  • DIGITAL MULTIMETER DESCRIPTION

    1. Range Control Button

    2. Data Hold Button

    3. AC/DC Current or Selecting Button

    4. Socket for transistor test

    5. Function switch/Power switch

    6. V//F Input Jack

    7. COM Input Jack

    8. mA/Cx Input Jack

    9. 10A Input Jack

    1. Range control button

    Range for AC/DC voltage, AC/DC current (A and mA only), resistance and frequency

    measuring can be selected manually or handled by autoranging. Push this button as following to

    choose range control mode and desired ranges.

  • 2. Data hold button When this button is pushed, the display will show the last reading and D-H symbol will appear

    until pushing it again.

    Data holding will be cancelled automatically when the function switch is rotated.

    3. AC/DC current or Selecting button Push this button to select AC or DC current measuring function when the function switch is set

    at A, mA, A positions.

    Push this button to select measuring function when the function switch is set at

    position.

    4. Input jacks This device has four input jacks that are protected against limits overload. During use, connect

    the black test lead to the COM jack and the red test lead as shown below:

    Function Red lead connection Input limits

    DCV/ACV V / / F 1000V DC or 750V rms AC

    kHz V / / F 250V DC or rms AC

    / V / / F 250V DC or rms AC

    A / mA mA / Cx 300mA DC or rms AC

    nF / F mA / Cx 300mA fuse protected

    A 10A 10A DC or rms AC

    A / mA ranges are protected by fuses.

  • SPECIFICATIONS

    1. General Display: LCD, 3260 counts max.

    Max. Voltage

    Between Terminals

    and Herat Ground

    1000V DC o 750V rms AC

    Fuse Protection A, mA: F 300 mA / 250V

    Measurement

    Method Dual slope integration A/D converter

    Ranging Method Auto/Manual

    Overrange Indication OL display

    Polarity Indication - display

    Power supply: 9V battery, NEDA 1604 o 6F22

    Operating: 0 to 40 C (32F to 104F) Temperature

    Storage: -10C to 50C (10F to 122F) Operating

    conditions: Humidity < 80%

    Dimension: 91 x 189 x 31.5 mm

    Weight: 310 gr.

    2. DC Voltage Range Resolution Accuracy

    326mV 0.1mV 0.5% of rdg 2 digits

    3.26V 1mV 0.8% of rdg 2 digits

    32.6V 10mV 0.8% of rdg 2 digits

    326V 0.1V 0.8% of rdg 2 digits

    1000V 1V 0.8% of rdg 2 digits

    Input impedance: 10M, more than 100M at 326mV range.

  • 3. AC Voltage

    Range Resolution Accuracy

    3.26V 1mV 1.2% of rdg 3 digits

    32.6V 10mV 0.8% of rdg 3 digits

    326V 0.1V 1.2% of rdg 3 digits

    1000V 1V 1.2% of rdg 3 digits

    Input impedance: 10M.

    Frequency range: 40 to 400Hz, 40 to 200Hz for 3.26V range.

    Response: average, calibrated in rms of sine wave.

    4. DC Current Range Resolution Accuracy Burden Voltage

    326A 0.1A 1.2% of rdg 3 digits 0.5mV / A

    3260A 1A 1.2% of rdg 3 digits 0.5mV / A

    32.6mA 10A 1.2% of rdg 3 digits 8.0mV / mA

    326mA 0.1mA 1.2% of rdg 3 digits 8.0mV / mA

    10A 10mA 4.0% of rdg 3 digits 0.02V / A

    Overload Protection: F 300mA fuse for A and mA ranges.

    5. AC Current Range Resolution Accuracy Burden Voltage

    326A 0.1A 1.5% of rdg 5 digits 0.5mV / A

    3260A 1A 1.8% of rdg 5 digits 0.5mV / A

    32.6mA 10A 1.8% of rdg 5 digits 8.0mV / mA

    326mA 0.1mA 2.0% of rdg 5 digits 8.0mV / mA

    10A 10mA 5.0% of rdg 7 digits 0.02V / A

    Overload protection: F 300mA fuse for A and mA ranges.

    Frequency range: 40 to 400 Hz

    Response: average, calibrated in rms of sine wave.

  • 6. Resistance Range Resolution Accuracy

    326 0.1 1.0% of rdg 3 digits

    3.26k 1 1.0% of rdg 1 digits

    32.6 k 10 1.0% of rdg 1 digits

    326 k 100 1.0% of rdg 1 digits

    3.26M 1k 1.0% of rdg 1 digits

    32.6M 10k 2.0% of rdg 2 digits

    Maximun open circuit voltage: 1.3V

    7. Capacitance Range Resolution Accuracy

    326nF 0.1nF 5.0% of rdg 10 digits

    32.6F 10nF 5.0% of rdg 10 digits

    8. Frequency Range Resolution Accuracy

    32.6kHz 10Hz 1.5% of rdg 10 digits

    150kHz 100Hz 3.0% of rdg 10 digits

    Sensitivity: 500mV rms up to 50kHz, 1V rms for 50kHz to 150kHz.

    BATTERY AND FUSE REPLACEMENT If the sign appears on the LCD display, it indicates that the battery should be replaced.

    Remove screws on the back cover and open the case. Replace the exhausted battery with a

    new one.

    Fuse rarely need replacement, and when blows, it is almost always as a result of the operators

    error. Open the case as mentioned above and take the PCB assembly out form the case.

    Replace the blown fuse rating specified.

    Warning: Before attempting to open the case, be sure that test leads have been disconnected

    from measurement circuit to avoid electric shock hazard.

    For protection against fire, replace fuses only with specified ratings: F1: F 300mA/250V.

  • OPERATING INSTRUCTION 1. Measuring voltage

    Connect the black test lead to the COM jack and the red test lead to the V / / F jack.

    Set the function switch at V V ~ range to be used and connect test leads across the

    source or load under measurement.

    Read LCD display. The polarity of red connection will be indicated when making a DC

    measurement.

    2. Measuring current Connect the black test lead to the COM jack and the red test lead to the mA/Cx jack for

    a maximum of 300mA. For a maximum of 10A, move the red lead to the A jack.

    Set the function switch at A, mA or A range to be used and push / ~ button to select

    DCA or ACA measuring mode.

    Connect test leads in series with the load in which the current is to be measured.

    Read LCD display. The polarity of red lead connection will be indicated when making a

    DC measurement.

    3. Measuring resistance Connect the black test lead to the COM jack and the red test lead to the V / / F jack

    (NOTE: The polarity of the red lead connection is positive +)

    Set the function switch at range to be used and connect test leads across the resistance under measurement.

    NOTE:

    For resistance above 3.26M, the device may take a few seconds to stabilize reading. This is normal for high resistance measuring.

    When the input is not connected, i.e. al open circuit, the figure OL will be displayed for the over-range condition

    When checking in-circuit resistance, be sure the circuit under test has all power removed and all capacitors are fully discharged.

  • 4. Measuring capacitance

    Connect the black test lead to the COM jack and the red test lead to the mA/Cx jack. (NOTE: The polarity of the lead connection is positive +)

    Set the function switch al nF or F position to be used. Connect test leads across the capacitor under measurement and be sure that the

    polarity of connection is observed. NOTE:

    When checking in-circuit capacitance, be sure that the circuit has all power removed

    and all capacitor are fully discharged.

    The range control mode in capacitance measurement is manual ranging and only two

    ranges (326nF, 32.6F) are provided.

    If the range control button is used in this function measuring, decimal points may be at

    incorrect positions.

    At the nF range, when the capacitor to be measured is not connected to test leads, the

    LCD may not read zero, but a few counts. These counts have to be subtracted from

    measuring results.

    5. Measuring frequency Connect the black test lead to the COM jack and the red test lead to the V / / F jack.

    Set the function switch at the kHz position and connect test leads across the source or

    load under measurement.

    NOTE:

    The input voltage should be between 200mV and 10V rms AC. If the voltage is more

    than 10V rms, reading may be out of the accuracy range.

    6. Continuity & Diode test Connect the black test lead to the COM jack and the red test lead to the V / / F jack

    (NOTE: The polarity of the red lead connection is positive +)

    Set the function switch at position and push the button to select continuity

    or diode test mode.

    In continuity testing, if continuity exists (i.e., resistance less than about 50 ), built-in

    buzzer will sound.

    If diode test mode is selected, connect the red and black leads to the anode and

    cathode of diode under test. The forward voltage drop of this diode in V will be

    displayed.

  • 7. Transistor test Set the function switch al hFE position.

    Identity whether the transistor is NPN or PNP type and locate emitter, base and

    collector lead. Insert leads of the transistor to be tested into proper holes of the testing

    sockets on the front panel.

    LCD display will show the approximate hFE value at the test condition of base current

    10A and Vce 3.2V.

    PRELIMINARY When using the multimeter, the user must observe all normal safety rules concerning:

    o Protection against the dangers of electrical current. o Protection of the meter against misuse.

    Full compliance with safety standards can be guaranteed only if used with test leads

    supplied. If necessary, they must be replaced with the same model or same electric ratings.

    Measuring leads must be in good condition.

    DURING USE Never exceed the protection limit values indicated in specifications for each of

    measurement.

    When the device is linked to a measurement circuit, do not touch unused terminals.

    When the value scale to be measured is unknown beforehand, set the range selector at

    the highest position.

    Before rotating the range selector to change functions, disconnect test leads from the

    circuit under test.

    When carrying out measurements on TV or switching power circuits, always remember that

    there may be high amplitude voltages pulses at test points which can damage the meter.

    Never perform resistance measurements on live circuits.

    Never perform capacitance measurements unless the capacitor to be measured has been

    discharged fully.

    Always be careful when working with voltages above 60V DC or 30V AC rms. Keep fingers

    behind the probe barriers while measuring.

  • MAINTENANCE Before opening the device, always disconnect test leads from all sources of electric current.

    For continue protection against fire, replace fuse only with specified voltage and current ratings: F1:

    F 300mA/250V.

    If any faults or abnormalities are observed, the multimeter can not the be used any more and it has

    to be checked out.

    Never use the meter unless the back cover is in place and fastened fully.

    To clean the meter, use a damp cloth and mild detergent only, do not use abrasives or solvents on

    it.

  • MULTMETRO DIGITAL

    PARTES DEL MULTMETRO

    1. Botn para el control del rango

    2. Botn Data Hold

    3. Indicador de seal

    4. Botn de seleccin AC/DC

    5. Interruptor de funciones

    6. Entrada V//F

    7. Entrada COM

    8. Entrada mA/Cx

    9. Entrada 10A

    1. Botn de control de rango El rango para la medida del voltaje AC/DC, corriente AC/DC (solo A y mA), resistencia y

    frecuencia debe ser seleccionado manualmente. Pulsar el botn, eligiendo el rango deseado.

  • 2. Botn Data hold Cuando el botn est pulsado, la pantalla visualizar el ltimo valor ledo y el smbolo D-H

    deber aparecer hasta que lo pulses otra vez.

    La funcin data hold se cancelar automticamente cuando se rote el interruptor de funciones.

    3. Botn de seleccin o corriente AC/DC Pulsar dicho botn para seleccionar corriente AC o DC cuando el interruptor est cambiado a

    las posiciones A, mA, A.

    Pulsar dicho botn para seleccionar la funcin de medida cuando el interruptor est

    cambiado a la posicin .

    4. Entradas Este multmetro posee cuatro entradas que estn protegidas contra la sobrecarga de los lmites

    establecidos. Durante su uso, conectar el cable negro a la entrada COM y el rojo segn lo

    visualizado a continuacin:

    Funcin Conexin del cable rojo Lmites de entrada

    DCV/ACV V / / F 1000V DC o 750V rms AC

    kHz V / / F 250V DC o rms AC

    / V / / F 250V DC o rms AC

    A / mA mA / Cx 300mA DC o rms AC

    nF / F mA / Cx 300mA fusible protegido

    A 10A 10A DC o rms AC

    Los rangos A/mA estn protegidos por fusibles.

  • ESPECIFICACIONES TCNICAS

    1. Generales

    Pantalla: 3260 puntos max. LCD

    Max. Voltaje entre

    terminales y

    conexin a tierra

    1000V DC o 750V rms AC

    Proteccin elctrica A, mA: F 300 mA / 250V

    Mtodo de medida Convertidor A/D de integracin dual

    Mtodos del rango Auto/Manual

    Indicador estado por

    encima del rango OL en pantalla

    Indicador de

    polaridad - en pantalla

    Batera: 9V (Neda 1604 o 6F22)

    Operativa: 0- 40 C (32F-104F) Temperatura

    Almacenaje: -10C-50C (10F-122F) Condiciones de

    funcionamiento: Humedad: < 80%

    Tamao: 91 x 189 x 31.5 mm

    Peso: 310 gr.

    2. Voltaje DC

    Rango Resolucin Precisin

    326mV 0.1mV 0.5% (rdg 2 dgitos)

    3.26V 1mV 0.8% (rdg 2 dgitos)

    32.6V 10mV 0.8% (rdg 2 dgitos)

    326V 0.1V 0.8% (rdg 2 dgitos)

    1000V 1V 0.8% (rdg 2 dgitos)

    Impedancia de entrada: 10M, mas de 100M a 326mV de rango.

  • 3. Voltaje AC

    Rango Resolucin Precisin

    3.26V 1mV 1.2% (rdg 3 dgitos)

    32.6V 10mV 0.8% (rdg 3 dgitos)

    326V 0.1V 1.2% (rdg 3 dgitos)

    1000V 1V 1.2% (rdg 3 dgitos)

    Impedancia de entrada: 10M.

    Rango de frecuencia: 40 de 400Hz, 40 de 200Hz a 3.26V de rango.

    Respuesta: promedio, calibrado en rms

    4. Corriente DC Rango Resolucin Precisin Voltaje de carga

    326A 0.1A 1.2% (rdg 3 digitos) 0.5mV / A

    3260A 1A 1.2% (rdg 3 digitos) 0.5mV / A

    32.6mA 10A 1.2% (rdg 3 digitos) 8.0mV / mA

    326mA 0.1mA 1.2% (rdg 3 digitos) 8.0mV / mA

    10A 10mA 4.0% (rdg 3 digitos) 0.02V / A

    Proteccin de sobrecarga: fusible F 300mA para los rangos A y mA.

    5. Corriente AC Rango Resolucin Precisin Voltaje de carga

    326A 0.1A 1.5% (rdg 5 digitos) 0.5mV / A

    3260A 1A 1.8% (rdg 5 dgitos) 0.5mV / A

    32.6mA 10A 1.8% (rdg 5 dgitos) 8.0mV / mA

    326mA 0.1mA 2.0% (rdg 5 dgitos) 8.0mV / mA

    10A 10mA 5.0% (rdg 7 dgitos) 0.02V / A

    Proteccin de sobrecarga: fusible F 300mA para los rangos A y mA.

    Rango de frecuencia: 40 a 400 Hz

    Respuesta: promedio, calibrado en rms en ondas senoidales.

  • 6. Resistencia Rango Resolucin Precisin

    326 0.1 1.0% (rdg 3 dgitos)

    3.26k 1 1.0% (rdg 1 dgitos)

    32.6 k 10 1.0% (rdg 1 dgitos)

    326 k 100 1.0% (rdg 1 dgitos)

    3.26M 1k 1.0% (rdg 1 dgitos)

    32.6M 10k 2.0% (rdg 2 dgitos)

    Mximo voltaje del circuito abierto: 1.3V

    7. Condensadores Rango Resolucin Precisin

    326nF 0.1nF 5.0% (rdg 10 dgitos)

    32.6F 10nF 5.0% (rdg 10 dgitos)

    8. Frecuencia Rango Resolucin Precisin

    32.6kHz 10Hz 1.5% (rdg 10 dgitos)

    150kHz 100Hz 3.0% (rdg 10 dgitos)

    Sensibilidad: 500mV rms hasta 50kHz, 1V rms para 50kHz a 150kHz.

    BATERA Y CAMBIO DE FUSIBLE Si el smbolo aparece en la pantalla LCD, esto indica que la batera deber

    reemplazarse. Quitar los tornillos de la tapa de la batera y abrirla. Reemplazar la batera por

    otra nueva.

    El fusible raramente necesita ser reemplazado y casi siempre suele ser un error del multimetro.

    Abrir la tapa y coger el PCB. Reemplazar el fusible fundido segn el rango especificado.

    Atencin: Antes de abrir la tapa, estar seguros de que el cable est desconectado del circuito

    de medida para evitar problemas elctricos.

    Para protecciones contra el fuego, reemplazar los fusibles solo con los rangos especificados:

    F1: F 300mA/250V.

  • INSTRUCCIONES DE USO 1. Medida del voltaje

    Conectar el cable negro a la entrada COM y el cable rojo a la entrada V / / F.

    Seleccionar la funcin V V ~ y conectar el cable a travs de la fuente o debajo de la

    medida de carga.

    Leer en la pantalla LCD. La polaridad de la conexin roja debe estar indicada para

    hacer mediciones DC.

    2. Medida de corriente Conectar el cable negro a la entrada COM y el cable rojo a la entrada mA/Cx para un

    mximo de 300mA. Para un mximo de 10A, mover el cable rojo a la entrada A.

    Elegir la funcin de rango A, mA o A y pulsar el botn / ~ para seleccionar el modo

    de medida DCA o ACA.

    Conectar los cables en serie para medir la corriente.

    Leer en la pantalla LCD. La polaridad de la conexin roja debe estar indicada para

    hacer mediciones DC.

    3. Medida de resistencia Conectar el cable negro a la entrada COM y el cable rojo a la entrada V / / F (NOTA:

    La polaridad de la conexin es positiva +)

    Elegir la funcin y conectar los cables a la resistencia. NOTA:

    Para resistencia de 3.26M, el multimetro debe tomarse unos pocos segundos para estabilizar la lectura. Esto es normal para medida de altas resistencias.

    Cuando la entrada no est conectada, p.e., abrir el circuito, la figura OL debe visualizarse para condiciones de sobrecarga.

    Cuando se chequea la resistencia de los circuitos, estar seguro de que el circuito est desconectado, adems de todos los condensadores estn descargados.

  • 8. Medida de la capacidad de los condensadores

    Conectar el cable negro a la entrada COM y el rojo a la entrada mA/Cx. (NOTA: La polaridad de la conexin es positiva +)

    Elegir la funcin de la posicin nF o F. Conectar los cables alrededor del condensador y estar seguro de que la polaridad de la

    conexin se visualiza. NOTA:

    Cuando se chequea la capacidad de los condensadores del circuito, estar seguros de

    que todos los componentes del circuito estn quitados y todos los condensadores

    descargados.

    El modo de control de rango en medida de condensadores es manual y solo est

    provisto de dos rangos (326nF, 32.6F).

    Si el botn de control de rangos est siendo utilizado en esta funcin de medida, los

    puntos decimales estarn en posicin incorrecta.

    En el rango nF, cuando el condensador a medir no est conectado a los cables, la

    pantalla no debe mostrar cero. Dicho dato tiene que ser restado de los resultados

    medidos.

    9. Medida de frecuencia Conectar el cable negro a la entrada COM y el cable rojo a la entrada V / / F.

    Elegir la function kHz y conectar los cables alrededor de la carga. NOTA:

    El voltaje de entrada debera estar entre 200mV y 10V rms AC. Si el voltaje est ms

    de 10V rms, la lectura estar fuera de sus lmites.

    10. Continuidad y test de diodo

    Conectar el cable negro a la entrada COM y el cable rojo a la entrada V / / F (NOTA: La polaridad de la conexin es positiva +)

    Elegir la funcin y pulsar el botn para seleccionar la continuidad o el

    modo test de diodo.

    Para comprobar la continuidad, si la continuidad existe (p.e., resistencia menor a 50 ),

    el timbre deber sonar.

    Si el modo de test de diodo est seleccionado, conectar el cable rojo y el negro al

    ctodo y nodo del diodo. El voltaje ir disminuyendo, y deber visualizarse en

    pantalla.

  • 11. Test de transistor Elegir la funcin hFE.

    Aunque el transistor sea de tipo NPN o PNP, localizar el emisor, base y colector.

    Insertar los cables del transistor para chequear los agujeros adecuados del frontal del

    panel.

    La pantalla LCD deber visualizar el valor aproximado en hFE de la corriente base

    10A y Vce 3.2V.

    PRELIMINAR

    Cuando se usa el multmetro, el usuario tiene que seguir las siguientes normas de

    seguridad: o Proteccin contra daos de corriente elctrica. o Proteccin del multmetro contra usos incorrectos.

    Los estndares de seguridad deben estar garantizados solo si se usan los cables cagados.

    Si es necesario, tienen que reemplazarse por otro del mismo modelo. Los cables deben

    estar en buenas condiciones.

    DURANTE EL USO Nunca exceder los valores lmites de proteccin indicados en las especificaciones para

    cada medida.

    Cuando el multmetro est unido al circuito, no tocar los terminales.

    Cuando el valor de escala a medir es desconocido, posicionar el selector de rangos en la

    posicin ms alta.

    Antes cambiar el selector de rangos, desconectar los cables del circuito.

    Recordar siempre que amplios voltajes de corriente pueden daar el multmetro.

    Nunca medir resistencias en circuitos en tensin.

    Nunca realizar medidas de condensadores a menos que el condensador est descargado.

    Trabajar siempre con cuidado cuando se trabaja con voltajes alrededor de 60V DC o 30V

    AC rms. Colocar los dedos por debajo de la sonda protectora mientras se mide.

  • MANTENIMIENTO Antes de abrir el multmetro, siempre desconectar los cables de todas las cargas elctricas.

    Para continuar la proteccin contra el fuego, reemplazar el fusible solo con los rangos de corriente

    y voltaje especificados: F1: F 300mA/250V.

    Si se observa algn defecto o algo anormal, el multmetro puede que no se deba usar ms, y por

    ello hay que verificarlo.

    Nunca usar el multmetro a menos que la tapa de la batera est en su sitio y fijada correctamente.

    Para limpiar el multmetro, usar una trapo hmedo y un jabn suave, no usar abrasivos o

    disolventes sobre el.