Download - Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Transcript
Page 1: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Marcy Stutzman, Philip Adderley, Matt Poelker

Thomas Jefferson National Accelerator Facility

Newport News, VA 23601

Continued Work toward XHV

for the

Jefferson Lab Polarized Electron Source

Page 2: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Thomas Jefferson National Accelerator Facility

• 6 (12) GeV Electron accelerator

for Nuclear Physics

• 85-90% polarization, up to

~250μA beam (CW) to 3 (4)

experimental halls

• DC photoemission electron

source

• Strained superlattice

GaAs/GaAsP photocathode

• NEG and Ion pumps

• Residual gasses are ionized,

back-accelerated and degrade

the photocathode

• Future accelerators (CLIC, EIC,

ILC) require higher currents

Marcy Stutzman AVS 2011

Page 3: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Cryopumped gun project

• Investigate adding bakable cryopump into system

– Leybold Coolvac 2000 BL, special order

– Cryosorber panel can be chilled with LN2 during bakeout

– Isolation valve for regeneration

– Chamber: currently in heat treatment

• Can we measure improvement in vacuum due to

cryopump?

Characterize UHV/XHV gauges

Marcy Stutzman AVS 2011

Page 4: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Ionization gauge current contributions

Ireal: pressure dependent gas phase ions – species sensitive

Ix-ray: x-ray induced electron desorption from collector

– reduce by geometry

IESD: ions arriving at collector from electron stimulated

desorption (ESD) of molecules on the grid

– reduce by degassing grid

Iheating: pressure rise due to filament heating – species sensitive

– reduce by material selection, geometry, long duration

.. neutESDrayxinvESDheatingrayxrealmeasured IIIIIII

Marcy Stutzman AVS 2011

Page 5: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Deep UHV/XHV gauges

• Extractor gauge – available for decades

– x-ray limit reduced through geometry

– x-ray limit quote: 7.5x10-13 Torr

• Axtran gauge – Bessel box energy discrimination

– electron multiplier to assist in low current measurements

– Purchased, not yet installed

– Measurement limit quote: <7.5e-15 Torr

• Watanabe BBB (Bent Belt Beam) gauge – Newly designed (JVSTA 28 (2010) p. 486)

– Operates with Leybold IE540 controller

– 230° degree deflector (similar to Helmer)

– BeCu housing to reduce Iheating

– Manufacturer’s lower limit: 4x10-14 Torr

extractor

BBB

Axtran

Marcy Stutzman AVS 2011

Page 6: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Gauge characterization chamber

• Heat treated twice

– 400°C 10 days

• Outgassing (Q)

– 3x10-14 Torr·L/s·cm2

• Q following 250°C bake

– 6.3x10-14 Torr·L/s·cm2

• 3800 cm2, 12L

• Pumping

– 4 WP1250 NEGs, 60%

1300 L/s

– 40 L/s ion pump

(behind right angle

valve)

• Extractor Gauge

• BBB Gauge

• 2 Leybold IE540

controllers

• 2 Keithley

electrometers

• UHV ion pump

power supply

• Diagnostic cross

with RGA and ion

pump

• NEG activation

flange Predicted pressure

2x10-13 Torr

Marcy Stutzman AVS 2011

Page 7: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Linearity between gauges

1.E-12

1.E-11

1.E-10

1.E-09

1E-12 1E-11 1E-10 1E-09

BB

B (

To

rr)

Extractor (Torr)

cooling runs heating runs

BBB and Extractor compared

vs. pressure

– Leybold IE 540 control

– Keithley Electrometer

Pressure varied in chamber by

heating NEGs or chilling

Conversion to Torr using

manufacturer calibration

factor / sensitivity

– Depends on species

– Ionization energy

– Ionization current

– Geometry

Gauge responses linear response over decades,

possible deviation at lowest pressures

Marcy Stutzman AVS 2011

Page 8: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Sensitivity

e

i

IP

IS

Ion current

Emission current

0.E+00

2.E-08

4.E-08

6.E-08

8.E-08

1.E-07

0.E+00 1.E-06 2.E-06 3.E-06 4.E-06 5.E-06 6.E-06

BB

B c

urr

ent

(Am

ps)

SRG (Torr, Hydrogen)

hydrogen pressure rise

0

2

4

6

8

10

0.E+00 1.E-06 2.E-06 3.E-06 4.E-06 5.E-06 6.E-06

Sen

siti

vit

y (

1/T

orr

)

SRG measured hydrogen pressure

BBB sensitivity vs. SRG

BBB and Extractor

– 120V electron energy

– 1.6 mA emission current

– geometry, collection

efficiency vary

BBB vs. SRG data from previous setup

Marcy Stutzman AVS 2011

Page 9: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

BBB sensitivity calculated using extractor

0

2

4

6

8

10

12

14

1E-12 1E-11 1E-10 1E-09

BB

B S

ensi

tiv

ity (

1/T

orr

)

Extractor Pressure (Torr)

SRG Data: pressure 105 higher than our area of concern

Calculate BBB sensitivity from Extractor gauge pressure?

Calculated Extractor

pressure relies on

sensitivity of extractor

gauge.

e

i

IS

IP

Gauge sensitivity depends on

• Ionization energy

• Gas species

• Geometry

• Collection efficiency

same

constant

Marcy Stutzman AVS 2011

Page 10: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

relative sensitivity

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.E-14 1.E-13 1.E-12 1.E-11

S(B

BB

) /

S(E

xt.)

Extractor current (Amps)

ratio of manufacturer

quoted sensitivities:

1.21=8/6.6

Define sensitivity ratio

- should be constant EXT

BBB

EXT

BBB

I

I

S

S

Can the deviation from constant behavior be explained by gauge backgrounds?

Marcy Stutzman AVS 2011

Page 11: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Background current measurements

0

10

20

30

0 100 200 300 400 500 600

Co

llec

tor

curr

ent

(fA

)

Reflector voltage (V)

Extractor x-ray current

accounts for

3/5 of total measured signal

BBB signal of 35 fA with

background of -1 fA

-10

0

10

20

30

40

50

0 50 100 150 200 250 300 350

Co

llec

tor

Cu

rren

t (f

A)

Deflector Voltage (V)

Gauge backgrounds

measured at different

times, different pressures

Marcy Stutzman AVS 2011

Page 12: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Sensitivity ratio: x-ray limit correction

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

1.E-15 1.E-14 1.E-13 1.E-12 1.E-11

Sen

siti

vit

y B

BB

/ S

ensi

tiv

ity

Ex

tra

cto

r

Extractor current (Amps)

• Subtraction of x-ray

background for

extractor gauge

overcorrects

• What else?

– ESD limits

– Load due to gauges

– Small current

measurement errors

– nonlinearity in

gauge response?

Marcy Stutzman AVS 2011

Page 13: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Electron stimulated desorption

• Electrons can liberate elements

adsorbed on the grid

• If grid - filament potential equal to

electron energy, ESD difficult to

separate

• Methods to reduce ESD

– high energy electron bombardment

(degas mode)

– operate grid at elevated temperature

– grid material optimization (BBB)

– stabilize for months

– Axtran: energy analysis since

electron energy ≠ grid-filament

potential

... ESDheatingrayxrealmeasured IIIII

Marcy Stutzman AVS 2011

Page 14: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Current due to heating by filaments

Use one gauge to measure the

additional current generated by

other hot filament

... ESDheatingrayxrealmeasured IIIII

0.E+00

1.E-14

2.E-14

3.E-14

4.E-14

5.E-14

0 10 20 30 40 50 60 70 80 90 100

Ga

ug

e cu

rren

t (A

mp

s)

Time (hours)

BBB turned off

BBB

Extractor

ΔI (BBB) = 4.8 fA

ΔI (Extractor) = 5.6 fA

BeCu BBB housing should reduce effect

Difference minimized after 6 months?

Marcy Stutzman AVS 2011

Page 15: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

So what is our pressure?

0.0E+00

5.0E-13

1.0E-12

1.5E-12

2.0E-12

2.5E-12

0 5 10 15 20 25 30

Ca

lcu

late

d p

ress

ure

(T

orr

)

Current (fA)

extractor

BBB

x-ray background

BBB heating Ext. heating

... ESDheatingrayxrealmeasured IIIII

predicted BBB (Torr) Extractor

Measured 1.4 x10-12 1.9 x 10-12

Corrected 1.1 x 10-12 5x10-14

Predicted 6x10-13

* pressures nitrogen

equivalent, Svendor

Marcy Stutzman AVS 2011

Page 16: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Conclusions

• Pressure in our systems (nitrogen equivalent) corrected for

gauge effects is near 1x10-12 Torr

• BBB signal to noise good: Noise < 10% signal

• Extractor gauge: measurements at lowest pressures

dominated by background

• BBB and extractor agree very well above 1x10-11 Torr

• The BBB gauge should be able to quantify pressure

improvements in the bakable cryopump system.

Marcy Stutzman AVS 2011

Page 17: Continued Work toward XHV for the Jefferson Lab … AVS 2011.pdf– Compare gauge with electron multiplier to avoid some small signal measurement issues – ESD ion discrimination

Future work

• Verify outgassing rate of chamber

– Is our predicted pressure correct? Does measured pressure agree?

• Add Axtran gauge to the system

– Compare gauge with electron multiplier to avoid some small signal

measurement issues

– ESD ion discrimination capability

– Determine which of the BBB and extractor is deviating from linear

response

• Gauge calibration to verify / determine sensitivity

– Repeat comparison with SRG (hydrogen vs. nitrogen)

– UHV calibration at NIST, PTB…

– XHV calibration (its own research project!)

• Add cryopump to NEG / ion pumping system

– Can we use the cryopump alone to achieve better pressures?

– Can we better activate NEGs?

– Is NEG / Ion pumping system limiting our pressure?

Marcy Stutzman AVS 2011