Wako Diagnostics
Extensive linear range (10-500 g/mL)
No significant effects from common interferents
Simple, fast, ready to use reagents
Because microalbuminuria may be reversible if diabetes is well controlled, its early detection may be very beneficial. However, microalbuminuria is not generally measurable by the simple test ordinarily used to detect proteinuria because it lacks sensitivity. For the detection of microalbuminuria, RIA, ELISA and nephelometry assay are used. However they have complicated procedures and are time consuming.
Wakos Microalbumin is a turbidimetric immunoassay using polyclonal antibodies against human albumin and allows accurate and reproducible measurement of microalbuminuria with a simple procedure.
Catalog No. Product Name Pkg Size Storage
999-06001* Autokit Micro Albumin
Buffer 2 x 45 mL 2-10C
Antibody 1 x 15 mL 2-10C
995-06101* Autokit Micro Albumin Calibrator Set 5 conc x 2mL 2-10C
991-06201* Autokit Micro Albumin Control Set 2 x 2 conc x 2 mL 2-10C* Product is available as a special purchase only.
Turbidimetric Immunoassay
MicroalbuminFor In Vitro Diagnostic Use
Copyright 2017 Wako Life Sciences, Inc. Wako Life Sciences, Inc. 1025 Terra Bella Ave. Mountain View, CA 94043
E: [email protected] T: 650.210.9153 www.wakodiagnostics.com
Performance characteristics
PrincipleWhen a sample is mixed with the antibody solution in the buffer, albumin in the sample combines specifically with the anti-human albumin antibody to yield immune complexes resulting in increased turbidity. Since the degree of turbidity is proportional to the amount of albumin in the sample, the amount of albumin can be determined by measuring the turbidity.
Correlation
Prozone Phenomenon
WLS-1952-2
PrecisionWithin-run precision
Sample # Replicates Mean (g/mL) SD CV (%)
1 21 4.97 0.09 1.81
2 21 105.81 0.79 0.75
3 21 510.84 3.91 0.77
Total precisionNumber of assay days
Mean (g/mL) SD CV (%) SWT ST
21 4.90 0.14 2.86 0.23 0.23
21 511.21 7.68 1.50 7.65 9.78
AccuracyThe accuracy of this method was demonstrated by a recovery study.
No. Expected value (g/mL)Obtained value
(g/mL)Recovery
(%)
1 33.1 32.6 98.4
2 58.1 56.9 97.8
3 118.1 117.6 99.6
0 100 200 300 400 500 6000
100
200
300
400
500
600
y = 1.0179x - 0.9619r = 0.9992
Wak
o Au
toki
t Mic
ro A
lbum
in,
g/m
L
Wako Micro Albumin B, g/mL
0 5000 10000 15000 20000 25000 30000 350000
200
400
600
800
1000
1200
1400
Obt
aine
d M
icro
Alb
umin
con
c.,
g/m
L
Micro Albumin expected conc., g/mL
Prozone phenomenon not observed up to 10000 g/mL
InstrumentsVarious automated analyzer applications are available.
For more product information, please contact customer service at 1-877-714-1924.