  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    -AMILASE

    Cat. Volume N Testes

    027 120 mL 120

    MODO: CIN IMPRESSO: EXTENSA

    WL1: 405 INTERVALO CINTICO: 60

    WL2: - N INTERVALOS: 2

    BLANK: NO DA/min: 0.220

    BLANK- AMOSTRA / PAD: NAO % LIM. LIN: 20

    CUB. FLUXO: SIM DIR: INCR

    TEMP: 37 LIN. LIM. MIN: 0

    VOL. ASPIR: 800 LIN. LIM. MAX: 7000

    RETARDO: 30 ABS REAT. MIN: 0.000

    PADRO: - ABS REAT. MAX: 2.500

    FATOR: 10183 ABS PAD. MIN: -

    PAD: - ABS PAD. MAX: -

    P1: - VR/VN MIN: #

    UNID: U/L VR/VN MAX: #

    DEC: 0 OBS: Caso o modelo do equipamento no apresente a opo de inserir o valor do FATOR recomendado (10183), pelo fato de serem disponveis apenas 4 dgitos para a programao, pode ser inserido o valor do FATOR divido pela metade (5092). Dessa forma, o resultado das amostras dever ser multiplicado por 2.

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 1

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    ACID PHOSPHATASE

    Cat. Volume N Testes

    10660 45 mL 45

    Fosfatase cida Prosttica: MODO: CIN IMPRESSO: EXTENSA WL1: 405 INTERVALO CINTICO: 30 WL2: - N INTERVALOS: 2 BLANK: NO DA/min: - BLANK- AMOSTRA / PAD: NAO % LIM. LIN: - CUB. FLUXO: SIM DIR: INCR TEMP: 37 LIN. LIM. MIN: 0 VOL. ASPIR: 800 LIN. LIM. MAX: 74 RETARDO: 30 ABS REAT. MIN: 0.000 PADRO: NAO ABS REAT. MAX: 2.500 FATOR: 248 ABS PAD. MIN: - PAD: - ABS PAD. MAX: - P1: - VR/VN MIN: # UNID: U/L VR/VN MAX: #

    DEC: 1 Para obter o resultado de Fosfatase cida Prosttica deve-se subtrair o valor obtido na dosagem da Fosfatase cida Total pelo valor encontrado na dosagem de Fosfatase cida Prosttica. Fosfatase cida Total:

    MODO: CIN IMPRESSO: EXTENSA WL1: 405 INTERVALO CINTICO: 30 WL2: - N INTERVALOS: 2 BLANK: NO DA/min: - BLANK- AMOSTRA / PAD: NAO % LIM. LIN: - CUB. FLUXO: SIM DIR: INCR TEMP: 37 LIN. LIM. MIN: 0 VOL. ASPIR: 800 LIN. LIM. MAX: 74 RETARDO: 30 ABS REAT. MIN: 0.000 PADRO: NAO ABS REAT. MAX: 2.500 FATOR: 248 ABS PAD. MIN: - PAD: - ABS PAD. MAX: - P1: - VR/VN MIN: # UNID: U/L VR/VN MAX: #

    DEC: 1

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 2

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    ALBUMINA

    Cat. Volume N Testes

    001 500 mL 500

    MODO: PF IMPRESSO: - WL1: 620 INTERVALO CINTICO: - WL2: - N INTERVALOS: - BLANK: SIM DA/min: - BLANK- AMOSTRA / PAD: NO % LIM. LIN: - CUB. FLUXO: SIM DIR: - TEMP: 37 LIN. LIM. MIN: 0 VOL. ASPIR: 800 LIN. LIM. MAX: 6.00 RETARDO: 003 ABS REAT. MIN: 0.000 PADRO: SIM ABS REAT. MAX: 2.500 FATOR: @ ABS PAD. MIN: - PAD: NICO ABS PAD. MAX: - P1: * VR/VN MIN: # UNID: g/dL VR/VN MAX: #

    DEC: 2

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 3

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    ALKALINE PHOSPHATASE LIQUICOLOR

    Cat. Volume N Testes

    12017 100 mL 100

    MODO: CIN IMPRESSO: EXTENSA WL1: 405 INTERVALO CINTICO: 30 WL2: - N INTERVALOS: 4 BLANK: NO DA/min: - BLANK- AMOSTRA / PAD: NAO % LIM. LIN: 20 CUB. FLUXO: SIM DIR: INCR TEMP: 37 LIN. LIM. MIN: 0 VOL. ASPIR: 800 LIN. LIM. MAX: 700 RETARDO: 60 ABS REAT. MIN: 0.000 PADRO: NAO ABS REAT. MAX: 2.500 FATOR: 2757 ABS PAD. MIN: - PAD: - ABS PAD. MAX: = P1: - VR/VN MIN: # UNID: U/L VR/VN MAX: #

    DEC: 0

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 4

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    AMILASE

    Cat. Volume N Testes

    002 55 mL 100

    MODO: PF IMPRESSO: - WL1: 660 INTERVALO CINTICO: - WL2: - N INTERVALOS: - BLANK: SIM DA/min: - BLANK- AMOSTRA / PAD: NO % LIM. LIN: - CUB. FLUXO: SIM DIR: - TEMP: 37 LIN. LIM. MIN: - VOL. ASPIR: 800 LIN. LIM. MAX: - RETARDO: 003 ABS REAT. MIN: 0.000 PADRO: NAO ABS REAT. MAX: 2.500 FATOR: 1 ABS PAD. MIN: - PAD: - ABS PAD. MAX: - P1: - VR/VN MIN: # UNID: Abs VR/VN MAX: #

    DEC: 3

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 5

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    APOLIPOPROTEIN A1, APO A1

    Cat. Volume N Testes

    11101 60 mL 60

    MODO: PF UNID: mg/dL WL1: 340 DEC: 0 WL2: - IMPRESSO: - BLANK: YES INTERVALO CINTICO: - BLANK- AMOSTRA / PAD: NAO N INTERVALOS: - CUB. FLUXO: SIM DA/min: - TEMP: 37 % LIM. LIN: - VOL. ASPIR: 800 DIR: 0 RETARDO: 003 LIN. LIM. MIN: 40 PADRO: SIM LIN. LIM. MAX: 400 FATOR: @ ABS REAT. MIN: 0.000 PAD: NICO ABS REAT. MAX: 2.400 P1: 0.0 ABS PAD. MIN: - P2: */8 ABS PAD. MAX: - P3: */4 VR/VN MIN: # P4: */2 VR/VN MAX: # P5: */1

    A curva de calibrao deve ser obtida com 5 pontos. Diluir o calibrador Apolipoprotein A1/B Standard, Cat 11104, de acordo com a instruo de uso do kit.

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 6

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    APOLIPOPROTEIN B, APO B

    Cat. Volume N Testes

    11102 60 mL 60

    MODO: PF UNID: mg/dL WL1: 340 DEC: 0 WL2: - IMPRESSO: - BLANK: YES INTERVALO CINTICO: - BLANK- AMOSTRA / PAD: NAO N INTERVALOS: - CUB. FLUXO: SIM DA/min: - TEMP: 37 % LIM. LIN: - VOL. ASPIR: 800 DIR: 0 RETARDO: 003 LIN. LIM. MIN: 20 PADRO: SIM LIN. LIM. MAX: 800 FATOR: @ ABS REAT. MIN: 0.000 PAD: NICO ABS REAT. MAX: 2.400 P1: 0.0 ABS PAD. MIN: - P2: */8 ABS PAD. MAX: - P3: */4 VR/VN MIN: # P4: */2 VR/VN MAX: # P5: */1

    A curva de calibrao deve ser obtida com 5 pontos. Diluir o calibrador Apolipoprotein A1/B Standard, Cat 11104, de acordo com a instruo de uso do kit.

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 7

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    ANTI-STREPTOLISIN-O (ASO)

    Cat. Volume N Testes

    11251P 100 mL 100

    MODO: PF IMPRESSO: - WL1: 540 INTERVALO CINTICO: - WL2: - N INTERVALOS: - BLANK: YES DA/min: - BLANK- AMOSTRA / PAD: NAO % LIM. LIN: - CUB. FLUXO: SIM DIR: - TEMP: 37 LIN. LIM. MIN: 0 VOL. ASPIR: 800 LIN. LIM. MAX: 800 RETARDO: 003 ABS REAT. MIN: 0.000 PADRO: SIM ABS REAT. MAX: 2.500 FATOR: @ ABS PAD. MIN: - PAD: NICO ABS PAD. MAX: - P1: * VR/VN MIN: # UNID: UI/mL VR/VN MAX: #

    DEC: 0 A calibrao do teste deve ser realizada com ASO Standard, Cat. 11351.

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 8

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    AUTO-BILIRUBIN-D LIQUICOLOR

    Cat. Volume N Testes

    10741 375 mL 300

    MODO: PF IMPRESSO: - WL1: 546 INTERVALO CINTICO: - WL2: - N INTERVALOS: - BLANK: SIM DA/min: - BLANK- AMOSTRA / PAD: NO % LIM. LIN: - CUB. FLUXO: SIM DIR: - TEMP: 37 LIN. LIM. MIN: 0 VOL. ASPIR: 800 LIN. LIM. MAX: 10.00 RETARDO: 003 ABS REAT. MIN: 0.000 PADRO: SIM ABS REAT. MAX: 2.500 FATOR: @ ABS PAD. MIN: - PAD: NICO ABS PAD. MAX: - P1: * VR/VN MIN: # UNID: mg/dL VR/VN MAX: #

    DEC: 2

    O laboratrio dever validar a programao fornecida. Ler a instruo de uso do produto para informaes a respeito da metodologia, reagentes, amostras. @ Dados inseridos pelo operador # Calculado pelo analisador * Concentrao do calibrador

    BIO PLUS MARCA REGISTRADA DE SEUS FABRICANTES

    Rev.: 12/09 Page 9

  • BIO PLUS / BIO 2000

    PROTOCOLOS DE APLICAO

    AUTO-BILIRUBIN-T LIQUICOLOR

    Cat. Volume N Testes

    10742 375 mL 300

    MODO: PF IMPRESSO: - WL1: 546 INTERVALO CINTICO: - WL2: - N INTERVALOS: - BLANK: SIM DA/min: - BLANK- AMOSTRA / PAD: NO % LIM. LIN: - CUB. FLUXO: SIM DIR: - TEMP: 37 LIN. LIM. MIN: 0 VOL. ASPIR: 800 LIN. LIM. MAX: 30.00 RETARDO: 003 ABS REAT. MIN: 0.000 PADRO: SIM ABS REAT. MAX: 2.500 FATOR: @ ABS PAD. MIN