Classification of Resistance Measurements
• Low resistances (0-1) Ω
• Medium resistances (1-0.1M)Ω
• High resistances (≥ 0.1MΩ)
1
Medium Resistance Measurement Methods
(1) Voltmeter-Ammeter (2) Substitution(3) Ohmmeter (4) Bridge
3
Voltmeter-Ammeter Method
As we know that the internal resistances of the voltmeter and ammeter are large and small, respectively.
4
5
Voltmeter-Ammeter Method(1) Simple and requires only two meters(2) Subject to error
Substitution Method• Keep at position 1 and
vary the Rh up to finite value of current flow through the circuit.
• Shift to position 2 with the same value of Rh in previous case, change the S value until the same/previous current flow through the circuit.
7
Ohmmeter Method(1) Only one meter(2) Non-linear scale
Using Bridge Circuit
9
• Bridge Circuit is a null method, operates on the principle of comparison
• A known (standard) value is adjusted until it is equal to the unknown value
10
Wheatstone Bridge
Ratio arms (R1 , R2) & Standard arm (R3 )
11
Wheatstone Bridge Apparatus
The Ratio Control switches the ratio in decade steps
The remaining four step switches set the resistance of the standard arm
R1 , R2 10 Ω, 100 Ω, 1 kΩ 10 kΩR3 Min. step 1 Ω, Max value 11110 Ω
Example
12
13
14
Measurement Errors
Thermal EMF
15
Creates error in measurement of low resistance
16
Contact Resistance
On the micro scale, surface roughness limits surface to surface contact
Measurement using four terminals to avoid contact resistance
Standard resistance model with contact resistance
17
18
19
20
Figure below show the schematic diagram of a Wheatstone bridge with the bridge elements. The battery is 5 V and its internal resistance negligible. The galvanometer has a current sensitivity of 10 mm/ µA and an internal resistance of 100 Ω. Calculate the deflection of the galvanometer caused by the 5 Ω unbalance in arm BC.
Assignment
Ans: 33.2 mm
21
The galvanometer in the previous question is replaced by one with an internal resistance of 500 Ω and a current sensitivity of 1 mm/µA. Assuming that the minimum deflection that can be observed on this galvanometer is 1 mm, determine what is the minimum unbalance in the resistance that can be detected in arm BC.
Assignment Continued
Answers to Assignment-2
1) 2.48 Ω2) 1.06 %3) (a) 99.9 Ω, 0.195 Ω, 0.195 %
(b) (i) 80 (ii) 98 (c) (i) 68 (ii) 95
4) (a) 38.3 % (b) 68.3 % (c) 95.5 % (d) 15.7 %
5) For series: For parallel:
6) (a) 8.7 Ω (b) 7.1 Ω (c) 5.6 Ω
Rn R
n n