Human impedance variability and defibrillator test protocol

Why 50 Ω loads are not enough to test modern defibrillators

Page 2: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Defibrillators: life saving

• Convert certain heart arrhythmias to normal• Convert v-fib back to normal• Have to work first time, every time.

– Batteries may not be fully charged or faulty– Charging circuits may be faulty, may not deliver

enough energy– Must deliver proper energy to account for differing

patient sizes/impedances

Page 3: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Current vs. energy

• Current—not energy—defibrillates.• Successful defibrillation requires enough

current be delivered to the heart muscle during the shock.– Must transit through the chest/thorax and the

impedance it represents.– Body mass, skin resistance, tissue type and

amount all play a part in the chest/thorax impedance presented to the charge delivered by the defibrillator.

Page 4: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Transthoracic impedance

• Transthoracic impedance = the body’s resistance to current flow

• Human impedance variability has been shown to vary from 25 ohms to 180 ohms1

• Energy in respect to impedance is the determining factor to successful defibrillation—not energy alone.

Page 5: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Transthoracic impedance (cont.)

• Just a few known causes of differing impedances in human beings include:– Body mass– Age– Disease– Skin resistance– Tissue type and amount

Page 6: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Transthoracic impedance (cont.)

• Successful defibrillation requires sufficient current to the heart muscle.

• Defibrillation current is affected by transthoracic impedance– Modern (biphasic) defibrillators

measure impedance and adjust energy delivery accordingly

“Humans exhibit a wide range of transthoracic impedance, and defibrillators compensate for this range of impedance in different ways.”

-- McDaniel, Garret, Burke and Arzbaecher2

Page 7: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Schematic and formula

EH = I2 x RH x Time(Watt Seconds or Joules)Defibrillator

Page 8: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

How defibrillators account for human impedance variability

• Monophasic (older) defibrillators– Current flows only one direction

impedance not measured

Page 9: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

How defibrillators account for human impedance variability (cont.)

• Biphasic (modern) defibrillators– Current flows first in one direction,

then reverses and flows in opposite direction.

– Impedance measured and energy delivery is adjusted internally based on energy setting or (for AEDs) arryhthmia.

Page 10: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Biphasic waveforms

Q: Are all biphasic waveforms alike?

A: No. There are different biphasic waveforms based on different manufacturer’s specifications.

Rectilinear Biphasic (ZOLL)

Smart Biphasic(Philips)

Pulsed Biphasic(Schiller)

Page 11: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Clinical impedance examples

• Low impedance (50 Ω)– A 360-joule biphasic defibrillator delivers more current than

required, exposing patient to potentially harmful high-peak currents

• Average impedance (75 Ω)– A 360-joule biphasic defibrillator and a 200-joule rectilinear

defibrillator may be equally effective

• High impedance (> 100 Ω)– A 200-joule rectilinear shock delivers a higher average

current than a 360-joule biphasic defibrillator shock, therefore making it more effective at lower energy levels.

Page 12: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Impedance and defibrillation outcomes

• Too little or too much is no good– Too much peak current during shock can injure the heart– It’s the peak current (not energy) that can injure the heart

It is imperative the current delivered to the heart is matched to the individual impedance of each patient and not just the delivered energy.

Page 13: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Load testing: accounting for impedance variability

Q: Are 50 Ω test loads enough to ensure output conditions of modern-day defibrillators?– Do all of your hospital’s patients have the same

input impedance? No!

A: Testing beyond the 50 Ω load is necessary to ensure defibrillator inventory performance of modern defibrillators

Page 14: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

Load testing: accounting for impedance variability (cont.)

Q: What loads are recommended?

A: Section 6.8.3 of the IEC 60601-2-4 standard3 and AAMI DF80 standards4 require defibrillators to be tested on different resistance loads of 25, 50, 75, 100, 125, 150 and 175 ohms to ensure proper current is delivered to patients with different impedances.

Page 15: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

When to test with other loads

• Each hospital’s equipment history should be considered.

• Standard protocol recommends multiple impedance tested on incoming inspection and after a repair to charging circuit or applied parts.

• May also be helpful when troubleshooting “failed to deliver energy” complaints.

Page 16: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

The 7010 solution

• Designed to facilitate load testing for modern defibrillator technology quality assurance and standards compliance

• Selectable loads– 25 Ω– 50 Ω– 75 Ω– 100 Ω– 125 Ω– 150 Ω– 175 Ω– 200 Ω

• The only device available today capable of testing beyond 175 ohms for extreme impedance conditions.

Impulse 7010 Selectable Load Accessory

Page 17: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

References

1. American Heart Association. Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Care. Circulation Supplement. 2000; 102:8

2. McDaniel W.C., Garrett, M, Burke M.C., Arzbaecher R, Comparison of the Efficacy of Two Biphasic Waveform Defibrillators in a Model of Simulated Higher Impedance Patients, Engineering in Medicine and Biology Soceity, 2003. Proceedings of the 25th Annual International Conference of the IEEE. Volume I, issue 17-21, Sept. 2003 Page(s) 183-185 Vol. I

3. IEC 60601-2-4 Medical Electrical Equipment Part 2-4: Particular Requirements for the Safety of Cardiac Defibrillators, Section 6.8.3

4. ANSI/AAMI DF80:2003 Medical electrical equipment—Part 2-4: Particular Requirements for the Safety of Cardiac Defibrillators (Including Automated External Defibrillators)

Page 18: Human impedance variability and defibrillator test protocol Why 50 Ω loads are not enough to test modern defibrillators.

References (cont.)

Want to know more? Check these out:• Mittal S, Ayati S, Stein KM, Knight BP, Morady F, Schwartzman D, et al.

Comparison of a novel rectilinear biphasic waveform with a damped sine wave monophasic waveform for transthoracic ventricular defibrillation. ZOLL Investigators. J Am Coll Cardiol 1999; 34: 1595-601.

• Schneider T, Martens PR, Paschen H, Kuisma M, Wolcke B, Gliner BE, et al. Multicenter, randomized, controlled trial of 150-J biphasic shocks compared with 200- to 360-J monophasic shocks in the resuscitation of out-of-hospital cardiac arrest victims. Optimized Response to Cardiac Arrest (ORCA) Investigators. Circulation 2000; 102: 1780-7.

• Mittal S, Ayati S, Stein KM, Schwartzman D, Cavlovich D, Tchou PJ, et al. Transthoracic cardioversion of atrial fibrillation: comparison of rectilinear biphasic versus damped sine wave monophasic shocks. Circulation 2000; 101: 1282-7.

• Walker RG, Melnick SB, Chapman FW, Walcott GP, Schmitt PW, Ideker RE. Comparison of six clinically used external defibrillators in swine. Resuscitation 2003; 57: 73-83.

• Niemann JT, Walker RG, Rosborough JP. Ischemically Induced Ventricular Fibrillation (VF): A Comparison of Fixed and Escalating Energy Defibrillation. Acad Emerg Med 2003; 10: 454.