CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2,...
-
Upload
annabelle-hopkins -
Category
Documents
-
view
216 -
download
1
Transcript of CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2,...
CVD Diamond Sensor Studies
for the Beam Calorimeter of the ILC Detector
K. Afanaciev2, I.Emelianchik2, Ch. Grah1, E. Kouznetsova1,
W. Lange1, W. Lohmann1
1 DESY, Zeuthen2 NCPHEP, Minsk
Beam Calorimeter
BeamCal: (4-28) mrad
• fast beam diagnostics
• detection and measurement of high energetic electrons and photons at very small angles
-
e+
e-
Z0
μ-
~
χ0
μ+
~ +
χ0
e+
e-
e+
e-
+-
σ ~ 102 fb (SPS1a)
σ ~ 106 fb
Beam Calorimeter Requirements:
Diamond/Tungsten sandwich
~15000 e+e- per BX (30 – 60 TeV)
~10 MGy / year for some area
=>
• radiation hard material
• with small Moliere radius
ILC bunch:small sizehigh charge
-> beamstrahlung :
-> high energy deposition in the Beam Cal:
pCVD Sensor Measurements
Requirements:
- stability under irradiation
- linearity of response
Measurements:
Current-Voltage characteristics
Charge Collection EfficiencyQmeas/Qcreated = CCD/L
Qcreated(MIP) = 36 eh/mm
Irradiation tests (up to 100 Gy)
Linearity measurements
Samples:
Fraunhofer IAF, Element Six
12x12 mm2, 300-900 m
&
Gate
PA
discr
discr
delayADC
Sr90
PM1
PM2
diamond
scint.
pCVD SensorsIrradiation Measurements:
~ 0.15 Gy/h
~ 0.60 Gy/h
~50%
Hadronic beam, 3 & 5 GeV
(CERN PS)
Fast extraction mode
~104-108 / ~10 ns
ADC
Diamond (+ PA) Scint.+PMT&
signal gate
pCVD Sensorslinearity test:
10 ns
17 s
Linearity testwide intensity range:
+ absolute intensity measurement (Thermoluminescence dosimetry)
relative intensities estimated from PMT’s: ~(1- 600)
HV(PMT)
intensity
L
R
),(),(),(
),(),(
)tan(,
1)(
202
0
2200
2/12
HVQdHVQdd
dHVQ
ddHVQddHVQ
Rd
Dd
ddI
PMT
LED
PMT calibration
Back to the testbeam data…HV(PMT)
intensity
LR
+ absolute calibration for one of the runs
1 RI = (27.3+/-2.9) 103 MIP/cm2
Linearity of diamond response
E64 FAP2
•The covered range : the particle fluence up to 107 MIPs per bunch
•The obtained deviation from a linear response: ~30% - at the level of systematic errors of the fluence calibration
Conclusions