Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan , John Lipp, Thomas McMullen 28/03/2014

Transcript of Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen...

Page 1: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Status of Indium Bumped FE-I4BRichard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen28/03/2014

Page 2: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Introduction• 300μm CPII sensor Indium bump

bonded to FE-I4B ROC

• Set up on probe station in GLADD to readout using USBPix

• RX Delay and Analog and Digital scans done to test module was working

• Tuning performed on the module to ascertain number of disconnected/merged bumps and dead pixels

RX Delay

Analog

Digital

Page 3: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Disconnected pixels• Disconnected pixels• (criteria: Sigma > 0e and < 200e)

• This is the case for 6 out of 26880 pixels

Page 4: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Merged Bumps• Merged pixels• (criteria: Crosstalk occurs at 100V)

• This is the case for 106 out of 26880 pixels

Page 5: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Dead Pixels• As tuning was not ideal, this is

an approximate value by counting pixels with sigma value less than 1e from Threshold Scan @ 100V from long tuning attempt (see later)

• 21 pixels dead

Page 6: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Bump Yield• DISCONNECTED: 6 • MERGED: 106• DEAD: 21

• TOTAL: 133 out of 26880

• BUMP-YIELD: 99.505%

Page 7: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• The IBL tuning scheme

did not work fully as it uses FAST tunes and all parameter space was not scanned• 3000e threshold with

9ToT @ 16ke

• THRESHOLD: Slightly high compared to target. White pixels here are not dead pixels, but pixels which are not ‘reached’ using a fast tuning.

Page 8: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• Compared to SC-3072-4-8

(250μm x 50μm)

• THRESHOLD: 2975e

Page 9: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• Sigma on threshold with 0V

bias shows very few pixels with values around 120e (disconnected)

• Next slide has log Y plot to show pixels with low values of sigma

Page 10: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• Log Y Scale on sigma plot

Page 11: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• Compared to SC-3072-4-8

(250μm x 50μm) at 100V

• SIGMA: 119.6e

Page 12: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• TOT: This looks good!

Page 13: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• Compared to SC-3072-4-8

(250μm x 50μm)

• TOT: 8.99

Page 14: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• LONG tuning at 100V looked

to be problematic:• Threshold very high

• although note the pixels which were white in fast tune are occupied now

• Sigma very spread out

• Closer look at individual SCurves of pixels showed very poor fit giving both:• HIGH THRESHOLD!• LARGE SIGMA! HIGH THRESHOLD!

SCurve is evident – fit has failed

Page 15: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• LONG tuning at 100V looked

to be problematic:• Threshold very high• Sigma very spread out

• Closer look at individual SCurves of pixels showed very poor fit giving both:• HIGH THRESHOLD!• LARGE SIGMA!

LARGE SIGMA!Poor fit – sigma actually much smaller!

Page 16: Status of Indium Bumped FE-I4B Richard Bates, Craig Buttar, Kate Doonan, John Lipp, Thomas McMullen 28/03/2014.

Tuning• I am contacting Malte Beckhaus at CERN to see if there is a

method of improving the fit for the SCurves • With this issue, the threshold value and sigma value for each

pixel is wrong in this case