XC-THOR - Direct Conversion

2
XC-THOR PRODUCT DATASHEET MEDICAL - INDUSTRIAL - DENTAL INDUSTRIAL PHOTON COUNTING DETECTOR

Transcript of XC-THOR - Direct Conversion

Page 1: XC-THOR - Direct Conversion

XC-THORPRODUCT DATASHEET

MEDICAL - INDUSTRIAL - DENTAL

INDUSTRIAL PHOTON COUNTING DETECTOR

Page 2: XC-THOR - Direct Conversion

Pixel TechnologyPhoton CountingEnergy DiscriminationSingle / Dual EnergySensor material and thicknessCadmium Telluride (CdTe) : 0.75mm (standard energy)Cadmium Telluride (CdTe) : 2.0mm (high energy)Pixel size100 μmPixel Fill Factor100%Tile Gap100 μmX-ray Energy rating10kVp to 300kVp (higher energy solutions available)Active sensor area12.8mm, 25.6mm and 51.2mm wide80mm to 400mm lengthCount Rate RangeUp to 200 Mcps per mm2

Imaging Speed10,000+lines/s (full resolution)Maximum FramerateUp to 300fpsMTF>75 % @ 2lp/mm, >45 % @ 5lp/mmDQE (0) >75% (RQA5 spectra, 6μGy, SID 1m)Lag0% (after X-ray 6μGy)Ghosting<0.1% 1min after X-rays (6μGy)

Data interfaceGigabit EthernetPower supply24VDCExternal TriggerOpto coupled 5V, 12-24V with external resistor

Operating temperature range+5 to 35oC @ 30% to 85% humidityStorage temperature range+5 to 50oC @ 10% to 95% humidityHousingIP67 ratedDesigned to withstand 20G impacts

Software GUI includedFully flexible SDK includedSupported operating systemsWindows 7 onwards

The XC-THOR series detectors are specifically designed for applications requiring solutions that are robust and radiation hard. The direct-conversion CdTe technology of these detectors bring the benefits of high sensitivity to Industrial applications along with the image sharpness intrinsic to photon counting. With Dual-Energy capability the XC-Thor additionally enables selective energy imaging providing powerful removal of low energy scatter radiation. And the superior radiation hardness of CdTe compared to scintillator detectors sets new standards in detector lifetime.

This datasheet contains typical information specific to products manufactured at the time of its publication. Direct Conversion AB. reserves the right to amend the specifications at any time. Contents herein do not constitute a warranty. © Direct Conversion AB.

INDUSTRIAL PHOTON COUNTING DETECTORXC-THOR

ELECTRICAL INTERFACE

ENVIRONMENTAL CONDITIONS

SOFTWARE

IMAGING PARAMETERSAND PERFORMANCE

directconversion.comemail: [email protected] Call +46 8622 23 00