Methodology: Plethysmogram is a measure of optical impedance -- Obtained from tail cuff with...

1
Methodology: Plethysmogram is a measure of optical impedance -- Obtained from tail cuff with photo-sensor (Figure 1) -- Blood pressure α blood volume -- Blood volume α optical impedance -- Optical impedance α voltage -- voltage α pressure applied by tail cuff Speaker valve transduces voltage to pressure (Figure 1) -- increase in voltage = increased displacement of speaker diaphragm -- increase in displacement of speaker diaphragm = increase in leak occlusion -- increase in leak occlusion = increase in pressure in cuff Analog circuit alters the plethysmographic signal and controls the speaker (Figure 4) --adjustable offset - used to alter DC offset = non-oscillatory cuff pressure - controlled by adjusting the supply voltage to the power amplifier (LM386-4 self centers output to ½ supply voltage) --inversion - need in increase in cuff pressure for an increase in arterial volume - increase in arterial volume = increase optical impedance = decrease in plethysmogram - since the speaker valve increases pressure with increasing voltage, must invert plethysmographic signal -- adjustable gain - plethysmogram must be scaled to create large enough fluctuations in pressure to maintain constant arterial volume - must be adjustable because the magnitude of pressure oscillations will vary between subjects and over time within subjects -gain and inversion accomplished with an inverting amplifier (LM741) Open-loop mode -- sets up system: centers cuff pressure oscillations around mean arterial pressure -- set cuff-applied pressure to mean arterial pressure -- Marey's Law – plethysmographic signal will be maximal at mean arterial pressure (Figure 6) -- increase pressure until plethysmographic signal is maximal Closed-loop mode -- measures continuous BP: creates dynamic, pulsatile unloading of the arterial wall -- applied pressure already set at mean arterial pressure -- allow plethysmographic signal to control pressure oscillations -- scale plethysmographic signal until the unadulterated plethysmographic signal is minimal -- pressure fluctuations created by the inverted and amplified plethysmographic signal equal the intra- arterial blood pressure waveform. Non-invasive Continuous Blood Pressure Measurement in Mice Andrew McClellan 1 , Laura Miller 1 , Sachin Paranjape 2 , Andre’ Diedrich, M.D., Ph.D. 1,2 1 Department of Biomedical Engineering, Vanderbilt University, 2 Autonomic Dysfunction Center, Department of Medicine, Vanderbilt University Medical Center; Nashville, Tennessee Introduction: --Several transgenic and gene knockout mouse models have been developed to study human diseases, with research relying on clinical measurements of blood pressure (BP) regulation: --Hypertension --Disorders of the cardiovascular system --Dysfunctions of the autonomic nervous system (orthostatic intolerance, multiple system atrophy, neurally-mediated syncope) --Diabetes --Spectral analysis techniques in the frequency domain have been used to assess the function of blood pressure regulation and autonomic nervous system function using data from clinical tests: 1,2 --Pharmacological baroreflex and hemodynamic response tests --Behavorial studies --Exercise and food studies **Continuous beat-to-beat sampling of blood pressure is required, reporting a blood pressure waveform as a function of time (vs. single systolic/diastolic values) Current devices for BP measurement in mice: --Inflatable tail cuff with photo-sensor: uses Korotkoff method, not continuous measurements --Implantable catheter/RF transmitter: highly invasive, expensive, time- consuming --Arterial line: invasive, cumbersome, time-consuming Objective: We aim to adopt the Penaz method 3 for continuous BP measurement at the mouse tail. The Penaz method has been successfully implemented with measurement at the human finger (Finapres™), so we will investigate whether the method can be used with mouse physiology, where heart rates can be between 500-700 bpm, and blood pressures can be between 70- 200mmHg. We will use a mouse tail cuff with photo-sensor that is currently used to measure single systolic and diastolic BP values. Recommendations: -- The major shortcoming of the system lies in the transduction from voltage to pressure with the speaker. A higher wattage power amplifier and speaker could produce more sufficient mechanical oscillations in the speaker. Other types of valves such as a solenoid, could also be viable alternatives. Theory: To non-invasively measure blood pressure waveforms the Penaz, or volume clamp, method was used. With this method, the amount of pressure applied to the artery that is needed to maintain the mean arterial volume is the blood pressure waveform. This calls for dynamic pulsatile unloading of the arteries. Thus, during systole, when the blood pressure increases, arteries will slightly expand due to their elasticity. To maintain a constant arterial volume, a pressure is applied equal to that of the intra-arterial pressure. Likewise, during diastole, when blood pressure decreases, the applied pressure is lessened. By measuring the applied pressure necessary to maintain a constant arterial volume, a blood pressure waveform can be obtained. Voltage to Pressure Transduction WinDAQ Data Acquisition Speaker Valve Lab air source Pressure Source LM386-4 Power Amplification Adjustable Gain Inverting Amplifier w/ potentiometer (LM741) Signal Inversion Pressure Meter Tail cuff Harvard Rat Blood Pressure Monitor Plethysmogra m Implementati on Need References: 1. Iellamo F, et al.:(1996) Evaluation of reproducibility of spontaneous baroreflex sensitivity at rest and during laboratory tests J. Hypertension 1996; 4: 1099-1104 2. Dawson SL, et al.:(1997) The reproducibility of cardiac baroreceptor activity assessed non-invasively by spectral and sequence techniques Clin. Auton. Research; 7: 279-84 3. Penaz, J. (1973) Photoelectric measurement of blood pressure, volume and flow in the finger. Digest. 10th Int. Conf. Med. Biol. Eng. Dresden, 104 Tail Cuff Pressure Monitor Power Supply Speaker Valve Analog Circuit Computer/ DAQ Figure 1. Top: tail cuff with photo- sensor Bottom: mouse restrained in tail cuff, speaker valve. Figure 2. Continuous blood pressure measurement system. Figure 4. Analog circuit to modify (invert, amplify, offset) the plethysmographic signal for output to the speaker valve. Results: Conclusions: --The open-loop mode of our system was achieved. Marey’s Law was demonstrated with our system, as we were able to center the cuff pressure around a mouse’s mean arterial pressure according to the appearance of the maximum plethysmographic signal (Figure 6a). We were also able to alter the speaker offset for voltage centering, where a change DC voltage offset resulted in a proportional change in constant pressure (Figure 5b). --The concept of the closed loop mode was demonstrated. The system was able to roughly regulate cuff-applied pressure with the inversion of the plethysomographic signal at its same frequency, as described by the Penaz Method (i.e. counteracting the blood pressure to achieve constant arterial volume) (Figure 8). The amplitude of the cuff-applied pressure could be controlled by changing the gain in the circuit (Figure 6b). --However, the maximum mechanical oscillations produced by the speaker valve were only enough to produce oscillatory changes in the cuff- applied with a magnitude of 1 mmHg. Without pressure changes of the same value as the mean arterial pressure of a mouse, complete arterial unloading could not be achieved. However, the plethysmographic signal was decreased during measurement, suggesting partial arterial unloading. Additionally, the shape of the pressure wave, as transduced by the speaker valve, was altered. --The Penaz Method, successful in humans, has potential for adaptation to mouse physiology and would be a safer, more cost effective, and more ethical way to obtain continuous blood pressure measurements. Voltage to Pressure C alibration C urve y = 0.0006x R 2 = 1 0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 0.1 0 20 40 60 80 100 120 140 160 P ressure (m m Hg) Voltage (v) Figure 5a. Left: calibration curve from input voltage to speaker to cuff-applied pressure. Figure 5b. Right: Change in constant cuff-applied pressure with change in DC voltage offset. Figure 6. Demonstration of Maney’s Law. When cuff-applied pressure (top) reached the mouse’s mean arterial pressure, the plethysmographic signal (bottom) reached its maximum and then diminished when the vessel became fully occluded. Figure 7. Cuff-applied pressure over time as the circuit gain was decreased. Pressure amplitude changed proportionally to change in gain. Figure 3. Continuous blood pressure measurement system schematic. Harvard Plethysmogram & Pressure Monitor WinDAQ pressure signal plethysmogra m Analog Circuit Tail Cuff with Photosensor invert and amplify offset and amplify power Pressure Source Speaker Valve Figure 8. Simulated plethysmographic signal (10Hz, 22mV p-p , shown in red) and cuff-applied pressure (grey) vs time. Cuff-applied pressure is 1 mmHg. Time scale is 20ms/div. 1 mmHg Figure 9. Incomplete arterial unloading. Plethysmographic signal from mouse (red) and cuff-applied pressure (grey) vs. time. Cuff-applied pressure was not enough to provide constant volume to the artery and completely diminish plethysmographic signal, however it was decreased suggesting partial arterial unloading. Time scale is 20ms/div.

Transcript of Methodology: Plethysmogram is a measure of optical impedance -- Obtained from tail cuff with...

Page 1: Methodology:  Plethysmogram is a measure of optical impedance -- Obtained from tail cuff with photo-sensor (Figure 1) -- Blood pressure α blood volume.

Methodology: Plethysmogram is a measure of optical impedance

-- Obtained from tail cuff with photo-sensor (Figure 1)-- Blood pressure α blood volume-- Blood volume α optical impedance -- Optical impedance α voltage-- voltage α pressure applied by tail cuff

Speaker valve transduces voltage to pressure (Figure 1)-- increase in voltage = increased displacement of speaker diaphragm-- increase in displacement of speaker diaphragm = increase in leak occlusion-- increase in leak occlusion = increase in pressure in cuff

Analog circuit alters the plethysmographic signal and controls the speaker (Figure 4)--adjustable offset

- used to alter DC offset = non-oscillatory cuff pressure- controlled by adjusting the supply voltage to the power amplifier

(LM386-4 self centers output to ½ supply voltage)--inversion

- need in increase in cuff pressure for an increase in arterial volume- increase in arterial volume = increase optical impedance = decrease in plethysmogram- since the speaker valve increases pressure with increasing voltage, must invert plethysmographic signal

-- adjustable gain- plethysmogram must be scaled to create large enough fluctuations in pressure to maintain constant arterial volume- must be adjustable because the magnitude of pressure oscillations will vary between subjects and over time within subjects-gain and inversion accomplished with an inverting amplifier (LM741)

Open-loop mode -- sets up system: centers cuff pressure oscillations around mean arterial pressure-- set cuff-applied pressure to mean arterial pressure-- Marey's Law – plethysmographic signal will be maximal at mean arterial pressure (Figure 6)-- increase pressure until plethysmographic signal is maximal

Closed-loop mode-- measures continuous BP: creates dynamic, pulsatile unloading of the arterial wall-- applied pressure already set at mean arterial pressure-- allow plethysmographic signal to control pressure oscillations-- scale plethysmographic signal until the unadulterated plethysmographic signal is minimal-- pressure fluctuations created by the inverted and amplified plethysmographic signal equal the intra- arterial blood pressure waveform.

Non-invasive Continuous Blood Pressure Measurement in MiceAndrew McClellan1, Laura Miller1, Sachin Paranjape2, Andre’ Diedrich, M.D., Ph.D.1,2

1Department of Biomedical Engineering, Vanderbilt University, 2Autonomic Dysfunction Center, Department of Medicine, Vanderbilt University Medical Center; Nashville, Tennessee

Introduction:--Several transgenic and gene knockout mouse models have been developed to study human diseases, with research relying on clinical measurements of blood pressure (BP) regulation:

--Hypertension--Disorders of the cardiovascular system--Dysfunctions of the autonomic nervous system (orthostatic intolerance, multiple system

atrophy, neurally-mediated syncope)--Diabetes

--Spectral analysis techniques in the frequency domain have been used to assess the function of blood pressure regulation and autonomic nervous system function using data from clinical tests:1,2

--Pharmacological baroreflex and hemodynamic response tests--Behavorial studies--Exercise and food studies

**Continuous beat-to-beat sampling of blood pressure is required, reporting a blood pressure waveform as a function of time (vs. single systolic/diastolic values)

Current devices for BP measurement in mice:--Inflatable tail cuff with photo-sensor: uses Korotkoff method, not continuous measurements--Implantable catheter/RF transmitter: highly invasive, expensive, time-consuming--Arterial line: invasive, cumbersome, time-consuming

Objective:We aim to adopt the Penaz method3 for continuous BP measurement at the mouse tail. The Penaz method has been successfully implemented with measurement at the human finger (Finapres™), so we will investigate whether the method can be used with mouse physiology, where heart rates can be between 500-700 bpm, and blood pressures can be between 70-200mmHg. We will use a mouse tail cuff with photo-sensor that is currently used to measure single systolic and diastolic BP values.

Recommendations:-- The major shortcoming of the system lies in the transduction from voltage to pressure with the speaker. A higher wattage power amplifier and speaker could produce more sufficient mechanical oscillations in the speaker. Other types of valves such as a solenoid, could also be viable alternatives.

Theory:To non-invasively measure blood pressure waveforms the Penaz, or volume clamp, method was used. With this method, the amount of pressure applied to the artery that is needed to maintain the mean arterial volume is the blood pressure waveform. This calls for dynamic pulsatile unloading of the arteries. Thus, during systole, when the blood pressure increases, arteries will slightly expand due to their elasticity. To maintain a constant arterial volume, a pressure is applied equal to that of the intra-arterial pressure. Likewise, during diastole, when blood pressure decreases, the applied pressure is lessened. By measuring the applied pressure necessary to maintain a constant arterial volume, a blood pressure waveform can be obtained.

Voltage to Pressure Transduction

WinDAQData Acquisition

Speaker Valve

Lab air sourcePressure Source

LM386-4Power Amplification

Adjustable Gain

Inverting Amplifier w/ potentiometer (LM741)

Signal Inversion

Pressure Meter

Tail cuffHarvard Rat

Blood Pressure Monitor

Plethysmogram

ImplementationNeed

References: 1. Iellamo F, et al.:(1996) Evaluation of reproducibility of spontaneous baroreflex sensitivity at rest and during laboratory tests J. Hypertension 1996; 4: 1099-11042. Dawson SL, et al.:(1997)  The reproducibility of cardiac baroreceptor activity assessed non-invasively by spectral and sequence techniques Clin. Auton. Research; 7: 279-843. Penaz, J. (1973) Photoelectric measurement of blood pressure, volume and flow in the finger. Digest. 10th Int. Conf. Med. Biol. Eng. Dresden, 104

Tail Cuff

Pressure Monitor

Power Supply

Speaker Valve

Analog Circuit

Computer/DAQ

Figure 1. Top: tail cuff with photo-sensorBottom: mouse restrained in tail cuff, speaker valve.

Figure 2. Continuous blood pressure measurement system.

Figure 4. Analog circuit to modify (invert, amplify, offset) the plethysmographic signal for output to the speaker valve.

Results:

Conclusions:--The open-loop mode of our system was achieved. Marey’s Law was demonstrated with our system, as we were able to center the cuff pressure around a mouse’s mean arterial pressure according to the appearance of the maximum plethysmographic signal (Figure 6a). We were also able to alter the speaker offset for voltage centering, where a change DC voltage offset resulted in a proportional change in constant pressure (Figure 5b).

--The concept of the closed loop mode was demonstrated. The system was able to roughly regulate cuff-applied pressure with the inversion of the plethysomographic signal at its same frequency, as described by the Penaz Method (i.e. counteracting the blood pressure to achieve constant arterial volume) (Figure 8). The amplitude of the cuff-applied pressure could be controlled by changing the gain in the circuit (Figure 6b).

--However, the maximum mechanical oscillations produced by the speaker valve were only enough to produce oscillatory changes in the cuff-applied with a magnitude of 1 mmHg. Without pressure changes of the same value as the mean arterial pressure of a mouse, complete arterial unloading could not be achieved. However, the plethysmographic signal was decreased during measurement, suggesting partial arterial unloading. Additionally, the shape of the pressure wave, as transduced by the speaker valve, was altered.

--The Penaz Method, successful in humans, has potential for adaptation to mouse physiology and would be a safer, more cost effective, and more ethical way to obtain continuous blood pressure measurements.

Voltage to Pressure Calibration Curve

y = 0.0006xR2 = 1

0

0.01

0.02

0.03

0.04

0.05

0.06

0.07

0.08

0.09

0.1

0 20 40 60 80 100 120 140 160

Pressure (mmHg)

Vo

ltag

e (v

)

Figure 5a. Left: calibration curve from input voltage to speaker to cuff-applied pressure.

Figure 5b. Right: Change in constant cuff-applied pressure with change in DC voltage offset.

Figure 6. Demonstration of Maney’s Law. When cuff-applied pressure (top) reached the mouse’s mean arterial pressure, the plethysmographic signal (bottom) reached its maximum and then diminished when the vessel became fully occluded.

Figure 7. Cuff-applied pressure over time as the circuit gain was decreased. Pressure amplitude changed proportionally to change in gain.

Figure 3. Continuous blood pressure measurement system schematic.

Harvard Plethysmogram & Pressure Monitor WinDAQ

pressure signal

plethysmogram

Analog CircuitTail Cuff with Photosensor invert and

amplifyoffset and

amplify power

Pressure Source

Speaker Valve

Figure 8. Simulated plethysmographic signal (10Hz, 22mVp-p, shown in red) and cuff-applied pressure (grey) vs time. Cuff-applied pressure is 1 mmHg. Time scale is 20ms/div.

1 mmHg

Figure 9. Incomplete arterial unloading. Plethysmographic signal from mouse (red) and cuff-applied pressure (grey) vs. time. Cuff-applied pressure was not enough to provide constant volume to the artery and completely diminish plethysmographic signal, however it was decreased suggesting partial arterial unloading. Time scale is 20ms/div.