CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2,...

11
CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2 , I.Emelianchik 2 , Ch. Grah 1 , E. Kouznetsova 1 , W. Lange 1 , W. Lohmann 1 1 DESY, Zeuthen 2 NCPHEP, Minsk

Transcript of CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2,...

Page 1: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

CVD Diamond Sensor Studies

for the Beam Calorimeter of the ILC Detector

K. Afanaciev2, I.Emelianchik2, Ch. Grah1, E. Kouznetsova1,

W. Lange1, W. Lohmann1

1 DESY, Zeuthen2 NCPHEP, Minsk

Page 2: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

Beam Calorimeter

BeamCal: (4-28) mrad

• fast beam diagnostics

• detection and measurement of high energetic electrons and photons at very small angles

-

e+

e-

Z0

μ-

~

χ0

μ+

~ +

χ0

e+

e-

e+

e-

+-

σ ~ 102 fb (SPS1a)

σ ~ 106 fb

Page 3: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

Beam Calorimeter Requirements:

Diamond/Tungsten sandwich

~15000 e+e- per BX (30 – 60 TeV)

~10 MGy / year for some area

=>

• radiation hard material

• with small Moliere radius

ILC bunch:small sizehigh charge

-> beamstrahlung :

-> high energy deposition in the Beam Cal:

Page 4: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

pCVD Sensor Measurements

Requirements:

- stability under irradiation

- linearity of response

Measurements:

Current-Voltage characteristics

Charge Collection EfficiencyQmeas/Qcreated = CCD/L

Qcreated(MIP) = 36 eh/mm

Irradiation tests (up to 100 Gy)

Linearity measurements

Samples:

Fraunhofer IAF, Element Six

12x12 mm2, 300-900 m

&

Gate

PA

discr

discr

delayADC

Sr90

PM1

PM2

diamond

scint.

Page 5: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

pCVD SensorsIrradiation Measurements:

~ 0.15 Gy/h

~ 0.60 Gy/h

~50%

Page 6: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

Hadronic beam, 3 & 5 GeV

(CERN PS)

Fast extraction mode

~104-108 / ~10 ns

ADC

Diamond (+ PA) Scint.+PMT&

signal gate

pCVD Sensorslinearity test:

10 ns

17 s

Page 7: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

Linearity testwide intensity range:

+ absolute intensity measurement (Thermoluminescence dosimetry)

relative intensities estimated from PMT’s: ~(1- 600)

HV(PMT)

intensity

L

R

Page 8: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

),(),(),(

),(),(

)tan(,

1)(

202

0

2200

2/12

HVQdHVQdd

dHVQ

ddHVQddHVQ

Rd

Dd

ddI

PMT

LED

PMT calibration

Page 9: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

Back to the testbeam data…HV(PMT)

intensity

LR

+ absolute calibration for one of the runs

1 RI = (27.3+/-2.9) 103 MIP/cm2

Page 10: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

Linearity of diamond response

E64 FAP2

Page 11: CVD Diamond Sensor Studies for the Beam Calorimeter of the ILC Detector K. Afanaciev 2, I.Emelianchik 2, Ch. Grah 1, E. Kouznetsova 1, W. Lange 1, W. Lohmann.

•The covered range : the particle fluence up to 107 MIPs per bunch

•The obtained deviation from a linear response: ~30% - at the level of systematic errors of the fluence calibration

Conclusions