Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf ·...

37
Mass Analyzers Double Focusing Magnetic Sector Quadrupole Mass Filter Quadrupole Ion Trap Linear Time-of-Flight (TOF) Reflectron TOF Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS)

Transcript of Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf ·...

Page 1: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Mass Analyzers

• Double Focusing Magnetic Sector• Quadrupole Mass Filter• Quadrupole Ion Trap• Linear Time-of-Flight (TOF)• Reflectron TOF• Fourier Transform Ion Cyclotron

Resonance (FT-ICR-MS)

Page 2: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Mass Analyzers• Resolution

– R = m / Δm• Accuracy/Precision

– mass measurement accuracy/reproducibility• Transmission

– % of ions allowed through the analyzer• Mass Range

– Highest m/z that can be analyzed• Scan Speed

– How many spectra per unit of time

Page 3: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Double-Focusing Magnetic Sector

Page 4: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Double-Focusing Magnetic SectorMagnetic Sector

mq

r2B2

2V=

B = magnetic field strengthr = radius of curvature in magnetic fieldV = accelerating voltagem = ion massq = ion charge

All ions of the same m/z will have the same radius

Only if the Ion kinetic energy is constant

Page 5: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Double-Focusing Magnetic SectorElectric Sector

r 2EkqE=

Ek = ion kinetic energyr = radius of curvature in electric fieldE = magnitude of electric fieldq = ion charge

All ions exiting the electric sector have the same kinetic energy

Page 6: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Double-Focusing Magnetic Sector

AutoSpec Premier New Technology

The AutoSpec Premier has a dual GC interface to allow the use of two GC injectors with two GC columns installed for alternate or dual simultaneous injection. The pumping configuration and capacity allows change of the GC column without venting the instrument, reducing instrument downtime and increasing productivity. The dual GC interface of the AutoSpec Premier enables direct line-of-sight GC column installation into the source region, as well as giving an improved thermal variation profile.

Electronics development has enhanced the precision with which the accelerating voltage can be set during voltage selected ion recording (VSIR). This ultimately improves mass precision, which can lead to superior data quality, especially in circumstances where a background peak is close to the analytical mass.

A cold reference probe is included with the system enabling liquid introduction at room temperature. This gives superior lock mass stability over an extended time period, as well as increased functionality; the result is the ability to immediately switch between course and fine control to satisfy calibration and sample analysis requirements.

AutoSpec Premier’s patented double-focusing, tri-sector geometry.

Sensitivity

To achieve the highest sensitivity, HR/SIR acquisition is used. In this mode, at 10,000 resolution (10% valley definition), an injection of 100 fg of 2,3,7,8-TCDD will give a signal-to-noise ratio at 321.8936 Da of > 125:1 on raw data (no smoothing applied). Noise, in this case, is taken to be equivalent to 4 standard deviations (+/- 2 ). It is important to note that some other protocols for dioxin analysis quote signal-to-noise with the noise defined as 2 standard deviations (+/- 1 ). For this definition of noise, the signal-to-noise value is > 250:1.

Unmatched HR/SIR sensitivity on raw, unsmoothed data.

Stability

The AutoSpec Premier gives superb stability of response over an extended period of analysis.

This reduces the frequency of tuning and calibration necessary and increases productivity by reducing instrument downtime for routine maintenance.

Mass accuracy

Masses can be measured accurately to < 2 mDa or < 5 ppm RMS.

Dynamic range

The AutoSpec Premier offers > 5 orders of linear dynamic range enabling quantification over a wide concentration range and analysis of mixtures containing components at high and low concentrations in one analytical run.

Percent deviation of the natural dioxin and furan congeners, monitored using an EPA-1613 five function acquisition over 72 hours.

AutoTune

The optimization of instrument tuning is made rapid and easy with the AutoTune facility. A sophisticated tuning algorithm combined with automated ion optics gives optimum tuning in a matter of minutes. The duration of the AutoTune routine is minimized by ‘fast’ slits, which have a response time of <0.2 seconds regardless of the positional change; this allows for extremely rapid manipulation of the source and collector slits for optimum resolution and transmission.

Automated resolution verification

Regulatory requirements for some analyses require instrument resolution to be checked at regular intervals. Some US EPA Methods, for example, specify resolution checks every 12 hours during analysis. MassLynx software can automatically check instrument resolution after any sample in the Sample List, and will verify the resolution of all reference peaks in the mass window of the experiment. A hardcopy of these results can be provided, showing the peaks checked and the resolution measured for each. If the resolution of any of the reference peaks is outside user-specified tolerances, the sample list will be halted, preventing the loss of valuable samples.

Acquisition Modes

The following acquisition modes are supported:

• Voltage Selected Ion Recording (VSIR)

• Magnet Scanning

• Voltage Scanning

• B/E Product Ion scanning

• B2/E Precursor Ion Scanning

• Constant Neutral Loss Scanning

• MIKES

• Multiple Reaction Monitoring (MRM)

Unmatched performance

The AutoSpec Premier gives unmatched performance in a compact footprint—the ultimate in high resolution MS.

With the AutoSpec Premier, you will benefit from:

• Full control with MassLynx™ Software in a Microsoft Windows environment

• TargetLynx™ Application Manager included as standard

• Unique tri-sector (EBE) ion optics

• Extra wide gap magnet for higher transmission

• Ultra high resolution > 80,000

• >105 linear dynamic range

• Constant gain, long-life photomultiplier detection system

• Plug-in EI and CI sources for rapid changeover

• Dual GC Interface, column change without venting instrument

• Low maintenance high capacity diffusion pumps as standard. Optional turbomolecular pumping configurations.

• Integrated GC (Agilent 6890) and autosampler (Agilent 7683, CTC-GCPal and CombiPal) control

• AutoTune and automated resolution checking

• Optional novel split detector for the elucidation of isobaric interferences

Sources and Options

Plug-in ion sources are provided for Electron Impact (EI) and Chemical Ionization (CI), allowing rapid and convenient source change without venting the instrument. GC is most often the sample introduction mode of choice, but it is not necessary to physically remove the GC to utilize the solids probe or Desorption Chemical Ionization probe (DCI).

EBE Geometry

The AutoSpec Premier incorporates unique tri-sector (EBE geometry) ion optics, designed to give maximum sensitivity in all modes at both high and low resolution.

The EBE design has one electric sector positioned before and one after the magnetic sector. The first electric sector has demagnifying optics to give a high dispersion to magnification ratio, allowing high resolutions to be obtained with a wider source slit and higher sensitivity. The mass resolution is continuously variable to 80,000 (10% valley definition). The second electric sector reduces background noise, improves abundance sensitivity and automatically rejects metastable ion interferences.

Premier Magnet

The AutoSpec Premier magnet is fully laminated utilizing grain orientated rolled steel sheet with high permeability and low hysteresis. The extra-wide gap magnet design allows use of a wider flight tube, giving higher transmission and reducing susceptibility to contamination. The patented post-acceleration photomultiplier detection system enables detection of positive or negative ions without switching dynode voltages. This detector has single ion detection capability; and the long-term stability of the photomultiplier gain ensures excellent and reproducible long-term performance.

Ion Optics and Pumping

The ion optics are supported by a pneumatic suspension bench to isolate the analyzer from floor vibrations and maximize instrument performance.

The instrument is pumped with low maintenance, high capacity diffusion pumps (or optional turbomolecular pumps), with the analyzer double differentially pumped to ensure high abundance sensitivity. The vacuum system pumping, monitoring and protection is fully automatic.

Page 7: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Magnetic Sector

• Typically a voltage of 5-10kV is used to accelerate ions

• To obtain a full spectrum, magnetic field is scanned

• To obtain a HR scan, voltage is scanned at constant magnetic field

• To gain maximum sensitivity at one mass SIM scan is done– B and E are constant for one or more

masses

Page 8: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Advantages• Very High Resolution

(60,000)• High Accuracy (<5 ppm)• 10,000 Mass Range

Disadvantages• Very Expensive• Requires Skilled

Operator• Difficult to Interface to

ESI• Low resolution MS/MS

without multiple analyzers

Double-Focusing Magnetic Sector

Page 9: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Quadrupole Mass Filter

http://www.asms.org

Page 10: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

QuadrupoleE = U - Vcos(2πνt)

E = potential applied to the rodsU = DC potentialV = RF amplitudeν = RF frequencyt = time

Quadrupole is scanned at a constant U/V

Page 11: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Quadrupole Mass Filter

http://www.asms.org

Page 12: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Quadrupole

• Typically U varies from 500-2000 V• V varies from 0 - 3000V (-3000 to +3000)• Scanning U/V at a fixed ratio gives a full

scan– Higher values of U/V give higher resolution

• RF only (U=0) transmits all ions• Higher sensitivity through SIM scan

– Jumping to specific points on the U/V line

Page 13: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Advantages• Inexpensive• Easily Interfaced to

Many Ionization Methods

Disadvantages• Low Resolution (<4000)• Low Accuracy

(>100ppm)• MS/MS requires

multiple analyzers• Low Mass Range

(<4000)• Slow Scanning

Quadrupole Mass Filter

Page 14: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

HPLC Inlet

Nebulizer Gas lnlet

Nebulizer Needle Capillary

Skimmer

Octopoles

Drying Gas

Vacuum Partition

Lenses

Endcap

Main RF Auxillary RF(1/3 Frequency of Main RF)

Conversion Dynode

Electron Multiplier

Optimized Asymptote Angle Ring Electrode

Quadrupole Ion Trap

Page 15: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Quadrupole Ion Trap

Page 16: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Quadrupole Ion Trap

• Ions are injected into the trap and all ions are trapped

• RF and DC are scanned to sequentially eject ions for detection

• Specific ions can be trapped while others are ejected

• Ion velocity can be increased to induced fragmentation

Page 17: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Advantages• Inexpensive• Easily Interfaced to

Many Ionization Methods

• MS/MS in one analyzer

Disadvantages• Low Resolution (<4000)• Low Accuracy

(>100ppm)• Space Charging

Causes Mass Shifts• Low Mass Range

(<4000)• Slow Scanning

Quadrupole Ion Trap

Page 18: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Time-of-Flight (TOF)mv2

2zVs=

B = magnetic field strengthv = ion velocityVs = accelerating voltagem = ion massq = ion charge

All ions of the same m/z will have the same flight time

Only if the Ion kinetic energy is constant

= Ekdt = v t2 =

mz

d2

2Vs

Page 19: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Advantages• Extremely High Mass

Range (>1 MDa)• Fast Scanning

Disadvantages• Low Resolution (4000)• Low Accuracy

(>200ppm)• MS/MS not possible

Linear Time-of-Flight (TOF)

Page 20: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

TOF

• Ions are accelerated with 5-35 kV• Space focusing of source ions is

accomplished by delayed extraction• An electrostatic analyzer (reflectron) is

used correct for kinetic energy spread

Page 21: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Reflectron Time-of-Flight (MALDI-TOF)

P 2+

T arget Ref lec tor Ref lec torplate detector

Line

ar d

etec

tor

Drift region

P 1+

PCIS / De!ection

/IS

2G

roun

dp o

tent

ial

P 1+

P 1+

ePar nts andfragments

IS1

IS1

Page 22: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Reflectron Time-of-Flight (ESI-TOF)

Courtesy Bruker Datonics BioTOF user’s Manual

ESI Source

RF Ion Guide

Spray Chamber

Accelerator

Orthogonal Interface:

Drift Region

Detector

Flight Tube

Reflectron

Page 23: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Reflectron Time-of-Flight (TOF)

Page 24: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Advantages• High Resolution

(>20,000 in some models)

• High Accuracy (<3ppm)• 10,000 Mass Range• Fast Scanning >100 Hz

Disadvantages• Low Resolution for MS/

MS (PSD)

Reflectron Time-of-Flight (TOF)

Page 25: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS

Page 26: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MSv

2πrf=

B = magnetic field strengthv = ion velocityf = orbital frequencym = ion massq = ion charger= orbital radius

At constant B, orbital frequency is inversely related to m/z

Frequency is independent of kinetic energy

qvB =mv2

rvr

2πf==qB2πm

CentripitalForce

Circular Path

r and v drop out

Page 27: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS• Ions are all trapped radially by a

magnetic field (typically 3-15 T)• Axial trapping by DC potential• Ion radius is increased by RF pulse

– also brings orbits into phase• Orbiting ions induce RF current in

receiver plates– Image current is a composite of all

frequences in time domain• FFT gives frequency (mass) spectrum

Page 28: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MSActively shielded magnet

analyzer stage

transfer stagesource chamberESI source

ESI needle(atmosphere)

capillaryionguide

500 L/secturbo pump10-10 mbar

500 L/secturbo pump10-6 mbar

70 L/secturbo pump10-8 mbar

250 L/secturbo-dragpump10-4 mbar

5 L/secrotary vane10-1 mbar

cell

Page 29: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS

Page 30: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS

Page 31: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS

Page 32: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS

Page 33: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS

Page 34: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS

Page 35: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MS

1200 1400 1600 1800 m/z

Δm = 0.01933 a.u.1/Δm = 51.7308 a.u.

mass = 64428 a.u.

1278.3 1278.8 m/z

52+

52+

60+36+

Electrospray: Broadband Spectrum of Bovine SerumAlbumin (66kDa) 7.0T Actively Shielded Magnet

Page 36: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

FT-ICR-MSElectrospray: Deconvoluted Spectrum of BovineSerum Albumin (66kDa) 7.0T Actively Shielded Magnet

66410 66430 66450 m/z

Δm = 1.004 a.u.

Page 37: Mass Analyzers - University of Minnesotamslab.chem.umn.edu/class/lecture/1115_MassAnalyzers.pdf · 2006. 11. 22. · Mass Analyzers • Resolution – R = m / Δm • Accuracy/Precision

Advantages• Extremely High

Resolution (>500,000)• Very Good Accuracy (<1

ppm)• MS/MS in one analyzer

Disadvantages• Expensive• Requires

Superconducting Magnet

• Slow MS/MS

FT-ICR-MS