100% Aqueous Compatible - MAC-MODmac-mod.com/pdf/product-bulletins/HALO AQ-C18.pdf · Advantages of...
Transcript of 100% Aqueous Compatible - MAC-MODmac-mod.com/pdf/product-bulletins/HALO AQ-C18.pdf · Advantages of...
100% Aqueous Compatible
INTRODUCING HALO® 90Å AQ-C18Introducing the new HALO® AQ-C18, a new bonded phase that is available on the Fused-Core® 2.7 μm superficially porous particle design with a 0.5 μm shell and 90Å pores. The AQ-C18 is a C18 bonded phase prepared using a proprietary procedure that increases phase polarity, making the AQ-C18 an excellent alternative C18 to consider, especially in aqueous mobile phases. HALO AQ-C18 is very complimentary to classic HALO C18 and extends C18 usefulness to 100% aqueous mobile phases.
Advantages of the New AQ-C18: • Resistant to dewetting and compatible with 100% aqueous mobile phases • Different selectivity than HALO C18, offering another option to resolve difficult peak pairs • Retains polar molecules more than classic C18 phases under most mobile phase conditions
Figure 1. The unique polar modified bonded phase of HALO AQ-C18 enables it to be run in 100% aqueous mobile phase without experiencing loss in retention due to dewetting when pressure is relieved. The retention is nearly 100% maintained compared to the HALO C18 after the pump is stopped and restarted.
RESISTANCE TO DEWETTING
TEST CONDITIONS:Column: 4.6 x 50 mm Top: HALO 90Å AQ-C18, 2.7 µm Bottom: HALO 90Å C18, 2.7 µmPart Numbers: Top: 92814-422 Bottom: 92814-402
Mobile Phase: 100% 20 mM Potassium Phosphate buffer, pH 7Flow Rate: 2 mL/minTemperature: 30 °C
Detection: 254 nmInjection: 0.5 µLSample: (1) thiourea, (2) 5-fluorocytosine, (3) adenine and (4) thymine
advanced-materials-tech.com
Figure 3. The manufacturing process of HALO 90Å AQ-C18 is highly controlled, which results in a reproducible product as demonstrated by the lot-to-lot comparisons in Figure 3. The percent RSDs for the retention factors are all below 1.5% while the alpha values of adjacent peaks have percent RSDs lower than 1%.
Figure 4. Stability of the HALO 90Å AQ-C18 columns is demonstrated using pH 2 mobile phase conditions. The retention is maintained after 19,000 column volumes.
INCREASED RETENTION FOR POLAR COMPOUNDS — DIFFERENT SELECTIVITY
LOT-TO-LOT REPRODUCIBILITY
HIGH STABILITY AT LOW PH
TEST CONDITIONS:Column: 4.6 x 100 mm
Top: HALO 90Å AQ-C18, 2.7 µm
Bottom: HALO 90Å C18, 2.7 µm
Part Numbers: Top: 92814-622
Bottom: 92814-602
Mobile Phase A: Water
Mobile Phase B: Methanol
Isocratic: 50/50 A/B
Flow Rate: 1.3 mL/min
Temperature: 35 °C
Detection: 254 nm
Injection: 0.5 µL
Sample: (1) cinnamyl alcohol, (2) 4-bromoacetanilide, (3) nitrobenzene, (4) 3,4-dinitrotoluene, (5) anisole and (6) 2,4-dinitrotoluene
TEST CONDITIONS:Column: 4.6 x 100 mm
HALO 90Å AQ-C18, 2.7 µm
Part Number: 92814-622
Mobile Phase A: Water
Mobile Phase B: Methanol
Isocratic: 50/50 A/B
Flow Rate: 1.3 mL/min
Temperature: 35 °C
Detection: 254 nm
Injection: 0.5 µL
Sample: (1) cinnamyl alcohol, (2) 4-bromoacetanilide, (3) nitrobenzene, (4) 3,4-dinitrotoluene, (5) anisole and (6) 2,4-dinitrotoluene
TEST CONDITIONS:Column: 2.1 x 50 mm
HALO 90Å AQ-C18, 2.7 µm
Part Number: 92812-422
Mobile Phase A: 50 mM KCl/HCl, pH 2
Mobile Phase B: Acetonitrile
Isocratic: 60/40 A/B
Flow Rate: 1.0 mL/min
Pressure: 235 bar
Temperature: 60 °C
Detection: 254 nm
Injection: 0.5 µL
Sample: (1) amitriptyline and (2) naphthalene
Figure 2. The HALO AQ-C18 column exhibits increased retention of polar compounds along with a different elution order relative to the HALO C18.
3521 Silverside Road, Suite 1-KWilmington, DE 19810
HALO® and Fused-Core® are registered trademarks of Advanced Materials Technology, Inc.
ACT NOWContact your local distributor to be among
the first to experience these new HALO® columns!www.advanced-materials-tech.com/find-a-distributor/
HALO AQ-C18
Dimension - ID x length (mm) Part No.
Dimension - ID x length (mm) Part No.
2.1 x 20 92812-222 3.0 x 100 92813-622
2.1 x 30 92812-322 3.0 x 150 92813-722
2.1 x 50 92812-422 3.0 x 250 92813-922
2.1 x 75 92812-522 4.6 x 20 92814-222
2.1 x 100 92812-622 4.6 x 30 92814-322
2.1 x 150 92812-722 4.6 x 50 92814-422
2.1 x 250 92812-922 4.6 x 75 92814-522
3.0 x 20 92813-222 4.6 x 100 92814-622
3.0 x 30 92813-322 4.6 x 150 92814-722
3.0 x 50 92813-422 4.6 x 250 92814-922
3.0 x 75 92813-522
HALO AQ-C18 Guard Columns, 3/Pack
Dimension - ID x length (mm) Part No.
2.1 x 5 92812-122
3.0 x 5 92813-122
4.6 x 5 92814-122
Guard Column Holder (1) 94900-001