Electrical Tissue Property Imaging Jin Keun Seo Medical Imaging Lab. Computational Science & Engineering Yonsei University KAIST April 12, 2012

Transcript of Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property...

Page 1: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Electrical Tissue Property Imaging

Jin Keun Seo

Medical Imaging Lab.Computational Science & Engineering

Yonsei University

KAISTApril 12, 2012

Page 2: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Electrical Tissue Property Imaging: conductivity σ andpermittivity ε

Develop a robust system

S X = b subject to ...

X = σ + iωε to be imaged. Free to choose ω.b : measured dataS is a sensitivity matrix made by Maxwell’s equations. (Key: Needto build up a robust sensitivity matrix S satisfying RIP condition.)

Page 3: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

About X: Conductivity σ and Permittivity ε

J = σ E, E = −∇u

I = V/R = V σS/L

(V = |∇u|L, I = |J|S)

J = iωε EI = iωC V

(C = Q/V = εS/L)

In free space, ε0 ≈ 8.85× 10−12 and σ = 0.

Page 4: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

The system S (X) = b should be based on Ohm’s law:

J = (σ + iωε)E = −τ∇u (τ = σ + iωε)

∗ Admittivity τ of biological tissues may be anisotropic at a low frequency, but it maybecome an isotropic as ω increases.

Extra-CellularFluid

-Cl+Na

-Cl+Na

+Na -Cl

+ + + + + + +

+ + + + + + +

_ _ _ _ _ _ _

_ _ _ _ _ _ _~

+

v(t)

Cell Membrane

Intra-CellularFluid

sin) ( )(i t I tω=

, /V S LIτ τ==J E is scalarτ11 12 13

21 22 23

31 32 33

τ τ ττ τ ττ τ τ

τ⎡ ⎤⎢ ⎥= ⎢ ⎥⎢ ⎥⎣ ⎦

Page 5: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

About X: Admittivity τ of biological tissue changes with the angular fre-quency ω

[Gabriel S et al 1996b Phys. Med. Biol., Haemmerich et al 2003 Phys. Med. Biol.]

Page 6: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Making S X = b requires to produce J inside the body Ω.

Inject current (dc / ac) Induce current (ac only)E+

E− electrodes

Ω

τ, J, E, H

x

y

· zcoil

I sinωt

Ω

τ, J, E, H

x

y

· z

∇ · J = ∇ · (τE) = 0n · J|∂Ω ≈

I|E±|

(χE+ − χE−

) −∇2H = ∇ ln τ ×∇× H− iµ0ωτH∇× A = µ0H, ∇ · A = 0, −∇u = E + iωAn · (τ(∇u + iωAs))|∂Ω ≈ −n · (iωτAp)

∣∣∂Ω

∗ J, E, H are time-harmonic current density, electric field and magnetic fluxdensity.

∗ As : secondary magnetic potential & Ap : primary magnetic potential

∗ In R3 \ Ω, σ = 0, ε0 = 8.85× 10−12 and µ0 = 4π × 10−7.

Page 7: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

To set up S (X) = b

∇× E = −iωµH∇×H = τE, ∇ ·H = 0

∃ A s.t. µH = ∇× A & ∇ · A = 0∃ u s.t. −∇u = E + iωA

∇ · (τ∇u) = iω∇τ · A & −∇2H = ∇ ln τ × (∇×H)− iωµτH

τE(r)︸ ︷︷ ︸J

+ iωτA︸ ︷︷ ︸Jeddy

= −(σ∇ur − ωε∇ui)− i(ωε∇ur + σ∇ui)︸ ︷︷ ︸Jtotal = −τ∇u

−σ∇ur: sum of conduction current bydc or ac-E & conduction eddy currentby ac-H induced by ac-D

−ω2ε∇ur: sum of D by ac-E anddisplacement eddy current by ac-Hinduced by ac-D

ωε∇ui: sum of D by ac-E induced byac-H

−σ∇ui: the sum of conductioncurrent by ac-E induced by ac-H andthe conduction eddy current by ac-Hinduced by ac conduction current.

Page 8: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

About b. What is measurable quantity?

Inject current (dc / ac) Induce current (ac only)E+

E− electrodes

Ω

τ, J, E, H

x

y

· zcoil

I sinωt

Ω

τ, J, E, H

x

y

· z

∇ · (τ∇u) = iω∇τ · A & −∇2H = ∇ ln τ × (∇×H)− iωµτH

EIT (≤ 1MHz): Boundary voltage u|∂Ω using electrodes.MREIT (≤ 1kHz): Internal Hz using MRIMIT (≤ 10MHz): External magnetic field using coilsMREPT (128MHz at 3T MRI): Internal H+ = 1

2(Hx + iHy) using MRI

Page 9: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Electrical ImpedanceTomography

S X = b

Page 10: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

EIT system: S (X) = b.

Reciprocity Principle

uP(Q+, ω, t)− uP(Q−, ω, t) =1I

∫Ωτ∇uP · ∇uQdr = uQ(P+, ω, t)− uQ(P−, ω, t)

Here, ∇ · (τ∇uP) = 0 in Ω with BC n · (τ∇uP)|∂Ω = I(δ(· − P+)− δ(· − P−)).

ε1

ε4ε5

ε6

ε7

ε10

ε11

ε12 ε13

ε14

ε15

ε16

τ(r, ω, t)

Ω

i(t) = I sin(ωt)︸ ︷︷ ︸Nuemann data

P+

P−

Q+

Q−

v

i

stream and equipotential lines for 16 ch.phantom

uP(Q+, ω, t)− uP(Q−, ω, t)︸ ︷︷ ︸Dirichlet data

Calderon’s problem:Reconstruct τfrom the NtD datauP|∂Ω : P± ∈ ∂Ω

> Note that t and t are different. The time t is usedfor the time change of the impedance, while t isrelated to ω.

Page 11: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Structure of S X = b.

X = δσ : dynamic conductivity imagingApply Neumann data:∇ ·(σ(r)∇uPj(r)

)= 0 in Ω

σ(r) ∂∂n uPj(r) = I(δ(r− Pj)− δ(r− Pj+1)) on ∂Ω

Measure Dirichlet Data (V1, · · · ,VnE ):k − th comp. of Vj = Vj,k[σ] = uPj(Pk)− uPj(Pk+1).

Denote uj = uPj . Let σ0 be a background conductivity.Vj,k[σ]− Vj,k[σ0] = −

∫Ω(σ − σ0)∇uj · ∇u0

k dr.: Data

Page 12: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Structure of S X = b.

Discretizing Ω into np elements as Ω = ∪npn=1qn,

∫q1∇uj · ∇u(0)

1 dr · · ·∫

qnp∇uj · ∇u(0)

1 dr...

......∫

q1∇uj · ∇u(0)

nE dr · · ·∫

qnp∇uj · ∇u(0)

nE dr

︸ ︷︷ ︸

Sj

X = Vj[σ]− Vj[σ0]︸ ︷︷ ︸data

S1

· · · · · · · · ·...

· · · · · · · · ·SnE

︸ ︷︷ ︸

S

δσ1...

δσnp

︸ ︷︷ ︸

δσ

=

δV1· · ·...· · ·δVnE

︸ ︷︷ ︸

δV

Page 13: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Structure of S X = b.

=Sensitivity matrix Data vector

Conductivity to be imaged

Page 14: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Structure of S X = b.

Column vector⇔ pixel sensitivity of all applied currents.Row vector: sensitivity distribution for a fixed current-voltage. Leftfigures are eigenvectors of S∗S.

=

DataPixel position

−0.1

−0.05

0

0.05

0.1

−0.15

−0.1

−0.05

0

0.05

0.1

0.15

−0.08

−0.06

−0.04

−0.02

0

0.02

0.04

0.06

0.08

Basis of row space via SVD

Column vector <-> DataCheck RIP condition: columns

Page 15: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Is the linearized EIT robust?

S X = δV ⇒ X = (S∗S)−1S∗δV︸ ︷︷ ︸+ tSVD or regularization

where

S =

S1...

SnE

, X =

δσ1...

δσnp

, δV =

δV1...

δVnE

.

Here, Sj is the j-th block of the linearized sensitivity matrix:

Sj =

q1∇u(0)

j · ∇u(0)1 dr · · ·

∫qnp∇u(0)

j · ∇u(0)1 dr

......

...∫q1∇u(0)

j · ∇u(0)nE dr · · ·

∫qnp∇u(0)

j · ∇u(0)nE dr

.

Page 16: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Solving S X = b.

=Sensitivity matrix Data vector

Conductivity to be imaged

• In LM, we try to find a best linearcombination of the column vectorsof the sensitivity matrix whichproduces the data such that | |S1

j · · · Snpj

| |

...δτ...

= Dataj.

The reconstructed image of the LM relies roughly on tSVD

S ≈ Ut0Λt0V∗t0

and the expected reconstructed image is

X =

t0∑t=1

1λt〈Dataj,ut〉vt

Page 17: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Eigenvectors of S∗S and image X.

ill-posedness: low sensitivity to local perturbation at inner pixelaway from ∂Ω.X = (S∗S)−1S∗b : low resolution for inner region.

=

DataPixel position

≫ ≫

λ≈ 0

=

Page 18: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

About X = (S∗S)−1S∗b.

Check Restricted Isometry Property (RIP) condition: S is said tohave RIP of order m if ∃ δm ∈ (0, 1) s.t.

(1− δm)‖X∥∥2

2 ≤ ‖ SX ‖22 ≤ (1 + δm)‖X‖2

2, ∀‖X‖0 ≤ m

Mutual coherence depends mainly on pixel size (or resolution).] of electrodes may not help much for increasing mutualincoherence. Theoretical results of Calderon problem may not beapplied to practical setting (like Riemann mapping theorem).

=

DataPixel position

−0.1

−0.05

0

0.05

0.1

−0.15

−0.1

−0.05

0

0.05

0.1

0.15

−0.08

−0.06

−0.04

−0.02

0

0.02

0.04

0.06

0.08

Basis of row space via SVD

Column vector <-> DataCheck RIP condition: columns

Page 19: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Sensitivity matrix structure.

Data =∫

Ω τ ∇uPj · ∇uPk dr

=

∫near∂Ω

τ ∇uPj · ∇uPk︸ ︷︷ ︸high

sensitivity

dr +

∫innerregion

τ ∇uPj · ∇uPk︸ ︷︷ ︸very low

sensitivity

dr

InteriorRegion

InteriorRegion

ii

v

v

Static EIT data depends mainly on conductivity distribution

near boundary.

Page 20: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Examples - Boundary influence in EIT-data

In the case when δτ 6= 0 near ∂Ω, it is difficult to probe internalchanges because the data depends mainly on δτ near ∂Ω.

σ=1

σ=2

σ=1

σ=2

Model Reconstructed image

δτ=0

δτ=0

δτ=1

δτ=1

Page 21: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Achievable spatial resolution: N-channel EIT

Imagine that σ is the true conductivity. Define

Aσj := φ ∈ C1(R) : φ′ > 0, φ(uPjσ (Pk)) = uPj

σ (Pk), k = 1, 2, · · · ,N

where uPjσ is the solution of

∇ · (σ∇uPj

σ ) = 0 in Ω

σ ∂∂n uPj

σ |∂Ω = δ(· − P+j )− δ(· − P−j )

.

Observation. Every element of∩N

j=1σ

φ′(uPjσ )

: φ ∈ Aσj are solutions.

(Achievable resolution of N-channel EIT is somehowrelated to this figure and sensitivity.)

It appears that we can increase EIT-resolution byincreasing the number of electrodes. However,when ∂Ω is covered mostly by electrodes, EITcannot probe internal conductivity. (Why?) How can we improve EIT-image quality? Needcareful arrangement of electrodes.

Page 22: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Difference EIT for eliminating boundary influences

To extract internal impedance information from EIT data, we need to get rid of theboundary influences. The use of difference data enable us to cancel out the boundaryinfluences.

Differencedata

=

∫near∂Ω

δτ∇ua · ∇ub dr

︸ ︷︷ ︸Hope δτ=0

+

∫interior

region

δτ∇ua · ∇ub dr

(ua =

∑ajuPj , ub =

∑bjuPj

)Time difference EIT (tdEIT) : ∂

∂t Data −→ ∂∂tτ(r, ω, t)

Frequency difference EIT (fdEIT) : ∂∂ωData −→ ∂

∂ω τ(r, ω, t)

Recovering δτ from the difference data requires to control thesensitivity part ∇ua · ∇ub.

Page 23: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

What is visible for static electrical conducting bodyusing currently available EIT system?

Images provided by frequency difference EIT.

Sample

500 Hz 5 kHz 50 kHz 500 kHz

membranes

500 Hz

5 kHz

50 Hz

Page 24: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Functional Analysis Approach for SX = b.

In Factorization Method, we measure an unchangability index ateach pixel in the imaging domain by estimating a difficulty ofproducing n-th column of sensitivity matrix by a linear combination ofdata-matrix such that

Snj =

| |Data1 · · · DatanE

| |

hn

1...

hnnE

.

=Sensitivity matrix Data vector

Conductivity to be imaged

Page 25: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Sensitivity matrix based Factorization (S-FM).

Snj ⇔

| |Data1 · · · DatanE

| |

Square matrix

Page 26: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Rationale of S-FM

For characterizing Sj, we introduce φnj satisfying

∇ · ∇φnj = ∇ · (χqn∇u(0)

j ) in Ω

∇φnj · n = 0 on ∂Ω

Theorem: Denote Φnj := (Φn

j (1) · · · Φnj (nE))T where

Φnj (k) := φn

j (Pk)− φnj (Pk+1) for k = 1, · · · , nE.

Then Φnj = n-th column of Sj.

Proof. From the Neumann boundary condition of u(0)k ,∫

Ω∇φn

j · ∇u(0)k dr = φn

j (Pk)− φnj (Pk+1)

On the other hand, it follows from the property of φnj that∫

Ω∇φn

j · ∇u(0)k dr =

∫qn

∇u(0)j · ∇u(0)

k dr.

Page 27: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Main Observation

Let δσ = χD where D ⊂⊂ Ω. Replace n2E-dimensional data vector δV

by the nE × nE-matrix

δV := (δV1 · · · δVnE ).

For all pixel qn and j-th injection current

Φnj · (δV)−1Φn

j ≥ maxh∈RnE

∣∣∣∫qn∇u(0)

j · ∇u(0)h dr

∣∣∣2∫D |∇u(0)

h |2dr

where u(0)h (r) :=

∑nEk=1 hk u(0)

k (r).If the pixel qn lies inside the inclusion D then for all injectioncurrents ∣∣Φn

j · (δV)−1Φnj

∣∣ ≤∫

qn

∣∣∣∇u(0)j

∣∣∣2 dr.

Page 28: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Projective EIT system for higher resolution image ofthe conductivity distribution near the surface Γ ⊂ ∂Ω.

Two pairs of drivingelectrodes E±1 and E±2 toinject currents.

Many sensing electrodes tomeasure detailed voltageson ∂Ω

Experiments showed thatEIT-standard reconstructionalgorithms fail to providegood images.

Sensing Electrodes

: EJ Lee, Muugii, Woo, Seo (2011)

Page 29: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Image Υt: a projected tdEIT image of γt/γ

∇2SΥt(r) = ∇S ·

∫D∇2

SΦ(r, r′)[J1,2(r)]†J1,2(r′)∇(γt(r′)γ(r′)

)dr′︸ ︷︷ ︸

measurable quantity

, for r ∈ Γ

> ∇S is surface gradient. J1,2 =

(J1

x J1y J1

z

J2x J2

y J2z

), [J1,2]† is the pseudo-inverse.

Simulation

(a) Front view (b) Side view (c) Back view30 cm

27 cm

20 cm

13x11 cm

2x2 cm

1 cm

Front view

Side view

4 cm

backgroundγ = 1

γ = 8at t = 0

γ = 7at t = 0.1

γ = 1.3at t = 0

γ = 1.1at t = 0.1

Reconstructed image with thresholding0.05

-0.07-0.07

0.05

1x10-3 1x10-3

Page 30: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Breast EIT-Frequency difference projected imageProjection image[

∂φ log(γ∗)1

sinφ∂θ log(γ∗)

]= −

[∂φu1 1

sinφ∂θu1

∂φu2 1sinφ∂θu2

]−1 [∇2

s u1

∇2s u2

]> where γ∗ is function defined on Ωs satisfying γ∗|∂Ωs = γ0

Model & Simulation

Sensing Electrodes

Front Side Back

Page 31: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Electrical tissue property imaging using MRI:τ = σ + iωε

J = (σ + iωε)E & −∇2H =∇γγ× (∇×H)− iωµ0γH

EIT

MREIT

MREPT

Page 32: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREIT-way of feeling admittivity (τ ≈ σ since ωε ≈ 0)

Biot-Savart Law: (z = (0, 0, 1)T )

Hz(r)︸ ︷︷ ︸Data

=1

∫Ω

< r− r′,−

target︷ ︸︸ ︷σ(r′)∇u(r′)× z >|r− r′|3

dr′ + Harmonic term

Experiment Setting data target σAttaching ElectrodezH

Ω

∇ · (σ∇u) = 0 in Ω, n · (σ∇u)|∂Ω ≈ IE±

(χE+ − χE− )

Page 33: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

How to measure Bz. (1989, Joy et al, Toronto Group)

Using the MRI scanner, we obtain the following complex k-space datainvolving Bz information in the slice Ωz0 :

SI+(kx, ky) =

∫ ∫Ωz0

M(x, y, z0)ei(γBz(x,y,z0)Tc+δ(x,y,z0))ei(xkx+yky)dxdy

Application of Fourier transform to the k-space MR signal SI+ yields thefollowing complex MR image,MI+(x, y, z0):

MI+(x, y, z0) := M(x, y, z0) eiγBz(x,y,z0)Tc eiδ(x,y,z0).

RecessedCopper Electrode

Saline

Agar

RecessedCopper Electrode

(a)M(x, y, z0) (b)wrapped Bz (c) Bz

Page 34: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREIT using full components of H = (Hx,Hy,Hz): It

requires subject rotation inside MRI scanner. (Impractical)

Least square method(1994)

minσ

12‖∇ ×H + σ∇u ‖2

Finite element model True

1/σ

1/σ 1/σReconstructed

Poor spatial Resolution

[Zhang 1992], [Woo, Lee, Moon 1994], [Idel and Birgul

1995]

J−substitution method(2002)

∇ ·(|∇ ×H||∇u|

∇u)

= 0

True Reconstructed 1/σ 1/σ

High spatial resolution

[Kwon, Seo, Yoon, Woo 2002], [Hasanov, Ma, Nachman, Joy

2008]

Page 35: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Harmonic Bz Algorithm: Key Observations

Hz, z−component of the magnetic field H = (Hx,Hy,Hz) can trace achange of σ in the tangential direction to the equipotential surface:

∇2Hz = ∇lnσ · (σ∇u× z)

where Hz(r) = 14π

∫Ω

〈r−r′,−σ(r′)∇u(r′)×z〉|r−r′|3 dr′ + harmonic term

Hz is blind to the transversal change of σ in the direction ∇xyu.

To uniquely reconstruct σ from Bz, at least two currents is needed.

[Seo, Kwon, Yoon, Woo 2003 IEEE TBE], [Seo, Woo SIAM REVIEW 2011]

Page 36: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Harmonic Bz Algorithm

For any (x, y, z0) ∈ Ωz0 := (x, y, z) ∈ Ω : z = z0, the harmonic Bzalgorithm is based on

∇2xy lnσ(x, y, z0) = ∇xy ·

[ σ ∂u1[σ]∂y −σ ∂u1[σ]

∂x

σ ∂u2[σ]∂y −σ ∂u2[σ]

∂x

]−1 [∇2Hz,1∇2Hz,2

]

0.5

1

1.5

2

2.5

3

3.5

4

0.5

1

1.5

2

2.5

3

3.5

Canine Pelvis Experiment Human Knee Experiment

[Seo et al 2003 IEEE Trans. Boimed. Eng., Oh et al 2003 Phys. Med. Biol.]

Page 37: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREIT: Animal imaging experiment

Harmonic Bz-algorithm requires to solve the forward model in which weneed to know Ω and Neumann data (semi-)automatically.

Domain Ω: We use level set method or snake to extract ∂Ω fromMR-image automatically.Neumann boundary data: Very difficult to get automaticalgorithm to get Neumann condition. Use semi-automatic methodwith Level set.

Electrode

Lead Wire RF Coil

MRI Bore

B0x

y

z

Page 38: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

CoReHa: Software for MREIT

MREIT Matlab Toolkit(2006: C. J. Park, S. H. Lee, Kwon, Woo, Seo),CoReHA(Conductivity Reconstructor using Harmonic Algorithm 2008: G.W. Jun, C.O. Lee, Woo, Seo)

Page 39: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREIT: CoReHa

Segmentation Electrode Modeling Meshing(Level set method) (Neuman data) (FEM)

Denoising/HarmonicInpainting

[G.W. Jun, C. O. Lee, Woo, Seo 2008]

Page 40: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREIT: Animal imaging

Kidney

Spinal Cord

Liver

Spleen

Stomach

IntestinesCarbon-hydrogel

Electrode

Page 41: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREIT: Human imaging

FDA safety limit: 1mA. Reducing the amount of current⇒ LowSNR data⇒ Denoising process⇒ produce artifacts

Page 42: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Coefficient of Elliptic PDE in terms of Neumann data:macroscopic-view

The irresponsible (or precise) expression would be

∇ · ((1 + (10−8 − 1)χC)∇u) = 0 in Ω C := (thin film with holes)

M R I I m a g e

Saline

V e r y T hi n I ns ul a t i ng F i l m ha v i ng s m a l l ho l e s

Page 43: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Apparent conductivity & Homogenization

Considering the overall structure of J = −σ∇u due to Neumann datadeeply, a responsible expression is

∇ · (σ∇u) = 0 in Ω σ =

σ1 inside OHP filmσ2 outside OHP film

because the coefficient of elliptic PDE can be viewed as a bestobservable quantity σ which minimizes

Φ(σ) = supg∈A‖Jg + σ∇ug‖L2(Ω) + λ‖∇σ‖BV A ⊂ H−1/2(∂Ω)

Page 44: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Coefficient of elliptic PDE observed by MREIT

The apparent conductivity by MREIT changes with the size of holesize:

∇ · (σ∇u) = 0 in Ω σ =

σ1 inside OHP filmσ2 outside OHP film

Bz data & Reconstructed Conudctivity distribution using MREIT

Ø of hole Without hole Ø 1mm Ø 2mm Ø 3mm Ø 4mm Ø 5mm Ø 6mm Ø 10mm

Conductivity change(%) 0 0.94 1.39 1.86 2.36 2.92 3.14 4.35

Page 45: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREPT-way of feeling Admittivity (τ = σ + iωε, ω2π ≥ 100 MHz)

H+ = 12(Hx + Hy) provide information of τ = σ + iωε through

−∇2 H+(r)︸ ︷︷ ︸Data

=12

((x + iy)× (∇×H(r))) · ∇ ln τ︸︷︷︸target

− iωµ0 τ︸︷︷︸target

H+(r).

Experiment Setting ( )H +ℜ ( )H +ℑ Target σ

[Haacke1991]

Page 46: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Measurable quantity for MREPT

According to Faraday’s law and reciprocity principle [Hoult2000], the inducedRF-signal at the coil C is

ξ︸︷︷︸RFsignal

:=

∮C

EM · d` = −iωµ0

I

∫Ω

H(r) ·M(r) dr

RF excitation with the main field B0 = −B0z generates time-harmonic field H(called B1-field) that is influenced by τ = σ + iωε.

During RF excitation, M(r, t) precess according to Bloch equation∂∂t M(r, t) = M(r, t)× γ(B0 + B1(r, t)).

If we turn off RF field H, M (M(r, t) = <eiω0tM(r)) creates HM and EM that aredictated by ∇× EM = −iω0µ0(HM + M), ∇×HM = τEM

Page 47: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Measurable quantity for MREPT-continue 2

According to Faraday’s law and reciprocity principle, the inducedRF-signal at the coil C in the the presence of the magnetization M is

ξ︸︷︷︸RFsignal

:=

∮C

EM · d` = −iωµ0

I

∫Ω

H(r) ·M(r) dr

This reciprocity is based on

ξ =1I

∫R3

EM · Jcdr =

∫R3

EM · (∇×H− τE)dr

= −iω0µ0

∫R3

M ·Hdr +

∫R3∇ · (EM ×H− E×HM)dr

= −iω0µ0

∫R3

M ·Hdr + 0

where Jc is the filamentary current flow along the coil.[Scott 2003] [Hoult2000]

Page 48: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Measurable quantity for MREPT-continue 3

NMR RF signal ξ can be expressed as

ξ = 2C1

∫Ω

M0(r)H−(r)

(sin(C2α|H+(r)|

) H+(r)

|H+(r)|

)dr

The transverse field Hxy = Hxx + Hyy can be decomposed intoHxy =

Hx−iHy2 a+ +

Hx−iHy2 a− = H+a+ + H−a− where a+ = x− iy and a− = x + iy.

The transversal component Mxy = Mxx + Myy interact with H+ components and itcan be approximated by Mxy(r) ≈ C1M0(r)

(sin(C2α|H+(r)|

) H+(r)|H+(r)|

)a+

Since a− · a− = 0 and a− · a+ = 2, the above identity follows from

ξ︸︷︷︸RFsignal

:=

∮C

EM · d` = −iωµ0

I

∫Ω

H(r) ·M(r) dr

Page 49: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Measurable quantity for MREPT-continue 4

For each flip angle α > 0, we can measure the following signal:

ξα(r) := 2C1M0(r)H−(r)

(sin(C2α|H+(r)|

) H+(r)

|H+(r)|

)

One simple method to extract |H+(r)| would be to use ξα(r) and ξ2α(r) to get

ξ2α(r)

ξα(r)=

sin(2C2α|H+(r)|

)sin (C2α|H+(r)|) = 2 cos

(C2α|H+(r)|

)Assuming H+(r)

|H+(r)| ≈H−(r)|H−(r)| , we can extract H+(r)

|H+(r)| by

12

Argξα(r) = Arg

H+(r)

|H+(r)|

Page 50: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREPT Reconstruction Method

−∇2H =∇γγ× (∇×H)︸ ︷︷ ︸

refraction term

−iωµ0γH

Wen (2003) uses the assumption of local homogeneity of τ toget

τ(r) =i

ωµ0

∇2H+(r)

H+(r)

Katscher et al (2009) performed initial experiments on a standardclinical MRI system: For any disk Dδ(r0) ⊂ Ω where ∇τ ≈ 0,

τ =

∮∂Dr∇×H · d`

iµ0ω∫

DrH · dS

.

Nachman et al.(2008) develop τ = ∇2H·(∇×H)iωµ0H·(∇×H) .

Page 51: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREPT for locally homogeneous subject (∇τ ≈ 0)

For any vector a,

−∇2H = ∇ ln τ ×∇×H− iµ0ωτH ⇒ τ =1

iωµ0

a · ∇2Ha ·H

This direct formula can not be reliable at frequencies below 10MHz.Why? [Wave number] ×|subject size| 1 ⇒ ∇2H ≈ 0 in Ω.

Here, diam(Ω) = 1m, σ = 0.1, εr = 100.

Page 52: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREPT Reconstruction

Conductivity Permittivity

[Katscher et al (2009)]

Page 53: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

MREPT - Error analysis using the direct reconstructionformula τ ∗ = 1

iωµ0

∇2H+

H+ (Note: τ = τ∗ when ∇τ = 0)

Error = τ − τ∗ =

(1

iωµ0

∇2H+

H+− ∇

2H−

iωµ0H−

)[1−

H+ ∂∂z H−

H− ∂∂z H+

]−1

where H = (H+ + H−,−iH+ + iH−,Hz).

MREPT

MREIT

Phantom Model Reconstruction Horizontal axis Vertical axis

[Seo, D H Kim, E J Woo, ... IEEE TMI 2012]

Page 54: Electrical Tissue Property Imaging - Yonsei University · 2014-12-30 · Electrical Tissue Property Imaging: conductivity ˙and permittivity Develop a robust system S X = b subject

Dual Frequency: MREIT vs EPT

MREIT provides σ at frequencies below a few kHz whereas MREPTproduces σ + iωε at the Larmor frequency (about 128 MHz at 3 T)using H+ = 1

2(Hx + iHy) and

−∇2H =∇ττ× (∇×H)− iωµ0τH

−3

−2

−1

0

1

2

3

C = 0.129 ± 0.242B = 3.019 ± 0.529A = 0.901 ± 0.480

C = 0.122 ± 0.200B = 3.210 ± 0.520A = 1.146 ± 0.505

−3

−2

−1

0

1

2

3

0.8

0.9

1.0

1.1

1.2

[S/m] [S/m]0.6

0.8

1.0

1.2

1.4