Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure,...

47
Radiation Detection and Measurement Radiation Detect & Measure, summer 2005 ( RSM)

Transcript of Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure,...

Page 1: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detection andMeasurement

Radiation Detect & Measure, summer 2005 ( RSM)

Page 2: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Types of radiation relevant toNuclear Medicine

Particle Symbol Mass Charge

Electron e-, β - 511 keV/c2 -1

Positron e+, β+ 511 keV/c2 +1

Alpha α 3700 MeV/c2 +2

Photon γ variable energy none

Radiation Detect & Measure, summer 2005 ( RSM)

Page 3: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Charged particle ranges(α particles)

Loses energy in a more or less continuous slowing downprocess as it travels through matter.

The distance it travels (range) depend only upon its initialenergy and its average energy loss rate in the medium.

The range for an α particle emitted in tissue is on the orderof µm’s.

Radiation Detect & Measure, summer 2005 ( RSM)

Page 4: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Charged particle ranges(β particles)

Electrons or β particles have ranges that are quitevariable from one electron to the next, even for electronsof exactly the same energy in a specific absorbingmaterial.

This is because of the different types of scattering eventsthe β particle can encounter (i.e., scattering events,bremsstrahlung-producing collisions, etc.).

The β range is often given as the maximum distance themost energetic β can travel in the medium.

The range for β particles emitted in tissue is on the orderof mm’s.

Radiation Detect & Measure, summer 2005 ( RSM)

Page 5: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Interactions of high energy photons withmatter

Photoelectric effectphoton is absorbed

Compton scatteringpart of the energy of the photon is absorbedscattered photon continues on with lower energy

Pair productionpositron-electron pair is createdrequires photons above 1.022 MeV

Coherent (Rayleigh) scatteringphoton deflected with very little energy lossonly significant at low photon energies (<50 keV)

Radiation Detect & Measure, summer 2005 ( RSM)

Page 6: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Summary of interaction of chargedparticles and photons with matter

Charged particles have a very short range in tissue~ mm for beta particles~ µm for alpha particles

Alphas have a predictable, continuously slowing pathBetas have a more random pathPhotons have a longer range

interact very infrequently, depositing much or all oftheir energy in each interaction: range ~cm.Compton scatter is the dominant process in tissue equivalent materials for the energy range ofphotons for Nuclear Medicine imaging

Radiation Detect & Measure, summer 2005 ( RSM)

Page 7: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Basic radiation detector

Radiation Detect & Measure, summer 2005 ( RSM)

• Pulse• Current• Stored

Page 8: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation detectors used inNuclear Medicine

Radiation Detect & Measure, summer 2005 ( RSM)

Page 9: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Types of Radiation detectors• Counters

– Number of interactions– Pulse mode

• Spectrometers– Number and energy of interactions– Pulse mode

• Dosimeters– Net amount of energy deposited– Current mode

Radiation Detect & Measure, summer 2005 ( RSM)

Page 10: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Pulse mode versus current mode

• Pulse mode– Detect individual photons– Required for most imaging applications

• Current mode– Measures average rates of photon flux– Avoids dead-time losses

Radiation Detect & Measure, summer 2005 ( RSM)

Page 11: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Interaction Rate and Dead-time

Radiation Detect & Measure, summer 2005 ( RSM)

paralyzable non-paralyzable

From: The Essential Physics of Medical Imaging (Bushberg, et al)

Page 12: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Types of radiation detectors

• Gas-filled detectors

• Solid-state (semiconductor) detectors

• Organic liquid scintillators

• Inorganic scintillators

Radiation Detect & Measure, summer 2005 ( RSM)

Page 13: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Gas-filled Detectors

Ionizing event in airrequires about 34 eV

From: Physics in Nuclear Medicine (Sorenson and Phelps) Radiation Detect & Measure, summer 2005 ( RSM)

Page 14: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Gas-filled detectors(operates in three ranges)

• Geiger-Muller counters

• Proportional counters

• Ionization chambers– Radiation survey meters– Dosimeters (dose calibrator)

Radiation Detect & Measure, summer 2005 ( RSM)From: Physics in Nuclear Medicine (Sorenson and Phelps)

Page 15: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Ionization Chambers

From: Physics in Nuclear Medicine (Sorenson and Phelps) Radiation Detect & Measure, summer 2005 ( RSM)

ATOMLAB 200 Dose Calibrator

No amplificationNo dead-timeSignal = liberated chargeSettings for different isotopesCalibrations

Page 16: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Geiger-Muller counters

From: Physics in Nuclear Medicine (Sorenson and Phelps) Radiation Detect & Measure, summer 2005 ( RSM)

No energy infoLong dead-timeThin window probe

Page 17: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Semiconductor detectors

• Works on same principle as gas-filled detectors(i.e., production of electron-hole pairs insemiconductor material)

• Only ~3 eV required for ionization (~34 eV, air)• Usually needs to be cooled (thermal noise)• Usually requires very high purity materials or

introduction of “compensating” impurities thatdonate electrons to fill electron traps caused byother impurities

Radiation Detect & Measure, summer 2005 ( RSM)

Page 18: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Semiconductor detectors

• CdZnTe detectors - can operate atroom temperature

Radiation Detect & Measure, summer 2005 ( RSM)

Page 19: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Organic liquid scintillators(liquid scintillator cocktail)

• Organic solvent - must dissolve scintillator material andradioactive sample

• Primary scintillator (p-terphenyl and PPO)• Secondary solute (wave-shifter)• Additives (e.g., solubilizers)• Effective for measuring beta particles (e.g., H-3, C-14).

Radiation Detect & Measure, summer 2005 ( RSM)

Page 20: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Inorganic scintillators(physical characteristics)

Absorption of radiation lifts electrons from valence toconduction band

Impurities (activators) create energy levels within theband gap permitting visible light scintillations

Radiation Detect & Measure, summer 2005 ( RSM)

Page 21: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Inorganic scintillators(physical characteristics)

NaI(Tl) BGO LSO(Ce) GSO(Ce)

Density (gm/cm3) 3.67 7.13 7.4 6.71

EffectiveAtomic Number 51 75 66 59

AttenuationCoefficient(@ 511 keV, cm-1 ) 0.34 0.955 0.833 0.674

Light Output(photons/Mev) 40K ~8K ~30K ~20K

Decay Time 230 ns 300 ns 12 ns 60 ns40 ns

Wavelength 410 nm 480 nm 420 nm 430 nm

Index of Refraction 1.85 2.15 1.82 1.85

Hygroscopy yes no no no

Rugged no yes yes no

Radiation Detect & Measure, summer 2005 ( RSM)

Page 22: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Sample Spectroscopy System

Radiation Detect & Measure, summer 2005 ( RSM)From: Physics in Nuclear Medicine (Sorenson and Phelps)

From: The Essential Physics of Medical Imaging (Bushberg, et al)

Page 23: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Photomultiplier tube

Radiation Detectors, Summer 2005 ( RSM)From: Physics in Nuclear Medicine (Sorenson and Phelps)

Page 24: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Energy Resolution

Radiation Detect & Measure, summer 2005 ( RSM)From: Physics in Nuclear Medicine (Sorenson and Phelps)

Page 25: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Interactions of Photons witha Spectrometer

Radiation Detect & Measure, summer 2005 ( RSM)

A. PhotoelectricB. Compton + PhotoelectricC. ComptonD. Photoelectric with

characteristic x-ray escapeE. Compton scattered photon

from lead shieldF. Characteristic x-ray from lead

shield

From: The Essential Physics of Medical Imaging (Bushberg, et al)

Page 26: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Sample Spectrum (Cs-137)

Radiation Detect & Measure, summer 2005 ( RSM)

A. PhotopeakB. Compton continuumC. Compton edge

D. Backscatter peakE. Barium x-ray photopeakF. Lead x-rays

Detection efficiency(32 keV vs. 662 keV)

From: The Essential Physics of Medical Imaging (Bushberg, et al)

Page 27: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Sample Spectrum (Tc-99m)

Radiation Detect & Measure, summer 2005 ( RSM)

A. PhotopeakB. Photoelectric with

iodine K-shell x-rayescape

C. Absorption of lead x-rays from shield

Note absence ofCompton continuum

Why?

From: The Essential Physics of Medical Imaging (Bushberg, et al)

Page 28: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Sample Spectrum (In-111)

Radiation Detect & Measure, summer 2005 ( RSM)

source detector

From: Physics in Nuclear Medicine (Sorenson and Phelps)

Page 29: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Effects of Pulse Pileup

Radiation Detect & Measure, summer 2005 ( RSM)From: Physics in Nuclear Medicine (Sorenson and Phelps)

Page 30: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Calibrations• Energy calibration

– Adjust energy windows around a known photopeak– Often done with Cs-137 and Co-57

• Dose calibration (dose calibrator)– Measure activity of know reference samples (e.g.,

Cs-137 and Co-57)– Linearity measured by repeated measurements of a

decaying source (e.g., Tc-99m)

Page 31: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

What piece of equipment would you used tomeasure the activity of a pure beta emitter?

Radiation Detect & Measure, summer 2005 ( RSM)From: The Essential Physics of Medical Imaging (Bushberg, et al)

Page 32: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Raphex Question

Radiation Detect & Measure, summer 2005 ( RSM)

D58. The window setting used for Tc-99m is set with the center at 140 keV with a

width of +/-14 keV i.e., 20%. The reason for this is:

A. The energy spread is a consequence of the statistical broadening when

amplifying the initial energy deposition event in the NaI(Tl) crystal.

B. The 140 keV gamma ray emission of Tc-99m is not truly monoenergetic but

the center of a spectrum of emissions.

C. The higher and lower Gaussian tails are a consequence of compton scattering

within the patient.

D. The result of additional scattered photons generated in the collimator.

E. A consequence of patient motion during scanning.

Page 33: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Raphex Answer

Radiation Detect & Measure, summer 2005 ( RSM)

D58. The window setting used for Tc-99m is set with the center at 140 keV with a

width of +/-14 keV i.e., 20%. The reason for this is:

A . Photons, which impinge upon the crystal, lose energy by Compton

scattering and the photoelectric effect. Both processes convert the gamma

ray energy into electron energy. On average approximately one electron

hole pair is produced per 30 eV of g amma ray e nergy deposited in the

crystal. These electrons result in the release of visible ligh t when trapped

in the crystal. These light quanta are collected and amplified by

photomultiplier tubes. The statistical fluctuation in the number of light

quanta collected and their amplification is what causes the spread in the

detected energy peak, even when most of the Tc-99m photons deposit

exactly 140 keV in the NaI(Tl) crystal.

Page 34: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Counting Statistics

Page 35: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Sources of Error

• Systematic errors– Consistently get the same error

• Random errors– Radiation emission and detection are

random processes• Blunder

– operator error

Page 36: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Measures of Central Tendency• Mean

– Average value• Median

– Middlemost measurement (or value)– Less affected by outliers

Example: 8, 14, 5, 9, 12Mean = 9.6Median = 9

Page 37: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Measures of Variability• Variance

– Measure of variability:

• Standard deviation– Square root of variance

Page 38: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Poisson PDF

• Radioactive decay and detection arePoisson random processes

• Variance– Variance = mean = pN = x

• Standard deviation– Standard deviation = √ variance = √ pN

Page 39: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Confidence Intervals

99.7±3.0σ99.0±2.58σ95.0±1.96σ90.0±1.64σ68.3±1.0σ50.0±0.674σ

Probabilitythat mean iswithin interval(%)

Interval aboutmeasurement

Page 40: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Raphex Question

D70. How many counts must be collected in an instrument with zero background

to obtain an error limit of 1% with a confidence interval of 95%?

A. 1000

B. 3162

C. 10,000

D. 40,000

E. 100,000

Page 41: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Raphex Answer

D70. How many counts must be collected in an instrument with zero background

to obtain an error limit of 1% with a confidence interval of 95%?

D. A 95% confidence interval means the counts must fall within two standard

deviations (SD) of the mean (N). Error limit = 1% = 2 SD/N, but SD = N1/2. Thus 0.01 = 2(N1/2)/N = 2/ N1/2. Where [0.01]2 = 4/N and N = 40,000.

Page 42: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Propagation of Error

√ σ21 + σ2

2x1 - x2Subtraction of two numbers containingrandom errors

√ σ21 + σ2

2x1 + x2Addition of two numbers containingrandom errors

σ/cx/cDivision of a number with random errorby a number without random error

cσcxMultiplication of a number with randomerror by a number without random error

StandardDeviation

OperationDescription

Page 43: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Raphex question

G74. A radioactive sample is counted for 1 minute and produces 900 counts.

The background is counted for 10 minutes and produces 100 counts. The net

count rate and net standard deviation are about ____ and ____ counts.

A. 800, 28

B. 800, 30

C. 890, 28

D. 890, 30

E. 899, 30

Page 44: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Raphex answerG74. A radioactive sample is counted for 1 minute and produces 900 c ounts.

The background is cou nted for 10 minutes and p roduces 100 c ounts. The net

count rate and net standard deviation are about ____ and ____ counts/min.

D . The net count rate is:

[ ( Ns/ts) - (Nb/tb)] = [(900/1) - (100/10)] = 890.

The net standard deviation, is:

[ (Ns/t2s) - (Nb/t

2b)]

1/2 = [(900) + (1)] = 30.

Page 45: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Extra StuffBinomial process

• Trial can have only two outcomes

Page 46: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Binomial probability density function(PDF)

• N is total number of trials• p is probability of success• x is number of successes

Page 47: Radiation Detection and Measurement · 2014-03-09 · Raphex Question Radiation Detect & Measure, summer 2005 ( RSM) D58. The window setting used for Tc-99m is set with the center

Radiation Detect & Measure, summer 2005 ( RSM)

Binomial probability density functionmean and variance

• N is total number of trials• p is probability of success• x is mean, σ is standard deviation

If p is very small then:

Same as Poisson random process.