELSO Philips … Fluke Biomedical Impulse 6000D/7000DP...

8
Impulse 6000D/7000DP Defibrillator/External Pacer Analyzer Key features Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional) Lown, Edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility AED technology compatibility First-class measurement accuracy ± 1 % of reading + 0.1 J Intuitive user interface and backlight, easy-to-read display Portable, rugged, easy to carry Long-lasting, rechargeable battery Pacer brand selections Pacer input protected against defibrillator output (7000DP only) 10 independent ECG outputs that provide 12 lead combinations for standardized clinical signals Flexible heart-rate settings (1 BPM step) facilitate rate meter accuracy and alarm testing DSP-based measurements enable future firmware and waveforms upgrade Unique integrated posts for secure connections Two-year extended warranty (no-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center) Optional Ansur test automation software to standardize testing procedures, capture waveforms and test results, and print and document test results Designed, tested, and built to incomparable Fluke quality standards Technical Data The Impulse 6000D Defibrillator Analyzer and Impulse 7000DP Defibrillator/Transcutaneous Pacer Analyzer Test Systems are rugged, portable preci- sion test instruments that ensure proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment. The Impulse 6000D and Impulse 7000DP test capabilities encompass the spectrum of worldwide- established pulse shapes, showcase breakthrough AED technology compatibility, and outperform in accuracy and standards. Additionally, the Impulse 7000DP incorporates the tests and the extensive range of test loads and measurement algorithms needed to test external transcutaneous pacemakers. In conjunction with an Impulse 7000DP, the Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω for defibrillator performance testing. A standard USB interface enables computer control and data transfer, and optional Ansur PC-based automation software increases productivity by outfitting users with an easy-to-use method to standardize testing procedures and capture, print and document data. www.elso.sk riešenia na presné meranie

Transcript of Fluke Biomedical Impulse 7000DP - ELSO Philips … Fluke Biomedical Impulse 6000D/7000DP...

Impulse 6000D/7000DPDefibrillator/External Pacer Analyzer

Key features

• Impulse7010DefibrillatorSelectableLoadAccessoryprovidesmultipleloadsof25Ω,50Ω,75Ω,100Ω,125Ω,150Ω,175Ω,and200Ω tocomplywithIEC60601-2-4standard(optional)

• Lown,Edmark,trapezoidal,biphasicandpulsed-biphasicdefibrillationtechnologycompatibility• AEDtechnologycompatibility• First-classmeasurementaccuracy±1%ofreading+0.1J• Intuitiveuserinterfaceandbacklight,easy-to-readdisplay• Portable,rugged,easytocarry• Long-lasting,rechargeablebattery• Pacerbrandselections• Pacerinputprotectedagainstdefibrillatoroutput(7000DPonly)• 10independentECGoutputsthatprovide12leadcombinationsforstandardizedclinicalsignals• Flexibleheart-ratesettings(1BPMstep)facilitateratemeteraccuracyandalarmtesting• DSP-basedmeasurementsenablefuturefirmwareandwaveformsupgrade• Uniqueintegratedpostsforsecureconnections• Two-yearextendedwarranty(no-costextendedwarrantyavailableafterfirst-yearcalibrationatanyFluke

Biomedicalauthorizedservicecenter)• OptionalAnsurtestautomationsoftwaretostandardizetestingprocedures,capturewaveformsandtest

results,andprintanddocumenttestresults• Designed,tested,andbuilttoincomparableFlukequalitystandards

Technical Data

TheImpulse6000DDefibrillatorAnalyzerandImpulse7000DPDefibrillator/TranscutaneousPacerAnalyzerTestSystemsarerugged,portablepreci-siontestinstrumentsthatensureproperoperationandultimateperformanceofcriticallife-supportcardiac-resuscitationequipment.

TheImpulse6000DandImpulse7000DPtestcapabilitiesencompassthespectrumofworldwide-establishedpulseshapes,showcasebreakthroughAEDtechnologycompatibility,andoutperforminaccuracyandstandards.Additionally,theImpulse7000DPincorporatesthetestsandtheextensiverangeoftestloadsandmeasurementalgorithmsneededtotestexternaltranscutaneouspacemakers.

InconjunctionwithanImpulse7000DP,theImpulse7010DefibrillatorSelectableLoadAccessoryprovidesmultipleloadsof25Ω,50Ω,75Ω,100Ω,125Ω,150Ω,175Ω,and200Ωfordefibrillatorperformancetesting.AstandardUSBinterfaceenablescomputercontrolanddatatransfer,andoptionalAnsurPC-basedautomationsoftwareincreasesproductivitybyoutfittinguserswithaneasy-to-usemethodtostandardizetestingproceduresandcapture,printanddocumentdata.

www.elso.sk

riešenia na presné meranie

2FlukeBiomedical Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzer

General specificationsOperating temperature10°Cto40°C(50°Fto104°F)

Storage temperature-20°Cto60°C(-4°Fto140°F)

Humidity10%to90%non-condensing

DisplayLCDdisplay

CommunicationsUSBdeviceportforcomputercontrol

Modes of operationManualandremote

PowerInternalrechargeableNiMHbatterypackforninehours(typical)operationafterfullchargeorthebatterychargercanoperatetheanalyzerandchargethebatterysimultaneously

Battery charger100Vto240Vinput,15V/1.5Aoutput.Forbestperformance,thebatterychargershouldbeconnectedtoaproperlygroundedacreceptacle

EnclosureABSplastichousing

Dimensions(WxDxH)32cmx24cmx13cm(13inx9.5inx5in)

Weight3.02kg(6.6lb,0.1oz)

Safety standardsCE:IEC/EN61010-12ndEdition;Pollutiondegree2CAN/CSA-C22.2No61010-1;UL61010-1C-Tick:AustralianEMC

Electromagnetic compatibility standards (EMC)EuropeanEMC:EN61326-1

Defibrillator analyzer technical specificationsEnergy output measurementCompatible defibrillator waveshapesLown,Edmark,trapezoidal,dcbiphasic,andacpulsedbiphasic

Note:ACpulsedbiphasicwaveformhasnotbeenapprovedintheUnitedStates.

Autoranged measurement0.1Jto600J

Accuracy0.1Jto360J:±(1%ofreading+0.1J)360Jto600J:±(1%ofreading+0.1J),typical

Note:Forpulsedbiphasicdefibrillator,specifiedaccuracyis±(1.5%ofreading+0.3J)onbothranges.

Load resistanceResistance:50ΩAccuracy:±1%,non-inductive(<2µH)

Pulse trigger level20V

Pulse widthRange:1msto50msAccuracy:±0.1ms

VoltageRange:20Vto5000VAccuracy:±(1%ofreading+2V)

CurrentRange:0.4Ato100AAccuracy:±(1%ofreading+0.1A)

Tilt (biphasicandpulsedbiphasic)Range:1%to99%Accuracy:±1digit

Interphase delay (biphasicandpulsedbiphasic)Range:0.1msto9.9msAccuracy:±0.1ms

Frequency (pulsedbiphasiconly)Range:2000Hzto8000HzAccuracy:±1%ofreading

Duty cycle(pulsedbiphasiconly)Range:1%to99%Accuracy:±1digit

Sample rate250kHz(4µssample)

Maximum average power12W,equivalentto10defibpulsesof360Jevery5minutes

Scope outputAutorange:2000:1,400:1,and80:1dependingonrange

Waveform playback• Output:BNC• Outputimpedance:50Ω• Amplitudeaccuracy:±5%

Charge time measurementRange:0.1sto100sAccuracy:±0.05s,typical

Synchronization test (elective cardioversion)Delaytimemeasurement• Timingwindow:ECGR-wave

peaktothedefibpulsepeak• Range:-120msto380ms;

measurestimingfrom120mspriortotheR-wavepeaktoupto380msfollowingtheR-wavepeak

• Resolution:1ms• Accuracy:±1ms

www.elso.sk

riešenia na presné meranie

Normalsinuswaves:Lead#Ref.amp.I II III V1 V2 V3 V4 V5 V6

Lead I referencePerformancewavesandRwavedetection:Lead#Ref.amp.IIIIIIV1V2V3V4V5V6

Normalsinuswaves:Lead#Ref.amp.IIIIIIV1V2V3V4V5V6

ECG normal sinusRates:10BPMto360BPMin1BPMsteps

ECG high level output (BNC jack)Amplitude:• Range:0.5VpermVof

referenceleadsetting• Accuracy±5%Outputimpedance:50Ω

ECG on defibrillator input loadSameastheLeadIIamplitudebutlimitedto±4mV

ECG performance wavesSquarewave:2Hzand0.125HzTriangularwave:2Hzand2.5HzSinewaves:0.05,0.5,5,10,40,50,60,100,150,and200HzPulse:30BPMand60BPM,60mspulsewidth

R-wave detectionWaveform:Haver-triangle

Amplitude:0.05mVto0.45mVin0.05mVstepsand0.5mVto5mVin0.5mVstepsRate:30,60,80,120,200,and250BPMWidths:8,10,12ms,and20msto200msin10msstepsAccuracy:±(1%setting+1ms)

Noise immunityWave:SineLinefrequency:50Hzor60Hz(±0.5Hz)Amplitude:•Range:0.0mVto10mVin

0.5mVsteps•Accuracy:±5%

Transvenous pacer pulse simulationWidths• Range:0.1ms,0.2ms,0.5ms,

1ms,and2ms• Accuracy:±5%ofsettingAmplitudes:• Range:0(off)and±2mV,

±4mV,±6mV,±8mV,±10mV,±12mV,±14mV,±16mV,±18mV,±20mV,±50mV,±100mV,±200,±500,and±700mV

• Accuracy:±(10%setting+0.2mV)

Amplitude of transvenous pacer pulse simulation signals relative to amplitude setting (in percent)

Lead II referenceLead#Ref.amp.IIIIIIV1V2V3V4V5V6

Lead I referenceLead#Ref.amp.IIIIIIV1V2V3V4V5V6

ECGwaves• Normalsinusrhythm(NSR):

10BPMto180BPMin1BPMsteps

• Atrialfibrillation:Coarseandfine

• Monomorphicventriculartachycardia:120BPMto240BPMin5BPMsteps

• Asystole:Flatline

Automated defibrillator test ECG wavesNormalsinus:10BPMto300BPMin1BPMstepsVentricularfibrillation:CoarseandfineMonomorphicventriculartachycardia:120BPMto300BPMin5BPMstepsPolymorphicventriculartachycardia:5typesAsystole:Flatline

ECG wavesECG generalLeadconfiguration:12-leadsimulation;RA,LL,LA,RL,V1-6withindependentoutputsLeadtoleadimpedance:1000Ω(nominal)Rateaccuracy:±1%nominal

ECG amplitudesReferencelead:Selectable,LeadII(default)orLeadISettings:0.05mVto0.45mVby0.05mVstepsand0.5mVto5mVby0.5mVstepsAccuracy(allperformancewavesandnormalsinusRwaves):• LeadII....................................±2%• Allotherleads.....................±5%• Defibpaddles.......................±5%

Amplitude of ECG signals relative to amplitude setting (in percent)

Lead II referencePerformancewavesandRwavedetection:Lead# Ref.amp.I II III V1 V2 V3 V4 V5 V6

70%100%30%

100%100%100%100%100%100%

70%100%30%24%48%

100%120%112%80%

100%150%50%

100%100%100%100%100%100%

100%150%50%24%48%

100%120%112%80%

67%100%33%67%67%67%67%67%67%

100%150%50%

100%100%100%100%100%100%

Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzerFlukeBiomedical3

www.elso.sk

riešenia na presné meranie

4FlukeBiomedical Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzer

Arrhythmia selectionsPacerinteractive(7000DPonly)• Demand:30BPMto360BPMin

1BPMsteps• Asynchronous• Non-capture• Non-function• Threshold(interactivepacing

simulationonly):10mAto250mAin10mAsteps

Supraventricular• Artrialfibrillationcourse• Atrialfibrillationfine• Atrialflutter• Sinusarrhythmia• Missedbeat• Atrialtachycardia• Paroxysmalatrialtachycardia

(PAT)• Nodalrhythm• SupraventriculartachycardiaPremature• AtrialPAC• NodalPNC• PVC1leftventricle• PVC1LVearly• PVC1LVRonT• PVC2rightventricle• PVC2RVearly• PVC2RVRonT• MultifocalPVCsVentricular• PVCs6/min• PVCs12/min• PVCs24/min• Freqmultifocal• Trigeminy• Bigeminy• PairPVCs• Run5PVCs• Run11PVCs• Monomorphicventricular

tachycardia:120BPMto300BPMin5BPMsteps

• Polymorphicventriculartachycardia:1to5

• Ventricularfibrillation:coarseandfine

• Asystole

Conduction• 1°Block• 2°BlockTypeI• 2°BlockTypeII• 3°Block• RightbundlebranchblockRBBB• LeftbundlebranchblockLBBBTransvenousPacedwithselectablepacerspikeamplitudesandwidths• Atrial80BPM• Async75BPM• Demandwithfrequentsinus

beats• Demandwithoccasionalsinus

beats• AVsequential• Non-capture• Non-function

Selections for all waves in group

Atrial pacer pulseWidth:0.1,0.2,0.5,1,2msPolarity:+or-Amplitude:0(off),2to20(by2),50,100,200,500,700mV

Ventricular pacer pulseWidth:0.1,0.2,0.5,1,2msPolarity:+or-Amplitude:0(off),2to20(by2),50,100,200,500,700mV

R-wave detectionRate:30,60,80,120,200,250BPMWidth:8,10,12,20to200(by10)msAmplitude:0.05to0.45(by0.05),0.5to5(by0.5)mV

www.elso.sk

riešenia na presné meranie

Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzerFlukeBiomedical5

Transcutaneous pacemaker analyzer technical specifications(7000DP only)

Test load Selections

Defibrillator inputFixedload:50ΩAccuracy:±1%,non-inductive(<2µH)Powerrating:10defibpulsesof360Jevery5minutes

Pacemaker inputVariableload:50Ωto1500Ωin50Ω stepsAccuracy:±2%,non-inductive(<2µH)Powerrating:5Ω(average),40Ω(peak)@1000Ω

MeasurementsManufacturer specific algorithms• GEResponder(1500and1700)• MDE300(MedicalData

Electronics)• MedtronicERS/PhysioControl

LIFEPAK• MRL(MedicalResearch

Laboratory/WelchAllyn)• Philips/Agilent/HP• SchillerMedical• ZOLLMedical(plusageneralpurposeDefaultAlgorithmselection)

CurrentRange:4mAto250mAAccuracy:±1%ofreading+0.02mA

Pulse rateRange:5PPMto800PPMAccuracy:±0.5%ofreading+0.1PPM

Pulse widthRange:1msto100msAccuracy:±0.5%ofreading+0.01ms

EnergyRange:1µJto2JAccuracy:±4%ofreading+10µJ

Demand and asynchronous mode testInput pacer pulse rates30PPMto200PPM

ECG NSR waveRate:10BPMto300BPMin1BPMstepsAmplitude:1mVUnderdriverate:10BPMminimumOverdriverate:300BPMmaximum

Sensitivity testAutomatic interactive threshold detectionCompatiblepacerrates:30PPMto120PPM

ECG R waveWaveforms:Square,triangle,sineWidth:1msto19ms(by1ms),20msto95ms(by5ms),100msto300ms(by25ms)

Accuracy:±5%ofsettingAmplitude:0.05mVto0.95mV(by0.05mV),1mVto5mV(by0.5mV)Accuracy:±5%ofsetting

Refractory period testsPaced refractory period20msto500ms

Sensed refractory period15msto500ms

Accuracy±1ms

Pacer pulse rate20PPMto200PPM

ECG Waveform:TrianglewavePulsewidth:40msAmplitude:1mV

www.elso.sk

riešenia na presné meranie

6FlukeBiomedical Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzer

Impulse 7010 Defibrillator Selectable Load AccessoryGeneral specifications

Maximum voltage 5000V

Maximum continuous power 12W,equivalentto10defibpulsesof360Jevery5minutes

Inductance<2µH,@25Ω<3µH,@50Ω<4µH,@75 Ωand100Ω<5µH,@125Ω<6µH,@150Ω<7µH,@175Ω<8µH,@200Ω

TemperatureOperating:10°Cto40°C(50°Fto104°F)Storage:-20°Cto60°C(-4°Fto140°F)

Humidity10%to90%non-condensing

Dimensions (WxDxH)154mmx272mmx138.7mm(6.07inx10.71inx5.46in)

Weight (net)1.54kg(3lb6.2oz)

Safety classComplieswithEN61010-12ndEdition,ClassIIproduct

Safety and EMC marks

WarrantyTwo-yearextendedwarranty(no-costextendedwarrantyavailableafterfirst-yearcalibrationatanyFlukeBiomedicalauthorizedservicecenter)

Calibration intervalOne-year

Electrical specifications (for Load Accessory and Analyzer together)

Load settings25Ω,50Ω,75Ω,100Ω,125Ω,150Ω,175Ω,and200Ω±1%

AccuracyEnergy(allexceptpulsedbiphasic):2%ofreading+0.1Jwith25,75Ωthough200Ωloads,1%ofreading+0.1Jwith50ΩloadEnergy(pulsedbiphasic):2.5%ofreading+0.3Jwith25,75Ω though200Ωloads,1.5%ofreading+0.3Jwith50ΩloadVoltage:1%ofreading+2Vwith25Ωand50Ωloads,2%ofreading+2Vwith75Ωthrough200ΩloadsCurrent:2%ofreading+0.1Awith25Ωload,1%ofreading+0.1Awith50Ωthrough200Ωloads

N10140

www.elso.sk

riešenia na presné meranie

Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzerFlukeBiomedical7

Ordering information

Models2811928Impulse6000DDefibrillatorAnalyzer120V(US)3077031 Impulse6000DDefibrillatorAnalyzer(Schuko)3077046 Impulse6000DDefibrillatorAnalyzer(UK)3077054 Impulse6000DDefibrillatorAnalyzer(Japan)3085270 Impulse6000D DefibrillatorAnalyzer(Australia)3085281 Impulse6000DDefibrillatorAnalyzer(India)2811919 Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer120V(US)3077005 Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer(Schuko)3077010 Impulse7000DP Defibrillator/TranscutaneousPacemakerAnalyzer(UK)3077022 Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer(Japan)3085296Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer(Australia)3085308Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzer(India)3326874TA-IMP7KDPImpulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation120V(US)3326888TA-IMP7KDP-01Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(Schuko)3326895TA-IMP7KDP-02Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(UK)3326901TA-IMP7KDP-03Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(Japan)3326912TA-IMP7KDP-04Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(Australia)3326920TA-IMP7KDP-05Impulse7000DPDefibrillator/TranscutaneousPacemakerAnalyzerwithtestautomation(India)

Standard accessories1626219USBComputerCommunicationCable3028681UserManualCD3028662Getting-StartedGuideBattery Eliminator(countryspecific)2814980CarryingCase3156262DefibPaddleContactPlates

Optional accessories3091370 AnsurImpulse6000D/7000DPPlug-In3065489 MedtronicERS/Physio-Control(FASTPATCH)(setoftwo):4mmdefibrillatoradapters3065450 KimberlyClark/R2DaroxMRL/MDE/NK:4mmdefibrillatoradapters3065438 Internaldischargepaddlecontacts(setoftwo)3065477 MedtronicERS/Physio-Control(QUIKPACE)(setoftwo):4mmpaceradapters3065527 ZollMedicalNTP/PD1400:4mmpaceradapters3065461 MedtronicERS/Physio-Control(QUIKCOMBO):4mmdefib/paceradapters3065492 Philips/Agilent/HP(CODEMASTERSeries-Round):4mmdefib/paceradapters3065509 Philips/AgilentHEARTSTARTFR2/MRX:4mmdefib/paceradapters3065511 ZollPD-2200Multi-FunctionPD-Series,M-Series,M-SeriesCCT,AEDPROandAEDPlus™defib/paceradapters3065423 GEMarquette(RESPONDER1500/1700Series)(setoftwo):4mmdefib/paceradapters3158544 Impulse 7010DefibrillatorSelectableLoadAccessory

www.elso.sk

riešenia na presné meranie

Fluke Biomedical.Better products. More choices. One company.

Fluke Biomedical6045CochranRoadCleveland,OH44139-3303U.S.A.

Fluke Biomedical EuropeScienceParkEindhoven51105692ECSon,TheNetherlands

For more information, contact us:IntheU.S.A.(800)850-4608orFax(440)349-2307InEurope/M-East/Africa+31402675435orFax+31402675436Fromothercountries+1(440)248-9300orFax+1(440)349-2307Email:[email protected]:www.flukebiomedical.com

©2008-2010FlukeBiomedical.Specificationssubjecttochangewithoutnotice.PrintedinU.S.A.3/20103393108CD-EN-N

Modification of this document is not permitted without written permission from Fluke Corporation.

About Fluke BiomedicalFluke Biomedical is the world’s leading manufacturer of quality biomedical test and simulation products. In addition, Fluke Biomedical provides the latest medical imaging and oncology quality-assurance solutions for regulatory compliance. Highly credentialed and equipped with a NVLAP Lab Code 200566-0 accredited laboratory, Fluke Biomedical also offers the best in quality and customer service for all your equipment calibration needs.

Today, biomedical personnel must meet the increasing regulatory pressures, higher quality standards, and rapid technological growth, while performing their work faster and more efficiently than ever. Fluke Biomedical provides a diverse range of software and hardware tools to meet today’s challenges.

Fluke Biomedical Regulatory CommitmentAs a medical test device manufacturer, we recognize and follow certain quality standards and certifications when developing our products. We are ISO 9001 certified and our products are:• CE Certified, where required• NIST Traceable and Calibrated• UL, CSA, ETL Certified, where required• NRC Compliant, where required

8FlukeBiomedical Impulse6000D/7000DPDefibrillator/ExternalPacerAnalyzer

www.elso.sk

riešenia na presné meranie

maros