RPM2014 Shielding Workshop - AAPM

30
The UK approach to shielding x-ray rooms 0.1 μGy/mGy cm 0.05 μGy/mGy cm 0.02 μGy/mGy cm 0.01 μGy/mGy cm 0.005 μGy/mGy cm 0.002 μGy/mGy cm 0.001 μGy/mGy cm 0.0005 μGy/mGy cm 1 m +90° -90° 0° 0° Rear Front Gantry Gantry Colin Martin University of Glasgow Scotland, UK 1

Transcript of RPM2014 Shielding Workshop - AAPM

Page 1: RPM2014 Shielding Workshop - AAPM

The UK approach to

shielding x-ray rooms

0.1 μGy/mGy cm

0.05 μGy/mGy cm

0.02 μGy/mGy cm

0.01 μGy/mGy cm

0.005 μGy/mGy cm

0.002 μGy/mGy cm

0.001 μGy/mGy cm

0.0005 μGy/mGy cm

1 m

+90°

-90°

0°0°

Rear

Front

GantryGantry

Colin Martin

University of Glasgow

Scotland, UK

1

Page 2: RPM2014 Shielding Workshop - AAPM

The UK approach to shielding x-ray rooms

Dose limits and constraints

Occupancy

Radiation sources

Predicting scatter levels from kerma-area product (KAP)

Shielding radiography and fluoroscopy rooms

Primary for radiography

Simplifications for mammography and dental

CT

Scatter from DLP

Tertiary scatter

2

Page 3: RPM2014 Shielding Workshop - AAPM

Dose Constraints

ALARP solution (As Low As Reasonably Practicable)

Control Room & Surrounds - 0.3 mSv y-1

X-ray room - Employees - 20 mSv y-1

Surrounding area - Public dose limit - 1 mSv y-1

Dose Limits

Dose Limits and Constraints

SURROUNDS

X-RAY ROOM

CONTROL ROOM

3

Page 4: RPM2014 Shielding Workshop - AAPM

Air Kerma and effective dose

Easier to measure or calculate

Ratio: Effective Dose

Air Kerma

Air kerma is greater than E

for X-rays, so this is a

conservative approach

Use an Air Kerma Constraint

0.3 mGy y-1 (300 μGy y-1)4

Page 5: RPM2014 Shielding Workshop - AAPM

OccupancyFor how long are people exposed

Constraint: 300 μGy y-1

Dose to individual

Design limit Air kerma constraint / Occupancy factor (T)

300/ T μSv y-1

5

Page 6: RPM2014 Shielding Workshop - AAPM

Partial Occupancy Factor: 0.2 – 0.5 Staff rooms

Wards

Clinics

Reporting room

Occasional Occupancy Factor: 0.05 – 0.125 Corridors,

Stairways

Store rooms

Unattended waiting rooms

Unattended car park

Changing rooms

Toilets

Gardens

Full Occupancy Factor: 1

Control room

Reception area

Nursing station

Offices

Shops

Living quarters

Adjacent buildings

Occupancy Factors

6

Page 7: RPM2014 Shielding Workshop - AAPM

Radiation sources

Secondary scatter

All modalities

Leakage

Minimal significance

Primary

Radiography

Tertiary scatter

Consider for CT

7

Page 8: RPM2014 Shielding Workshop - AAPM

Scatter air kerma: source of most radiation

Varies with primary beam air kerma and beam size

Varies with angle and kV

Dependent on beam position, patient size

Scatter Kerma-Area Product (KAP)

Equation used to determine scatter air kerma Ks is:

Ks = S × KAP

S is a scatter factor

8

Page 9: RPM2014 Shielding Workshop - AAPM

Scatter

Varies with angle and kV

Scatter factor S = Ks / KAP

0

2

4

6

8

10

12

0 30 60 90 120 150 180

Angle of scatter

Sc

att

er

Fa

cto

r (S

) µ

Gy

(G

y.c

m2)-1 125 kVp

100 kVp

85 kVp

70 kVp

50 kVp

Williams JR (1996) Br J Radiol, 69,1032

10 Gy

4 Gy

3 Gy

8 Gy

Scatter

air kermaper Gy cm2

9

Page 10: RPM2014 Shielding Workshop - AAPM

0.0

0.2

0.4

0.6

0.8

1.0

03

06

09

01

20

15

01

80

Sc

att

eri

ng

an

gle

Relative dose

Incident beam parallel to barrierDistance to barrier shortest at an angle of 90⁰

Taking into account variation with angle and distance to

barrier: Maximum air kerma occurs at 117⁰

Variation in dose at barrier with angleWall

8 Gy

5 Gy

2 Gy

θ

X-ray beam

4 Gy

10 Gy

3 Gy

8 Gy

Relative dose

10

Page 11: RPM2014 Shielding Workshop - AAPM

Scatter factors (S) for fluoroscopy and radiography from

measurement and calculation

S = Ks / KAP

Smax = 0.031 x kV + 2.5 µGy (Gy cm2)-1 @ 1 m

85 kV: Smax = 5.1 µGy (Gy cm2)-1

For interventional beams with copper filtration

85 kV: Smax = 8 µGy (Gy cm2)-1

11

Page 12: RPM2014 Shielding Workshop - AAPM

Workload – in terms of KAP

(At least) Two approaches:

1. Predict clinical usage

• Typical KAP values per exam

2. Assume typical total KAP values

• Exceptional workload with 800 patients/wk

KAP: 500 Gy cm2

• Typical workload with 180 patients / wk

KAP: 150 Gy cm2

12

Page 13: RPM2014 Shielding Workshop - AAPM

Scatter Calculation

Scatter air kerma (Ks) = S×KAPAnn

d2

S = Scatter factor KAPAnn = Annual workload

d = Distance to barrier

Design criterion (C) = Dose constraint

Occupancy (T)Annual dose constraint = 300 µGy, T = 0.05 – 1.0

Transmission (B) = C

Ks13

Page 14: RPM2014 Shielding Workshop - AAPM

Empirical equations can be used to link shielding

thickness to broad beam transmission

B = [(1+/).exp.(x)-/]-1/

B = broad beam transmission

x = thickness of material

x = 1 ln B- + /

1 + /

Archer, Thornby & Bushong (1983)

Health Physics, 44, 507-17

1.0E-08

1.0E-07

1.0E-06

1.0E-05

1.0E-04

1.0E-03

1.0E-02

1.0E-01

1.0E+00

0 1 2 3

Tra

nsm

issio

n

Lead thickness (mm)

70 kV

90 kV

100 kV

14

Page 15: RPM2014 Shielding Workshop - AAPM

Primary beamTwo approaches

Detector Air Kerma (DAK) Method ~ 10 μGyand allow for lead equivalence of

cassette/bucky/table

Entrance Air Kerma (EAK) methodAdjust for inverse square law

A

DC

BA

DC

B

EAKDAK

Primary

incident

on wall

15

Page 16: RPM2014 Shielding Workshop - AAPM

Modalities where scatter per image

or exam can be used

Mammography 7.6 µGy per image

Intra-oral dental 0.5 µGy per image

Panoramic dental 0.65 µGy per exam

Dental cone beam 6-20 µGy per image

Intra-ora dental clinics

Gypsum Wallboard (plaster board) – 10-15 mm thick

Transmission of 40 mm is 0.1 (120 X-rays per week at 1 m)16

Page 17: RPM2014 Shielding Workshop - AAPM

Protecting a CT scanner roomScatter factors SCT linked to Dose Length Product (DLP)

Scatter factors of the form: SCT = Ks / DLP where values of Ks represent the scatter air kerma at 1 m from the iso-centre

for a particular direction.

Factors based on scatter measurements on CT scanners for 4 major vendors

Coefficients provide links

between scatter air kerma and

DLP derived from

measurements on CT scans of

anthropomorphic phantom.

17

Page 18: RPM2014 Shielding Workshop - AAPM

CT scanner dose distributions

0.1 μGy/mGy cm

0.05 μGy/mGy cm

Measured dose contour

0.02 μGy/mGy cm

0.01 μGy/mGy cm

0.005 μGy/mGy cm

0.002 μGy/mGy cm

0.001 μGy/mGy cm

0.0005 μGy/mGy cm

0.0002 μGy/mGy cm

1 m

Scatter from body scan with Philips MX8000

Scatter from head scan with GE Lightspeed 16

Central region shielded by gantry

Dose per DLP higher for body than head scans18

Page 19: RPM2014 Shielding Workshop - AAPM

Measured dose contour0.02 μGy/mGy cm

Calculated dose contour0.02 μGy/mGy cm

1 m

Factors SCTbody for

calculating scatter air

kerma from DLP

Scatter per DLP

0.3 μGy (mGy cm)-1

40o

20o

Scatter per DLP

0.36 μGy (mGy cm)-1

CT body scans

Rear

Front

The scatter factors

can be considered to

be independent of kV

19

Page 20: RPM2014 Shielding Workshop - AAPM

CT gantry provides protection equivalent

to a factor of 10

Measured dose contour0.02 μGy/mGy cm

Calculated dose contour0.02 μGy/mGy cm

1 m

3

2

Isocentre is

nearer front

of gantry.

Ratio 3:2.

40o

20o

Behind the gantry

Scatter per DLP is

0.04 μGy (mGy cm)-1

Rear

Front

20

Page 21: RPM2014 Shielding Workshop - AAPM

Scatter factorsAir kerma at 1 m from scanner isocentre at different

angles with respect to the scan plain

Negative angles - front of the gantry; positive angles - rear

Exam Sector of Angular range Scatter factor

CT scanner per unit DLP μGy (mGy cm)-1

Body Front -90º - -20º 0.36

Body Rear 40º - 90º 0.3

Body Gantry -20º - 40º 0.04

Head Front -90º - -20º & 0.14

and rear 40º - 90º

Head Gantry -20º - 40º 0.014

Apply inverse square law to air kerma value at 1 m 21

Page 22: RPM2014 Shielding Workshop - AAPM

Comparison of calculated and

measured scatter air kerma contours

Calculated contours --- are compared with measurements of scatter air kerma ---.

No significant difference found between 120 kV and 140 kV

0.3 μGy (mGy cm)-1

0.04 μGy (mGy cm)-1

0.36 μGy (mGy cm)-1

22

Page 23: RPM2014 Shielding Workshop - AAPM

Prediction of workload Workload obtained from audit of local practice ideally.

Consider body and head separately

Head group - all exams using a small field of view

Mean exam DLPs provide an indication of likely values.

Mean – 850 mGy cm; 3rd quartile - 900-1,000 mGy cm

Annual Workloads for 38 CT scanners - body & head scans

Mean DLP 3rd Quartile DLP Maximum DLP per annum per annum per annum(Gy cm) (Gy cm) (Gy cm)

Body 1,900 3,400 5,000 Head 1,300 2,500 3,600 Total 3,200 5,500 8,600

23

Page 24: RPM2014 Shielding Workshop - AAPM

Protecting the floors above and below

0.1 μGy/mGy cm

0.05 μGy/mGy cm

0.02 μGy/mGy cm

0.01 μGy/mGy cm

0.005 μGy/mGy cm

0.002 μGy/mGy cm

0.001 μGy/mGy cm

0.0005 μGy/mGy cm

1 m

Again gantry provides

some protection, so

the X-rays are incident

at angle θ = 30o.

θ

24

Page 25: RPM2014 Shielding Workshop - AAPM

Protecting the ceiling and floor Calculated for persons at vertical distance 0.5 m above

the next floor level and 1.0 m above floor below.

Protection afforded by gantry means that only angles of incidence (θ) > 30o need be considered.

If vertical distance is d, then distance from isocentre for calculation = d / cos θ

Oblique incidence, so use an

equivalent barrier thickness equal to

mean of actual thickness and that in

direction of scatter

Calculated thickness =x

Barrier thickness required

= x (1 + cos θ) / 2

θ

x

25

Page 26: RPM2014 Shielding Workshop - AAPM

The problem of tertiary scatter

b

Ceiling

height h

Tertiary

scatter

Barrier

height b

Limiting ray for

calculation of

shielding

30°

The scatter of X-rays from the ceiling over short barriers

is high enough to give staff a radiation dose

How do we calculate this dose?26

Page 27: RPM2014 Shielding Workshop - AAPM

Variation of tertiary scatter with barrier and

ceiling heights

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.5 2 2.5 3 3.5 4 4.5 5

Barrier height b (m)

Scatt

er

air

kerm

a p

er

DL

P (

uG

y / (

Gy c

m))

3 m

3.4 m

3.8 m

4.2 m

4.6 m

5 m

Ceiling Height (m)

Scatter level

depends on the

heights of the

ceiling slab and

the protective

barrier

Low ceiling

High ceiling

27

Page 28: RPM2014 Shielding Workshop - AAPM

Calculation of tertiary scatter levels

Equation: Ksec = (C – m.b) (DLPbody + DLPhead/2)

Where b is the height of the barrier and C and m are

constants dependent on the height of the ceiling.

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.5 2 2.5 3 3.5 4 4.5 5

Barrier height b (m)

Scatt

er

air

kerm

a p

er

DL

P (

uG

y / (

Gy c

m))

3 m

3.4 m

3.8 m

4.2 m

4.6 m

5 m

Ceiling Height (m) Linear equations fitted to

lines in graph and used

to estimate tertiary

scatter levels

Ceiling Scatter for 5,000

Gy cm workload and

2 m high barrier

Height (m) (mGy)

3.6 4.0

4.0 3.0

4.4 2.2

28

Page 29: RPM2014 Shielding Workshop - AAPM

Summary Set dose constraint and include occupancy

Scatter is the main component of stray radiation

Scatter levels can be calculated from the KAP

Predict KAP based on clinical workload or data in literature

Radiography requires shielding for both scatter and primary

Calculation of CT shielding based on DLP workload

Separate scatter factors for body and head

Attenuation afforded by gantry can be taken into account

Obliquity of scatter on ceiling can be taken into account

Tertiary scatter from ceiling slabs can be calculated

29

Page 30: RPM2014 Shielding Workshop - AAPM

Radiation Shielding for

Diagnostic Radiology

BIR - 2nd Edition

Authors:

David Sutton

Colin Martin

Jerry Williams

Debbie Peet

Dundee, Glasgow, Edinburgh & Guildford

Available from the online BIR bookshop

www.bir.org.uk

Thank you for your attention

30