QC Gamma Camera

23
Quality Control in Gamma Camera

description

A quality control for new equipment should start with an acceptance test to verify the equipment meets the specifications given by the vendor. The acceptance test should be performed according to accepted international standards and may require the use of instruments and phantoms not available in the department. The acceptance test forms the basis of the reference tests routinely performed in the department during the life-time of the equipment according to a schedule worked out by the medical physicist in cooperation with the nuclear medicine department. Certain parameters should be tested daily, others on weekly, monthly and yearly basis.

Transcript of QC Gamma Camera

Page 1: QC Gamma Camera

Quality Control

in Gamma

Camera

Page 2: QC Gamma Camera

Gamma Camera Imaging of

Radioactive Sources in Patients

Three major Components:

1. Collimator – localizes γ-ray source in patient

2. NaI(Tl)Crystal (single or multi-crystal) over width of patient stops the γ-rays.

3. Array of PMT’s – localizes γ-ray interaction in crystal

Page 3: QC Gamma Camera

Collimator Types

Page 4: QC Gamma Camera

Quality Control (QC)

The observation techniques and activities used to fulfill requirements for quality.

A minimum level of routine QC is required to ensure that nuclear medicine equipment is functioning properly.

These minimum QC tests are intended to detect problems before they impact on clinical patient studies.

Further tests may be required to trace the cause of the problem and to ensure that the equipment is performing properly after service or adjustment

Page 5: QC Gamma Camera

1 Visual Inspection

2 Background Radiation Levels and Contamination

3 Photopeak And Window Setting

4 Uniformity

5 Resolution

6 Whole Body Scan Resolution

7 Centre Of Rotation

Page 6: QC Gamma Camera

•Intrinsically • Specific to Crystal and

PMT’s

•Extrinsically •Includes the Collimator

Page 7: QC Gamma Camera

Visual Inspection

• A visual inspection may reveal obvious defects which may compromise the safety or the imaging efficacy of the system (e.g. frayed or damaged electrical cables).

• Of particular importance is the visual inspection of collimators on a regular basis and whenever collimators are changed.

• Signs of denting or scratching may indicate mechanical damage to the collimator, and stains may be a sign of possible contamination.

• Both of these may produce artefacts such as cold or hot spots on planar images and rings on SPECT images.

Page 8: QC Gamma Camera

Background Radiation Levels and

Contamination• High levels of background radiation may arise from “hot”

patients in the proximity of the imaging system or other sources of unshielded radiation.

• If high energy imaging agents are being used, the potential also exists for penetration through the back of the gamma camera, where shielding is thinner.

• Other sources of background radiation may include radioactive contamination on the floor, walls or even the detector itself.

• Background radiation, if it is of sufficient intensity, has the potential to seriously compromise any type of imaging.

• Even moderately elevated background levels have the potential to seriously degrade intrinsic uniformity or other intrinsic measurements.

Page 9: QC Gamma Camera

Photopeak And Window Setting

• Incorrect photopeak energy window setting(s) can degrade uniformity, reduce sensitivity or can increase the scatter contribution to the image.

• Peak settings should be checked and adjusted in a consistent manner and the settings should be recorded to detect long term drift in the settings.

• Sudden changes in peak setting indicate a possible fault in the camera and should be fully investigated and rectified if necessary before the camera is again used for clinical studies.

Page 10: QC Gamma Camera

Uniformity

• The uniformity or "flood" QC procedure checks that the response of the detector to a uniform irradiation is uniform within defined limits.

• It is one of the most basic QC tests of the gamma camera.

• Interpretation of clinical images taken with the gamma camera rely on the assumption that differences seen are due to differences in tracer distribution in the patient only and not differences introduced by the gamma camera.

Page 11: QC Gamma Camera

Resolution

• The purpose of resolution checks is to detect gradual, long term deterioration of resolution, rather than detecting abrupt changes.

• Inappropriate adjustments carried out during service may affect the resolution, without necessarily being apparent in the uniformity or other checks.

Page 12: QC Gamma Camera

Whole Body Scan Resolution

• To avoid loss of resolution in the scanning direction during whole body scans, the relative physical position between bed and detector has to be accurately synchronised with the electronic offset applied to the image data to form the whole body image.

• Both mechanical problems and drift or inappropriate adjustment of image offset or size can cause a loss of resolution for whole body scans, particularly in older systems.

Page 13: QC Gamma Camera

Centre Of Rotation

• The rotation axis (or centre of rotation) assumed by the reconstruction program has to accurately coincide with the mechanical axis of rotation to avoid loss of resolution and distortion in the reconstructed slices.

• Centre of rotation (COR) offsets are easily corrected during the reconstruction process.

• Centre of rotation offset can vary with collimator and as a function of detector rotation and radius of rotation.

• It is important to establish which factors affect COR offset on each particular camera and then make appropriate allowances for it.

Page 14: QC Gamma Camera

Recommended Frequency of QC

Tests DAILYGamma Camera

• Visual Inspection

• Background/Contamination Check

• Photopeak check and adjustment, if necessary. Depending on the camera, this may require check and adjustment of photopeaks for all radionuclides used with the particular camera.

• Uniformity check on gamma camera

Page 15: QC Gamma Camera

WEEKLYGamma Camera

• High Count Flood uniformity check on computer, particularly for SPECT systems.

Page 16: QC Gamma Camera

Extrinsic Background

Page 17: QC Gamma Camera

Extrinsic Uniformity

Page 18: QC Gamma Camera

ResultExtrinsic Background

• 120 sec

• Det 1 4418 ±20%

• Det 2 4366 ±20%

Extrinsic Uniformity

• 18 000 counts

• ≤6.0% UFOV

Page 19: QC Gamma Camera

Non-Uniformity from

Cracked/Broken Crystal

Crystal may cracked:

•From mechanical shock during collimator exchange.

•by thermal shock where the crystal temperature changes by more than 10 deg./hour.

Page 20: QC Gamma Camera

Non-Uniformity fromCollimator Structure

ArtifactsLarge Diameter Holes Irregular Lead Foil

Construction

Page 21: QC Gamma Camera

Non-Uniformity from

Cracked/Broken CrystalCrushed Lead

Septa Lead Foil

Separation

Page 22: QC Gamma Camera

References

• Minimum QC for Nuclear Medicine – ANZSNM Technical Standards Subcommittee

• http://www.who.int/medicines/areas/quality_safety/quality_assurance/en/

Page 23: QC Gamma Camera

THANK YOU