20th February 2007 Fergus Wilson, RAL 1 Experimental Particle Physics Particle Interactions and Detectors

Transcript of Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors...

Page 1: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 1

Experimental Particle Physics

Particle Interactions and Detectors

Page 2: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 2

How do we detect Particles?Particle Types

Charged (e-/K-/π-)Photons (γ)Electromagnetic (e-)Hadronic (K-/π-/μ-)Muonic (μ-)Gravitons !

Interaction with matterIonisation LossRadiation LossPhoton AbsorptionElectromagnetic ShowersHadronic ShowersCherenkov RadiationTransition Radiation

In general, we measure the energy lost as the particle passes through a medium.

Page 3: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 3

Transverse slice through CMS detectorClick on a particle type to visualise that particle in CMS

Press “escape” to exit

Page 4: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 4

Page 5: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 5

Which particles interact with which subdetector?

YesYesHadronicCalorimeter

YesCherenkov

YesMuon Detector

YesYesEM Calorimeter

YesYesTransitionRadiation

YesYesYesTracking

PhotonNeutralHadron(π0)

MuonCharged Hadron(K+/π+)

ElectronDetector

(caveat: some particles leave a small signal in a subdetector e.g. muon in EM calorimeter)

Page 6: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 6

Charged Particle DetectorsPhysics

IonisationMean Energy LossFluctuationsCherenkov LightTransition Radiation

Detectors1. Emulsion2. Bubble Chambers3. Scintillation Detectors4. Wire Chambers5. Multi Wire Proportional

Chambers (MWPC)6. Geiger Muller7. Solid State Devices8. Time Projection (TPC)9. Resistive Plate Counters

(RPC)10. Limited Streamer Tubes (LST)11. Cherenkov12. Transition Radiation (TRD)

Page 7: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 7

Ionisation and Atomic ExcitationHeavy Charged particles interact with electrons in material as they passCan be calculated: The Bethe-Bloch EquationOk for energies between 6 MeV and 6 GeVFunction only of β (approximately)

⎥⎦

⎤⎢⎣

⎡−−=−

2)(2ln

21 2

2max

222

22 βγδβγβ

β ITcm

AZKq

dxdE e

222max 2 γβcmT e≈

Maximum energy loss in single collision

Density correctionIonisation Constant for

material

Constant

≈1/2Stopping

Power

Page 8: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 8

Stopping Power1/β2 ln(β2)

Ionisation Constant

Page 9: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 9

Mean Energy Loss in different materials

⎥⎦

⎤⎢⎣

⎡−≈− 2

2

222

22 2ln βγβ

β Icm

AZKq

dxdE e

Low energy ~ 1/β2

High energy ~ ln γ

Minimum atγ≈3

Distance units:

g cm-2

Page 10: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 10

Energy FluctuationsBethe-Block only gives mean, not most probableLarge high energy tail – δ raysLandau distribution:

δ-rays : electrons produced by the Bethe-Block equation that have sufficient energy to ionize further atoms through subsequent interactions on their own.

Page 11: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 11

Particle Identification by Energy Loss

LAB momentum, GeV/c-110 1 10

corr

ecte

d D

CH

dE

/dx

400

600

800

1000

1200

1400

DCH dE/dx for various particle types, Run 3, data

Results from a Time Projection Chamber

(PEP4/9)

Results from a Drift Chamber

(BaBar)

p

Page 12: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 12

Ionisation Detectors

Ionisation used to detect particles in different ways:1. Observe physical or chemical change due to ions2. Detect energy from recombination - scintillation3. Collect and measure free charges - electronic

Page 13: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 13

EmulsionsExpose film to particles and developNatural radioactivity was discovered this wayStill occasionally used for very high precision, low rateexperimentsSimilar technique in etched plastics

CHORUS (neutrinos)

Page 14: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 14

Bubble Chambers (1960s-1970s)Ionisation trail nucleates bubbles in superheated liquidLiquid H2 (or similar) close to boiling pointSuddenly reduce pressure.Fire beam into chamberTake photoCloud chamber similar: ions nucleate condensation in saturated vapour

BEBC

Page 15: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 15

Scintillation DetectorsDetect photons from electronic

recombination of ionsOrganic (plastic)Inorganic (crystal or glass)

doping normally required

Not very efficient ~1 photon/100eVLight carried to sensitive photodetectorsFast, cheap and flexible

Page 16: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 16

Wire Chambers

Free electrons will be attracted to anodeElectric field near thin wire increasesSecondary ionisation may start to occur

avalanche!

A typical gas detector will have ~20 primary ions per cm created by a track.

+V

e-

e-

e-

e-

Page 17: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 17

Gas Amplification

Full charge collectionStart of avalanche region

Maximum gain ~107 Avalanche fills volume

Arcing

Geiger Muller Tube

Resistive Plate Chambers

Proportional Chambers

Page 18: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 18

Geiger RegionGeiger CounterSpark Chamber

short bias pulse->localise breakdownStreamer Chamber

Large volume, transparent electrodes

Page 19: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 19

MWPC

Need better idea for large volume coverage at high ratesMulti-Wire Proportional Chamber Fast

Ion Drift Velocity ~ 50 km/s (50 μm/ns)Resolution ~pitch/√12x from anodey from ions at segmented cathode plane

Page 20: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 20

Drift Chambers

Electron drift speed depends on electric field and gasTime delay of hit gives distancefrom sense anodeExtra wires can be used to separate drift and avalanche regionsTypical values:

drift distance ~cmdrift time ~μsprecision ~100 μm

Page 21: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 21

Stereo Readout

Good z resolutionNeed readout along lengthGhost hits

Good pattern recognitionReadout from endsPoor z resolution

Page 22: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 22

BaBar Drift Chamber

Open Cell Drift Chamber2.8 m longGas volume ~ 5.6 m3

7100 anode wiresAxial and stereo~50,000 wires in total

Page 23: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 23

Time Projection ChamberIngredients:

GasE.g.: Ar + 10 to 20 % CH4

E-fieldE ~ 100 to 200 V/cm

B-fieldas big as possible to measure momentumto limit electron diffusion

Wire chamberto detect projected tracksTiming gives z measurement

Long drift distances~ metres

y

z

x

E

B drift

chargedtrack

wire chamber to detect projected tracks

gas volume with E & B fields

Page 24: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 24

Detector with TPC

Page 25: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 25

General considerations for Wire Chambers

Gas, voltage and geometry must be chosen carefully.precision, amplification, avalanche characteristics...Chambers can be damaged.

External magnetic field influences behaviour.Must be measured and understood.

MWPC:fast, reliableoften used for triggering

Drift/TPC:large volume, reasonably precisehigh incident fluxes can cause “short circuit”long readout time

Need other solution for high rates and/or extreme precision

Page 26: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 26

Silicon Strip Detector

Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal thickness (0.3mm), and have very fast ( 0.000000025 second) time response.Silicon, a semiconductor, can be fabricated in two forms; n type, with a surplus of electron sites in the crystal lattice, and p type, with a deficit of electron sites in the crystal lattice.The majority of silicon detectors consist of n type bulk material. The back face has an aluminium contact over the complete surface. The front face has p type silicon strips implanted in the surface. These p type strips aluminium strips on their surface. The aluminium strips are separated from their associated p type silicon strips by a thin insulator. An electric field is applied between the p strips and the back face.When a charged particle passes through a silicon detector it creates ionisation in the bulk of the silicon. This frees electrons from the atoms of the silicon and leaving these atoms with an electron vacancy. These vacancies are referred to as "holes". The "holes" "drift" in the electric field towards the negatively charged p type strips. The electrons "drift" towards the positively charged back plane.When the "holes" reach the p type strip they are collected and induce a measurable charge on the associated aluminium strip. Thealuminium strips are connected to sensitive electronic read out channels.By recording which electronic channel fired, it is possible to determine where the charged particle passed through the detector.

Page 27: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 27

Solid State DetectorsDetect ionisation charges in solids

high density → large dE/dx signalmechanically simplecan be very precise

Semiconductorssmall energy to create electron-hole pairssilicon extremely widely used

band gap 1.1 eVmassive expertise and capability in electronics industry

Resistorsplastic – cheapdiamond – robust, rad. hardGermanium – can be made thick

Usually Silicon !

Page 28: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 28

Cerenkov DetectorCerenkov Radiation

speed of light in medium = c/nn = refractive indexcharged particles produce light “shock waves” if v>c/nlight cone cosθ = c/vn = 1/(nβ)“eerie blue glow”Useful for separating pions and kaons

1 2 (m)

θC

mirror

Beampipe

Track

Photodetectors 3 3 0 mrad

C F 4 10

LHCb

Page 29: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 29

Transition Radiation Detector

An energetic charged particle moving through matter momentarily polarizes the material nearby. If the particle crosses a boundary where the index of refraction changes, the change in polarization gives rise to the emission of electromagnetic transition radiation. About one photon is emitted for every 100 boundaries crossed. Transition radiation is emitted even if the velocity of the particle is less than the light velocity of a given wavelength, in contrast to Cerenkov radiation. Consequently, this radiation can take place in the x-ray region of the spectrum where there is no Cerenkov radiation, because the index of refraction is less than one. At each interface between materials, the probability of transition radiation increases with the relativistic gamma factor. Thus particles with large γγgive off many photons, and small γ give off few. For a given energy, this allows a discrimination between a lighter particle (which has a high γand therefore radiates) and a heavier particle (which has a low γ and radiates much less).Useful for separating pions and electrons

GLAST

Page 30: Experimental Particle Physics - Science and Technology ... · Particle physics needs detectors which can determine the position of particles with an accuracy of 0.01 mm, have minimal

20th February 2007 Fergus Wilson, RAL 30

Next Time...

More interactions and detectors