ST Optical Jumper Cord(Simplex)

(Cord diameter : φ0.9mm Type)

TECHNICAL SPECIFICATIONS

p. 1

Page 2: ST Optical Jumper Cord(Simplex)fascoinfrasistema.com.mx/modulos/productos/pdf/ST Optical Jumpe… · Permissible deviation in dimensions without tolerance indication is in accordance

ww

w.f

asc

o.c

om

.mx

Paseo de la Reforma No. 295 Piso Mezzanine Colonia Cuauhtémoc Delegación Cuauhtémoc Ciudad de México, México CP. 06500

(55) 5533 8246 [email protected]

1. PROVISION

1.1 Application Limit

These specifications apply to the ST optical jumper cord.

2. ORDER INFORMATION

Order information of the ST optical jumper cord in Table 1

Table 1 Information of Order

p. 2

Page 3: ST Optical Jumper Cord(Simplex)fascoinfrasistema.com.mx/modulos/productos/pdf/ST Optical Jumpe… · Permissible deviation in dimensions without tolerance indication is in accordance

ww

w.f

asc

o.c

om

.mx

Paseo de la Reforma No. 295 Piso Mezzanine Colonia Cuauhtémoc Delegación Cuauhtémoc Ciudad de México, México CP. 06500

(55) 5533 8246 [email protected]

3. GENERAL SPECIFICATIONS

3.1 Parts and Materials

Parts and the materials of ST connector are shown in Table 2

Table 2 Parts and Materials of ST Jumper cord

Cord Φ 0.9mm

1*5

1~99

1*

1*

Q’ty

MT-RJ Housing Set

MT-Ferrule

Parts Name

9

8

7

No.

1*4

1*3

1*

Ceramic Ferrule

ST Housing Set

SC Housing Set

FC Housing Set

LC Housing Set

MU Housing Set6

2/1*2

21

Q’tyParts NameNo.

* In case of hybrid type jumper cords

3.2 Physical Dimensions

Figure 1 shows end face of ferruleFigure 2 geometry structure of ST endFigure 3 insertion Loss and Measurement system Figure 4 return Loss and Measurement systemFigure 5 shows the assembled state of ST Optical jumper cord(ST/PC-ST/PC). Figure 7 shows the assembled state of ST Optical jumper cord(ST/PC-SC/PC). Figure 9 shows the assembled state of ST Optical jumper cord(ST/PC-ST/PC).

Figure 10 shows the assembled state of ST Optical jumper cord(ST/PC-LC/PC). Figure 12 shows the assembled state of ST Optical jumper cord(ST/PC-MU/PC).

3.3 General Tolerances

Permissible deviation in dimensions without tolerance indication is in accordance with JIS

B 0405 class m, as shown in Table3.

p. 3

Page 4: ST Optical Jumper Cord(Simplex)fascoinfrasistema.com.mx/modulos/productos/pdf/ST Optical Jumpe… · Permissible deviation in dimensions without tolerance indication is in accordance

ww

w.f

asc

o.c

om

.mx

Paseo de la Reforma No. 295 Piso Mezzanine Colonia Cuauhtémoc Delegación Cuauhtémoc Ciudad de México, México CP. 06500

(55) 5533 8246 [email protected]

Table 3 General Tolerance (JIS B 0405 class m)

±0 .312030

±0 .2306

±0 .163

±0 130.5

Basic size step (mm)

Over UnderPermissible deviation (mm)

4. OPTICAL PERFORMANCE

Multi Mode

<0.3dB, Typical 0.2dB

> 20dB,Typical 30dB

<0.3dB, Typical 0.2dB

PC : > 45dB, Typical 48dB

UPC : > 55dB, Typical 58dB

Characteristics

Insertion Loss

Return Loss

Single Mode

Table 4 Optical performance of ST end

5. END FACE INSPECTION

Check the polishing with a microscope or magnifier(X800)

Large Pits or Internal chips Deep scratches

Small Pits

Core

Figure 1 End face inspection

p. 4

Page 5: ST Optical Jumper Cord(Simplex)fascoinfrasistema.com.mx/modulos/productos/pdf/ST Optical Jumpe… · Permissible deviation in dimensions without tolerance indication is in accordance

ww

w.f

asc

o.c

om

.mx

Paseo de la Reforma No. 295 Piso Mezzanine Colonia Cuauhtémoc Delegación Cuauhtémoc Ciudad de México, México CP. 06500

(55) 5533 8246 [email protected]

7. INSERTION AND RETURN LOSS MEASUREMENT SYSTEM The test

samples are measured in the following system

The insertion loss(I.R) is defined as Insertion Loss = -10P0

Measuring instruments : RX3000 JDS Backreflection and Loss test set

P1

Master AdaptorMaster Plug

Light source

Optical PowerMeter

Measured Plug

P0

P1

Photo Detector

Figure 3 Insertion Loss and Measurement system

Backreflection Meter

Measurement hybrid jumper

Backreflection Meter

Termination point for BR

Termination point for BR0

Measurement hybrid jumper

Figure 4 Return Loss and Measurement system

p. 5

Page 6: ST Optical Jumper Cord(Simplex)fascoinfrasistema.com.mx/modulos/productos/pdf/ST Optical Jumpe… · Permissible deviation in dimensions without tolerance indication is in accordance

ww

w.f

asc

o.c

om

.mx

Paseo de la Reforma No. 295 Piso Mezzanine Colonia Cuauhtémoc Delegación Cuauhtémoc Ciudad de México, México CP. 06500

(55) 5533 8246 [email protected]

8.

Table 5 Reliability characteristics of ST end

Item

Impact Test

Cable Retention

Durability

Vibration

Twist Test

Flex Test

Temperature Life

Humidity

Low Temperature

Temperature Cycling

Condition(GR-326)

1.5 m × 8drops

0.25Kg, 90゜

200 times

0~55Hz/min x 2hours

0.9Kg x(X:1.5,Y:3)

0.9Kgf x 100 times

85℃ × 14days

75℃ , 95%RH ×14days

-10℃ × 4days

-40 ~ 75 ℃ 8hours/cycle 42cycles

Typical properties

Insertion Loss ≦ 0.2

Δi ≦ 0.2

Δi ≦ 0.2

Δi ≦ 0.2

Δi ≦ 0.2

Δi ≦ 0.2

Δi ≦ 0.2

Δi ≦ 0.2

Δi ≦ 0.2

Δi ≦ 0.2

λ=1300 nm Δi : Deviation (dB)

9. SHIPPING INSPECTION

Insertion loss, return loss for ST jumper cord are inspected before shipping and the

inspection results are enclosed.

10. PACKING

The product is packed to prevent damage during shipment.

RELIABILITY CHARACTERISTICS

p. 6

Page 7: ST Optical Jumper Cord(Simplex)fascoinfrasistema.com.mx/modulos/productos/pdf/ST Optical Jumpe… · Permissible deviation in dimensions without tolerance indication is in accordance

ww

w.f

asc

o.c

om

.mx

Paseo de la Reforma No. 295 Piso Mezzanine Colonia Cuauhtémoc Delegación Cuauhtémoc Ciudad de México, México CP. 06500

(55) 5533 8246 [email protected]

Figure 5 shows the assembled state of ST Optical jumper cord(ST/PC-ST/PC).

Figure 6 shows the assembled state of ST Optical jumper cord(ST/PC-SC/PC).

p. 7

Page 8: ST Optical Jumper Cord(Simplex)fascoinfrasistema.com.mx/modulos/productos/pdf/ST Optical Jumpe… · Permissible deviation in dimensions without tolerance indication is in accordance

ww

w.f

asc

o.c

om

.mx

Paseo de la Reforma No. 295 Piso Mezzanine Colonia Cuauhtémoc Delegación Cuauhtémoc Ciudad de México, México CP. 06500

(55) 5533 8246 [email protected]

Figure T shows the assembled state of ST Optical jumper cord(ST/PC-FC/PC).

Figure U shows the assembled state of ST Optical jumper cord(ST/PC-iC/PC).

p. 8

Page 9: ST Optical Jumper Cord(Simplex)fascoinfrasistema.com.mx/modulos/productos/pdf/ST Optical Jumpe… · Permissible deviation in dimensions without tolerance indication is in accordance

ww

w.f

asc

o.c

om

.mx

Paseo de la Reforma No. 295 Piso Mezzanine Colonia Cuauhtémoc Delegación Cuauhtémoc Ciudad de México, México CP. 06500

(55) 5533 8246 [email protected]

Figure V shows the assembled state of ST Optical jumper cord(ST/PC-jU/PC).

p. 9