Status of dE/dx Calibration

27
Status of dE/dx Calibration Yuri Fisyak October 17, 2002

description

Status of dE/dx Calibration. Yuri Fisyak October 17, 2002. Outlook. Why ? What ? Where ? Conclusions. Why ?. Last calibration was done in March, 2002: σ (dE/dx)/(dE/dx) = 8.2% for 76 cm track STAR CDR (p. 4C-33) σ (dE/dx)/(dE/dx) = 0.47 N -0.46 (Ph) -0.32 , P = 1atm - PowerPoint PPT Presentation

Transcript of Status of dE/dx Calibration

Page 1: Status of dE/dx Calibration

Status of dE/dx Calibration

Yuri FisyakOctober 17, 2002

Page 2: Status of dE/dx Calibration

Outlook

• Why ?• What ?• Where ?• Conclusions

Page 3: Status of dE/dx Calibration

Why ?

• Last calibration was done in March, 2002:– σ(dE/dx)/(dE/dx) = 8.2% for 76 cm track– STAR CDR (p. 4C-33)

• σ(dE/dx)/(dE/dx) = 0.47 N-0.46(Ph)-0.32, P = 1atm• σInner = 14.3%, h = 1.15 cm, N = 12;• σOuter = 7.7%, h = 1.95 cm, N = 32;• σ = 6.8% for 76 cm track in TPC

– H.Bichsel simulation: • σ = 7.3 % for 11 * 1.2 cm + 31 * 2.0 cm

– I am not happy that ~1% is missing

Page 4: Status of dE/dx Calibration

Why (cont.) ?• Comparison new and old tcl• H.Bichsel’s calculations reproduce data only

qualitatively. Is this due to calibration procedure ?• H.Bichsel claims that we have non linearity in dE

measurement. Can we check this ?• Can dE/dx calibration be done in one pass? Can we

move it into fast online ?

Page 5: Status of dE/dx Calibration

What does calibration include?• Applying pad correction obtained from pulser data (Fabrice did this and I will not talk about)• For good clusters (used in fit, no overlaps) • For good global tracks (No. fit points ≥ 30, Track length in TPC > 40 cm)• Z =log[(dE/dx)measured/(dE/dx)predicted for π]

•Fit Z-distribution with Gauss(μ,σ) + pol3 in +/-3σ range. It is supposed that we have ~80% π and μ corresponds to π.

•μ should not depend on

•Time, Pressure

•Sector, row

•Drift distance

•…

Page 6: Status of dE/dx Calibration

What does prediction mean ?

0.45 GeV/c

•Hans Bichsel made presentation about his model during the last collaboration meeting. He is preparing STAR TN on this subject.

•PAI model (Photo Absorption model for Ionization energy loss, V.Grishin V.Ermilova,S.Kotelnikov, NIM A309(1991) 476) gives the same predictions but PAI simulation program (provided by P.Nevski) has problem at low β (it was not tuned for this range). For this reason I will use Bichsel’s calculations.

Page 7: Status of dE/dx Calibration

9.6% @ 76 cm

Resolution before calibration

Page 8: Status of dE/dx Calibration

What does March calibration mean?

• March calibration included:– The same procedure as for Year 1 data– As prediction it was used Sirrf– Calibration was done for all tracks (no restriction on

momentum)– Time dependence : overall gain correction factor each

few hours (1-4)– Sector and pad row correction– Drift distance correction– Result : σ = 9.6 % → σ = 8.2 %

Page 9: Status of dE/dx Calibration

Red – 40% truncation

Green – 30% truncation

Black - fit

8.2% @ 76 cm

Page 10: Status of dE/dx Calibration

9.7%

9.9% =>9.8% after setting convolution flag

New tcl a little bit worse than old one

Page 11: Status of dE/dx Calibration

New calibration• Calibration is based on tracks with 0.4 < p < 0.5 GeV/c

(~MIP for pions: βγ = p/m = 4).• Calibration has been done for new tcl only.• As prediction it was used Bichsel’s calculation with dx

dependence (see next slide) i.e. Z =log[(dE/dx)measured/(dE/dx)predicted for π(βγ,dx)]

• This calibration gives σ = 8.8% (instead of 8.2% obtained in March because it was done only for 0.4<p<0.5GeV/c but resolution is obtained for all momenta).

Page 12: Status of dE/dx Calibration
Page 13: Status of dE/dx Calibration

Calibration done for p in [0.4,0.5] GeV/c with new tcl

σ = 8.8% @76 cm

Page 14: Status of dE/dx Calibration

Check of ADC nonlinearityFor uniquely identified tracks:• σ < 15% and• v = log[(dEdx)/(dE/dx)J],

where J = [e,π,K,p,d]• |v| < 3σ for only J, and • |(dE/dx)J - (dE/dx)k| >5σ, for

J≠KPlot predicted dE versus

measured dEInner

Outer

It is seen clean nonlinearity besides saturation (with offset ~1keV).

It is not clear what is origin for this nonlinearity:

•ADC, it has to be checked with pulser data

•tcl ? due to threshold effects?

Page 15: Status of dE/dx Calibration

Time dependenceGain variation versus time. ~5%

Pressure variation versus time

Oxygen concentration variation versus time

Page 16: Status of dE/dx Calibration

Time dependence (cont.)

• Drift distance correction: e+AdO

where A = 1.75e-6 (1/cm/ppmO2),

provided by A.Lebedev (ALEPH data) d – drift distance (cm), O – concentration of Oxygen (ppm)

Page 17: Status of dE/dx Calibration

Pressure

•Both Drift distance and Pressure correction allow to remove time dependence.

•TPC Gain Monitor cannot be used (at least now) because its results depend on magnetic field.

Page 18: Status of dE/dx Calibration

< 1 %

Page 19: Status of dE/dx Calibration

After pad level correction from pulser calibration it still exists gain variation versus sector and row

Page 20: Status of dE/dx Calibration

Dependence of gain correction versus drift distance (after accounting absorption) for Inner rows has opposite behavior with respect to expected. ADC nonlinearity is overcorrected !?

Page 21: Status of dE/dx Calibration

7.4% @ 76 cm

Page 22: Status of dE/dx Calibration

For both + and -

Page 23: Status of dE/dx Calibration

For positive only

Page 24: Status of dE/dx Calibration

Negative only

Page 25: Status of dE/dx Calibration

Bichsel shapes

Inner

fit by φ(μ+(1+σ)z), where φ(z) is Bichsel shape and z = log(dE/dEmost probable);

Page 26: Status of dE/dx Calibration

Outer

Both inner and outer rows are reasonably well described by Bichsel shape.

Page 27: Status of dE/dx Calibration

Conclusions

• In sense of dE/dx old and new tcl give comparable results.• It is observed significant nonlinearity in dE with respect to H.Bichsel

calculation.• Accounting this nonlinearity allows to improve resolution by ~1%

(8.2%=>7.4%) which is matched with our expectation.• To understand the nonlinearity is necessary to get pulser data with

different signal amplitudes.• If the linearity will be understood than it will be possible to move

dE/dx calibration in fast online.• With new calibration the Bichsel shape of dE/dx distribution

describes data pretty well. Thus we can claim that model calculations gives quantitative agreement with data.