Applications of the Prominence RF-20Axs Fluorescence Detector ...

LAAN-A-LC-E184 No.L406 High Performance Liquid Chromatography Applications of the Prominence RF-20Axs Fluorescence Detector (Part 5) Analysis of Voglibose with Postcolumn Derivatization System Voglibose is a diabetes drug which inhibits the activity of α-glucosidase. HPLC post-column derivatization is specified as the test method for voglibose in the Japanese Pharmacopeia, Fifteenth Edition (voglibose purity test, quantitation of voglibose in pharmaceutical tablets). Here we introduce an example of voglibose tablet testing as specified in the Japanese Pharmacopeia, Fifteenth Edition, using the Prominence post-column derivatization system with the Prominence RF-20Axs fluorescence detector. Fig. 1 shows the flow diagram of this post-column derivatization system. The Japanese Pharmacopeia specifies that after mixing with the reaction solution (taurine / sodium periodate solution) at about 100 °C, the solution is to be cooled at a constant temperature of about 15 °C. However, since the RF-20Axs is equipped with cell temperature control, we adjusted the cell temperature to 15 °C. Not only does this eliminate the requirement for a temperature-controlled cooling bath, it can also provide improved accuracy and reduce fluorescence quenching at elevated temperatures. Fig. 2 *3 shows the chromatogram obtained following injection of 50 μL of the voglibose standard solution *1 (250 μg/L *2 , prepared using mobile phase), and Table 1 *4 shows the analytical conditions. *1: The voglibose standard was provided by Sawai Pharmaceutical Co., Ltd. *2: The Japanese Pharmacopeia specifies a standard solution concentration of 40 mg/L in the quantitative method. *3: The peak in the vicinity of 5 minutes in Fig. 2 and Fig. 4 originates from the sample solvent. *4: The Japanese Pharmacopeia indicates use of a column with an inner diameter of 4 mm, but here analysis was conducted using a column having a 4.6 mm inner diameter. n Analysis of Standard Solution 1. Mobile Phase 2. Pump for Mobile Phase (LC-20AD) 3. Autosampler (SIL-20AC) 4. Column Oven (CTO-20AC) 5. Column 6. Reaction Reagent 7. Pump for Reaction Reagent (LC-20AD) 8. Chemical Reaction Box (CRB-6A) 9. Reaction Coil 10. Cooling Coil 11. Detector (RF-20Axs) 12. Data Processor waste 1 3 4 5 6 7 8 9 10 11 12 2 5.0 4.0 3.0 2.0 1.0 0.0 0.0 5.0 10.0 15.0 20.0 min mV Peak 1. Voglibose (250 μg/L) *2 (Blank) 1 Fig. 1 Flow Diagram Fig. 2 Chromatogram of Voglibose Standard (250 μg/L, 50 μL injected) <Separation> Column : Shim-pack CLC-NH2 (M) (150 mm L. × 4.6 mm I.D., 5 µm) Mobile Phase : (Sodium) phosphate buffer (pH 6.5) / Acetonitrile = 300 / 600 (v/v) Flow Rate : 0.8 mL/min Column Temp. : 25 ˚C Injection Volume : 50 µL <Detection> Reaction Reagent : Taurine / Sodium periodate aq. solution Flow Rate : 0.8 mL/min Reaction Coil : PTFE, 20 m L. × 0.5 mm I.D. Reaction Temp. : 100 ˚C Cooling Coil : PTFE, 2 m L. × 0.3 mm I.D. Cooling Temp. : 25 ˚C (CTO-20AC) 15 ˚C (RF-20Axs) Detection : RF-20Axs Ex. at 350 nm, Em. at 430 nm Cell Temp. : 15 ˚C Table 1 Analytical Conditions

Transcript of Applications of the Prominence RF-20Axs Fluorescence Detector ...

Page 1: Applications of the Prominence RF-20Axs Fluorescence Detector ...

LAAN-A-LC-E184

No.L406High Per formance Liquid Chromatography

Applications of the Prominence RF-20Axs Fluorescence Detector (Part 5)Analysis of Voglibose with Postcolumn Derivatization System

Voglibose is a diabetes drug which inhibits the activity of α-glucosidase. HPLC post-column derivatization is specified as the test method for voglibose in the Japanese Pharmacopeia, Fifteenth Edition (voglibose purity test, quantitation of voglibose in pharmaceutical

tablets). Here we introduce an example of voglibose tablet testing as specif ied in the Japanese Pharmacopeia, Fifteenth Edition, using the Prominence post-column derivatization system with the Prominence RF-20Axs fluorescence detector.

Fig. 1 shows the flow diagram of this post-column derivatization system. The Japanese Pharmacopeia specifies that after mixing with the reaction solution (taurine / sodium periodate solution) at about 100 °C, the solution is to be cooled at a constant temperature of about 15 °C. However, since the RF-20Axs is equipped with cell temperature control, we adjusted the cell temperature to 15 °C. Not only does this eliminate the requirement for a temperature-controlled cooling bath, it can also provide improved accuracy and reduce fluorescence quenching at elevated temperatures.

Fig. 2*3 shows the chromatogram obtained following injection of 50 μL of the voglibose standard solution*1 (250 μg/L*2, prepared using mobile phase), and Table 1*4 shows the analytical conditions.*1: The voglibose standard was provided by Sawai Pharmaceutical

Co., Ltd.*2: The Japanese Pharmacopeia specifies a standard solution

concentration of 40 mg/L in the quantitative method.*3: The peak in the vicinity of 5 minutes in Fig. 2 and Fig. 4 originates

from the sample solvent.*4: The Japanese Pharmacopeia indicates use of a column with an

inner diameter of 4 mm, but here analysis was conducted using a column having a 4.6 mm inner diameter.

n Analysis of Standard Solution

1. Mobile Phase2. Pump for Mobile Phase (LC-20AD)3. Autosampler (SIL-20AC)4. Column Oven (CTO-20AC)5. Column6. Reaction Reagent

7. Pump for Reaction Reagent (LC-20AD)8. Chemical Reaction Box (CRB-6A)9. Reaction Coil10. Cooling Coil11. Detector (RF-20Axs)12. Data Processor

waste

1

3

4

5

67

8

9

1011

12

2

5.0

4.0

3.0

2.0

1.0

0.0

0.0 5.0 10.0 15.0 20.0 min

mV■ Peak

1. Voglibose (250 μg/L)*2

(Blank)

1

Fig. 1 Flow Diagram Fig. 2 Chromatogram of Voglibose Standard (250 μg/L, 50 μL injected)

<Separation>Column : Shim-pack CLC-NH2 (M)

(150 mm L. × 4.6 mm I.D., 5 µm)Mobile Phase : (Sodium) phosphate buffer (pH 6.5) / Acetonitrile

= 300 / 600 (v/v)Flow Rate : 0.8 mL/minColumn Temp. : 25 ˚C Injection Volume : 50 µL

<Detection>Reaction Reagent : Taurine / Sodium periodate aq. solutionFlow Rate : 0.8 mL/minReaction Coil : PTFE, 20 m L. × 0.5 mm I.D.Reaction Temp. : 100 ˚C Cooling Coil : PTFE, 2 m L. × 0.3 mm I.D.Cooling Temp. : 25 ˚C (CTO-20AC) → 15 ˚C (RF-20Axs)Detection : RF-20Axs Ex. at 350 nm, Em. at 430 nmCell Temp. : 15 ˚C

Table 1 Analytical Conditions

Page 2: Applications of the Prominence RF-20Axs Fluorescence Detector ...

SHIMADZU CORPORATION. International Marketing Division3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan Phone: 81(3)3219-5641 Fax. 81(3)3219-5710Cable Add.:SHIMADZU TOKYO

No.L406

Fig. 3 shows the calibration curve obtained from analysis of voglibose standard solutions with concentrations of 2-250 μg/L (50 μL injected). Excellent linearity was obtained, with an R2 value greater than 0.9999.

We conducted system suitability testing for voglibose tablets as specified in the Japanese Pharmacopeia, Fifteenth Edition. The left-hand chromatogram in Fig. 5 was obtained using a test solution containing lactose and voglibose. The 8.2 peak resolution clearly satisfies the official criterion value (4 or greater). The overlaid

chromatograms at the right in Fig. 5 were obtained using 50 μL injections of 40 mg/L voglibose standard solution. The peak area repeatability for these voglibose analyses was 0.16 % RSD (n = 6), satisfying the criterion value of 2.0 % or less.

Fig. 4 shows the chromatogram obtained following injection of 50 μL of a 2 μg/L voglibose standard solution. The voglibose peak area repeatability (n = 5) for this analysis was 1.4 % RSD.Use of the RF-20Axs permits micro-level analysis of voglibose at high sensitivity and with high accuracy.

n Linearity

n System Suitability Test

n Analysis of Voglibose at High Sensitivity

■ Peaks1. Lactose (4 g/L)2. Voglibose (40 mg/L)

17.5

15.0

12.5

10.0

7.5

5.0

2.5

0.0

mV

1

2

0.0 5.0 10.0 15.0 20.0 25.0 min

■ Peak1. Voglibose (40 mg/L)

45

40

35

30

25

20

15

10

5

0

mV

0.0 5.0 10.0 15.0 20.0 25.0 min

1

Fig. 5 System Suitability Test / System Performance (Left), System Repeatability (Right)

125000

100000

75000

50000

25000

0

Concentration (μg/L)

Peak

Are

a

0 50 100 150 200 250

100

90

80

70

60

50

40

30

20

10

0

-10

-20

1

μV

(Blank)

0 5.0 10.0 15.0 20.0 min

■ Peak1. Voglibose (2 μg/L)

Fig. 3 Calibration Curve (2-250 μg/L, 50 μL injected) Fig. 4 Chromatogram of Voglibose Standard at High Sensitivity (2 μg/L, 50 μL injected)

[References]The Japanese Pharmacopeia, Fifteenth Edition (Edited by Pharmaceutical and Medical Device Regulatory Science Society of Japan)