Download - Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Transcript
Page 1: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Spectroscopy

Infrared Spectroscopy (IR)Fourier Transform Infrared

(FTIR)

Page 2: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

The IR Region

• These frequencies match the frequencies of covalent bond stretching and bending vibrations. Infrared spectroscopy can be used to find out about covalent bonds in molecules.

– type of bonds are present

– some structural information

Page 3: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 3

The IR Region• Infrared radiation

λ = 2.5 to 17 μm

ν = 4000 to 600 cm-1

• Units are wave numbers, or cm-1, the reciprocal of the wavelength in centimeters. Wave numbers are proportional to frequency and energy.

Page 4: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Energy Levels: Basic Ideas

Page 5: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 5

Stretching Frequencies

• Frequency decreases with increasing atomic weight.

• Frequency increases with increasing bond energy.

Page 6: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 6

Vibrational ModesNonlinear molecule with n atoms usually has 3n -

6 fundamental vibrational modes.

Page 7: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 7

Fingerprint of Molecule

• Whole-molecule vibrations and bending vibrations are also quantitized.

• No two molecules will give exactly the same IR spectrum (except enantiomers).

• Simple stretching: 1600-3500 cm-1.• Complex vibrations: 600-1400 cm-1, called the

“fingerprint region.” =>

Page 8: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 8

IR-Active and Inactive

• A polar bond is usually IR-active.• A non-polar bond in a symmetrical

molecule will absorb weakly or not at all.

Page 9: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 9

Carbon-Carbon Bond Stretching

• Stronger bonds absorb at higher frequencies:– C-C 1200 cm-1

– C=C 1660 cm-1

– CC 2200 cm-1 (weak or absent if internal)• Conjugation lowers the frequency:

– isolated C=C 1640-1680 cm-1

– conjugated C=C 1620-1640 cm-1

– aromatic C=C approx. 1600 cm-1 =>

Page 10: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 10

Carbon-Hydrogen Stretching

• Bonds with more s character absorb at a higher frequency.– sp3 C-H, just below 3000 cm-1 (to the right)– sp2 C-H, just above 3000 cm-1 (to the left)– sp C-H, at 3300 cm-1

=>

Page 11: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 11

An Infrared Spectrometer

=>

Page 12: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

FTIR (Fourier Transform Infrared)

• FTIR spectrometer obtains an infrared spectra by first collecting an interferogram of a sample signal using an interferometer,

• FTIR then performs a Fourier Transform on the interferogram to obtain the spectrum. – Fourier transform defines a relationship between

a signal in time domain and its representation in frequency domain.

Page 13: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

FTIR (Fourier Transform Infrared)

• An interferometer is an instrument that uses the technique of superimposing (interfering) two or more waves, to detect differences between them.

• Michelson Interferometer.– A mirror moves at a fixed rate. Its position is determined accurately by

counting the interference fringes of a collocated Helium-Neon laser. – The interferometer splits a beam of radiation into two paths having

different lengths, and then recombines them. – A detector measures the intensity variations of the exit beam as a

function of path difference.

Page 14: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Schematic of Michelson Interferometer

Page 15: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 15

Summary of IR Absorptions

=>=>

Page 16: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 16

An Alkane IR Spectrum

=>

Page 17: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 17

An Alkene IR Spectrum

=>

Page 18: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 18

An Alkyne IR Spectrum

=>

Page 19: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 19

O-H and N-H Stretching

• Both of these occur around 3300 cm-1, but they look different.– Alcohol O-H, broad with rounded tip.– Secondary amine (R2NH), broad with one sharp

spike.– Primary amine (RNH2), broad with two sharp

spikes.– No signal for a tertiary amine (R3N) =>

Page 20: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 20

An Alcohol IR Spectrum

=>

Page 21: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 21

An Amine IR Spectrum

=>

Page 22: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 22

Carbonyl Stretching

• The C=O bond of simple ketones, aldehydes, and carboxylic acids absorb around 1710 cm-1.

• Usually, it’s the strongest IR signal.• Carboxylic acids will have O-H also.• Aldehydes have two C-H signals around

2700 and 2800 cm-1. =>

Page 23: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 23

A Ketone IR Spectrum

=>

Page 24: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 24

An Aldehyde IR Spectrum

=>

Page 25: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 25

O-H Stretch of a Carboxylic Acid

This O-H absorbs broadly, 2500-3500 cm-1, due to strong hydrogen bonding.

=>

Page 26: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 26

Variations in C=O Absorption

• Conjugation of C=O with C=C lowers the stretching frequency to ~1680 cm-1.

• The C=O group of an amide absorbs at an even lower frequency, 1640-1680 cm-1.

• The C=O of an ester absorbs at a higher frequency, ~1730-1740 cm-1.

• Carbonyl groups in small rings (5 C’s or less) absorb at an even higher frequency. =>

Page 27: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 27

An Amide IR Spectrum

=>

Page 28: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 28

Carbon - Nitrogen Stretching

• C - N absorbs around 1200 cm-1.• C = N absorbs around 1660 cm-1 and is

much stronger than the C = C absorption in the same region.

• C N absorbs strongly just above 2200 cm-

1. The alkyne C C signal is much weaker and is just below 2200 cm-1 . =>

Page 29: Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)

Chapter 12 29

A Nitrile IR Spectrum

=>