Download - SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

Transcript
Page 1: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

SCOPE SERVICES OF ISO/IEC 17025

ELECTROMAGNETIC – DC/LOW FREQUENCY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

DC Voltage - Source 0 mV (Shorted) >0 mV to < 330 mV 330 mV to < 3.3 V

3.3 V to < 33 V 33 V to < 330 V 330 V to 1020 V

50 V to 1 kV >1 kV to 6 kV >6 kV to 10 kV

0.070 μV 24 μV/V+2.0 μV 14 μV/V+2.8 μV 15 μV/V+24 μV

22 μV/V+0.18 mV 22 μV/V+1.8 mV 3.7 mV/V +2.5 V 3.6 mV/V +2.5 V 3.5 mV/V +2.6 V

DC Voltage - Measure 0 mV to < 200 mV 200 mV to < 2 V

2 V to < 20 V 20 V to < 200 V 200 V to 1000 V

0 V to 15 V > 15 V to 30 V > 30 V to 60 V > 60 V to 150 V > 150 V to 300 V > 300 V to 600 V

8.5 μV/V+0.71 μV 7.1 μV/V+1.1 μV 7.1 μV/V+6.3 μV 12 μV/V+76 mV

12 μV/V+0.63 mV 2.4 mV/V+54 mV 2.4 mV/V+0.11 V 2.4 mV/V+0.21 V 2.4 mV/V+0.54 V 2.4 mV/V+1.1 V 2.4 mV/V+2.1 V

Page 2: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

DC Voltage – Measure (Cont.) 100 V to 1 kV > 1 kV to 6 kV > 6 kV to 10 kV 1 kV to <20 kV 20 kV to 35 kV

>35 kV to 40 kV

3.7 mV/V+2.5 V 3.6 mV/V+2.5 V 3.5 mV/V+2.6 V 20 mV/V+0.11 V 10 mV/V+0.11 V 20 mV/V+0.11 V

DC Current – Source 0 mA (Opened) > 0 mA to < 0.33 mA 0.33 mA to < 3.3 mA 3.3 mA to < 33 mA 33 mA to < 330 mA 330 mA to < 1.1 A

1.1 A to < 3 A 3 A to < 11 A 11 A to 20.5 A

0.58 nA 0.18 mA/A+24 nA 0.12 mA/A+59 nA

0.12 mA/A+0.30 μA 0.12 mA/A+3.1 μA 0.24 mA/A+47 μA 0.44 mA/A+48 μA

0.58 mA/A+0.91 mA 1.2 mA/A+1.2 mA

DC Clamp-On Ammeters 20.5 A to < 150 A 150 A to 1025 A

5.8 mA/A+0.21 A 5.9 mA/A+1.5 A

DC Current - Measure

0 μA to < 200 μA 200 μA to < 2 mA 2 mA to < 20 mA

20 mA to < 200 mA 200 mA to < 2 A

0.12 mA/A+5.8 nA 0.12 mA/A+7.5 nA 0.12 mA/A+75 nA 0.12 mA/A+1.3 μA 0.24 mA/A+25 μA

0 A to 0.5 A > 0.5 A to 1 A > 1 A to 2 A > 2 A to 5 A > 5 A to 10 A

> 10 A to 20 A

2.4 mA/A+2.9 mA 2.4 mA/A+6.0 mA 2.4 mA/A+7.3 mA 2.4 mA/A+18 mA 2.4 mA/A+37 mA 2.4 mA/A+76 mA

Page 3: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

DC Current – Measure (Cont.)

0 A to 30 A >30 A to 50A

>50 A to 100A >100 A to 500 A >500 A to 1000A

3.5 mA/A 2.4 mA/A 2.4 mA/A 2.4 mA/A 5.8 mA/A

DC/AC cutoff current - Measure DC: 0 mA to 100 mA AC:@ 50 Hz, 60 Hz 0.02 mA to 100 mA

2.0 mA/A 7.5 mA/A

AC Voltage – Source @ 10 Hz to 45 Hz 1 mV to < 33 mV

33 mV to < 330 mV 0.33 V to < 3.3 V 3.3 V to < 33 V

@ > 45 Hz to 10 kHz 1 mV to < 33 mV

33 mV to < 330 mV 0.33 V to < 3.3 V 3.3 V to < 33 V

@ > 10 kHz to 20 kHz 1 mV to < 33 mV

33 mV to < 330 mV 0.33 V to < 3.3 V 3.3 V to < 33 V

0.95 mV/V+7.1 μV 0.37 mV/V+9.6 μV 0.35 mV/V+60 μV

0.36 mV/V+0.77 mV

0.23 mV/V+7.1 μV 0.19 mV/V+9.5 μV 0.18 mV/V+70 μV

0.18 mV/V+0.71 mV

0.28 mV/V+7.1 μV 0.21 mV/V+9.5 μV 0.23 mV/V+70 μV

0.28 mV/V+0.71 mV

Page 4: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

AC Voltage – Source (cont.)

@ > 20 kHz to 50 kHz 1 mV to < 33 mV

33 mV to < 330 mV 0.33 V to < 3.3 V 3.3 V to < 33 V 33 V to < 330 V

@ > 50 kHz to 100 kHz 1 mV to < 33 mV

33 mV to < 330 mV 0.33 V to < 3.3 V 3.3 V to < 33 V 33 V to < 330 V

1.2 mV/V+7.1 μV 0.42 mV/V+9.5 μV 0.36 mV/V+60 μV

0.41mV/V+0.71 mV 0.36 mV/V+7.0 mV

4.1 mV/V+14 μV 0.94 mV/V+38 μV

0.82 mV/V+0.15 mV 1.1 mV/V+1.9 mV 2.4 mV/V+58 mV

@ > 100 kHz to 500 kHz 1 mV to < 33 mV

33 mV to < 330 mV 0.33 V to < 3.3 V @ 45 Hz to 1 kHz 33 V to < 330 V 330 V to 1020 V

@ > 1 kHz to 10 kHz 33 V to < 330 V

@ > 10 kHz to 20 kHz 33 V to < 330 V

@ > 1 kHz to 5 kHz 330 V to 1020 V

@ > 5 kHz to 10 kHz 330 V to 1020 V

9.3 mV/V+58 μV 2.4 mV/V+81 μV

2.9 mV/V+0.70 mV

0.23 mV/V+2.8 mV 0.35 mV/V+12 mV

0.24 mV/V+7.2 mV

0.30 mV/V+7.1 mV

0.30 mV/V+12 mV

0.35 mV/V+12 mV

Page 5: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

AC Voltage – Source (cont.) @ 50 Hz, 60 Hz 50 V to 1 kV

> 1 kV to 6 kV > 6 kV to 10 kV

6.2 mV/V+2.7 V 6.0 mV/V+2.5 V 5.9 mV/V+2.6 V

AC Voltage - Measure @ 20 Hz to 40 Hz 10 mV to < 200 mV

200 mV to < 2 V 2 V to < 20 V

20 V to < 200 V 200 V to 1000 V

@ > 40 Hz to 100 Hz 10 mV to < 200 mV

200 mV to < 2 V 2 V to < 20 V

20 V to < 200 V 200 V to 1000 V

@ > 100 Hz to 2 kHz 10 mV to < 200 mV

200 mV to < 2 V 2 V to < 20 V

20 V to < 200 V 200 V to 1000 V

@ > 2 kHz to 10 kHz 10 mV to < 200 mV

200 mV to < 2 V 2 V to < 20 V

20 V to <200 V 200 V to 1000 V

0.27 mV/V+4.7 μV 0.20 mV/V+25 μV

0.20 mV/V+0.24 mV 0.24 mV/V+4.7 mV 0.46 mV/V+50 mV

0.25 mV/V+4.7 μV 0.18 mV/V+25 μV

0.18 mV/V+0.24 mV 0.18 mV/V+2.5 mV 0.46 mV/V+28 mV

0.25 mV/V+2.5μV 0.16 mV/V+24 μV

0.16 mV/V+0.24 mV 0.16 mV/V+2.5 mV 0.69 mV/V+29 mV

0.25 mV/V+4.7 μV 0.18 mV/V+24 μV

0.18 mV/V+0.25 mV 0.18 mV/V+2.6 mV 0.69 mV/V+29 mV

Page 6: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS UNCERTAINTY(±)]

AC Voltage – Measure (cont.)

@ > 10 kHz to 30 kHz 10 mV to < 200 mV

200 mV to < 2 V 2 V to < 20 V

20 V to < 200 V 200 V to 1000 V

@ > 30 kHz to 100kHz 10 mV to < 200 mV

200 mV to < 2 V 2 V to < 20 V

20 V to < 200 V 200 V to 1000 V

@ > 100 kHz to 300 kHz 200 mV to < 2 V

2 V to < 20 V 20 V to < 200 V

@ > 300 kHz to 1 MHz 2 V to < 20 V

20 V to < 200 V

0.47 mV/V+9.3 μV 0.29 mV/V+47 μV

0.29 mV/V+0.47 mV 0.29 mV/V+4.7 mV 1.5 mV/V+24 mV

0.82 mV/V+24 μV

0.58 mV/V+0.24 mV 0.58 mV/V+2.4 mV 0.58 mV/V+24 mV 3.6 mV/V+24 mV

3.5 mV/V+2.4 mV 3.5 mV/V+1.2 mV 4.8 mV/V+12 mV

12 mV/V+1.2 mV 18 mV/V+12 mV

@ 45 Hz to 66 Hz 1 V to 15 V

> 15 V to 30 V > 30 V to 60 V

> 60 V to 150 V > 150 V to 300 V > 300 V to 600 V

@ > 66 Hz to 1 kHz 1 V to 15 V

> 15 V to 30 V > 30 V to 60 V

> 60 V to 150 V > 150 V to 300 V > 300 V to 600 V

1.8 mV/V+34 mV 1.8 mV/V+62 mV 1.8 mV/V+0.13 V 1.8 mV/V+0.34 V 1.8 mV/V+0.62 V 1.8 mV/V+1.3 V

3.5 mV/V+62 mV 3.5 mV/V+0.13 V 3.5 mV/V+0.25 V 3.5 mV/V+0.62 V 3.5 mV/V+1.3 V 3.5 mV/V+2.5 V

Page 7: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

AC Voltage – Measure (cont.) @ 50 Hz to 60 Hz 500 V to 1 kV > 1 kV to 6 kV > 6 kV to 10 kV

@ 50 Hz to 60 Hz 1 kV to <20 kV 20 kV to 35 kV

>35 kV to 40 kV

6.2 mV/V+2.7 mV 6.0 mV/V+2.5 V 5.9 mV/V+2.6 V

50 mV/V+0.11 V 50 mV/V+0.15 V 50 mV/V+0.20 V

AC Current - Source @ 10 Hz to 20 Hz 29 μA to < 330 μA 330 μA to < 3.3 mA 3.3 mA to < 33 mA 33 mA to < 330 mA @ > 20 Hz to 45 Hz 29 μA to < 330 μA 330 μA to < 3.3 mA 3.3 mA to < 33 mA 33 mA to < 330 mA @ > 45 Hz to 1 kHz 29 μA to < 330 μA 330 μA to < 3.3 mA 3.3 mA to < 33 mA 33 mA to < 330 mA 330 mA to < 1.1 A 1.1 A to < 3 A

2.4 mA/A+0.12 μA 2.4 mA/A+0.19 μA 2.1 mA/A+2.5 μA 2.1 mA/A+25 μA 1.8 mA/A+0.12 μA 1.5 mA/A+0.19 μA 1.1 mA/A+2.5 μA 1.1 mA/A+25 μA 1.5 mA/A+0.12 μA 1.2 mA/A+0.20 μA 0.48 mA/A+2.5 μA 0.48 mA/A+27 μA 0.59 mA/A+0.12 mA 0.70 mA/A+0.14 mA

Page 8: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

AC Current – Source (cont.) @ > 1 kHz to 5 kHz 29 μA to < 330 μA 330 μA to < 3.3 mA 3.3 mA to < 33 mA 33 mA to < 330 mA 330 mA to < 1.1 A

1.1 A to < 3 A 3 A to < 11 A 11 A to 20.5 A

@ > 5 kHz to 10 kHz 29 μA to < 330 μA 330 μA to < 3.3 mA 3.3 mA to < 33 mA 33 mA to < 330 mA 330 mA to < 1.1 A

1.1 A to < 3 A @ > 10 kHz to 30 kHz

29 μA to < 330 μA 330 μA to < 3.3 mA 3.3 mA to < 33 mA 33 mA to < 330 mA @ 10 Hz to 45 Hz 330 mA to < 1.1 A

1.1 A to < 3 A

3.5 mA/A+0.18 μA 2.4 mA/A+0.26 μA 0.93 mA/A+2.6 μA 1.2 mA/A+59 μA 7.0 mA/A+1.2 mA 7.0 mA/A+1.2 mA 35 mA/A+2.7 mA 35 mA/A+5.9 mA

9.3 mA/A+0.24 μA 5.8 mA/A+0.37 μA 2.4 mA/A+3.8 μA 2.4 mA/A+0.12 μA 2.9 mA/A+5.8 mA 2.9 mA/A+5.8 mA

19 mA/A+0.47 μA 12 mA/A+0.71 μA 4.7 mA/A+4.8 μA

4.7 mA/A+0.24 mA

2.1 mA/A+0.12 mA 2.1 mA/A+0.15 mA

@ > 45 Hz to 100 Hz 3 A to < 11 A 11 A to 20.5 A

@ > 100 Hz to 1 kHz 3 A to < 11 A 11 A to 20.5 A

0.71 mA/A+2.5 mA 1.4 mA/A+5.9 mA

1.2 mA/A+2.6 mA 1.8 mA/A+6.0 mA

Page 9: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

AC Clamp-On Ammeters @ 45 Hz to 65 Hz 20.5 A to < 150 A 150 A to 1025 A

@ > 65 Hz to 440 Hz 20.5 A to < 150 A

@ > 65 Hz to 100 Hz 150 A to 1025 A

6.7 mA/A+0.35 A 6.7 mA/A+1.4 A

13 mA/A+0.35 A

12 mA/A+1.4 A

AC Current – Measure @ 10 Hz to 5 kHz 10 μA to < 200 μA 200 μA to < 2 mA 2 mA to < 20 mA

20 mA to < 200 mA @ 10 Hz to 1 kHz

200 mA to < 2 A @ > 1 kHz to 5 kHz

200 mA to < 2 A

0.36 mA/A+24 nA

0.35 mA/A+0.24 μA 0.35 mA/A+2.4 μA 0.35 mA/A+24 μA

0.70 mA/A+0.47 mA

2.4 mA/A+0.93 mA

@ 45 Hz to 66 Hz 1 mA to 0.5 A > 0.5 A to 1 A > 1 A to 2 A > 2 A to 5 A > 5 A to 10 A

> 10 A to 20 A @ 66 Hz to 1 kHz

1 mA to 0.5 A > 0.5 A to 1 A > 1 A to 2 A > 2 A to 5 A > 5 A to 10 A

> 10 A to 20 A

1.8 mA/A+1.1 mA 1.8 mA/A+2.4 mA 1.8 mA/A+4.6 mA 1.8 mA/A+12 mA 1.8 mA/A+24 mA 1.8 mA/A+51 mA

3.5 mA/A+2.1 mA 3.5 mA/A+4.3 mA 3.5 mA/A+8.4 mA 3.5 mA/A+23 mA 3.5 mA/A+46 mA 3.5 mA/A+96 mA

Page 10: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

AC Current – Measure (Cont.)

@ DC to 1 kHz > 1 A to 30 A

@ > 1 kHz to 5 kHz > 1 A to 30 A

3.5 mA/A

5.8 mA/A

Resistance - Source Resistance ( 4 Wire ) 0 Ω (Shorted)

> 0 Ω to < 11 Ω 11 Ω to < 33 Ω 33 Ω to < 110 Ω

110 Ω to < 330 Ω 330 Ω to < 1.1 kΩ 1.1 kΩ to < 3.3 kΩ 3.3 kΩ to < 11 kΩ 11 kΩ to < 33 kΩ 33 kΩ to < 110 kΩ

Resistance ( 2 Wire ) 110 kΩ to < 330 kΩ 330 kΩ to < 1.1 MΩ 1.1 MΩ to < 3.3 MΩ 3.3 MΩ to < 11 MΩ 11 MΩ to < 33 MΩ 33 MΩ to < 110 MΩ 110 MΩ to < 330 MΩ 330 MΩ to 1100 MΩ

0.63 μ Ω

61 μ Ω / Ω +1.2 m Ω 53 μ Ω / Ω +1.8 m Ω 42 μ Ω / Ω +1.7 m Ω 35 μ Ω / Ω +2.4 m Ω 34 μ Ω / Ω +2.4 m Ω 34 μ Ω / Ω +2.4 m Ω 34 μ Ω / Ω +2.4 m Ω 34 μ Ω / Ω +0.24 Ω 34 μ Ω / Ω +0.24 Ω

38 μ Ω / Ω +7.8 Ω 39 μ Ω / Ω +3.8 Ω 71 μ Ω / Ω +48 Ω

0.16 m Ω / Ω +0.46 k Ω 0.30 m Ω / Ω +3.3 k Ω 0.59 m Ω / Ω +4.6 k Ω 3.5 m Ω / Ω +0.23 M Ω 18 m Ω / Ω +0.59 M Ω

Resistance – Source 0.1 to 1 Ω 1 to 10 Ω

10 to 100 Ω 100 to 1000 Ω

1 to 10 kΩ 10 to 100 kΩ

0.12 m Ω/Ω+2.0 mΩ 0.12 m Ω/Ω +2.0 mΩ 0.12 m Ω/Ω +2.0 mΩ 0.12 m Ω/Ω +2.0 mΩ

0.12 m Ω/Ω 0.12 m Ω/Ω

Page 11: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Resistance,- Source High Resistance,

Insulation Resistance,

Test voltage @ ± (10 to 5000) V 10 kΩ 100 kΩ 1 MΩ

5 MΩ 10 MΩ

20 MΩ 100 MΩ

1 GΩ 10 GΩ

0.18 kΩ 1.8 kΩ 18 kΩ 90 kΩ

0.18 MΩ 0.36 MΩ 1.8 MΩ

0.014 GΩ 0.19 GΩ

Surface resistivity/resistance meter, Electrostatic Resistance

Test voltage @ ± (Up to 100) V 1 MΩ 5 MΩ

10 MΩ 20 MΩ 100 MΩ

1 GΩ 10 GΩ

18 kΩ 90 kΩ

0.18 MΩ 0.36 MΩ 1.8 MΩ

0.014 GΩ 0.19 GΩ

Resistance - Measure 0 Ω to < 20 Ω 20 Ω to < 200 Ω 200 Ω to < 2 kΩ 2 kΩ to < 20 kΩ

20 kΩ to < 200 kΩ 200 kΩ to < 2 MΩ 2 MΩ to < 20 MΩ

Test voltage @ 10 V 20 MΩ to < 200 MΩ 200 MΩ to < 2 GΩ

18 μ Ω / Ω +27 μ Ω 13 μ Ω / Ω +0.12 m Ω 11 μ Ω / Ω +1.1 m Ω

11 μ Ω / Ω +12 m Ω 11 μ Ω / Ω +0.14 Ω 17 μ Ω / Ω +5.0 Ω

35 μ Ω / Ω +0.17 k Ω

0.35 m Ω / Ω +13 k Ω 3.5 m Ω / Ω +1.1 M Ω

Page 12: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Resistance-Measure Insulation Resistance

Test voltage @ up to 5 kV 1 k Ω to 2 k Ω

>2 k Ω to 20 k Ω >20 k Ω to 200 k Ω >200 k Ω to 2 M Ω >2 M Ω to 20 M Ω

>20 M Ω to 200 M Ω >200 M Ω to 2 G Ω >2 G Ω to 20 G Ω

>20 G Ω to 200 G Ω

18 μ Ω / Ω +27 μ Ω

13 μ Ω / Ω +0.12 m Ω 11 μ Ω / Ω +1.1 m Ω 11 μ Ω / Ω +12 m Ω 11 μ Ω / Ω +0.14 Ω 17 μ Ω / Ω +5.0 Ω

35 μ Ω / Ω +0.17 k Ω 0.35 m Ω / Ω +13 k Ω 3.5 m Ω / Ω +1.1 M Ω

Capacitance - Source @ 5 kHz 0.22 nF to < 0.40 nF

@ 1 kHz 0.40 nF to < 1.1 nF 1.1 nF to < 3.3 nF 3.3 nF to < 11 nF 11 nF to < 33 nF

33 nF to < 110 nF 110 nF to < 330 nF

6.1 mF/F+12 pF

6.2 mF/F+13 pF 5.9 mF/F+14 pF 3.0 mF/F+13 pF 3.0 mF/F+59 pF 3.0 mF/F+59 pF

3.0 mF/F+0.58 nF

Capacitance – Source (Cont.)

@ 100 Hz 330 nF to < 1.1 μF 1.1 μF to < 3.3 μF 3.3 μF to < 11 μF 11 μF to < 33 μF

@ 50 Hz 33 μF to < 110 μF

3.0 mF/F+1.4 nF 3.0 mF/F+6.8 nF 3.0 mF/F+13 nF 4.7 mF/F+68 nF

5.4 mF/F+0.13 μF

Page 13: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Capacitance - Source @ DC 110 μF to < 330 μF 330 μF to < 1.1 mF 1.1 mF to < 3.3 mF 3.3 mF to < 11 mF 11 mF to < 33 mF 33 mF to 110 mF

5.4 mF/F+0.68 μF 5.3 mF/F+1.3 μF 5.3 mF/F+6.8 μF 5.3 mF/F+13 μF 8.7 mF/F+68μF

13 mF/F+0.14 mF Test Frequency @ 120 Hz, 1 kHz, 10

kHz, 100 kHz, 1 MHz, 5 MHz,10 MHz,

13 MHz 1 pF 10 pF 100 pF 1 nF 10 nF 100 nF

1.2 fF 12 fF

0.12 pF 1.2 pF 1.2 pF 1.2 pF

Inductance - Source Test Frequency @ 120 Hz, 1 kHz, 10 kHz, 1 mH 10 mH

1.2 mH/H 1.2 mH/H

Capacitance - Measure @ 0.1 kHz to 10 kHz 10 pF to < 100 pF

100 pF to < 1000 pF 1 nF to < 10 nF

10 nF to < 100 nF 100 nF to < 1000 nF

1.2 mF/F 1.2 mF/F 1.2 mF/F 1.2 mF/F 1.2 mF/F

Page 14: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Capacitance – Measure (cont.) > 1 kHz to 100 kHz 100 pF to < 1000 pF

1 nF to < 10 nF 10 nF to < 100 nF

100 nF to < 1000 nF

1.2 mF/F 1.2 mF/F 1.2 mF/F 1.2 mF/F

Inductance - Measure @ 0.1 kHz to 10 kHz 0.01 μH to < 0.1 mH

0.1 mH to < 1 mH 1 mH to < 10 mH

10 mH to < 100 mH 100 mH to < 1 H

1 H to 10 H

1.2 mH/H 1.2 mH/H 1.2 mH/H 1.2 mH/H 1.2 mH/H 1.2 mH/H

DC Power - Source 1 W to < 336.6 W (33 mV to 1020 V,

0.33 mA to < 330 mA) 336.6 W to < 3.06 kW

(33 mV to 1020 V, 0.33 A to < 3 A)

3.06 kW to 20.91 kW (33 mV to 1020 V,

3 A to 20.5 A)

20.91 kW to < 153 kW (33 mV to 1020 V, 20.5 A to < 150 A) 153 kW to< 1.046 MW (33 mV to 1020 V, 150 A to < 1025 A)

0.27 mW/W+76 mW

0.26 mW/W+0.76 W

0.81 mW/W+0.76 W

6.6 mW/W+7.6 W

6.6 mW/W+7.6 W

Page 15: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

AC Power - Source Single Phase, PF = 1 @ 45 Hz to 65 Hz 1 W to < 9.18 W

(330 mV to 1020 V, 3.3 mA to < 9 mA)

9.18 W to < 33.66 W (330 mV to 1020 V, 9 mA to < 33 mA)

33.66 W to < 91.8 W (330 mV to 1020 V, 33 mA to < 90 mA)

91.8 W to < 336.6 W (330 mV to 1020 V, 90 mA to < 330 mA) 336.6 W to < 918 W (330 mV to 1020 V, 0.33 A to < 0.90 A) @ 45 Hz to 65 Hz

918 W to < 2.24kW (330 mV to 1020 V, 0.90 A to < 2.2 A)

1.5 mW/W+7.6 mW

1.1 mW/W+7.6 mW

1.5 mW/W+7.6 mW

1.1 mW/W+76 mW

1.4 mW/W+76 mW

1.2 mW/W+0.76 W

2.24 kW to < 4.59 kW (330 mV to 1020 V,

2.2 A to < 4.5 A) 4.59 kW to 20.91 kW (330 mV to 1020 V,

4.5 A to 20.5 A)

1.5 mW/W+0.76 W

1.3 mW/W+0.76 W

Page 16: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

AC Power – Source (Cont.) Single Phase, PF = 1 @ 45 Hz to 65 Hz

20.91 kW to < 153 kW (33 mV to 1020 V, 20.5 A to < 150 A)

153 kW to < 1.046 MW (33 mV to 1020 V, 150 A to < 1025 A)

13 mW/W+7.6 W

13 mW/W+7.6 W

Charge Plate Monitor -5 kV to -1 kV -1 kV to 0 V 0 V to 1 kV 1 kV to 5 kV

12 mV/V+2.6 V 12 mV/V+2.5 V 12 mV/V+2.5 V 12 mV/V+2.6 V

Tachometer, RPM – Measure Non-Contact

(0 to 100) rpm (>100 to 1000) rpm

(>1000 to 120000) rpm

0.0058 rpm 0.058 rpm 0.58 rpm

Stroboscopes (0 to 1000) rpm (>1000 to 100000) rpm

0.058 rpm 0.58 rpm

Timer/Sweep time /Stopwatch

0.5 s to 60 s >1 min to 2 min >2 min to 3 min >3 min to 4 min >4 min to 5 min >5 min to 10 min >10 min to 30 min >30 min to 60 min >60 min to 90 min

>90 min to 120 min

2.0 ms 10 ms 10 ms 10 ms 10 ms 10 ms 10 ms 10 ms 10 ms 10 ms

Page 17: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Sound level meter 94 dB 114 dB

0.15 dB 0.15 dB

Phase Meter 0 ° to 360 ° 0.010 °

Gauss Meter, Tala Meter 957 G 2.96 kG

0.12 G 0.12 G

Earth Ground Continuity, Ground bond

Test current max 30 A 0.1 Ω 0.2 Ω 0.5 Ω

0.58% 0.58% 0.58%

Pulse - Source Pulse - Measure

Transformer ratio-Source

1 count to 500000 count 1 count to 500000 count

1 turn to 100000 turn

0.58 pulse 0.58 pulse

11 ppm

Frequency - Source

0.01 Hz to < 120 Hz 120 Hz to < 1.2 kHz 1.2 kHz to < 12 kHz 12 kHz to <120 kHz

120 kHz to <1.2 MHz 1.2 MHz to < 2 MHz

2.9 ppm+59 μHz 2.9 ppm+0.59 mHz 2.9 ppm+5.8 mHz 2.9 ppm+58 mHz 2.9 ppm+0.59 Hz 2.9 ppm+5.8 Hz

20 Hz to 100 Hz >100 Hz to 1 kHz >1 kHz to 10 kHz

>10 kHz to 100 kHz >100 kHz to 1 MHz >1 MHz to 10 MHz

>10 MHz to 20 MHz 100 μHz to 100 Hz >100 Hz to 1 kHz >1 kHz to 10 kHz

>10 kHz to 100 kHz >100 kHz to 1 MHz

2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 10 ppm 10 ppm 10 ppm 10 ppm 10 ppm

Page 18: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Frequency – Source (Cont.) >1 MHz to 15 MHz

20 Hz to 100 Hz >100 Hz to 1 kHz >1 kHz to 10 kHz

>10 kHz to 100 kHz >100 kHz to 1 MHz >1 MHz to 10 MHz

>10 MHz to 100 MHz >100 MHz to 1GHz

>1GHz to 3 GHz >3GHz to 6 GHz

10 ppm 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012

Frequency – Measure

Period - Measure

0.1 Hz to 10 Hz >10 Hz to 100 Hz >100 Hz to 1 kHz >1 kHz to 10 kHz

>10 kHz to 100 kHz >100 kHz to 1 MHz >1 MHz to 10 MHz

>10 MHz to 100 MHz >100 MHz to 1 GHz

>1 GHz to 5 GHz >5 GHz to 10 GHz >10 GHz to 20 GHz

10 s to 0.2 ns

2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012 2.0 x1012

Page 19: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Oscilloscopes Square Wave Signal Impedance : 50 Ω @ 10 Hz to 10 kHz Impedance : 1 MΩ @ 10 Hz to 1 kHz

DC Bandwidth

@ 50 kHz to 300 MHz

1.8 mVp-p to< 2.2 Vp-p

1.8 mVp-p to<105 Vp-p

0 mV to 200 V 6 mVp-p to 12 mVp-p

>12 mVp-p to 5.5 Vp-p

2.9 mV/V+100 μV

2.9 mV/V+100 μV

0.58 mV/V 0.50 dB 0.40 dB

@ >300 MHz to 500 MHz 6 mVp-p to 12 mVp-p >12 mVp-p to 5.5 Vp-p

0.60 dB 0.50 dB

@ >500 MHz to 6000 MHz 6 mVp-p to 12 mVp-p >12 mVp-p to 5.5 Vp-p

0.60 dB 0.60 dB

Rise Time/Fall Time 10% to 90%, 90% to 10%

29 ps 29 ps

Horizontal Deflection : Time Mark 5 s 2 s 1 s

0.5 s 0.2 s 0.1 s

50 ms 20 ms 10 ms 5 ms 2 ms 1 ms

0.5 ms 0.2 ms 0.1 ms

5.8 ms/s 2.4 ms/s 1.2 ms/s 0.61 ms/s 0.26 ms/s 0.15 ms/s 87 μs/s 52 μs/s 41 μs/s 35 μs/s 32 μs/s 30 μs/s 30 μs/s 30 μs/s 30 μs/s

Page 20: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Horizontal Deflection : Time Mark (Cont.)

50 μs 20 μs 10 μs 5 μs 2 μs 1 μs

0.5 μs 0.2 μs 0.1 μs 50 ns 20 ns 10 ns 5 ns 2 ns 1ns

30 μs/s 30 μs/s 30 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s 29 μs/s

Amplitude Modulation - Source Frequency Carrier

(150 kHz to 10 MHz)

Modulation Rate : 20 Hz to 50 Hz

AM 5 to 30 % AM > 30 to 99 % Modulation Rate

: > 50 Hz to 10 kHz AM 5 to 30 %

AM > 30 to 99 %

3.5 % of rdg. +0.018 % 3.5 % of rdg. +0.18 %

2.4 % of rdg. +0.018 % 2.4 % of rdg. +0.18 %

Page 21: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Amplitude Modulation - Source (cont.)

Frequency Carrier (>10 MHz to 1300 MHz)

Modulation Rate : 20 Hz to 50 Hz AM 5 to 30 %

AM > 30 to 99 % Modulation Rate

: > 50 Hz to 50 kHz AM 5 to 30 %

AM > 30 to 99 % Modulation Rate

: > 50 kHz to 100 kHz AM 5 to 30 %

AM > 30 to 99 %

3.5 % of rdg. +0.018 % 3.5 % of rdg. +0.18 %

1.2 % of rdg. +0.018 % 1.2 % of rdg. +0.18 %

3.5 % of rdg. +0.018 % 3.5 % of rdg. +0.18 %

Frequency Modulation - Source Frequency Carrier

(250 kHz to 10 MHz)

FM Deviation ≤40 kHz Modulation Rate : 20 Hz to 10 kHz

2.4 %+0.014 kHz

Frequency Modulation - Source (Cont.)

Frequency Carrier (250 kHz to 10 MHz)

FM Deviation ≤400 kHz Modulation Rate : 20 Hz to 50 kHz 50 kHz to 100 kHz

5.8 %+0.14 kHz 1.2 %+0.14 kHz

Phase Modulation - Source Frequency Carrier

(250 kHz to 10 MHz) Frequency Carrier

( >10 MHz to 1300 MHz)

Phase Deviation ≤400 Radians

Modulation Rate : 200 Hz to 10 kHz

Phase Deviation ≤400 Radians

Modulation Rate : 200 Hz to 20 kHz

4.7 %+0.14 Radian

3.5 %+0.14 Radian

Page 22: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Distortion/Sinad/ Signal to noise - Source

Distortion

@ 20 Hz to 20 kHz (0.01% to 100%) 0 dB to -80 dB

@ > 20 kHz to 100 kHz (0.01% to 100%) 0 dB to -80 dB

0.38 dB

0.38 dB

Sinad Signal to Noise

@ 20 Hz to 20 kHz 0 dB to 80 dB

@ > 20 kHz to100 kHz 0 dB to 80 dB

20 Hz to100 kHz 0 dB to 80 dB

0.38 dB

0.38 dB

0.38 dB

Amplitude Modulation - Measure Frequency Carrier

(150 kHz to 10 MHz)

Modulation Rate : 20 Hz to 50 Hz AM 5 to 30 %

AM > 30 to 99 % Modulation Rate

: > 50 Hz to 10 kHz AM 5 to 30 %

AM > 30 to 99 %

3.5 % of rdg. +0.018% 3.5 % of rdg. +0.18%

2.4 % of rdg. +0.018% 2.4 % of rdg. +0.18%

Page 23: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Frequency Carrier (>10 MHz to 1300 MHz)

Modulation Rate : 20 Hz to 50 Hz AM 5 to 30 %

AM > 30 to 99 % Modulation Rate

: > 50 Hz to 50 kHz AM 5 to 30 %

AM > 30 to 99 % Modulation Rate

: > 50 kHz to 100 kHz AM 5 to 30 %

AM > 30 to 99 %

3.5 % of rdg. +0.018% 3.5 % of rdg. +0.18%

1.2 % of rdg. +0.018% 1.2 % of rdg. +0.18%

3.5 % of rdg. +0.018% 3.5 % of rdg. +0.18%

Frequency Modulation - Measure Frequency Carrier

(250 kHz to 10 MHz) Frequency Carrier

( >10 MHz to 1300 MHz)

FM Deviation ≤40 kHz Modulation Rate : 20 Hz to 10 kHz

FM Deviation ≤40 kHz Modulation Rate : 20 Hz to 50 Hz Modulation Rate

: > 50 Hz to 100 kHz

2.4 %+0.014 kHz

5.8 %+0.14 kHz

1.2 %+0.14 kHz

Phase Modulation - Measure Frequency Carrier

(250 kHz to 10 MHz) Frequency Carrier

( >10 MHz to 1300 MHz)

Phase Deviation ≤400 Radians

Modulation Rate : 200 Hz to 10 kHz

Phase Deviation ≤400 Radians

Modulation Rate : 200 Hz to 20 kHz

4.7 %+0.14 Radian 3.5 %+0.14 Radian

Page 24: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Distortion/Sinad/ Signal to noise - Measure

Distortion

@ 20 Hz to 20 kHz (0.01% to 100%) 0 dB to -80 dB

@ > 20 kHz to 100 kHz (0.01% to 100%) 0 dB to -80 dB

1.2 dB

2.4 dB

Sinad Signal to Noise

@ 20 Hz to 20 kHz 0 dB to 80 dB

@ > 20 kHz to100 kHz 0 dB to 80 dB

20 Hz to100 kHz 0 dB to 80 dB

1.2 dB

2.4 dB

1.2 dB

RF Power-Source (8 to - 70) dBm

10 Hz to 20 MHz 0.12 dB

RF Tuned Power- Measure

2.5 MHz to 1.3 GHz 0 to -10 dB

<-10 to -20 dB <-20 to -30 dB <-30 to -40 dB <-40 to -50 dB <-50 to -60 dB <-60 to -70 dB <-70 to -80 dB <-80 to -90 dB

<-90 to -100 dB

0.18 dB 0.19 dB 0.19 dB 0.20 dB 0.21 dB 0.21 dB 0.22 dB 0.22 dB 0.22 dB 0.30 dB

Page 25: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

RF Absolute Power- Measure 100 kHz to 4.2 GHz 25 dBm to 35 dBm 10 MHz to 6 GHz

-60 dBm to <-10 dBm -10 dBm to <0 dBm 0 dBm to 20 dBm

100 kHz to 2.6 GHz -10 to 30 dBm

<-20 dBm to -10 dBm

0.051 dB

0.15 dB 0.13 dB 0.10 dB

0.024 dB 0.036 dB

RF Power- Source 9 kHz to 6 GHz (13dBm to-139dB)

DC to 12.4 GHz >12.4 GHz to 18 GHz

100kHz to 2.6GHz

1.2 dB 0.30 dB 0.70 dB 0.020 dB

Noise Figure - Measure Tuning Frequency: 10 MHz to <100 MHz 100 MHz to 1.8 GHz

Noise Figure and Gain: 0.1 dB to 20 dB

Noise Source Drive: 28 VDC

2.0 x 10-12 2.0 x 10-12

0.20 dB

0.10 mV/V

RF Power meter 3.16 μW 10 μW

31.6 μW 100 μW

0.316 mW 1 mW

3.16 mW 10 mW

31.6 mW 100 mW

0.010 μW 0.015 μW 0.10 μW 0.15 μW

0.0010 mW 0.0015 mW 0.010 mW 0.015 mW 0.10 mW 0.15 mW

Page 26: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Temperature-Simulation Resistance Temperature Detector

(Pt100 (385))

- 200 °C to - 80 °C

>- 80 °C to 0 °C >0 °C to 100 °C

>100 °C to 300 °C >300 °C to 400 °C >400 °C to 630 °C >630 °C to 800 °C

0.082 °C 0.082 °C 0.10 °C 0.12 °C 0.13 °C 0.16 °C 0.28 °C

Temperature-Simulation Thermocouple

(With cold junction compensation)

Type K

- 200 °C to - 100 °C >- 100 °C to - 25 °C >- 25 °C to 120 °C >120 °C to 1000 °C

>1000 °C to 1372 °C

0.23 °C 0.21 °C 0.20 °C 0.21 °C 0.21 °C

Temperature-Simulation Thermocouple

(With cold junction compensation)

Type J

Type T

- 210 °C to - 100 °C >- 100 °C to - 30 °C >- 30 °C to 150 °C >150 °C to 760 °C >760 °C to 1200 °C - 250 °C to - 150 °C > - 150 °C to 0 °C >0 °C to 120 °C

>120 °C to 400 °C

0.22 °C 0.20 °C 0.20 °C 0.20 °C 0.21 °C 0.33 °C 0.20 °C 0.19 °C 0.19 °C

Page 27: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Temperature-Simulation (Cont.) Thermocouple (With cold junction

compensation) Type R

Type S

Type E

Type N

0 °C to 250 °C

>250 °C to 400 °C >400 °C to 1000 °C

>1000 °C to 1767 °C

0 °C to 250 °C >250 °C to 400 °C >400 °C to 1000 °C

>1000 °C to 1767 °C

-250 °C to -100 °C >-100 °C to -25 °C >-25 °C to 350 °C >350 °C to 650 °C >650 °C to 1000 °C

- 200 °C to - 100 °C >- 100 °C to - 25 °C >- 25 °C to 120 °C >120 °C to 410 °C >410 °C to 1300 °C

0.48 °C 0.41 °C 0.40 °C 0.40 °C

0.48 °C 0.41 °C 0.41 °C 0.42 °C

0.28 °C 0.20 °C 0.20 °C 0.20 °C 0.21 °C

0.27 °C 0.21 °C 0.21 °C 0.21 °C 0.21 °C

Page 28: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Temperature-Simulation Thermocouple (Without cold

junction compensation) Type K

Type J

- 200 °C to - 100 °C >- 100 °C to - 25 °C >- 25 °C to 120 °C >120 °C to 1000 °C

>1000 °C to 1372 °C

- 210 °C to - 100 °C >- 100 °C to - 30 °C >- 30 °C to 150 °C >150 °C to 760 °C >760 °C to 1200 °C

0.14 °C 0.10 °C 0.090 °C 0.10 °C 0.11 °C

0.13 °C 0.090 °C 0.090 °C 0.090 °C 0.10 °C

Simulation (Cont.)

Thermocouple (Without cold

junction compensation)

Type T

Type R

Type S

- 250 °C to - 150 °C

>- 150 °C to 0 °C

>0 °C to 120 °C

>120 °C to 400 °C

0 °C to 250 °C

>250 °C to 400 °C

>400 °C to 1000 °C

>1000 °C to 1767 °C

0 °C to 250 °C

>250 °C to 400 °C

>400 °C to 1000 °C

>1000 °C to 1767 °C

0.29 °C

0.11 °C

0.090 °C

0.090 °C

0.33 °C

0.20 °C

0.19 °C

0.19 °C

0.32 °C

0.21 °C

0.20 °C

0.21 °C

Page 29: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Simulation (Cont.) Thermocouple (Without cold

junction compensation) Type E

Type N

-250 °C to -100 °C >-100 °C to -25 °C >-25 °C to 350 °C >350 °C to 650 °C >650 °C to 1000 °C

- 200 °C to - 100 °C >- 100 °C to - 25 °C >- 25 °C to 120 °C >120 °C to 410 °C >410 °C to 1300 °C

0.21 °C 0.090 °C 0.090 °C 0.080 °C 0.090 °C

0.20 °C 0.11 °C 0.11 °C 0.10 °C 0.11 °C

Temperature-Measure Resistance Temperature Detector

(Pt100 (385))

- 200 °C to - 80 °C

>- 80 °C to 0 °C >0 °C to 100 °C

>100 °C to 300 °C >300 °C to 400 °C >400 °C to 630 °C >630 °C to 800 °C

0.012 °C 0.012 °C 0.013 °C 0.021 °C 0.023 °C 0.028 °C 0.033 °C

Page 30: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Temperature-Measure (Cont.) Thermocouple (Without cold

junction compensation) Type K

Type J

Type T

Type R

- 200 °C to - 100 °C >- 100 °C to - 25 °C >- 25 °C to 120 °C >120 °C to 1000 °C

>1000 °C to 1372 °C

- 210 °C to - 100 °C >- 100 °C to - 30 °C >- 30 °C to 150 °C >150 °C to 760 °C >760 °C to 1200 °C

- 250 °C to - 150 °C

>- 150 °C to 0 °C >0 °C to 120 °C

>120 °C to 400 °C

0 °C to 250 °C >250 °C to 400 °C >400 °C to 1000 °C

>1000 °C to 1767 °C

0.10 °C 0.060 °C 0.060 °C 0.060 °C 0.070 °C

0.090 °C 0.060 °C 0.060 °C 0.050 °C 0.060 °C

0.20 °C 0.070 °C 0.060 °C 0.060 °C

0.24 °C 0.15 °C 0.13 °C 0.12 °C

Page 31: SCOPE SERVICES OF ISO/IEC 17025 - ISOCAL TechnologySCOPE SERVICES OF ISO/IEC 17025 ELECTROMAGNETIC – DC/LOW FREQUENCY PARAMETER / EQUIPMENT RANGE CALIBRATION & MEASUREMENT CAPABILITY

PARAMETER /

EQUIPMENT RANGE

CALIBRATION & MEASUREMENT

CAPABILITY

[EXPRESSED AS

UNCERTAINTY(±)]

Temperature-Measure (Cont.) Thermocouple (Without cold

junction compensation) Type S

Type E

0 °C to 250 °C >250 °C to 400 °C >400 °C to 1000 °C

>1000 °C to 1767 °C

-250 °C to -100 °C >-100 °C to -25 °C >-25 °C to 350 °C >350 °C to 650 °C >650 °C to 1000 °C

0.24 °C 0.15 °C 0.14 °C 0.14 °C

0.14 °C 0.050 °C 0.060 °C 0.050 °C 0.060 °C

Thermocouple (Without cold junction compensation)

Type N

- 200 °C to - 100 °C >- 100 °C to - 25 °C >- 25 °C to 120 °C >120 °C to 410 °C >410 °C to 1300 °C

0.14 °C 0.080 °C 0.070 °C 0.060 °C 0.060 °C