MODULE 3 MEASUREMENT OF RESISTANCE, POWER,...

of 335 /335
MODULE 3 MEASUREMENT OF RESISTANCE, POWER, POWER FACTOR AND ENERGY 1

Embed Size (px)

Transcript of MODULE 3 MEASUREMENT OF RESISTANCE, POWER,...

  • MODULE 3 MEASUREMENT OF RESISTANCE, POWER,

    POWER FACTOR AND ENERGY

    1

  • Measurement of resistance

    Measurement of low resistance

    (upto 1 ohm)

    Measurement of medium resistance

    (1 to 0.1M )

    Measurement of high resistance

    (greater than 0.1M )

    Measurement of earth resistance

    2

  • Measurement of low resistance

    Measurement of low resistance

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

    3

  • Measurement of low resistance

    4

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Ammeter Voltmeter method

    =

    =

    So to find resistance

    Measure potential across the resistance using voltmeter

    Measure current through the resistance using Ammeter

    5

  • Ammeter Voltmeter method

    6

  • Ammeter Voltmeter method

    Connection (a)

    Voltmeter resistance is infinite

    7

  • Ammeter Voltmeter method

    Connection (a)

    If ammeter resistance is zero, then measured value is the actual value of unknown resistance

    If Rx>>>RA Ammeter effect

    becomes negligible

    8

  • Ammeter Voltmeter method

    Connection(b)

    Voltmeter reading is the true voltage across resistance

    Ammeter reads total current which is sum of current through resistance and voltmeter

    9

  • Ammeter Voltmeter method

    10

  • Ammeter Voltmeter method

    11

  • Ammeter Voltmeter method

    Actual value of resistance is measured if (Rm/RV)=0

    Which means voltmeter resistance is infinite

    If Rx

  • Measurement of low resistance

    13

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Potentiometer method

    14

  • Potentiometer method

    R-rheostat

    B- battery

    S- standard resistance

    X- unknown resistance

    15

  • Potentiometer method

    Circuit is connected as shown in figure

    Rheostat R- regulate current

    Voltage drop across standard resistance and unknown resistance is measured with the help of potentiometer (Vs and Vx )

    =

    =

    16

  • Potentiometer method

    Advantage

    High accuracy

    17

  • Measurement of low resistance

    18

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Kelvin Double Bridge method

    19

  • Kelvin Double Bridge(contd)

    Why it is called double bridge??

    it is because it incorporates the second set of ratio arms as shown

    20

  • Kelvin Double Bridge(contd)

    In this the ratio arms p and q are used to connect the galvanometer at the correct point between j and k to remove the effect of connecting lead of electrical resistance t.

    Under balance condition voltage drop between a and b (i.e. E) is equal to F (voltage drop between a and c)

    21

    http://www.electrical4u.com/electrical-resistance-and-laws-of-resistance/http://www.electrical4u.com/voltage-or-electric-potential-difference/

  • Kelvin Double Bridge(contd)

    22

  • Kelvin Double Bridge(contd)

    23

  • Kelvin Double Bridge(contd)

    24

  • Kelvin Double Bridge(contd)

    It is used to measure resistances as low as 0.00001

    25

  • Measurement of low resistance

    26

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Ohm meter method

    An ohmmeter is an electrical instrument that measures electrical resistance, the opposition to an electric current.

    Micro-ohmmeters (microhmmeter or microohmmeter) make low resistance measurements.

    Megohmmeters (aka megaohmmeter or in the case of a trademarked device Megger) measure large values of resistance.

    The unit of measurement for resistance is ohms ().

    27

    http://en.wikipedia.org/wiki/Electricityhttp://en.wikipedia.org/wiki/Measuring_instrumenthttp://en.wikipedia.org/wiki/Electrical_resistancehttp://en.wikipedia.org/wiki/Electric_currenthttp://en.wikipedia.org/wiki/Meggerhttp://en.wikipedia.org/wiki/%CE%A9

  • Ohm meter method

    Instead of measuring current and voltage , if one quantity is kept constant, then resistance is proportional to other quantity

    Principle of Ohmmter

    If current is kept constant, then resistance is proportional to voltmeter reading connected across the resistance

    28

  • Measurement of low resistance

    29

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Series type Ohm meter

    30

  • Series type Ohm meter

    31

  • Series type Ohm meter

    Current flowing through the meter depends on unknown resistance

    Meter deflection is proportional to value of resistance

    32

  • Series type Ohm meter

    To mark zero on the scale

    A-B is shorted

    R2 is adjusted so that current (Ifsd) through the meter gives full scale deflection

    This position is marked as zero

    33

  • Series type Ohm meter

    To mark infinity

    A-B is opened

    No current flows throgh the circuit

    So pointer does not deflect

    This point is marked as infinity

    Rh is half scale resistance marking

    34

  • Series type Ohm meter

    For half scale deflection,

    Since total resistance

    presented to battery is 2Rh

    35

  • Series type Ohm meter

    To produce full scale deflection, current required is

    36

  • Series type Ohm meter

    37

  • Series type Ohm meter

    38

  • Measurement of low resistance

    39

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Shunt type Ohmmeter

    40

  • Shunt type Ohmmeter

    Consists of battery in series with adjustable resistance R1 and meter

    Switch- provided to disconnect the battery when instrument is not in use

    Unknown resistance is connected in parallel with meter hence the name shunt type Ohmmeter

    41

  • Shunt type Ohmmeter

    A-B shorted Entire current passes though

    short, hence meter reads zero Pointer position is marked as

    zero

    A-B opened Entire current passes through

    the meter and the deflection of the pointer is marked as infinity.

    A-B some unknown resistance Meter will show

    intermediate readings

    42

  • Shunt type Ohmmeter

    AB is opened

    Current flowing through meter is:

    =

    1 +

    43

  • Shunt type Ohmmeter

    If AB is shorted

    Current flowing through the meter:

    = 0 mperes

    44

  • Shunt type Ohmmeter

    If unknown resistance is connected across AB

    At node 1 = +

    =

    1 +

    +

    45

    =( + )

    ( + )1+

  • Shunt type Ohmmeter

    Current through meter is

    =

    +

    46

    =(+)

    1(+)+ *

    +

  • Shunt type Ohmmeter

    For full scale deflection, = 0

    = 0 Amperes

    47

  • Shunt type Ohmmeter

    For half scale deflection,

    =

    Current through meter is:

    =

    21+

    48

  • Measurement of low resistance

    49

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Crossed coil Ohmmeters

    These instruments are also called ratiometers

    It consists of two rigidly fixed coils at an angular seperation of 900

    Indication depends on ratio of currents through two coils

    50

  • Measurement of low resistance

    51

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Fixed Magnet Moving Coil Ohmmeter

    52

  • Fixed Magnet Moving Coil Ohmmeter

    The two moving coils moves in permanent magnetic field.

    53

  • Fixed Magnet Moving Coil Ohmmeter

    Torque produced is:

    54

  • Fixed Magnet Moving Coil Ohmmeter

    The two torques produced acts in opposite directions

    55

  • Fixed Magnet Moving Coil Ohmmeter

    At equilibrium,

    56

  • Fixed Magnet Moving Coil Ohmmeter

    Current I1 is proportional to voltage drop across unknown resistance X

    1 =

    57

  • Fixed Magnet Moving Coil Ohmmeter

    That means deflection

    tan =1

    2

    That means deflection depends on value of X

    58

  • Fixed Magnet Moving Coil Ohmmeter

    If X is not in circuit then pointer will set parallel to voltage coil which means X= infinity ohms

    If X is shorted, maximum current flows through current coil , which will set the pointer parallel to current coil, which means X= zero ohms

    59

  • Measurement of low resistance

    60

    Me

    asu

    re

    me

    nt

    of

    low

    r

    esi

    st

    an

    ce

    Ammeter Voltmeter method

    Potentiometer method

    Kelvin Double Bridge method

    Ohm meter method

    Series type

    Shunt type

    Crossed coil Ohm meters

    Fixed magnet moving coil Ohmmeter

    Crossed coil moving magnet Ohmmeter

  • Crossed coil Moving magnet type Ohm meter

    61

  • Crossed coil Moving magnet type Ohm meter

    Consists of :

    Two fixed coils

    Pivoted magnetic needle attached with a pointer

    62

  • Crossed coil Moving magnet type Ohm meter

    Case 1) Terminal ab is opened

    No current flows through current coil

    Bottom portion of the magnetic needle moves in the direction of pressure coil

    Pointer shows resistance is zero ohms

    63

  • Crossed coil Moving magnet type Ohm meter

    Case 2) Terminal ab is shorted

    Current flowing through current coil is maximum

    Bottom portion of magnetic needle move towards current coil

    Resistance shown by the magnetic needle is zero

    64

  • Crossed coil Moving magnet type Ohm meter

    Case 3) If unknown resistance X is connected across ab

    Deflection of the magnetic needle will be proportional to ratio of currents through pressure coil and current coil

    65

  • Measurement of medium resistance

    Measurement of medium resistance

    Ammeter Voltmeter method

    Substitution method

    Wheatstone bridge method

    Carey Foster Bridge method

    Ohmmeter method

    66

  • Measurement of medium resistance

    Measurement of medium resistance

    Ammeter Voltmeter method

    Substitution method

    Wheatstone bridge method

    Carey Foster Bridge method

    Ohmmeter method

    67

  • Ammeter Voltmeter method

    Same as measuring low resistance

    68

  • Measurement of medium resistance

    Measurement of medium resistance

    Ammeter Voltmeter method

    Substitution method

    Wheatstone bridge method

    Carey Foster Bridge method

    Ohmmeter method

    69

  • Substitution method

    Method 1

    R is variable resistance

    X is unknown resistance

    First X is put in the circuit, value of current is noted

    Then X is removed and value of current is noted

    70

  • Substitution method

    Method 2

    Initially switch is at position 1, current is measured

    Then switch put to position 2, current is measured

    71

  • Measurement of medium resistance

    Measurement of medium resistance

    Ammeter Voltmeter method

    Substitution method

    Wheatstone bridge method

    Carey Foster Bridge method

    Ohmmeter method

    72

  • Wheatstones Bridge

    73

  • Wheatstones Bridge(contd)

    For measuring accurately any electrical resistance Wheatstone bridge is widely used.

    There are two known resistors, one variable resistor and one unknown resistor connected in bridge form as shown below.

    By adjusting the variable resistor the current through the Galvanometer is made zero.

    When the electric current through the galvanometer becomes zero, the ratio of two known resistors is exactly equal to the ratio of adjusted value of variable resistance and the value of unknown resistance.

    In this way the value of unknown electrical resistance can easily be measured by using a Wheatstone Bridge.

    74

  • Wheatstones Bridge(contd)

    It is a four arms bridge circuit where arm AB, BC, CD and AD are consisting of electrical resistances P, Q, S and R respectively.

    Among these resistances P and Q are known fixed electrical resistances and these two arms are referred as ratio arms.

    An accurate and sensitive Galvanometer is connected between the terminals B and D through a switch S2

    75

  • Wheatstones Bridge(contd)

    The voltage source of this Wheatstone bridge is connected to the terminals A and C via a switch S1 as shown.

    A variable resistor S is connected between point C and D.

    The potential at point D can be varied by adjusting the value of variable resistor S.

    76

    http://www.electrical4u.com/ideal-dependent-independent-voltage-current-source/

  • Wheatstones Bridge(contd)

    Suppose current I1 and current I2are flowing through the paths ABC and ADC respectively. If we vary the electrical resistance value of arm CD the value of current I2 will also be varied as the voltage across A and C is fixed.

    77

  • Wheatstones Bridge(contd)

    If we continue to adjust the variable resistance one situation may comes when voltage drop across the resistor S that is I2.S is becomes exactly equal to voltage drop across resistor Q that is I1.Q.

    Thus the potential at point B becomes equal to the potential at point D hence potential difference between these two points is zero hence current through galvanometer is nil.

    Then the deflection in the galvanometer is nil when the switch S2 is closed.

    78

  • Wheatstones Bridge(contd)

    79

  • Wheatstones Bridge(contd)

    Now potential of point B in respect of point C is nothing but the voltage drop across the resistor Q.

    80

  • Wheatstones Bridge(contd)

    Again potential of point D in respect of point C is nothing but the voltage drop across the resistor S .

    81

  • Wheatstones Bridge(contd)

    Equating, equations (i) and (ii) we get,

    82

  • Wheatstones Bridge(contd)

    The electrical resistances P and Q of the Wheatstone bridge are made of definite ratio such as 1:1; 10:1 or 100:1 known as ratio arms and S the rheostat arm is made continuously variable from 1 to 1,000 or from

    1 to 10,000

    Wheat stones bridge can be used for measurement of resistances upto 100000ohms

    83

  • Wheatstones bridge mthod

    Limitations

    While measuring low resistance, resistance of leads and contacts become significant resulting in error

    While measuring high resistances, galvanometer becomes insensitive to imbalance

    Change in resistance of bridge arms due to heating effect

    84

  • Measurement of medium resistance

    Measurement of medium resistance

    Ammeter Voltmeter method

    Substitution method

    Wheatstone bridge method

    Carey Foster Bridge method

    Ohmmeter method

    85

  • Carey Foster Bridge

    86

  • Carey Foster Bridge

    In the diagram to the right, X and Y are resistances to be compared.

    P and Q are nearly equal resistances, forming the other half of the bridge.

    The bridge wire EF has a jockey contact D placed along it and is slid until the galvanometer G measures zero.

    The thick-bordered areas are thick copper busbars of almost zero resistance.

    87

    http://en.wikipedia.org/wiki/Busbar

  • Carey Foster Bridge

    Place a known resistance in position Y.

    Place the unknown resistance in position X.

    Adjust the contact D along the bridge wire EF so as to null the galvanometer.

    This position (as a percentage of distance from E to F) is l1.

    88

  • Carey Foster Bridge

    Swap X and Y. Adjust D to the new null point. This position is l2.

    If the resistance of the wire per percentage is , then the resistance difference is the resistance of the length of bridge wire between l1 and l2:

    89

  • Carey Foster Bridge

    90

  • Carey Foster Bridge

    Two resistances to be compared, X and Y, are connected in series with the bridge wire.

    Thus, considered as a Wheatstone bridge, the two resistances are X plus a length of bridge wire, and Y plus the remaining bridge wire.

    The two remaining arms are the nearly equal resistances P and Q, connected in the inner gaps of the bridge.

    91

  • Carey Foster Bridge

    Let l1 be the null point D on the bridge wire EF in percent.

    is the unknown left-side extra resistance EX

    is the unknown right-side extra resistance FY

    is the resistance per percent length of the bridge wire:

    92

  • Carey Foster Bridge

    93

  • Carey Foster Bridge

    Equations 1 and 2 have the same left-hand side and the same numerator on the right-hand side, meaning the denominator on the right-hand side must also be equal

    94

  • Carey Foster Bridge

    95

  • Carey Foster Bridge method

    Advantages high accuracy, unaffected by thermoelectric emfs

    Limitations unsuitable for measuring low resistances

    96

  • Measurement of medium resistance

    Measurement of medium resistance

    Ammeter Voltmeter method

    Substitution method

    Wheatstone bridge method

    Carey Foster Bridge method

    Ohmmeter method

    97

  • Ohm meter method

    Same as measuring low resistance using ohm meter

    98

  • Measurement of high resistance

    Measurement of high resistance

    Direct deflection method

    Megger method

    Loss of charge method

    Mega Ohm bridge method

    99

  • Problems associated with measurement of high resistances

    1) Leakage that occurs over and around the component or specimen under test , or over the binding posts by which the component is attached to the instrument or within the instrument

    - If not controlled resistance measured will be a combined effect

    - Effect of leakage paths on measurements can be removed by Guard circuit

    100

  • Problems associated with measurement of high resistances

    2) Electrostatic effect stray charges appear in the measuring circuit- so errors

    3) Absorption by insulating materials- current falls fairly and steeply in the beginning and gradually thereafter

    4) Resistance of insulating materials- falls rapidly with increasing temperature

    5) Galvanometer employed should be highly sensitive 6) Voltage supply of 100V or more should be applied

    depending on breakdown voltage of the component 7) One point of the circuit should be effectively

    grounded obtaining definite ratios in the potential distribution

    101

  • Guard circuit

    102

  • Guard circuit

    Case (a) without guard circuit

    High resistance is measured by Ammeter- Voltmeter method

    Micro ammeter carries sum of leakage current and resistance current (IR+IL)

    So reading in the ammeter will not be accurate due to error caused by leakage current.

    103

  • Guard circuit

    Case (b)- With guard

    Guard terminal surrounds the resistance entirely and is connected to battery side of micro ammeter

    So current through micro ammeter is IR only and hence resistance can be measured accurately

    104

  • Simple guard circuit

    105

  • Guarded wheat stones bridge

    106

  • Measurement of high resistance

    Measurement of high

    resistance

    Direct deflection

    method

    Megger method

    Loss of charge method

    Mega Ohm bridge method

    107

  • Direct deflection method

    108

    Measurement of insulation resistance of a cable

  • Direct deflection method

    Case (a) cables with metal sheath

    Galvanometer G measures the current between core and metal sheath

    Leakage currents over the surface of insulating material are carried by the guard wire wound on the insulation and does not flow through the insulation

    The ration of voltage applied between the core and metal sheath and current flowing between them(galvanometer deflection) gives insulation resistance of the cable

    109

  • Direct deflection method

    Case (b)

    Cable is immersed in water for at least 24hrs, so that it enters pores of the cable

    Initially galvanometer should be shunted, if possible it should be connected in series with a high resistance(MegaOhms)

    Leakage current flows through guard wire

    Ration of voltmeter reading to galvanometer deflection gives the value of insulation resistance

    110

  • Direct deflection method

    Limitations

    Galvanometer should be highly sensitive

    Galvanometer should be prevented from initial inrush of currents

    Battery should be at least 500V and its emf should remain constant

    111

  • Measurement of high resistance

    Measurement of high

    resistance

    Direct deflection method

    Megger method

    Loss of charge method

    Mega Ohm bridge method

    112

  • MEGGER METHOD

    113

  • MEGGER METHOD

    Working principle: Electromagnetic induction

    When a current carrying conductor is placed in a magnetic field it experiences force whose magnitude depends on strength of current and magnetic field

    114

  • MEGGER METHOD

    It consists of three coils:

    Current coil or deflection coil

    Pressure or control coil

    Compensating coil

    Coils are mounted on a central shaft and is free to rotate along a C shaped structure

    115

  • MEGGER METHOD

    It also consists of

    Permanent magnet to Ohmmeter and dc generator

    116

  • MEGGER METHOD

    The coils are connected to the circuit with flexible leads called ligaments

    Current coil is connected in series with resistance R1 connected to T2 and one generator terminal

    In the event of short, R1 protects the current coil

    117

  • MEGGER METHOD

    Potential coil

    One end connected to compensating coil in series with protective resistance R2

    Other end is connected to generator

    Compensating coil is used for better scale operations

    118

  • MEGGER METHOD

    When the test terminals T1 T2 is open, no current flows through current coil

    Some current flows through voltage coil, so the pointer move towards infinity ohm reading

    119

  • MEGGER METHOD

    If test terminals are shorted, high current passes through current coil and the scale reading will be zero ohms

    120

  • MEGGER METHOD

    When a high resistance is connected between T1 and T2 deflection of pointer is proportional to ratio of currents through pressure coil and current coil

    Guard terminal is provided to eliminate leakage current

    121

  • MEGGER METHOD APPLICATIONS

    Measurement of high resistance

    Measurement of insulation resistance

    Used for testing continuity, pointer shows full deflection if continuity is there between two points in a circuit

    122

  • Measurement of high resistance

    Measurement of high

    resistance

    Direct deflection method

    Megger method

    Loss of charge method

    Mega Ohm bridge method

    123

  • Loss of charge method

    124

  • Loss of charge method

    Step 1: Capacitor C is charged by battery- by keeping switch in position 1

    Step 2: Capacitor C is discharged via Rx and Rleak

    125

  • Loss of charge method

    Step 3: Time (t) taken for the potential difference to fall from V1 to V2 is noted during discharge.

    =

    +

    126

  • Loss of charge method

    At the time of discharge:

    =

    =

    =

    127

  • Loss of charge method

    =

    =

    128

  • Loss of charge method

    =

    0

    21

    2

    1=

    2

    1= ( )

    From the above

    expression Reff can be determined

    129

  • Loss of charge method

    The test is then repeated with Rleak only.

    So value of Rleak is found and from the expression of resistance RX that is unknown value of resistance is found.

    130

  • Measurement of high resistance

    131

    Measurement of high

    resistance

    Direct deflection method

    Megger method

    Loss of charge method

    Mega Ohm bridge method

  • Mega Ohm Bridge Method

    132

  • Mega Ohm Bridge Method

    It consits of:

    Power supply

    Bridge members

    Amplifier

    Indicating instrument

    133

  • Mega Ohm Bridge Method

    Sensitivity for high resistance is obtained by :

    1) using high voltages of

    500V or 1000V

    2) Use of sensitive null

    indicating arrangement

    such as high gain

    amplifier with CRO or

    electronic voltmeter

    134

  • Mega Ohm Bridge Method

    Dial R2 is calibrated as

    1- 10 100 1000M

    Dial R2 1-10 is in logarithmic scale

    Unknown resistance is

    3 =142

    Junction of arms R1and R2 is brought out on the main panel and designated as Guard Terminal

    135

  • EARTHING AND MEASUREMENT OF EARTH RESISTANCE

    What is meant by earthing?

    How will you measure earth resistance?

    136

  • EARTHING

    The connection of electrical machinery/equipment to a general mass of earth, with a conducting material of low resistance is called earthing or grounding

    The conducting material used is known as earth electrode

    137

  • ADVANTAGES OF USING EARTH ELECTRODE

    All parts of the electrical equipment will be at zero potential.

    Leakage current flows through low resistance path provided by the earth electrode so human protection

    Voltage spikes/ current spikes due to lightning or short circuits or other faults will easily get dissipated to earth.

    138

  • ADVANTAGES OF USING EARTH ELECTRODE

    In the case of three phase system, neutral is earthed which helps to maintain line voltage constant

    For telephone and traction work, earthing acts as return path. So cost of cable and cast of such cable is avoided.

    Earth electrode ensures low resistance path and hence able to carry leakage currents without deterioration.

    139

  • MEASUREMENT OF EARTH RESISTANCE

    Measurement of earth resistance

    Fall of potential method

    Megger earth tester

    140

  • MEASUREMENT OF EARTH RESISTANCE

    Measurement of earth

    resistance

    Fall of potential

    method

    Megger earth tester

    141

  • FALL OF POTENTIAL METHOD

    142

  • FALL OF POTENTIAL METHOD

    Potential E is applied

    Current I circulates through the E and Q

    Voltage between E and P is noted

    143

  • FALL OF POTENTIAL METHOD

    Pattern of current flow through earth:

    144

  • FALL OF POTENTIAL METHOD

    Current diverge from E

    Current converge at Q

    Current density is high near E and Q

    Near electrodes, voltmeter reads high, where as between the electrodes

    145

  • FALL OF POTENTIAL METHOD

    The potential V rises near E and Q due to current density

    In the middle section V remains constant

    146

  • FALL OF POTENTIAL METHOD

    Value of earth resistance is given by:

    =

    147

  • FALL OF POTENTIAL METHOD

    The measurement of VEP is done at various points between E and Q

    The potential variation curve is shown in fig

    Resistance RE is determined when the potential curve is absolutely flat To get accurate reading,

    distance between P and Q should be large

    148

  • FALL OF POTENTIAL METHOD

    Variation of resistance with distance is shown

    149

  • MEASUREMENT OF EARTH RESISTANCE

    Measurement of earth

    resistance

    Fall of potential method

    Megger earth tester

    150

  • MEGGER EARTH TESTER

    151

  • MEGGER EARTH TESTER

    It consists of

    DC generator

    Current reverser

    Rectifier

    Current coil

    Potential coil

    Electrodes E,P and Q

    152

  • MEGGER EARTH TESTER

    Current reverser and rectifier have L type commutators

    These are mounted on the shaft and rotated with handle

    Two brushes of commutator are arranged so that it they make contact alternately with each segment of the commutator

    153

  • MEGGER EARTH TESTER

    Other two brushes of commutator are placed in such away that they always make contact with the commutator

    154

  • MEGGER EARTH TESTER

    Earth tester consists of the terminals P1, C1, P2,C2

    P1 andd C1 is shorted and connected to common point E

    P2 and C2 are connected to auxiliary electrodes P and Q respectively

    155

  • MEGGER EARTH TESTER

    The ratio of voltage sensed by voltage coil and current passing through current coil, directly gives the value of earth resistance RE

    Deflection of the pointer gives RE

    It can be used for dc purposes only, but to measure ac reverser and rectifier is used

    156

  • MEGGER EARTH TESTER

    AC current through soil prevents back emf in the soil due to electrolytic action

    157

  • MEASUREMENT OF POWER

    Measurement of power is done by wattmeters

    Wattmeter is a combination of Ammeter and Voltmeter.

    So it contains current coil and voltage coil(pressure coil)

    158

  • MEASUREMENT OF POWER

    159

    Wattmeters

    Dynamometer type

    Induction type

    Electrostatic type

  • Wattmeter

    160

    Current coil carries load current

    Pressure coil carries current proportional to voltage

    Inductance of pressure coil should be minimum to avoid phase lag between current and voltage

  • MEASUREMENT OF POWER

    161

    Wattmeters

    Dynamometer type

    Induction type

    Electrostatic type

  • Dynamometer type wattmeter

    162

  • Dynamometer type wattmeter

    Fixed coil current flowing is proportional to load current

    Moving coil- current flowing is proportional to load voltage

    163

  • Dynamometer type wattmeter

    Strength of magnetic field- proportional to currents through two coils

    164

  • Dynamometer type wattmeter

    165

    V- supply voltage

    I load current

    R- resistance of moving

    coil circuit

  • Dynamometer type wattmeter

    Fixed coil current: =

    Moving coil current:

    =

    Deflecting torque:

    =

    166

  • Dynamometer type wattmeter

    For dc circuit deflecting torque is proportional to power

    For ac circuit deflecting torque is proportional to voltage, current and power factor

    =

    167

  • Dynamometer type wattmeter

    1

    2

    M- mutual inductance between

    two coils

    Instantaneous torque is given by:

    = 12

    168

  • Dynamometer type wattmeter

    Instantaneous voltage in pressure coil is:

    = 2

    Instantaneous current through pressure coil is:

    =2

    = 2

    - resistance of

    pressure coil

    169

  • Dynamometer type wattmeter

    Current through current coil lags voltage by an angle

    = 2 I sin

    170

  • Dynamometer type wattmeter

    Instantaneous torque is given by:

    171

  • Dynamometer type wattmeter

    Average deflecting torque is given by:

    172

  • Dynamometer type wattmeter

    Control torque is =

    K- spring constant

    - final steady state deflection

    At balance position

    = ()

    173

  • Dynamometer type wattmeter

    At balance condition:

    174

  • Shape of scale of dynamometer wattmeter

    175

  • Shape of scale of dynamometer wattmeter

    Deflection is proportional to power measured

    and scale is uniform since

    is constant.

    Wattmeters are designed such that

    remains over 40 to 50 degree on each side

    of zero mutual inductance position.

    M varies linearly in this zone with respect to

    176

  • Shape of scale of dynamometer wattmeter

    If zero mutual inducatnce position is kept in the middle then M varies linearly for deflection upto 80 to 100 degrees

    177

  • Shape of scale of dynamometer wattmeter

    178

  • Dynamometer type wattmeter

    Ranges:

    Current coil: 0.25A to 100A

    Pressure coil : 5V to 750V

    179

  • Dynamometer type wattmeter

    Dynamometer type wattmeter

    Suspended coil torsion type

    Pivoted coil indicating type

    180

  • Suspended coil torsion type dynamometer wattmeter

    181

  • Suspended coil torsion type dynamometer wattmeter

    The moving, or voltage, coil is suspended from a torsion head by a metallic suspension which serves as a lead to the coil.

    This coil is situated entirely inside the current or fixed coils and the winding in such that the system is a static.

    Errors due to external magnetic fields are thus avoided. The torsion heads carries a scale, and when in use, the

    moving coil is bought back to the zero position by turning this head; the number of divisions turned through when multiplied by a constant for the instrument gives the power.

    Eddy currents are eliminated as far as possible by winding the current coils of standard wire and by using no metal parts within the region of the magnetic field of the instrument.

    182

  • Suspended coil torsion type dynamometer wattmeter

    The mutual inductance errors are completely eliminated by making zero position of the coil such that the angle between the planes of moving coil and fixed coil is 90 degree. i.e. the mutual inductance between the fixed and moving coil is zero.

    The elimination of pivot friction makes possible the construction of extremely sensitive and accurate electrodynamic instruments of this pattern.

    183

  • Pivoted coil indicating type dynamometer wattmeter

    184

  • Pivoted coil indicating type dynamometer wattmeter

    In these instruments, the fixed coil is wound in two halves, which are placed in parallel to another at such a distance, that uniform field is obtained.

    The moving coil is wound of such a size and pivoted centrally so that it does not project outside the field coils at its maximum deflection position.

    The springs are pivoted for controlling the movement of the moving coil, which also serves as currents lead to the moving coil.

    185

  • Pivoted coil indicating type dynamometer wattmeter

    The damping is provided by using the damping vane attached to the moving system and moving in a sector-shaped box.

    The reading is indicated directly by the pointer attached to the moving system and moving over the calibrated scale.

    The eddy current errors, within the region of the magnetic field of the instrument, are minimized by the use of non-metallic parts of high resistivity material.

    186

  • Electrodynamometer type wattmeter

    Advantages:

    1) In dynamometer type wattmeter, the scale of the instrument is uniform (because deflecting torque is proportional to the true power in both DC as well as AC and the instrument is spring controlled.)

    2) High degree of accuracy can be obtained by careful design; hence these are used for calibration purposes.

    187

  • Electrodynamometer type wattmeter

    Disadvantages:

    1) The error due to the inductance of the pressure coil at low power factor is very serious (unless special features are incorporated to reduce its effect)

    2) In dynamometer type wattmeter, stray field may affect the reading of the instrument. To reduce it, magnetic shielding is provided by enclosing the instrument in an iron case.

    188

  • Errors in dynamometer type wattmeter

    Errors

    Due to pressure coil inductance

    Due to pressure coil capacitance

    Due power loss in pressure coil and current coil

    Due to Eddy current

    Due to friction

    Due to temperature

    Due to stray fields

    189

  • Errors in dynamometer type wattmeter

    Errors

    Due to pressure coil inductance

    Due to pressure coil capacitance

    Due power loss in pressure coil and current coil

    Due to Eddy current

    Due to friction

    Due to temperature

    Due to stray fields

    190

  • Error due to pressure coil inductance

    Current through pressure coil will be in phase with voltage applied

    Due to inductance of pressure coil, current in pressure coil lags behind supply voltage

    So there will be error in reading

    191

  • Error due to pressure coil inductance

    192

  • Error due to pressure coil inductance

    Pressure coil current lagging behind voltage by an angle

    193

  • Error due to pressure coil inductance

    Let power factor be lagging

    194

  • Error due to pressure coil inductance

    195

  • Error due to pressure coil inductance

    196

  • Error due to pressure coil inductance

    197

  • Error due to pressure coil inductance

    If power factor is leading:

    198

  • Error due to pressure coil inductance

    199

  • Error due to pressure coil inductance

    200

  • Compensation for Error due to pressure coil inductance

    201

  • Compensation for Error due to pressure coil inductance

    202

  • Compensation for Error due to pressure coil inductance

    203

  • Errors in dynamometer type wattmeter

    Errors

    Due to pressure coil inductance

    Due to pressure coil capacitance

    Due power loss in pressure coil and current coil

    Due to Eddy current

    Due to friction

    Due to temperature

    Due to stray fields

    204

  • Error due to pressure coil capacitance

    Pressure coil has capacitance as well as inductance

    Capacitance is due to inter-turn capacitance in the high values of series resistor.

    The effect produced is same as that of inductance circuit except that pressure coil currents leads applied voltage

    So wattmeter will read low on lagging power factors of load, by increasing angle between load and voltage coil currents

    205

  • Error due to pressure coil capacitance

    Effect of frequency: vary angle between V and pressure coil current, the angle is increasing with frequency.

    If inductive reactance of pressure coil = capacitive reactance of pressure coil, there will be no error

    206

  • Errors in dynamometer type wattmeter

    Errors

    Due to pressure coil inductance

    Due to pressure coil capacitance

    Due power loss in pressure coil and current coil

    Due to Eddy current

    Due to friction

    Due to temperature

    Due to stray fields

    207

  • Error due to power loss in pressure coil and current coil

    Let Rp be the resistance of pressure coil

    Rc be resistance of current coil

    208

  • Error due to power loss in pressure coil and current coil

    There are two methods of connecting wattmeter

    209

  • Error due to power loss in pressure coil and current coil

    Voltage applied to the pressure coil = voltage across load + voltage drop across current coil

    So wattmeter measures power loss in current coil in addition to power consumed by load

    210

  • Error due to power loss in pressure coil and current coil

    211

  • Error due to power loss in pressure coil and current coil

    Current coil carries current to the load and current to pressure coil

    So wattmeter reads:

    power consumed by load + power loss in pressure coil

    212

  • Error due to power loss in pressure coil and current coil

    213

  • Error due to power loss in pressure coil and current coil

    Power loss in pressure coil is less comapared to current coil

    Connection (a) is preferred for small currents

    Where as vice versa for high currents

    214

  • Compensation for Error due to power loss in pressure coil

    Compensating coil is used to eliminate error due to current coil carrying pressure coil current in addition to load current

    Compensating coil is coincident and identical to current coil

    If compensating coil is connected in series with current coil, current passed through two coils produces resultant magnetic field = zero

    215

  • Compensation for Error due to power loss in pressure coil

    Compensating coil is connected in series with pressure coil so that magnetic field opposes that of current coil and neutralize pressure coil component of current in current coil

    So if no load current flows through instrument, deflection will be zero

    216

  • Compensation for Error due to power loss in pressure coil

    217

  • Errors in dynamometer type wattmeter

    Errors

    Due to pressure coil inductance

    Due to pressure coil capacitance

    Due power loss in pressure coil and current coil

    Due to Eddy current

    Due to friction

    Due to temperature

    Due to stray fields

    218

  • Error due to eddy currents

    Eddy currents are induced in solid metal parts of instrument by alternating magnetic field

    The phase of induced emf will be 90 degree behind inducing flux

    Eddy currents are practically in phase with its emf and it sets up a magnetic field which is combined with that of current coil

    So a resultant magnetic field is produced which is less than current coil alone and which lags behind current coil by small angle

    219

  • Error due to eddy currents

    Eddy current error cannot be easily calculated

    If metal parts are more in instrument it is significantly high

    220

  • Errors in dynamometer type wattmeter

    Errors

    Due to pressure coil inductance

    Due to pressure coil capacitance

    Due power loss in pressure coil and current coil

    Due to Eddy current

    Due to friction

    Due to temperature

    Due to stray fields

    221

  • Error due to friction

    Deflecting torque is very less in the instrument

    So there is frictional error

    To reduce frictional error

    Weight of moving parts should be reduced

    Greater care must be taken on pivoting

    222

  • Errors in dynamometer type wattmeter

    Errors

    Due to pressure coil inductance

    Due to pressure coil capacitance

    Due power loss in pressure coil and current coil

    Due to Eddy current

    Due to friction

    Due to temperature

    Due to stray fields

    223

  • Error due to temperature

    As temperature changes

    Resistance of pressure coil changes

    Stiffness of springs changes

    The above effects are opposite in action

    So they neutralize

    If pressure coil composed of copper and of resistance alloy having negligible resistance, temperature coefficient is 1: 10

    224

  • Errors in dynamometer type wattmeter

    Errors

    Due to pressure coil inductance

    Due to pressure coil capacitance

    Due power loss in pressure coil and current coil

    Due to Eddy current

    Due to friction

    Due to temperature

    Due to stray fields

    225

  • Error due to stray fields

    Dynamometer wattmeter has weak operating field

    It is affected by stray magnetic fields resulting in errors

    So these instruments are shielded against effect of stray magnetic fields

    Lamination sheets are used in portable lab equipments, steel cases for switch board instruments

    Precision type is not shielded to keep down eddy current errors

    226

  • Low power factor electrodynamometer wattmeter

    Ordinary electrodynamometer wattmeter are not suitable for measuring power at low power factors due to:

    1) Small deflecting torque on the moving system even when pressure coil and current coil are excited.

    2) Introduction of large error due to inductance of pressure coil

    227

  • Low power factor electrodynamometer wattmeter

    So features incorporated in lpf meters are:

    228

  • Low power factor electrodynamometer wattmeter

    229

  • Low power factor electrodynamometer wattmeter

    230

  • Low power factor electrodynamometer wattmeter

    231

  • POWER FACTOR METER

    232

  • POWER FACTOR METER

    But due to errors in meters used, the method mentioned is not accurate

    So a meter is necessary to measure power factor directly which is called power factor meter

    233

  • POWER FACTOR METER

    Construction of power factor meter is similar to that of wattmeter

    Current circuit carries current or fraction of current whose power factor needs to be measured

    Voltage coil is split into two parts

    Inductive

    Non inductive

    234

  • POWER FACTOR METER

    The currents in two paths of are proportional to voltage across the circuit

    Deflection depends on current through current circuit and currents in two branches of voltage circuit

    235

  • TYPES OF POWER FACTOR METERS

    Power factor meters

    Electrodynamometer type

    Moving iron type

    Rotating field type

    Alternating field type

    236

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER

    237

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER

    Construction is same as that of wattmeter

    F1-F2 fixed coils connected in series

    A-B moving coils with their axes in quadrature

    A-B moves together and carries the pointer showing power factor

    238

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER

    F1-F2 carries main current

    If current is high, then a portion of current is passed through it

    Magnetic field produced around F1-F2 is proportional to current

    239

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER

    Coil A is connected in series with R

    Coil B is connected in series with L

    Values of R and L are adjusted so that coils A and B carries equal currents at normal frequency

    At normal frequency

    R = L

    240

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER

    Current in coil A is in phase with supply voltage due to R

    Current in coil B is in quadrature with supply voltage due to L

    Current in coil B is frequency dependent due to L

    Current in coil A is frequency independent due to R

    241

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER

    Current in coils A and B produces magnetic fields of equal strength displaced by 90 degrees

    242

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER- working

    X-X uniform magnetic field produced by fixed coils

    Due to interaction of magnetic fileds produced by each coils, torque is developed

    Torque experienced by coil A and coil B are opposite to each other

    Pointer attains equilibrium when two torques are equal

    243

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER- working

    244

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER- working

    245

  • ELECTRODYNAMOMETER TYPE POWER FACTOR METER- working

    So angular position taken up by moving coil is equal to system power factor angle

    Operation of meter is depended only on frequency and wave form

    246

  • MOVING IRON POWER FACTOR METER

    247

    Disadvantage: Accuracy is less

  • MOVING IRON POWER FACTOR METER TYPES

    Rotating field type

    Alternating field type

    248

  • ROTATING FIELD TYPE MOVING IRON POWER FACTOR METER

    249

  • ROTATING FIELD TYPE MOVING IRON POWER FACTOR METER

    Consists of three fixed coils F1,F2 and F3 displaced by 1200

    These coils are fed from three CTs

    Coil F1- phase R

    Coil F2- phase Y

    Coil F3 phase B

    250

  • ROTATING FIELD TYPE MOVING IRON POWER FACTOR METER

    Coil Q- kept in middle of F1,F2 and F3 and is connected across any two lines of the supply through a series resistance

    251

  • ROTATING FIELD TYPE MOVING IRON POWER FACTOR METER

    Inside coil Q there is a short pivoted iron rod

    The rod carries two sector shaped vanes I1, I2

    The rod also carries damping vanes and the pointer

    252

  • ROTATING FIELD TYPE MOVING IRON POWER FACTOR METER Coil Q and soft iro system

    produces alternating flux

    This flux interacts with fluxes produced by F1,F2 and F3

    Due to R current through coil Q is in phase with supply

    So deflection of moving system is proportional to power factor angle

    253

  • ROTATING FIELD TYPE MOVING IRON POWER FACTOR METER Coils F1, F2, F3

    displaced by 120 degrees produces rotating field by induction motor action

    It keeps the system rotating

    Due to high resistivity iron parts this motion is reduced by reducing induced currents

    254

  • ROTATING FIELD TYPE MOVING IRON POWER FACTOR METER

    This meter can be used for balanced loads

    It is also called

    Westinghouse power factor meter

    255

  • ALTERNATING FIELD TYPE MOVING IRON POWER FACTOR METER

    256

  • ALTERNATING FIELD TYPE MOVING IRON POWER FACTOR METER

    Spindle carries pointer, damping vanes and three moving irons

    The moving irons are sector shaped

    Moving iron have 120 degrees with respect to each other

    Q1,Q2,Q3- iron sectors

    257

  • ALTERNATING FIELD TYPE MOVING IRON POWER FACTOR METER

    Iron sectors are magnetized by voltage coils P1,P2 and P3

    Current coil is divided into two equal parts F1 and F2

    Current coil carries one of the three line currents

    258

  • ALTERNATING FIELD TYPE MOVING IRON POWER FACTOR METER When connected in

    circuit, moving system moves and attains a position in which mean torque in one of the iron pieces is neutralized by other two

    At this position deflection is proportional to phase angle between currents and voltages in three phase system

    259

  • ALTERNATING FIELD TYPE MOVING IRON POWER FACTOR METER

    The voltage coils are at different levels hence resultant flux is alternating

    This instrument is also called Nalder-Lipman power factor meter

    260

  • ENERGY

    261

  • ENERGY

    262

  • ENERGY METER TYPES

    263

    Energy meter

    Single phase

    Three phase

  • SINGLE PHASE ENERGYMETER CONSTRUCTION

    264

  • SINGLE PHASE ENERGYMETER CONSTRUCTION- PARTS

    265

    Driving system

    Moving system

    Braking system and

    Registering system.

  • Driving system

    consists of two electromagnets, called shunt magnet and series magnet, of laminated construction.

    266

  • Driving system

    A coil having large number of turns of fine wire is wound on the middle limb of the shunt magnet.

    This coil is known as pressure or voltage coil and is connected across the supply mains.

    This voltage coil has many turns and is arranged to be as highly inductive as possible.

    In other words, the voltage coil produces a high ratio of inductance to resistance.

    This causes the current, and therefore the flux, to lag the supply voltage by nearly 900.

    267

  • Driving system

    An adjustable copper shading rings are provided on the central limb of the shunt magnet to make the phase angle displacement between magnetic field set up by shunt magnet and supply voltage is approximately 90degrees.

    The copper shading bands are also called the power factor compensator or compensating loop

    268

  • Driving system

    The series electromagnet is energized by a coil, known as current coil which is connected in series with the load so that it carry the load current.

    The flux produced by this magnet is proportional to, and in phase with the load current.

    269

  • Moving system

    The moving system essentially

    consists of a light rotating aluminium disk mounted on a vertical spindle or shaft.

    The shaft that supports the aluminium disk is connected by a gear arrangement to the clock mechanism on the front of the meter to provide information that consumed energy by the load

    270

  • Moving system

    The time varying (sinusoidal) fluxes produced by shunt and series magnet induce eddy currents in the aluminium disc.

    The interaction between these two magnetic fields and eddy currents set up a driving torque in the disc.

    The number of rotations of the disk is therefore proportional to the energy consumed by the load in a certain time interval and is commonly measured in killowatt-hours (Kwh).

    271

  • Braking system

    Damping of the disk is provided by a small permanent magnet, located diametrically opposite to the a.c magnets.

    The disk passes between the magnet gaps.

    272

  • Braking system

    The movement of rotating

    disc through the magnetic field crossing the air gap sets up eddy currents in the disc that reacts with the magnetic field and exerts a braking torque.

    By changing the position of the brake magnet or diverting some of the flux there form, the speed of the rotating disc can be controlled.

    273

  • Registering or Counting system

    The registering or counting system essentially consists of gear train, driven either by worm or pinion gear on the disc shaft, which turns pointers that indicate on dials the number of times the disc has turned.

    274

  • Registering or Counting system

    The energy meter thus determines and adds together or integrates all the instantaneous power values so that total energy used over a period is thus known.

    Therefore, this type of meter is also called an integrating meter

    275

  • Working/operation of single phase energy meter

    276

  • Working/operation of single phase energy meter

    Induction instruments operate in alternating-current circuits and they are useful only when the frequency and the supply voltage are approximately constant

    The rotating element is an aluminium disc, and the torque is produced by the interaction of eddy currents generated in the disc with the imposed magnetic fields that are produced by the voltage and current coils of the energy meter.

    277

  • Working/operation of single phase energy meter

    Let us consider a sinusoidal flux (t) is acting perpendicularly to the plane of the aluminium disc, the direction of eddy current (Ie) by Lenzs law is indicated in figure

    278

  • Working/operation of single phase energy meter

    279

    =0 since reactance of Al dsc is zero

  • Working/operation of single phase energy meter

    So in all induction type meters, two eddy currents are there so that resultant torque will be there

    280

  • Working/operation of single phase energy meter

    281

  • Working/operation of single phase energy meter

    Current coil produces two fluxes in opposite directions

    So torques produced by the interaction of eddy current due to voltage and current coil is opposite at two points.

    Hence the disc will start to rotate

    282

  • Derivation of Torque equation

    Phasor Diagram

    283

  • Derivation of Torque equation

    284

  • Derivation of Torque equation

    285

  • Derivation of Torque equation

    286

  • Derivation of Torque equation

    287

  • ENERGYMETER CONSTANT

    288

  • SOURCES OF ERRORS IN SINGLE PHASE ENERGYMETER

    Incorrect magnitude of fluxes due to abnormal voltages and load currents

    Incorrect phase relation of fluxes due to defective lagging, abnormal frequencies, changes in iron loss

    Unsymmetrical magnetic structure disc rotates when pressure coils alone is excited

    289

  • SOURCES OF ERRORS IN SINGLE PHASE ENERGYMETER

    Changes in resistance of disc due to change in temperature

    Changes in strength of drag magnets due to temperature and ageing

    Phase angle errors due to lowering of power factor

    Abnormal deflection of moving parts

    Badly distorted waveform

    Changes in retarding torque of the disc

    290

  • ERRORS IN SINGLE PHASE ENERGY METER

    Erro

    rs

    Phase error

    Frictional Error

    Creeping

    Speed Error

    Temperature error

    Overload compensation

    Voltage compensation

    291

  • Phase error

    An error due to incorrect adjustment of the position of shading band results an incorrect phase displacement between the magnetic flux and the supply voltage (not in quadrature)

    This is tested with 0.5 p.f. load at the rated load condition

    292

  • Compensation for phase error

    By adjusting the position of the copper shading band in the central limb of the shunt magnet this error can be eliminated.

    293

  • Compensation for phase error

    A lag coil is placed as shown in figure so that phase angle between Voltage and flux is exactly 90 degrees

    This kind of compensation is called

    Power factor adjustment or Quadrature adjustment

    Or Inductive load adjustment

    294

  • Frictional Error

    Frictional forces at bearings and registering mechanism give rise to unwanted braking torque on the disc

    So additional driving torque is required

    295

  • Compensation for frictional error

    The two shading bands on the limbs are adjusted to create this extra torque.

    This adjustment is done at low load (at about 1/4th of full load at unity p.f.).

    296

  • Creeping Error

    In some meters a slow but continuous rotation is seen when pressure coil is excited but with no load current flowing.

    This slow revolution records some energy.

    This is called the creep error.

    This slow motion may be due to

    (a) incorrect friction compensation,

    (b) stray magnetic field

    (c) for over voltage across the voltage coil.

    297

  • Compensation for creeping error

    This can be eliminated by drilling two holes or slots in the disc on opposite side of the spindle.

    When one of the holes comes under the poles of shunt magnet, the rotation being thus limited to a maximum of 180 degrees

    298

  • Compensation for creeping error

    In some cases, a small piece of iron tongue or vane is fitted to the edge of the disc.

    When the position of the vane is adjacent to the brake magnet, the attractive force between the iron tongue or vane and brake magnet is just sufficient to stop slow motion of the disc with full shunt excitation and under no load condition.

    299

  • Speed Error

    Due to the incorrect position of the brake magnet, the braking torque is not correctly developed.

    This can be tested when meter runs at its full load current alternatively on loads of unity power factor and a low lagging power factor

    300

  • Compensation for speed error

    The speed can be adjusted to the correct value by varying the position of the braking magnet towards the centre of the disc or away from the centre and the shielding loop.

    If the meter runs fast on inductive load and correctly on non-inductive load, the shielding loop must be moved towards the disc.

    On the other hand, if the meter runs slow on non-inductive load, the brake magnet must be moved towards the center of the disc.

    301

  • Temperature Error

    Energy meters are almost inherently free from errors due to temperature variations.

    Temperature affects both driving and braking torques equally (with the increase in temperature the resistance of the induced-current path in the disc is also increases) and so produces negligible error.

    A flux level in the brake magnet decreases with increase in temperature and introduces a small error in the meter readings

    302

  • Compensation for temperature error

    This error is frequently taken as negligible, but in modern energy meters compensation is adopted in the form of flux divider on the brake magnet

    303

  • Overload compensation

    When the disc rotates in the field of series magnetic field under load conditions, it cuts series flux and dynamically induced emfs is produced on the disc

    This produces eddy currents on the disc

    Due to interaction of eddy current flux and series magnet flux, braking torque is produced

    This is proportional to square of current

    304

  • Overload compensation

    This braking torque is called self braking torque and if load is high it causes serious errors in the instrument

    To minimize this braking torque, full load speed of the disc is limited to 40rpm

    The current coil series flux is kept minimum with respect to shunt coil flux

    305

  • Overload compensation

    Practically an overload compensating device in the form of saturable magnet shunt is used

    At high loads, shunt saturates and diverts some series magnetic flux.

    This compensates for self braking torque

    306

  • Voltage compensation

    When supply voltage varies, energymeter causes errors

    This is due to:

    Non linear magnetic characteristics of shunt magnet core

    Braking torque is proportional to square of supply voltage

    307

  • Voltage compensation

    Voltage compensation is provided by saturable magnetic shunt

    It diverts a large proportion of flux into the active path when supply voltage increases

    This is done by increasing side limb reluctance and providing holes in side limbs

    308

  • ADVANTAGES OF INDUCTION TYPE ENERGYMETER

    309

  • DISADVANTAGES OF INDUCTION TYPE ENERGYMETER

    310

  • THREE PHASE ENERGYMETER

    Three phase system

    Four wire Three element energy meter

    Three wire Two element energy meter

    311

  • THREE PHASE THREE ELEMENT ENERGYMETER

    312

  • THREE PHASE THREE ELEMENT ENERGYMETER

    It consists of three elements

    Each element is similar to that of single phase energy meter

    Pressure coils are P1,P2 and P3

    Current coils are C1, C2 and C3

    All elements are mounted in a vertical line in common case and have a common spindle, gearing and recording mechanism

    313

  • THREE PHASE THREE ELEMENT ENERGYMETER

    The coils are connected in such a manner that the net torque produced is equal to sum of torques produced by each element

    These are employed for three phase four wire system, where fourth wire is the neutral wire

    Current coils are connected in series with the lines where as pressure coils are connected in parallel across line and neutral

    314

  • THREE PHASE THREE ELEMENT ENERGYMETER

    One unit of three phase energy meter is cheaper than three individual units

    Due to interaction of eddy currents between all elements, errors are produced which are reduced by suitable adjustments

    315

  • THREE PHASE TWO ELEMENT ENERGYMETER

    316

  • THREE PHASE TWO ELEMENT ENERGYMETER

    It is provided with two discs for an element

    Shunt magnet is carrying pressure coil

    Series magnet is carrying current coil

    Pressure coils are connected in parallel

    Current coils are connected in series

    Torque is produced in same manner as that of single phase energy meter

    The total torque on registering mechanism is sum of torques on two discs

    317

  • ELECTRONIC ENERGY METER

    318

  • ELECTRONIC ENERGY METER

    Average power = mean product of instantaneous voltage across the load and instantaneous current through load

    Potential divider- for making voltage to required level

    Voltage is scaled in the required range using voltage scaling device

    Current scaling device scales load voltage which is proportional to load current

    319

  • ELECTRONIC ENERGY METER

    Both scaled voltages are connected to voltage and current multiplier unit

    Voltage and current multiplier unit outputs current as a result of product of ac voltage and current

    The current is proportional to instantaneous power applied to voltage controlled oscillator

    VCO works on the principle of constant current charging capacitor

    320

  • ELECTRONIC ENERGY METER

    VCO basically voltage to frequency converter Output of VCO is square wave The frequency of square wave is proportional to

    output current of VCO So power dependent current and frequency

    dependent current decides the value of consumed energy

    ADC (analog to digital converter) converts analog signal to digital signal

    Display unit displays energy in watt-hour

    321

  • ADVANTAGES OF ELECTRONIC ENERGYMETER

    High sensitivity

    No frictional losses

    Less loading effect

    Low load, full load, creeping adjustments are not required

    High frequency range

    High accuracy of 1%

    322

  • MAXIMUM DEMAND METER

    ASSIGNMENT

    What is maximum demand meter?

    Explain the following meters with neat figures

    1) Merz price maximum demand indicator

    2) Thermal type maximum demand indicator

    3) Digital Maximum demand indicator

    323

  • TRIVECTOR METER

    324

  • TRIVECTOR METER

    325

  • TRIVECTOR METER

    326

  • TRIVECTOR METER

    Ratchet coupling is linked to main common register shaft to which final drive from each gear system is connected

    Shaft is always driven by direct drive which has the maximum speed

    At that time all other slower shafts are idle on ratchets.

    327

  • TRIVECTOR METER

    As power factor changes, other gear drive drives the shaft at higher speed and drive shifts to another ratchet

    For a given V-I product, speed of kWh meter varies as power factor varies

    328

  • TRIVECTOR METER VARIATION OF PERCENTAGE SPEED vs

    PHASE ANGLE

    329

  • TOD METER

    Electric company supplies electricity to various loads such as domestic, industrial and commercial purposes

    These loads varies over various time periods

    For some time load is maximum and for sometime load is minimum

    The hours in which load is maximum is called peak load hours

    330

  • TOD METER

    The hours in which load is minimum is called off-peak hours

    During on-peak hours company has to generate more power to supply for the demand.

    This causes difficulties

    331

  • TOD METER

    Time of delay rate is special service offered by electric company that allows consumer to take advantage of lower electricity price during a certain time period

    Consumer can save money being on TOD rate

    To lessen the load during a particular time of a day, company offers special rate to the consumers who are willing shift the load or portion of the load to off-peak hours

    332

  • TOD METER

    A special metering arrangement is done to measure energy consumption during different time zones of the day including on-paek and off-peak hours

    Trivector meter itself is provided with capability required

    It is provided with a time of delay registser (TOD register) which is capable of being progarammed during off-peak and on-peak hours

    333

  • TOD METER

    TOD meters are time- of delay meter which is suitable of recording and indicating consumption during specific time periods of the day

    Trivector meter with such an arrangement is called TOD meter

    334

  • TOD METER

    335