02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 1 Abstract NEUTRON...

12
02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER WITH POSSIBILITY OF γ- BACKGROUND DISCRIMINATION Y. I. Vinogradov, A.M. Shvetsov, O. P. Vikhlyantsev, A.V. Kuryakin Russian Federal Nuclear Center, All-Russian Research Institute of Experimental Physics, Sarov, Russia, Email: [email protected] The neutron spectrometer with proportional recoil protons counter was developed in VNIIEF and used for a number of years [1, 2]. The bottom range of measurements of the spectrometer in mixed n-γ fields has been limited by energy 0.5 MeV The carried out researches [3] have shown possibility of discrimination of impulses of the counter from neutrons and γ -quanta by their shape and expansions of a range of measurement of neutron spectra in the field of low energies to tens keV. Then it has been developed hardware and the spectrometer software, allowing to register amplitude of an impulse of the proportional counter and a maximum growth rate of an impulse on forward front and to write down them in ListMode (a mode of consecutive record of parametres of impulses). The program part includes a number of the service possibilities allowing to adjust registration equipment during measurements (in on-line mode), to estimate visually the quality of discrimination of impulses from neutrons and γ‑quanta and to receive hardware spectra. The complex of the programs is used for off-line mode analysis of the received data, which provides processing of the list of the received data with discrimination of impulses, caused by neutrons and γ‑quanta, construction of corresponding hardware spectra, calculation of response matrix of the proportional gas counter and reconstruction of spectra. Research of possibilities of the spectrometer with n-g discrimination, spent on 252 Cf source, has shown that the bottom range of measurements of neutron spectra decreases till 60-80 keV. Reference list: 1. Kuvshinov M. I., Shvetsov A. M., Egorov V. P. et al. Measurement of neutron spectra for spherical critical assemblies of 235 U(90 %), 235 U(36 %), and 239 Pu(98 %) // Proceedings International Conference on the Physics of Reactors PHYSOR96, Mito, Ibaraki, Japan, September 16–20, 1996 / Japan Atomic Energy Research Institute Inc., Japan. 1996. Vol. 2. P.338‑346. 2. Shvetsov A. M., Egorov V. P., Fomushkin E. F. et al. Measurements of leakage neutrons spectra for spherical assemblies of 235 U(90 %), 235 U(36 %), and 239 Pu(98 %) // Proceedings International Conference on Nuclear Data for Science and Technology, Trieste, Italy, May 18–23, 1997 / Italian Phis. Soc. Inc., Bologna, Italy. 1997. Part 2. P.1359–1361. 3. A. M. Shvetsov, Yu. I. Vinogradov, O. P. Vikhlyantsev, A. V. Kuryakin. Investigation of the possibility of γ-background discrimination in a neutron spectrometer with a proportional counter of recoil protons. // Bulletin of the Russian Academy of Sciences: Physics, 2007, т.71, № 12, с.1748-1753.

Transcript of 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 1 Abstract NEUTRON...

Page 1: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 1

AbstractNEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER WITH

POSSIBILITY OF γ-BACKGROUND DISCRIMINATIONY. I. Vinogradov, A.M. Shvetsov, O. P. Vikhlyantsev, A.V. Kuryakin

Russian Federal Nuclear Center, All-Russian Research Institute of Experimental Physics, Sarov, Russia, Email: [email protected]

• The neutron spectrometer with proportional recoil protons counter was developed in VNIIEF and used for a number of years [1, 2]. The bottom range of measurements of the spectrometer in mixed n-γ fields has been limited by energy 0.5 MeV

• The carried out researches [3] have shown possibility of discrimination of impulses of the counter from neutrons and γ -quanta by their shape and expansions of a range of measurement of neutron spectra in the field of low energies to tens keV.

• Then it has been developed hardware and the spectrometer software, allowing to register amplitude of an impulse of the proportional counter and a maximum growth rate of an impulse on forward front and to write down them in ListMode (a mode of consecutive record of parametres of impulses). The program part includes a number of the service possibilities allowing to adjust registration equipment during measurements (in on-line mode), to estimate visually the quality of discrimination of impulses from neutrons and γ‑quanta and to receive hardware spectra.

• The complex of the programs is used for off-line mode analysis of the received data, which provides processing of the list of the received data with discrimination of impulses, caused by neutrons and γ‑quanta, construction of corresponding hardware spectra, calculation of response matrix of the proportional gas counter and reconstruction of spectra.

• Research of possibilities of the spectrometer with n-g discrimination, spent on 252Cf source, has shown that the bottom range of measurements of neutron spectra decreases till 60-80 keV.

Reference list:1. Kuvshinov M. I., Shvetsov A. M., Egorov V. P. et al. Measurement of neutron spectra for spherical critical assemblies of 235U(90 %), 235U(36 %), and 239Pu(98 %) //

Proceedings International Conference on the Physics of Reactors PHYSOR96, Mito, Ibaraki, Japan, September 16–20, 1996 / Japan Atomic Energy Research Institute Inc., Japan. 1996. Vol. 2. P.338‑346.

2. Shvetsov A. M., Egorov V. P., Fomushkin E. F. et al. Measurements of leakage neutrons spectra for spherical assemblies of 235U(90 %), 235U(36 %), and 239Pu(98 %) // Proceedings International Conference on Nuclear Data for Science and Technology, Trieste, Italy, May 18–23, 1997 / Italian Phis. Soc. Inc., Bologna, Italy. 1997. Part 2. P.1359–1361.

3. A. M. Shvetsov, Yu. I. Vinogradov, O. P. Vikhlyantsev, A. V. Kuryakin. Investigation of the possibility of γ-background discrimination in a neutron spectrometer with a proportional counter of recoil protons. // Bulletin of the Russian Academy of Sciences: Physics, 2007, т.71, № 12, с.1748-1753.

Page 2: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 2

Neutron detector• Designed and used in VNIIEF for number of years• Proportional gas counter, ~14% efficiency , ΔE/E~3.7%• Range: 0.5÷8 MeV without, or 0.05÷8 MeV with n-γ discrimination• ~ 3 kV voltage, 3 bar methane (CH4 + some 3He uses for calibration)• CAMAC based data acquisition, n spectra reconstruction software

Construction of proportional counter:1 – glass isolation; 2 - cathode, Ø 75 mm

× 1 m;3 – anode, Ø40 µm; 4 – guard pipes; 5 – spring for anode;6 - flanges, 7 – fluoroplastic gaskets; 8 – details from kovar (Fe+Co+Ni alloy)

Page 3: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 3

Principle of energy calibration• Gas counter contains some 3He for calibration

• Use 3He(n,p)T reaction with peak E=0.764 MeV

Thermal peak

Channel, k

N,C

ount

s

)(764,0

)(0

kkkk

kE

k0 Δk

Page 4: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 4

Why need n-γ discrimination• γ influent to low range spectra (E < 0.5 MeV)• γ influence is difficult to calculate theoretically• Low energy range limited to 0.5 MeV due to n+γ• Some kind of n-γ discrimination required for E < 0.5 MeV

n+γ

n(clean)

neutron counter spectra

E,MeV

N,c

ounts

Page 5: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 5

Principle of n-γ discrimination• Neutrons produce recoil protons• γ–quanta produce electrons• Hardware p, e pulses have different shapes

(front timing)• tfront, (dE/dt)max/Emax criteria can be used for n-

γ separation

Emax

(dE/dt)max

time

Hardware pulse from counter channel

0,05 0,10 0,15 0,20 0,25 0,30 0,35 0,40 0,45 0,500,0

0,1

0,2

0,3

0,4

0,5

0,6

U, В

t, *10-4 с

Typicalhardwaresignal fromgas detector

tfront

Typical shape of pulse front registered with TDS-3000, 50 MHz

Page 6: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 6

n-γ discrimination criteriaPreliminary study show that:• tfront criteria is not good for n-γ• (dE/dt)max/Emax criteria is good

γ-quantumneutrons

N,c

oun

ts

Page 7: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 7

n-γ discrimination efficiency and range• Fit [E,E+ΔE] slice spectra with 2 log-normal distributions• Find Sγ, Sn and cross (mutual overlap) ΔS areas• Criteria for n-γ low range is ΔS/Sn·100% < 5%• This range depends on Nγ/Nn

[70÷80] keVslice spectra

Low energy n-γ range (by 5% level) depends on Nγ/Nn ratio.

Nγ/Nn= 2.5

Nγ/Nn= 0.7

Nγ/Nn= 0.3

Nγ/Nn= 0.2

Page 8: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 8

Low energy n-γ range vs Nγ/NnE

n-γ(

0.05

),MeV

Low energy n-γ range (by 5% level) depends on Nγ/Nn ratio.

Page 9: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 9

n-γ discrimination tests• n-γ discrimination tested with mixed n-γ and clear γ sources• Low energy range without n-γ discrimination ~ 0.5 MeV• Low energy range with n-γ discrimination ~ 60÷80 keV (depend on Nγ/Nn)

γ-quantum

neutrons

γ-quantum

(dE/dt)max/Emax, µs-1

E,M

eV

E,M

eV

n+γ – source (252Cf) γ – source (СОСГИ-М )

Page 10: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 10

n-γ discrimination hardware• CAMAC based data acquisition system designed

SAPACounter

FA

ADC

CC

PC

In.1

In.2

DU

Gate1

Gate1

ST

SD

CA

MA

C b

us

HV

С1In.1

Gate1

In.2

Gate2

С2In.1

Gate1

In.2

Gate2

OR

HV – high voltage unit (Canberra 3106D), PA – charge sensitive 10 ns preamplifier (POLON 1105A), SA – spectrometry 2 µs amplifier (Canberra 2025), DU – discriminator unit (Canberra 2035A), SD – logic shaper and delay (Canberra 2055), FA – fast amplifier (Canberra 2111), ST – stretcher (ORTEC 242), ADC – 13 bit analog to digital converter (ORTEC AD 413A), C1,C2 – counter (POLON 420A), OR – output register (POLON OR350), CC – CAMAC controller (PKK4, VNIIEF designed), PC – personal computer for data acquisition with CRW-DAQ control software by VNIIEF.

Page 11: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 11

Online software (data acquisition)• CRW-DAQ package designed in VNIIEF, Sarov• Registered in Russian Federation registry • Data acquisition in List mode, 2-parameter events• CAMAC based data acquisition, up to 60 kHz rate

γ

n

(dE/dX) max

E max

Page 12: 02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, kouriakine@mail.ru 1 Abstract NEUTRON SPECTROMETER BASED ON PROPORTIONAL RECOIL PROTONS COUNTER.

02.07.2010 Alexey Kuryakin, RFNC-VNIIEF, Sarov, Russia, [email protected] 12

Offline software• Preliminary spectrum processing

– Energy calibration by 3He(n,p)T reaction

– n-γ discrimination, background subtraction

• Response matrix calculation (SUPREMAC)– Uses NDX, ENDF/B-VI nuclear constants database

– Uses Monte-Carlo for response matrix calculation

– Optimized for performance

• Neutron spectrum reconstruction (UFONIM)– Ill-posed problem (Fredholm equations of 1st kind)

– Uses iteration method by Tarasko

– Randomization method to calculate stat. errors

• Designed in VNIIEF

i

m

jjij SUG

1

G - response matrix,U – n-spectrum to find,S – hardware spectrum.

ij

mjij

ikimk

mk UG

GSUU )(

)()1(